Literature DB >> 32256887

Atrial signal amplitude predicts atrial high-rate episodes in implantable cardioverter defibrillator patients: Insights from a large database of remote monitoring transmissions.

Massimo Zecchin1, Francesco Solimene2, Antonio D'Onofrio3, Gabriele Zanotto4, Saverio Iacopino5, Carlo Pignalberi6, Valeria Calvi7, Giampiero Maglia8, Paolo Della Bella9, Fabio Quartieri10, Antonio Curnis11, Mauro Biffi12, Alessandro Capucci13, Fabrizio Caravati14, Gaetano Senatore15, Matteo Santamaria16, Fabio Lissoni17, Michele Manzo18, Massimiliano Marini19, Massimo Giammaria20, Antonio Rapacciuolo21, Gianfranco Sinagra1, Daniele Giacopelli22, Alessio Gargaro22, Ennio C Pisanò23.   

Abstract

BACKGROUND: Parameters measured during implantable cardioverter defibrillator (ICD) implant also depend on bioelectrical properties of the myocardium. We aimed to explore their potential association with clinical outcomes in patients with single/dual-chamber ICD and cardiac resynchronization therapy defibrillator (CRT-D).
METHODS: In the framework of the Home Monitoring Expert Alliance, baseline electrical parameters for all implanted leads were compared by the occurrence of all-cause mortality, adjudicated ventricular arrhythmia (VA), and atrial high-rate episode lasting ≥24 hours (24 h AHRE).
RESULTS: In a cohort of 2976 patients (58.1% ICD) with a median follow-up of 25 months, event rates were 3.1/100 patient-years for all-cause mortality, 18.1/100 patient-years for VA, and 9.3/100 patient-years for 24 h AHRE. At univariate analysis, baseline shock impedance was consistently lower in groups with events than without, with a 40 Ω cutoff that better identified high-risk patients. However, at multivariable analysis, the adjusted-hazard ratios (HRs) lost statistical significance for any endpoint. Baseline atrial sensing amplitude during sinus rhythm was lower in patients with 24 h AHRE than in those without (2.45 [IQR: 1.65-3.85] vs 3.51 [IQR: 2.37-4.67] mV, P < .01). The adjusted HR for 24 h AHRE in patients with atrial sensing >1.5 mV vs those with values ≤1.5 mV was 0.52 (95% CI: 0.33-0.83), P = .006.
CONCLUSIONS: Although lower baseline shock impedance was observed in patients with events, the association lost statistical significance at multivariable analysis. Conversely, low sinus rhythm atrial sensing (≤1.5 mV) measured with standard transvenous leads could identify subjects at high risk of atrial arrhythmia.
© 2020 The Authors. Journal of Arrhythmia published by John Wiley & Sons Australia, Ltd on behalf of the Japanese Heart Rhythm Society.

Entities:  

Keywords:  cardiac resynchronization therapy; impedance; implantable cardioverter defibrillator; pacing threshold; sensing

Year:  2020        PMID: 32256887      PMCID: PMC7132187          DOI: 10.1002/joa3.12319

Source DB:  PubMed          Journal:  J Arrhythm        ISSN: 1880-4276


INTRODUCTION

During implant of implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT‐Ds), pacing threshold, impedance, and sensing amplitude are routinely assessed for all implanted leads. Their monitoring during follow‐up with regular in‐office visits or remote control is then used for the surveillance of integrity and functioning of leads. Beyond technical aspects, several factors may influence these measurements, including properties of the myocardial tissue surrounding the lead electrodes. , Prior studies suggested an association between temporal changes in some of these parameters and clinical events, such as heart failure functional class changes or cardiac arrhythmias occurrence. , , However, it is unknown whether their values at implant could have a clinical relevance as a systematic analysis of association between baseline measurements and clinical outcomes has never been performed. The aim of the present study was to explore whether baseline electrical parameters routinely measured during implant have an association with long‐term mortality or incidence of atrial and ventricular arrhythmias (VAs) in ICD/CRT‐D recipients.

METHODS

The present analysis was performed in the framework of the Home Monitoring Expert Alliance (HMEA), an independent scientific project based on a nationwide repository of data generated by remote monitoring (RM) of cardiac implantable electronic devices (CIEDs) during ordinary medical practice. A total of 41 Italian sites, listed in the Appendix, provided data for this analysis. All included patients provided written informed consent before RM activation.

Objective and patient selection

We aimed to investigate whether baseline electrical parameters routinely measured during CIED implant could show an association with long‐term mortality or incidence of atrial and VAs. All ICD and CRT‐D recipients registered in the HMEA database were selected for the present analysis. Devices were implanted from January 2007 to March 2017.

Data collection and analysis endpoints

Baseline data were collected at the time of device implant. They included patient characteristics and electrical parameters routinely measured for any implanted lead during ICD/CRT‐D implant procedures. Pacing impedance, capture threshold, and sensing amplitude were collected for atrial, right ventricle (RV), and left ventricle (LV) leads. Electrical parameters were usually obtained in bipolar configuration, expect for a minority (2.7%) of unipolar LV leads. In addition, high‐voltage shock impedance, measured between distal coil and case, was reported for all RV leads. Atrial parameters were not included in the present analysis if the patient was in atrial fibrillation (AF) at implant. All CIEDs were manufactured by the same company (Biotronik) and follow‐up data were automatically and daily generated by the RM system (Home Monitoring; Biotronik), which provided device diagnostics and intracardiac electrogram (IEGM) recordings of all atrial and ventricular arrhythmic episodes. Atrial high‐rate episodes (AHREs) were recorded and transmitted based on a rate criterion set at 200 bpm (standard setting of devices). The atrial sensitivity adapted automatically on an ongoing basis to the measured amplitude of the atrial activity. The lowest sensing threshold that can be reached was 0.1 mV. VAs were automatically classified by the discrimination algorithms of the device. The median detection cutoff programed for the first VA zone was 158 (IQR: 150‐171) beats/min with a counter of 28 (IQR: 26‐40) beats to detect. In order to exclude false episodes from the analysis, the first VA detected by the device was adjudicated by a three‐member board who visually reviewed IEGM recordings. A two‐stage adjudication process was used: a first stage used an algorithm of objective electrophysiological criteria such as ventricular cycle The endpoints of the analysis were time to all‐cause death and to first postimplant adjudicated VA and AHRE lasting ≥24 hours (24 h AHRE). All‐cause mortality was estimated after site staff confirmed death status of patients with interrupted RM transmissions and no evidence of device replacement. For the AHRE analysis, the 24 hour duration threshold was used because it could be the sign of impaired atrial tissue and has recently been reported as the duration associated with an increased risk of ischemic stroke or systemic embolism. Short‐lasting atrial arrhythmias seem to have less clinical significance and an immediate anticoagulation in these patients is unlikely to result in reduction of the risk of stroke. Only patients who had atrial diagnostics capability were selected for this endpoint. In order to investigate whether some of the baseline electrical parameters could be a marker of endpoint occurrence, variables that had significant differences in the descriptive analysis were used to stratify event rates by value classes. The endpoints were then compared between the two subgroups defined using the value that maximized the difference in event rate as the cutoff value.

Statistical analysis

We described the selected population by using all‐cause death, VA, and 24 h AHRE occurrence as grouping criteria. Binary and categorical variables were reported as percentages of available data, and continuous variables as median (interquartile range [IQR]). Baseline between‐group comparisons were performed with the Wilcoxon signed‐rank test for continuous variable, Pearson chi‐squared or Fisher's tests for noncontinuous variables, as appropriate. Event rates were reported as the number of events divided by the amount of person‐time observed; the 95% confidence intervals (CIs) were calculated by means of the Poisson distribution within an iterative procedure. Kaplan‐Meier curves were generated and compared between groups using adjusted and unadjusted proportional hazard models. Adjusting covariates were age, sex, presence of hypertension, diabetes, ischemic cardiomyopathy, history of AF, and CHA2DS2‐VASC score. Statistical significance was defined as P < .05. All statistical analyses were performed using the version 11E of STATA software (StatCorp LB).

RESULTS

Population

A total of 2976 patients were included in the present analysis, 827 (27.8%) implanted with a single‐chamber ICD, 902 (30.3%) with a dual‐chamber ICD, and 1247 (41.9%) with a CRT‐D device. Baseline characteristics are reported in Table 1.
TABLE 1

Patient characteristics

 TotalSurvivorsDeceased P * Free from VAVA P * Free from 24 h AHRE24 h AHRE P *
Number of patients2976284922720229541998500
Follow‐up (mo)25 [12‐44]25 [12‐44]25 [12‐44].5521 [10‐37]37 [22‐51]<.0124 [12‐42]26 [14‐47].01
Age (y)70 [61‐77]69 [60‐77]75 [69‐80]<.0170 [31‐77]69 [61‐77].4369 [60‐76]74 [67‐79]<.01
Sex (female)19.4%19.4%19.8%.8620.9%16.4%.0120.9%16.0%.02
NYHA functional class
I‐II71.3%73.5%48.2%<.0172.1%69.6%.5672.7%66.1%<.01
III‐IV27.5%25.5%49.4%26.8%29.1%27.3%33.9%
LVEF, %30 [27‐35]30 [28‐35]30 [25‐35].0130 [28‐35]30 [25‐35].9730 [27‐35]30 [28‐35].95
QRS duration (ms)120 [100‐140]120 [100‐140]130 [120‐150]<.01120 [100‐140]120 [100‐140].89120 [100‐143]130 [106‐145].05
Device type
Single chamber ICD27.8%28.2%23.3%.0127.9%27.0%.83<.01
Dual‐chamber ICD30.3%30.9%22.9%30.2%33.0%57.5%46.7%
CRT‐D41.9%40.9%53.7%42.0%40.0%42.5%53.3%
Comorbidities
Hypertension52.4%52.7%48.4%.2653.1%51.0%.3552.6%49.5%.26
Diabetes23.6%23.1%29.8%.0425.2%20.2%.0123.7%24.1%.88
Stroke/TIA8.7%7.9%11.3%.108.0%8.7%.538.8%7.9%.56
Chronic kidney disease13.0%11.7%27.5%<.0113.1%12.6%.7312.6%15.1%.19
History of heart failure23.4%22.9%32.5%.0224.6%20.5%.0422.2%22.8%.82
CHA2DS2‐VASC class
0‐110.2%10.9%0.6%<.019.4%12.0%.3910.9%6.4%<.01
216.2%17.0%7.0%17.3%14.0%17.1%13.1%
323.4%23.3%24.0%23.3%23.4%23.3%21.3%
424.9%24.4%30.4%24.1%26.4%24.3%23.2%
≥525.3%24.4%38.0%25.9%24.2%24.4%36.0%
Cardiomyopathy
Ischemic50.8%49.8%61.8%.0151.0%50.3%.7451.4%53.8%.41
Dilated idiopathic35.2%35.6%30.4%.1534.6%36.3%.4335.6%33.9%.52
Valvular7.7%7.6%8.5%.666.8%9.7%.015.9%8.6%.07
Other5.8%6.0%2.1%.246.5%4.2%.044.9%5.9%.85
Documented arrhythmias
Ventricular fibrillation7.8%8.2%3.2.026.8%9.9%.017.6%6.3%.37
Sustained VT16.1%15.8%19.9%.1412.3%24.5%<.0115.0%13.8%.55
History of AF21.6%20.9%29.1%.0120.1%24.7%.019.5%49.8%<.01
Paroxysmal AF7.0%6.7%9.9% 7.1%6.7% 7.0%13.2% 
Persistent/permanent AF14.6%14.2%19.2% 13.0%18.0% 2.5%36.6% 

Data are reported as median [interquartile range] or percentage.

Only devices with atrial sensing capability were included in the 24 h AHRE analysis.

Abbreviations: AF, atrial fibrillation; AHRE, atrial high‐rate episode; CRT‐D, cardiac resynchronization therapy‐defibrillator; ICD, implantable cardioverter defibrillator; LVEF, left ventricular ejection fraction; TIA, transient ischemic attack; VA, ventricular arrhythmias; VT, ventricular tachycardia.

Wilcoxon signed‐rank test or Pearson χ2.

Patient characteristics Data are reported as median [interquartile range] or percentage. Only devices with atrial sensing capability were included in the 24 h AHRE analysis. Abbreviations: AF, atrial fibrillation; AHRE, atrial high‐rate episode; CRT‐D, cardiac resynchronization therapy‐defibrillator; ICD, implantable cardioverter defibrillator; LVEF, left ventricular ejection fraction; TIA, transient ischemic attack; VA, ventricular arrhythmias; VT, ventricular tachycardia. Wilcoxon signed‐rank test or Pearson χ2. During a median follow‐up of 25 [IQR: 12‐44] months, there were 227 (7.6%) all‐cause deaths (event rate: 3.1/100 patient‐years). As expected, survivors were younger, had lower New York Heart Association (NYHA) functional class and QRS duration, and higher left ventricular ejection fraction. Adjudicated VAs were found in 954 (32.0%) patients (event rate: 18.1/100 patient‐years), more frequently in males and in subjects implanted for secondary prevention and with history of AF. Among the 2498 patients implanted with devices with atrial sensing capability, 500 (20.0%) developed 24 h AHRE (event rate: 9.3/100 patient‐years). The group with atrial arrhythmia had higher age, lower female prevalence and, as expected, a very significant proportion (49.8%) of patients with history of AF before enrollment (despite in sinus rhythm at implant).

Baseline electrical parameters and all‐cause death

Some of the baseline electrical parameters were statistically different between survivors and deceased patients (Table 2). Atrial sensing (3.38 [IQR: 2.30‐4.61] mV vs 2.63 [IQR: 1.74‐3.86] mV, P < .01), LV pacing impedance (621 [IQR: 498‐749] Ω vs 579 [IQR: 460‐695] Ω, P = .01), and LV sensing (12.0 [IQR: 8.32‐16.5] Ω vs 10.7 [IQR: 8.09‐12.8] Ω, P = .02) were slightly higher in survivors. A more marked difference was observed in the shock impedance with a median value of 61 [IQR: 52‐69] Ω in the survivors compared to 51 [IQR: 41‐61] Ω in deceased subjects (P < .01).
TABLE 2

Baseline electrical parameters by groups

 TotalSurvivorsDeceased P * Free from VAVA P * Free from 24 h AHRE24 h AHRE P *
Atrial pacing impedance (Ω) 554 [487‐645]554 [489‐644]542 [463‐673].53554 [491‐637]552 [476‐668].84554 [488‐646]549 [480‐641].39
Atrial threshold (V) 0.80 [0.60‐1.02]0.80 [0.60‐1.02]0.80 [0.60‐1.02].930.85 [0.64‐1.06]0.80 [0.60‐1.00] .03 0.80 [0.60‐1.03]0.81 [0.65‐1.02].16
Atrial sensing (mV) 3.33 [2.26‐4.58]3.38 [2.30‐4.61]2.63 [1.74‐3.86] <.01 3.44 [2.30‐4.66]3.18 [2.19‐4.34] .01 3.51 [2.37‐4.67]2.45 [1.65‐3.85] <.01
RV pacing impedance (Ω)524 [469‐595]524 [470‐595]515 [452‐605].38524 [470‐594]524 [468‐603].69522 [470‐592]516 [464‐592].35
RV threshold (V)0.60 [0.50‐0.77]0.60 [0.50‐0.77]0.61 [0.50‐0.92].140.60 [0.50‐0.77]0.60 [0.49‐0.77].940.61 [0.50‐0.78]0.60 [0.50‐0.80].86
RV sensing (mV)11.7 [8.48‐15.9]11.7 [8.48‐16.1]11.4 [9.16‐14.7].4112.1 [8.63‐16.5]11.1 [8.24‐14.3] <.01 11.9 [8.65‐16.3]11.0 [8.19‐14.6] <.01
LV pacing impedance (Ω)616 [494‐743]621 [498‐749]579 [460‐695] .01 606 [486‐741]633 [522‐750] .03 610 [492‐738]648 [525‐793] .01
LV threshold (V)1.01 [0.70‐1.42]1.02 [0.71‐1.42]0.98 [0.62‐1.48].541.01 [0.73‐1.44]1.01 [0.66‐1.35].261.10 [0.75‐1.59]1.01 [0.80‐1.50].70
LV sensing (mV)11.8 [8.25‐15.9]12.0 [8.32‐16.5]10.7 [8.09‐12.8] .02 12.2 [8.25‐16.9]11.1 [8.33‐14.4] .01 12.0 [8.29‐16.8]11.6 [8.90‐14.6].17
Shock impedance (Ω)61 [51‐69]61 [52‐69]51 [41‐61] <.01 62 [53‐70]56 [46‐65] <.01 62 [52‐70]56 [46‐66] <.01

Data are reported as median [interquartile range].

Only devices with atrial sensing capability were included in the 24 h AHRE analysis.

Abbreviations as listed in Table 1.

Wilcoxon signed‐rank test. Bold values if p < .05.

Excluding patients in atrial fibrillation at implant.

Baseline electrical parameters by groups Data are reported as median [interquartile range]. Only devices with atrial sensing capability were included in the 24 h AHRE analysis. Abbreviations as listed in Table 1. Wilcoxon signed‐rank test. Bold values if p < .05. Excluding patients in atrial fibrillation at implant.

Baseline electrical parameters and ventricular arrhythmias

Patients without adjudicated VA had slightly higher atrial threshold (0.85 [IQR: 0.64‐1.06] vs 0.80 [IQR: 0.60‐1.00] V, P = .03), atrial signal amplitude (3.44 [IQR: 2.30‐4.66] mV vs 3.18 [IQR: 2.19‐4.34] mV, P = .01), RV signal amplitude (12.1 [IQR: 8.63‐16.5] mV vs 11.1 [IQR: 8.24‐14.3] mV, P < .01), and LV signal amplitude (12.2 [IQR: 8.25‐16.9] mV vs 11.1 [IQR: 8.33‐14.4] mV, P = .01) as compared to patients who experienced VAs. Shock impedance was still different between groups confirming a lower value in patient who experienced this endpoint (56 [IQR: 46‐65] Ω vs 62 [IQR: 53‐70] Ω, P < .01). On the other hand, baseline LV pacing impedance was lower in subjects free from VA (606 [486‐741] Ω vs 633 [522‐750] Ω, P = .03). Table 2 depicts the entire analysis.

Baseline electrical parameters and atrial arrhythmias

When considering 24 h AHRE, few baseline electrical parameters showed differences between groups. Lower values of atrial signal amplitude (2.45 [IQR: 1.65‐3.85] mV vs 3.51 [IQR: 2.37‐4.67] mV, P < .01) and shock impedance (56 [IQR: 46‐66] Ω vs 62 [IQR: 52‐70] Ω, P < .01) were observed in patients who experienced 24 h AHRE. RV signal amplitude and LV pacing impedance values had minor differences (Table 2).

Markers of arrhythmia occurrence and prognosis

At the descriptive analysis, the baseline shock impedance showed consistent differences for all study endpoints with lower median values in patients who experienced death, VA, and AHRE. Table 3 reported the event rates in subgroups according to their baseline shock impedance value. Patients with shock impedance ≤40 Ω had a higher incidence of all study endpoints: 6.0 (95% CI: 4.5‐7.9)/100 patient‐years for all‐cause death, 22.8 (95% CI: 18.5‐27.2)/100 patient‐years for VA, and 10.8 (95% CI: 8.4‐13.8)/100 patient‐years for 24 h AHRE occurrence. Figure 1 shows Kaplan‐Meier curves: at 6 years, all‐cause mortality for this subgroup was 36.2% (95% CI: 27.0%‐47.4%) with 67.8% (95% CI: 59.1%‐76.2%) VA and 44.6% (95% CI: 35.6%‐54.8%) 24 h AHRE incidence. However, the adjusted HRs between patients with shock impedance >40 Ω vs those with values ≤40 Ω were not statistically significant (all‐cause mortality: 0.70 [0.45‐1.07], P = .102; VA: 0.78 [0.60‐1.01], P = .062; 24 h AHRE: 0.94 [0.61‐1.45], P = .797). Table 4 reports the association of the adjusting covariates of the multivariate models.
TABLE 3

Event rates of death, VA, and 24 h AHRE occurrence by baseline shock impedance subgroups

Shock impedance subgroupDeathsEvent rateVAEvent rate24 h AHREEvent rate
≤40 Ω506.0 (4.5‐7.9)11922.8 (18.5‐27.2)6710.8 (8.4‐13.8)
>40 and ≤50 Ω613.5 (2.7‐4.5)21318.3 (15.9‐20.9)1008.6 (6.9‐10.4)
>50 and ≤60 Ω532.8 (2.1‐3.6)25918.4 (16.2‐20.8)1379.9 (8.3‐11.7)
>60 and ≤70 Ω452.4 (1.8‐3.3)24817.8 (15.7‐20.2)1248.8 (7.3‐10.5)
>70 Ω181.9 (1.1‐2.9)11514.6 (12.0‐17.5)729.2 (7.2‐11.6)
Total2273.1 (2.7‐3.6)95418.1 (16.9‐19.3)5009.3 (8.5‐10.2)

Event rates are expressed as events/100 patient‐years (95% confidence interval).

Only devices with atrial sensing capability were included in the 24 h AHRE analysis.

Abbreviations as listed in Table 1.

FIGURE 1

Kaplan‐Meier curves of all‐cause mortality (A), VA (B), and 24 h AHRE (C) occurrence free rates by ≤40 and >40 Ω baseline shock impedance. AHRE, atrial high‐rate episode; VA, ventricular arrhythmia

TABLE 4

Association between baseline shock impedance >40 Ω and all‐cause mortality, VA and 24 h AHRE incidence

 All‐cause mortalityVA24 h AHRE
HR95% CI P HR95% CI P HR95% CI P
Unadjusted model
Shock impedance >40 Ω0.490.36‐0.68<.0010.700.58‐0.85<.0010.610.45‐0.83.001
Adjusted model
Shock impedance >40 Ω0.700.45‐1.07.1020.780.60‐1.01.0620.940.61‐1.45.797
Adjusting covariates
Age1.031.01‐1.05.0041.000.99‐1.01.3841.010.99‐1.03.184
Sex (female)1.160.76‐1.76.4890.770.62‐0.95.0160.560.37‐0.84.005
Hypertension0.760.54‐1.07.1130.960.81‐1.14.6300.750.56‐1.00.050
Diabetes1.581.10‐2.25.0120.840.69‐1.02.1290.960.70‐1.32.803
Ischemic cardiomyopathy1.390.93‐2.07.1110.910.76‐1.10.3470.970.71‐1.34.869
History of AF1.471.03‐2.12.0341.301.08‐1.57.0063.722.80‐4.94<.001
CHA2DS2‐VASC score1.150.98‐1.36.0931.050.97‐1.14.2551.151.00‐1.31.042

Univariate and multivariate Cox proportional hazard models.

Only devices with atrial sensing capability were included in the 24 h AHRE analysis.

Abbreviations: AF, atrial fibrillation; AHRE, atrial high‐rate episode; CI, confidence interval; HR, hazard ratio; VA, ventricular arrhythmias.

Event rates of death, VA, and 24 h AHRE occurrence by baseline shock impedance subgroups Event rates are expressed as events/100 patient‐years (95% confidence interval). Only devices with atrial sensing capability were included in the 24 h AHRE analysis. Abbreviations as listed in Table 1. Kaplan‐Meier curves of all‐cause mortality (A), VA (B), and 24 h AHRE (C) occurrence free rates by ≤40 and >40 Ω baseline shock impedance. AHRE, atrial high‐rate episode; VA, ventricular arrhythmia Association between baseline shock impedance >40 Ω and all‐cause mortality, VA and 24 h AHRE incidence Univariate and multivariate Cox proportional hazard models. Only devices with atrial sensing capability were included in the 24 h AHRE analysis. Abbreviations: AF, atrial fibrillation; AHRE, atrial high‐rate episode; CI, confidence interval; HR, hazard ratio; VA, ventricular arrhythmias. The incidence of 24 h AHRE was significantly different according to baseline atrial signal amplitude measured in sinus rhythm (Table 5). Subjects with atrial signal ≤1.5 mV showed an event rate of 24 h AHRE of 23.3 (95% CI: 18.4‐29.2)/100 patient‐years, with an incidence of 37.1% (30.5%‐44.6%) and 60.8% (40.7%‐81.3%) at 2 and 6 years, respectively (Figure 2A). The adjusted HR of events in patients with atrial signal >1.5 mV vs those with values ≤1.5 mV was 0.52 (95% CI: 0.33‐0.83), P = .006 (unadjusted HR 0.43 [95% CI: 0.31‐0.61], P < .001).
TABLE 5

Event rates of 24 h AHRE occurrence by baseline atrial sensing (all patients in sinus rhythm at implant)

Atrial sensing subgroup24 h AHREEvent rate
≤1.5 mV7623.3 (18.4‐29.2)
>1.5 and ≤2.5 mV1029.6 (7.8‐11.6)
>2.5 and ≤3.5 mV656.2 (4.8‐7.9)
>3.5 and ≤4.5 mV464.4 (3.3‐5.9)
>4.5 mV544.7 (3.5‐6.2)
Total3437.4 (6.7‐8.3)

Event rates are expressed as events/100 patient‐years (95% confidence interval).

Only devices with atrial sensing capability. Patients in atrial fibrillation at implant were excluded.

Abbreviations as listed in Table 1.

FIGURE 2

Kaplan‐Meier curves of AHRE lasting >24 h occurrence free rates by ≤1.5 and >1.5 mV baseline atrial sensing for all patients (A) and after excluding patients with history of atrial fibrillation (B). Patients in atrial fibrillation at implant were excluded. AHRE, atrial high‐rate episode

Event rates of 24 h AHRE occurrence by baseline atrial sensing (all patients in sinus rhythm at implant) Event rates are expressed as events/100 patient‐years (95% confidence interval). Only devices with atrial sensing capability. Patients in atrial fibrillation at implant were excluded. Abbreviations as listed in Table 1. Kaplan‐Meier curves of AHRE lasting >24 h occurrence free rates by ≤1.5 and >1.5 mV baseline atrial sensing for all patients (A) and after excluding patients with history of atrial fibrillation (B). Patients in atrial fibrillation at implant were excluded. AHRE, atrial high‐rate episode After excluding patients with a known previous history of AF, atrial signal >1.5 mV confirmed to be significantly associated with a lower risk of 24 h AHRE during follow‐up as shown in Figure 2B (adjusted HR: 0.51 [95% CI: 0.31‐0.85], P = .010).

DISCUSSION

In the present analysis on about 3000 ICD and CRT‐D patients, we found some associations between electrical parameters at implant and long‐term clinical outcomes. Baseline shock impedance values were lower in patients with atrial and VAs and in those who died during follow‐up. A cutoff of 40 Ω identified a subgroup with a particularly high incidence of events; however, the association was not significant if adjusted by other patients' characteristics. Conversely, the higher incidence of atrial arrhythmias in patients with baseline atrial sensing in sinus rhythm ≤1.5 mV compared to >1.5 mV was statistically significant even after adjustment with patient characteristics. This parameter could be used as a potential marker of underlying atrial tissue disease, potentially identifying patients who may benefit from an intensive monitoring approach which can be provided by daily RM.

Predictors of death and ventricular arrhythmias

Electrical data obtained during device implant can be influenced by several factors, as lead heart contact, lead position, and lead characteristics. However, bioelectrical properties can also be modified by other factors, including ischemia or the presence of fibrosis. In addition, shock impedance, which is calculated between the distal part of the lead (ventricular coil) and the ICD, depends on the conduction characteristics of the thorax including the whole heart and lungs. Therefore, low values can be observed when electrical conduction is favored, as in the case of fluid overload. As a result, deceased and VA/AHRE patients showed lower baseline shock impedance in our study, probably as a consequence of increased lung congestion and more severe heart failure symptoms even at implant. However, we were not able to detect a significant association in multivariable models between baseline values and events occurrence. The association between decreased shock impedance and heart failure or VA was shown in several studies, but the temporal relationship between the events is still unclear. We also found differences in the left ventricular pacing impedance, which is a near‐field measurement, as calculated mainly in bipolar configuration. Patients who died during follow‐up had lower baseline values than survivors. This result may reflect a higher percentage of patients with the LV lead located in an ischemic zone since infarct scar showed lower electrical impedance than the normal myocardium. , Finally, lower signal amplitudes for both right and LV leads were found in patients with death and VA events. This is not surprising as cytopenia and fibrosis are associated with lower signals and more advanced heart disease. , However, it should be noted that these parameters were statistically different only at the univariate analysis and the difference between groups was very small and with questionable clinical significance.

Predictors of AHRE

The difference of atrial signal amplitude at implant between patients with and without AHRE later detected during follow‐up was more striking. It is well known that in the atria low signal amplitude is associated with the presence of scar, AF recurrences, and heart failure. , However, most data were obtained during atrial mapping for AF ablation, while data are lacking on how intraoperative atrial signal amplitude during sinus rhythm can predict atrial tachyarrhythmias after device implant. In our study, patients with atrial signal amplitude lower than 1.5 mV at implant had a risk of 24 h atrial arrhythmias of 23.3/100 patient‐years during follow‐up, while for patients with higher atrial signal values the overall event rate was 6.2/100 patient‐years. This association was significant even if adjusted by other patients' characteristics and excluding patients with history of atrial arrhythmias before implant. AF is often asymptomatic and the identification of high‐risk patients is still an open issue. Recent data showed an incidence of 24 h AHRE in patients without AF history at high risk for thromboembolic events (CHA2DS2‐VASC score ≥ 5) of 7.7% and 40.4% at 2 and 6 years, respectively. In our analysis, at 2 and 6 years, 29.9% and 42.6% of patients with baseline low atrial signal developed this arrhythmia. In this scenario, atrial signal amplitude in sinus rhythm could be a useful marker to identify subjects more likely to develop atrial arrhythmias who may benefit from an intensive monitoring approach, which can be provided by daily RM. More intriguing is the relationship found between AHRE and shock impedance, confirming the association between atrial arrhythmias and severity of heart disease, despite the difference, although statistically significant, was clinically quite negligible.

Limitations

This study is an observational retrospective analysis suffering from all the known limitations of this design. Leads were placed according to clinical practice without specific recommendations and were not verified with fluoroscopy images, excluding the use of a variable lead location as adjusting covariate in our models. However, the large sample size of the database in terms of patients and sites is an important strength of this analysis tempering potential biases. As the HMEA database is based on the Home Monitoring system, all devices included in the present analysis were made by Biotronik and this could have an impact on the detection algorithms of AHRE, signals, and impedance measurements. Atrial high‐rate episodes were not adjudicated potentially including far‐field artifact and noise. However, the impact of the adjudication has been shown to be less relevant when using relatively long thresholds for diagnosis. The positive predictive value of AHRE increased to 98.2% when the threshold duration was prolonged to 24 h as in our analysis. Finally, device programing was not uniform reflecting ordinary medical practice. As a relevant proportion of the devices included in our cohort were implanted before 2014 when more aggressive antitachycardia settings and shorter detections were largely used, the rate of VA may be higher as compared to contemporary cohorts.

CONCLUSIONS

Shock impedance values were lower in patients who experienced death and both atrial and VAs during follow‐up. However, the association was not significant if adjusted by other patients' characteristics. Conversely, subjects with atrial signal amplitude below 1.5 mV showed a significant higher risk of atrial arrhythmias as compared to those with >1.5 mV, potentially revealing the presence of a more impaired atrial tissue.

CONFLICT OF INTERESTS

DG and AG are employees of BIOTRONIK Italia; the remaining authors have no conflict of interests for this article. Appendix Click here for additional data file.
  16 in total

1.  Positive predictive value of device-detected atrial high-rate episodes at different rates and durations: an analysis from ASSERT.

Authors:  Elizabeth S Kaufman; Carsten W Israel; Girish M Nair; Luciana Armaganijan; Syamkumar Divakaramenon; Georges H Mairesse; Axel Brandes; Eugene Crystal; Otto Costantini; Roopinder K Sandhu; Ratika Parkash; Stuart J Connolly; Stefan H Hohnloser; Jeff S Healey
Journal:  Heart Rhythm       Date:  2012-03-20       Impact factor: 6.343

2.  Coronary sinus signal amplitude predicts left atrial scarring.

Authors:  Philipp Attanasio; Daniel Qaiyumi; Robert Röhle; Alexander Wutzler; Erdal Safak; Bogdan Muntean; Leif-Hendrik Boldt; Burkert Pieske; Wilhelm Haverkamp; Martin Huemer
Journal:  Acta Cardiol       Date:  2017-12-22       Impact factor: 1.718

3.  Organizational model and reactions to alerts in remote monitoring of cardiac implantable electronic devices: A survey from the Home Monitoring Expert Alliance project.

Authors:  Gabriele Zanotto; Antonio D'Onofrio; Paolo Della Bella; Francesco Solimene; Ennio C Pisanò; Saverio Iacopino; Cristina Dondina; Daniele Giacopelli; Alessio Gargaro; Renato P Ricci
Journal:  Clin Cardiol       Date:  2018-12-15       Impact factor: 2.882

4.  Electrical impedance properties of normal and chronically infarcted left ventricular myocardium.

Authors:  D Schwartzman; I Chang; J J Michele; M S Mirotznik; K R Foster
Journal:  J Interv Card Electrophysiol       Date:  1999-10       Impact factor: 1.900

5.  Prognostic implications of atrial fibrillation in heart failure with reduced, mid-range, and preserved ejection fraction: a report from 14 964 patients in the European Society of Cardiology Heart Failure Long-Term Registry.

Authors:  Barak Zafrir; Lars H Lund; Cecile Laroche; Frank Ruschitzka; Maria G Crespo-Leiro; Andrew J S Coats; Stefan D Anker; Gerasimos Filippatos; Petar M Seferovic; Aldo P Maggioni; Manuel De Mora Martin; Lech Polonski; José Silva-Cardoso; Offer Amir
Journal:  Eur Heart J       Date:  2018-12-21       Impact factor: 29.983

6.  Clinical Implications of Brief Device-Detected Atrial Tachyarrhythmias in a Cardiac Rhythm Management Device Population: Results from the Registry of Atrial Tachycardia and Atrial Fibrillation Episodes.

Authors:  Steven Swiryn; Michael V Orlov; David G Benditt; John P DiMarco; Donald M Lloyd-Jones; Edward Karst; Fujian Qu; Mara T Slawsky; Melanie Turkel; Albert L Waldo
Journal:  Circulation       Date:  2016-10-18       Impact factor: 29.690

7.  Endocardial infarct scar recognition by myocardial electrical impedance is not influenced by changes in cardiac activation sequence.

Authors:  Gerard Amorós-Figueras; Esther Jorge; Concepción Alonso-Martin; Daniel Traver; Maria Ballesta; Ramon Bragós; Javier Rosell-Ferrer; Juan Cinca
Journal:  Heart Rhythm       Date:  2017-12-27       Impact factor: 6.343

8.  Electrical remodeling of the atria in congestive heart failure: electrophysiological and electroanatomic mapping in humans.

Authors:  Prashanthan Sanders; Joseph B Morton; Neil C Davidson; Steven J Spence; Jitendra K Vohra; Paul B Sparks; Jonathan M Kalman
Journal:  Circulation       Date:  2003-09-02       Impact factor: 29.690

9.  Does the CHA2DS2-VASc score reliably predict atrial arrhythmias? Analysis of a nationwide database of remote monitoring data transmitted daily from cardiac implantable electronic devices.

Authors:  Giovanni Rovaris; Francesco Solimene; Antonio D'Onofrio; Gabriele Zanotto; Renato P Ricci; Tiziana Mazzella; Saverio Iacopino; Paolo Della Bella; Giampiero Maglia; Gaetano Senatore; Fabio Quartieri; Mauro Biffi; Antonio Curnis; Valeria Calvi; Antonio Rapacciuolo; Matteo Santamaria; Alessandro Capucci; Massimo Giammaria; Andrea Campana; Fabrizio Caravati; Daniele Giacopelli; Alessio Gargaro; Ennio C Pisanò
Journal:  Heart Rhythm       Date:  2018-03-01       Impact factor: 6.343

10.  Duration of device-detected subclinical atrial fibrillation and occurrence of stroke in ASSERT.

Authors:  Isabelle C Van Gelder; Jeff S Healey; Harry J G M Crijns; Jia Wang; Stefan H Hohnloser; Michael R Gold; Alessandro Capucci; Chu-Pak Lau; Carlos A Morillo; Anne H Hobbelt; Michiel Rienstra; Stuart J Connolly
Journal:  Eur Heart J       Date:  2017-05-01       Impact factor: 29.983

View more
  1 in total

1.  How to get the optimal defibrillation lead parameters using myocardial perfusion scintigraphy in patients with coronary artery disease.

Authors:  Tariel A Atabekov; Roman E Batalov; Svetlana I Sazonova; Sergey N Krivolapov; Mikhail S Khlynin; Anna I Mishkina; Konstantin V Zavadovsky; Antonio Curnis; Sergey V Popov
Journal:  Int J Cardiovasc Imaging       Date:  2021-06-07       Impact factor: 2.357

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.