Yoichi Takaya1, Teiji Akagi1, Koji Nakagawa1, Rie Nakayama1, Takashi Miki1, Nobuhisa Watanabe2, Norihisa Toh1, Hiroshi Ito1.
Abstract
BACKGROUND: Septal malalignment is related to erosion and device embolization in transcatheter closure of atrial septal defect (ASD), but limited information is available.Entities:
Mesh:
Year: 2020 PMID: 32206044 PMCID: PMC7072101 DOI: 10.1155/2020/6090612
Source DB: PubMed Journal: J Interv Cardiol ISSN: 0896-4327 Impact factor: 2.279
Figure 1Transesophageal echocardiography showing septal malalignment. The septum primum attached to the aorta is malaligned toward the left atrial side (solid arrow). The septum primum is separated from the septum secundum (dotted arrow). The defect surface of the septum primum (solid line) is different from that of the septum secundum (dotted line). The distance of separation between the septum primum and the septum secundum is measured (double arrow).
Clinical characteristics.
| Malalignment (+) ( | Malalignment (−) ( |
| |
|---|---|---|---|
| Age (years) | 52 ± 19 | 46 ± 24 | 0.02 |
| Female | 104 (57%) | 144 (62%) | 0.28 |
| Aortic rim deficiency | 173 (94%) | 135 (58%) | <0.001 |
| Pulmonary-to-systemic blood flow ratio | 2.4 ± 0.9 | 2.1 ± 0.7 | <0.001 |
| Pulmonary arterial pressure (mmHg) | 18 ± 7 | 18 ± 7 | 0.94 |
Data are presented as mean ± standard deviation.
Figure 2Frequency of septal malalignment. The frequency of septal malalignment according to age is shown.
Figure 3Severity of septal malalignment. The distance of separation between the septum primum and the septum secundum in patients with septal malalignment is shown.
ASD diameter and device size in 166 patients with septal malalignment.
| ASD diameter at the septum primum (mm) | 19 ± 6 |
| ASD diameter at the septum secundum (mm) | 16 ± 6 |
| Amplatzer Septal Occluder device | 103 (62%) |
| Device size (mm) | 22 ± 7 |
| Occlutech Figulla Flex II device | 63 (38%) |
| Device size (mm) | 24 ± 5 |
Data are presented as mean ± standard deviation. ASD, atrial septal defect.
Figure 4Difference between the ASD diameter and the device size in patients with septal malalignment. The difference between the ASD diameter measured at the septum primum and Amplatzer Septal Occluder device size (a) and the Occlutech Figulla Flex II device size (b). ASD, atrial septal defect.