A new statistical methodology overcame the defects of the Bland-Altman method.

Patrick Taffé1, Patricia Halfon2, Matthieu Halfon3.   

Abstract

BACKGROUND AND OBJECTIVES: The Bland and Altman's limits of agreement (LoA) method is the most commonly used statistical method to assess the bias and precision of a new measuring device (it has been cited over 40,000 times as of March 2019). What is less known is that the LoA method can be dramatically misleading.
METHODS: A new statistical methodology, which circumvents these deficiencies, has recently been published and made available in the R and Stata statistical packages. We aimed at introducing and illustrating with a small data set on blood pressure (BP) measurements, taken by two different oscillometric devices, the use of this new methodology to a clinical audience.
RESULTS: For diastolic BP, the LoA method was particularly misleading, as it identified differential and proportional biases of opposite signs compared with the new methodology. Regarding systolic BP, the LoA method strongly overestimated both the differential and proportional biases, for both devices.
CONCLUSION: The LoA method may be dramatically misleading and does not allow one to estimate the precision of each measurement method. We recommend the use of the newly developed statistical methodology instead.
Copyright © 2020 Elsevier Inc. All rights reserved.

Entities:  

Keywords:  Bias; Blood pressure; Limits of Agreement; Oscillometric device; Precision

Mesh:

Year:  2020        PMID: 32247027     DOI: 10.1016/j.jclinepi.2020.03.018

Source DB:  PubMed          Journal:  J Clin Epidemiol        ISSN: 0895-4356            Impact factor:   6.437


  3 in total

1.  Agreements between mean arterial pressure from radial and femoral artery measurements in refractory shock patients.

Authors:  Hemmawan Wisanusattra; Bodin Khwannimit
Journal:  Sci Rep       Date:  2022-05-25       Impact factor: 4.996

2.  An Extension of the Bland-Altman Plot for Analyzing the Agreement of More than Two Raters.

Authors:  Sören Möller; Birgit Debrabant; Ulrich Halekoh; Andreas Kristian Petersen; Oke Gerke
Journal:  Diagnostics (Basel)       Date:  2021-01-01

3.  Smartwatch-Based Blood Pressure Measurement Demonstrates Insufficient Accuracy.

Authors:  Maarten Falter; Martijn Scherrenberg; Karen Driesen; Zoë Pieters; Toshiki Kaihara; Linqi Xu; Enrico Gianluca Caiani; Paolo Castiglioni; Andrea Faini; Gianfranco Parati; Paul Dendale
Journal:  Front Cardiovasc Med       Date:  2022-07-11
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.