Literature DB >> 24038861

Echocardiographic predictors of cardiac erosion after Amplatzer septal occluder placement.

Zahid Amin1.   

Abstract

UNLABELLED: The risk of erosion after Amplatzer septal occluder (ASO) device placement in atrial septal defects is well described. Aortic rim deficiency and use of over-sized device increase the risk of erosion. This study attempts to describe device characteristics, anatomical features and echocardiographic predictors that increase the risk of erosion.
METHODS: From 2005 through 2012, 12 new cases, with nine confirmed and three suspected device erosions where pre-procedural, intra-procedural, and/or post-procedural echocardiograms were available and, were reviewed. Following parameters were evaluated: ASD location (high or low), rims deficiency and consistency, septal mal-alignment, dynamic nature of the defect; device edge relationship toward the transverse sinus (TS), atrial free wall tenting and the size of the defect compared with the size of the device used for closure.
RESULTS: We found poor posterior rim consistency, aortic rim absence (in multiple views) and absent aortic rim at O degree in 100% of the patients. Septal mal-alignment and dynamic ASD was present in nearly 50% of the cases. The device was over-sized in three patients only. A 26-mm device was the most common device that resulted in erosion. In cases, where patient had experienced bloody pericardial effusion and the device was in place, device tenting in the TS was observed. Surgical explantation of the device confirmed presence of erosion in all cases.
CONCLUSION: Aortic rim absence in multiple views, poor posterior rim consistency, septal mal-alignment, and dynamic ASD appear to be factors where erosion risk increases significantly. A thorough assessment of the device edge by echocardiography in short-axis may show device tenting of the atrial free wall into the TS. This finding should be a strong indictor to recommend surgical removal of device after occurrence of pericardial effusion.
Copyright © 2013 Wiley Periodicals, Inc.

Entities:  

Keywords:  complications pediatric cath/intervention; patent foramen ovale/atrial septal defect; pediatric interventions

Mesh:

Year:  2013        PMID: 24038861     DOI: 10.1002/ccd.25175

Source DB:  PubMed          Journal:  Catheter Cardiovasc Interv        ISSN: 1522-1946            Impact factor:   2.692


  20 in total

1.  Initial Experience with the Nit-Occlud ASD-R: Short-Term Results.

Authors:  Mustafa Orhan Bulut; Ilker Kemal Yucel; Mehmet Kucuk; Sevket Balli; Evic Zeynep Basar; Ahmet Celebi
Journal:  Pediatr Cardiol       Date:  2016-06-09       Impact factor: 1.655

2.  Very late erosion of Amplatzer occluder device resulting in Cardiac tamponade after 15 years.

Authors:  Jan-Erik Guelker; Ruben Jansen; Kolja Sievert; Horst Sievert; Stefan Bertog
Journal:  Clin Res Cardiol       Date:  2018-02-26       Impact factor: 5.460

Review 3.  State-of-the-Art Atrial Septal Defect Closure Devices for Congenital Heart.

Authors:  Michael L O'Byrne; Daniel S Levi
Journal:  Interv Cardiol Clin       Date:  2019-01

4.  Transcatheter Closure of Atrial Septal Defects using the GORE(®) Septal Occluder in Children Less Than 10 kg of Body Weight.

Authors:  Tariq Abu-Tair; Christiane M Wiethoff; Jascha Kehr; Wlodzimierz Kuroczynski; Christoph Kampmann
Journal:  Pediatr Cardiol       Date:  2016-02-19       Impact factor: 1.655

Review 5.  ASD Closure in Structural Heart Disease.

Authors:  Dominik M Wiktor; John D Carroll
Journal:  Curr Cardiol Rep       Date:  2018-04-17       Impact factor: 2.931

6.  Increasing propensity to pursue operative closure of atrial septal defects following changes in the instructions for use of the Amplatzer Septal Occluder device: An observational study using data from the Pediatric Health Information Systems database.

Authors:  Michael L O'Byrne; Russell T Shinohara; Elena K Grant; Joshua P Kanter; Matthew J Gillespie; Yoav Dori; Jonathan J Rome; Andrew C Glatz
Journal:  Am Heart J       Date:  2017-07-19       Impact factor: 4.749

7.  The influence of deficient retro-aortic rim on technical success and early adverse events following device closure of secundum atrial septal defects: An Analysis of the IMPACT Registry®.

Authors:  Michael L O'Byrne; Matthew J Gillespie; Kevin F Kennedy; Yoav Dori; Jonathan J Rome; Andrew C Glatz
Journal:  Catheter Cardiovasc Interv       Date:  2016-05-18       Impact factor: 2.692

Review 8.  Transcatheter device closure of atrial septal defects: more to think about than just closing the hole.

Authors:  Michael L O'Byrne; Andrew C Glatz; Matthew J Gillespie
Journal:  Curr Opin Cardiol       Date:  2018-01       Impact factor: 2.161

9.  Prevalence of deficient retro-aortic rim and its effects on outcomes in device closure of atrial septal defects.

Authors:  Michael L O'Byrne; Andrew C Glatz; Sherzana Sunderji; Aswathi E Mathew; David J Goldberg; Yoav Dori; Jonathan J Rome; Matthew J Gillespie
Journal:  Pediatr Cardiol       Date:  2014-05-14       Impact factor: 1.655

Review 10.  Transcatheter closure of atrial septal defect: principles and available devices.

Authors:  Se Yong Jung; Jae Young Choi
Journal:  J Thorac Dis       Date:  2018-09       Impact factor: 2.895

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.