Sebastian Primpke1, Richard K Cross2, Svenja M Mintenig3, Marta Simon4, Alvise Vianello4, Gunnar Gerdts1, Jes Vollertsen4. 1. Alfred Wegener Institute, Helmholtz Centre for Polar and Marine Research, Biologische Anstalt Helgoland, Kurpromenade, Helgoland. 2. Pollution Science Area, UK Centre for Ecology and Hydrology, Oxfordshire, UK. 3. Copernicus Institute of Sustainable Development, Utrecht University, The Netherlands. 4. Department of the Built Environment, Aalborg University, Aalborg, Denmark.
Abstract
Microplastics (MP) are ubiquitous within the environment, but the approaches to analysis of this contaminant are currently quite diverse, with a number of analytical methods available. The comparability of results is hindered as even for a single analytical method such as Fourier transform infrared spectroscopy (FT-IR) the different instruments currently available do not allow a harmonized analysis. To overcome this limitation, a new free of charge software tool, allowing the systematic identification of MP in the environment (siMPle) was developed. This software tool allows a rapid and harmonized analysis of MP across FT-IR systems from different manufacturers (Bruker Hyperion 3000, Agilent Cary 620/670, PerkinElmer Spotlight 400, and Thermo Fischer Scientific Nicolet iN10). Using the same database and the automated analysis pipeline in siMPle, MP were identified in samples that were analyzed with instruments with different detector systems as well as optical resolutions and the results discussed.
Microplastics (MP) are ubiquitous within the environment, but the approaches to analysis of this contaminant are currently quite diverse, with a number of analytical methods available. The comparability of results is hindered as even for a single analytical method such as Fourier transform infrared spectroscopy (FT-IR) the different instruments currently available do not allow a harmonized analysis. To overcome this limitation, a new free of charge software tool, allowing the systematic identification of MP in the environment (siMPle) was developed. This software tool allows a rapid and harmonized analysis of MP across FT-IR systems from different manufacturers (Bruker Hyperion 3000, Agilent Cary 620/670, PerkinElmer Spotlight 400, and Thermo Fischer Scientific Nicolet iN10). Using the same database and the automated analysis pipeline in siMPle, MP were identified in samples that were analyzed with instruments with different detector systems as well as optical resolutions and the results discussed.
Small plastic particles, called microplastics (MP) are currently recognized as a
potential risk for environmental and human health.[1,2] The near-ubiquitous
contamination of the environment, from terrestrial soils and air to the freshwater
and marine environments has raised the profile of this topic in recent years,
leading to a wealth of methods and approaches for sampling and analyzing MP in
environmental matrices.[3-11] In general, these particles
are defined as <5 mm in size while a lower size limit and a standard definition
of MP has yet to be agreed on.[7,12] Subcategories distinguishing
between large MP (5 mm–500 µm) and small MP (500–1 µm)[13] are often used, reflecting practical considerations during the full
analytical procedure.The analytical procedure to identify particles can be divided into three steps,
starting with sampling for MP, followed by sample extraction and finally,
identification and quantification. Each of these steps has its challenges (cf.
Lusher et al.,[3] Brander et al.,[4] Primpke et al.,[6] and Cowger et al.[14]). Additionally, there is an increasing awareness for quality assurance and
quality control (QA/QC) to successfully and reliably identify MP in environmental
samples.[15,16] Individual steps for QA/QC are currently discussed within this
overarching special issue.[4] While QA/QC is important for the quality of the results, the intercomparison
of studies is further hampered by a missing harmonization of the three steps:
sampling, sample extraction, and identification. Especially for the chemical
analysis, a plethora of different methods and software are used to interpret
generated data.[5,6,14] Among the
spectroscopy-based techniques, Fourier transform infrared spectroscopy (FT-IR) is
considered most suitable to analyze MP from 5 mm to ∼10 µm, even if it requires
different acquisition modes according to the particle size.[6] Micro-FT-IR (µFT-IR) imaging[17-20] permits scanning of the whole
sample, avoiding the manual sorting/point-and-shoot steps that are otherwise
necessary, which are a source of bias as the analysis becomes operator
dependent.[21,22] On the other hand, FT-IR imaging for MP analysis produces a
huge amount of data (large areas are scanned at a very high spatial resolution)
which are difficult to manage using the commercial software provided by the
instrument manufacturers.[4] Challenges handling this data, together with the lack of a suitable software
tool, not only impedes a reliable workflow but also makes it difficult, if not
impossible, to compare between studies performed using FT-IR Imaging.[14] While initial studies comparing various analytical methods have been conducted,[23] a comparison between studies using the same analytical technique but on
different instruments is missing for MP research. One reason is the lack of a
suitable software tool to perform such a comparison. Such a study, however, is
essential to reproduce and compare results within meta-studies. Moreover, it allows
the harmonized analysis of MP in the future, as the advantages and disadvantages of
the currently available high-throughput analysis pipelines for individual systems
can be determined.To achieve this goal, a new software tool is presented and intercalibrated with
existing studies.[24] Its performance is evaluated via the analysis against existing reference data
sets and comparison of the achieved results. Furthermore, the tool allows the
analysis of data generated by different FT-IR imaging techniques from various
manufacturers. This is demonstrated through the application of the software to data
sets from state of the art instruments from four major manufacturers, namely
Agilent, Bruker, PerkinElmer, and ThermoFisher Scientific, with different types of
detection modes ranging from focal plane array (FPA), linear arrays to
single-element detectors. This analysis is followed by a short performance
evaluation of the assigned spectra. The software and corresponding reference spectra databases[24] are available free of charge via the internet and can be used by everyone for
the analysis of FT-IR data.
Materials and Methods
Sample Extraction and Analysis Using the Agilent System
Sample preparation started with an effluent sample (Ryaverket wastewater
treatment plant [WWTP], Götheborg, Sweden) as collected using filtration with a
custom filtration device containing a ø100 mm stainless steel filter (10 µm mesh
size).The material collected on the steel filter mesh was treated to extract MP using a
method derived by Löder et al.[13] and modified by Liu et al.[22] The filter was sonicated into filtered Milli-Q water containing 5% (w/v)
sodium dodecyl sulfate (SDS) to detach the solids and left stirring (100 rpm) at
50 ℃ for 48 h. The resulting suspension was then filtered onto a 47 mm steel
filter (10 µm mesh). The particles trapped were re-suspended and first incubated
with proteolytic enzymes (Alcalase, Novozymes, Denmark) in a Tris buffer
(pH 8.2, stirring at 100 rpm, and 50 ℃ for 48 h), filtered onto a 47 mm steel
filter, and the solids were then removed from the filter surface. A second
enzymatic treatment was performed using cellulolytic enzymes (Cellulase blend
and Viscozyme, Sigma-Aldrich) in acetate buffer (pH 4.8, stirring at 100 rpm and
50 ℃ for 48 h) to eliminate the majority of the organic fraction of the sample
matrix. The remaining undigested matter was filtered onto a 47 mm steel filter,
and the solids were again removed from the filter surface. The gathered solids
were oxidized using Fenton reaction (hydrogen peroxide catalyzed by Fe(II) at
∼20 ℃ for 24 h). After a further filtration on a 47 mm steel filter, the solids
were removed from the filter surface and recovered. MP were further separated
from the inorganic particles in a zinc chloride solution (1.7 g cm−3)
using a glass separatory funnel. After discharging the settled material, the
supernatant was filtered (47 mm steel filter) and the material was recovered
following the same procedure described for the previous steps using 50% (v/v)
ethanol. The extracted MP were transferred in 10 mL glass headspace vial, the
liquid was evaporated at 55˚C, and finally, 5 mL 50% (v/v) ethanol solution were
added to obtain a known volume in the vial.In order to minimize MP contamination deriving from the equipment used for
sampling and sample preparation, all lab tools were flushed with filtered
(1.2 µm) Milli-Q water three times before use. Tools made of glass or metal or
which were coated with PTFE were used instead of plastic whenever possible.
Sample containers were covered with aluminum foil to reduce airborne
contamination, and steel filters were muffled at 500 ℃ before usage.The µFT-IR analysis was performed using a FPA-based µFT-IR imaging technique
provided by a Cary 620 µFT-IR microscope from Agilent Technologies (USA) coupled
with a Cary 670 FT-IR spectroscope. The instrument was equipped with a 128 × 128
FPA/mercury–cadmium–telluride (MCT) imaging detector (FPA-MCT-imaging detector).
The analysis was carried out in transmission mode, using a 15 × Cassegrain
(visible IR) objective-condenser system which produces 5.5 µm pixel resolution
on the FPA detector. An aliquot of the sample (600 µL) was deposited onto a
ø13 mm × 2-mm-thick zinc selenide (ZnSe) transmission window. A background scan
was collected before each sample scan on a clean window at 8 cm−1
spectral resolution, using 120 co-added scans in the spectral range of
3750–850 cm−1. Subsequently, an area of 14 × 14 tiles was scanned
on the samples window, using 30 co-added scans, and the same settings as for the
background scan. The analyzed area covered the entire active surface of the
windows (diameter of 10 mm, area 78.5 mm2), recording the spectra of
all deposited particles.
Sample Extraction and Analysis Using the PerkinElmer System
The sample represents a composite sample taken at 30-min intervals across a 24-h
period, directly sampling from the effluent of a WWTP. The auto-sampler filtered
this water through a woven stainless steel cylindrical filter cartridge (27.8 cm
long, nominal pore size 10 µm, ∼500 cm2 filter area; Wolftechnik,
Germany). The concentrated sample was transferred from the filter into
dispersion for processing in the lab. Processing in order to “clean” the sample
in preparation for µFT-IR analysis consisted of two steps: a Fenton's reaction
to chemically degrade organic material and enzyme digestion to remove cellulosic
and proteinaceous material.The Fenton's reaction used a Fe(II) solution (0.05 M
FeSO4°7 H2O, Fischer Scientific, USA, >98% purity)
acidified with 0.2% sulfuric acid (H2SO4, AnaTaR, 98.07%
purity) and 30% hydrogen peroxide (H2O2, Fisher
Scientific). The reaction was allowed to exhaust itself overnight before the
sample was filtered and re-dispersed for enzymatic digestion. The enzyme
digestion steps utilized cellulase in a pH 5, phosphate-buffered saline solution
(MP Biomedicals, USA) incubated at 50 ℃ for 48 h and trypsin at 37 ℃ for 30 min
(Sigma-Aldrich, Germany). The final concentrated sample was dispersed and stored
in 50% ethanol (Sigma-Aldrich, Germany) before depositing onto 25 mm diameter
5.0 µm pore size silver membrane filters (Sterlitech, USA) for µFT-IR analysis.
All reagents were filtered through a 1.2 µm glass-fiber filter before use, and
all processing took place in the Microflow Biological Safety Cabinet, fitted
with HEPA filters to control for airborne microplastic contamination during the
processing of samples.For the µFT-IR analysis, a PerkinElmer Spotlight 400 FT-IR microspectrometer with
MCT detector was used for the analysis of the sample. The silver filter
containing the processed sample was mounted on a glass slide and held in place
with a clamped stainless steel O-ring. The spectrometer collected spectra in the
range between 4000 and 700 cm−1 in reflectance mode. A background
spectrum was collected for each sample from a blank area of the silver filter at
a spectral resolution of 8 cm−1 and pixel size of 25 µm. A total of
90 scans were taken per pixel with an interferometer speed of
2.2 cm s−1. An optical image was first collected by tiling single
field of view images together to cover an area of approximately 13 mm × 13 mm. A
smaller mapping area for the FT-IR spectrum of 11.6 mm × 11.6 mm was selected
(92 % of the filter area), due to constraints on the file size that could be
generated by the PerkinElmer SpectrumIMAGE software. The FT-IR mapping was
performed with the same parameters as that of the background scan, but at four
scans per pixel. Atmospheric correction was performed on the resulting .fsm
file.
Sample Extraction and Measurement Using the ThermoFisher Scientific
System
The sample was taken by filtering water at the effluent of a WWTP over a
stainless sieve (ø20 cm, ThermoFisher Scientific) having a mesh size of 20 µm.
The concentrated sample was exposed successively to SDS (one day, 5%, Serva
Electrophoresis GmbH, Germany), potassium hydroxide (five days, 10%, Carl Roth
GmbH, Germany), and hydrogen peroxide (two days, 32%, Carl Roth GmbH, Germany).
During all steps, the sample was incubated in an oven with a temperature set at
35 ℃. In between steps, the sample was filtered over a ø47 mm stainless steel
filter with a mesh size of 20 µm. Inorganic particles were removed by performing
a density separation using a zinc chloride (ZnCl2, Carl Roth GmbH,
Germany) solution with a density of 1.6 g cm−3. Subsequently, sample
residues were filtered on an aluminum oxide filter (Anodisc 25 mm, Whatman, UK)
which was then dried at 35 ℃ for several days.In order to minimize MP contamination, all chemicals were filtered through
stainless steel filters with a mesh size of 10 µm. Additionally, all lab
equipment was thoroughly rinsed before usage, and the lab surfaces cleaned with
ethanol (30%, Carl Roth GmbH and Co. KG, Germany) and Milli-Q water. Whenever
possible, plastic equipment was reduced by tools made of glass or metal, and
when finishing sample handling these were immediately covered with aluminum foil
to reduce airborne contamination.For the µFT-IR analysis, an FT-IR microscope equipped with a single MCT detector
(Nicolet iN10, ThermoFisher Scientific, USA) was used. For the measurements, an
Anodisc filter was placed on a calcium fluoride crystal (EdmundOptics, Germany).
About one third of each filter was mapped in transmission mode, with one scan
recorded per pixel, the aperture size set at 50 × 50 µm, the spectral resolution
as 16 cm−1, and the spatial resolution at 20 µm. A background scan
using the same settings was conducted on a blank area of the same filter.
Sample Extraction and Measurement Using the Bruker System
For this comparison study, a data set of a sample investigated in previous
studies[25,26] was chosen. The sample was from the effluent of the WWTP
Holdorf. Sample location and further sampling information are available within
the previous studies.[25,26] Briefly summarized, the sample was directly taken from the
effluent of the WWTP. The sample was processed by the enzymatic digestion
protocol of Löder et al.[13] and subsequently concentrated onto an Anodisc filter (25 mm diameter,
0.2 µm pore size, GE Whatman). The sample was placed and covered by a
BaF2 window prior to measurement on a custom-made sample holder
as described in detail in Primpke et al.[25]The µFT-IR measurement was performed using a Bruker TENSOR II spectrometer, which
is connected to a Hyperion 3000 µFT-IR-microscope (Bruker Optics GmbH,
Ettlingen, Germany). The spectra were collected using a 64 × 64 FPAMCT detector
as described in literature.[19] Prior to measurement, a visual image of the sample surface was recorded.
The FT-IR measurements were performed using 15 × Cassegrain objective, in the
spectral range of 3600–1250 cm−1 with 4 × 4 binning at a spectral
resolution of 8 cm−1 and six coadded scans. With this setup, a pixel
size of 11.1 × 11.1 µm per spectra was achieved. All data were collected using
Bruker software OPUS 7.5 (Bruker Optics GmbH, Germany).
Sample Preparation of Algae Samples
Ecotoxicity tests were carried out with the microalga Raphidocelis
subcapitata with modifications according to “Water
Quality–Freshwater Algal Growth Inhibition Test with Unicellular Green Algae,
ISO Standard 8692, 2004”. The test included five replicates of the control
sample, in which algae were not exposed to the toxicant, and triplicates of
algae exposed to five different concentrations of the tested toxicant. The
replicates were combined after the toxicity test, and algae from all treatments
were preserved with Lugol's iodine solution for their infrared analysis.For the infrared analysis described in Kansiz et al.,[27] 2 mL of the preserved control sample was prepared. The cells were
centrifuged, and the residues of the preservative and growth media were washed
off to prevent interference with algae spectra. The purified cells were
resuspended in 200 µL deionized water, and the entire volume was deposited on a
13 mm diameter, 1-mm-thick CaF2 transmission window, and dried in a
vacuum desiccator.A Cary 620 µFT-IR microscope from Agilent Technologies (USA) coupled with a Cary
670 FT-IR spectroscope was used for the FPA-based µFT-IR imaging analysis of the
algae cells. The analysis was carried out in transmission mode using the
15 × Cassegrain objective in high magnification mode to create a mosaic with
1.1 μm pixel resolution on the FPA. A background scan was collected on a clean
window in the range of 3750–850 cm−1 with an 8 cm−1
spectral resolution applying 256 co-added scans. An area of 6 × 6 tiles was
scanned on the sample window following the background scan with the same
parameters applying 240 co-added scans per pixel.
Software and Database
The software siMPle is the combination of the software MPhunter presented in Liu et al.[22] and the automated analysis of Primpke et al.[21] It is written in Delphi using Windows 10 as the operating system, which is
available at www.simple-plastics.eu, where reference databases for FT-IR and
Raman spectroscopy are also provided (Fig. 1).
Figure 1.
The software graphical user interface with a loaded reference (blue) and
a sample spectrum (orange) using the Match single spectrum function of
siMPle on a spectrum assigned as polyvinylchloride using ESM1.csv.
The software graphical user interface with a loaded reference (blue) and
a sample spectrum (orange) using the Match single spectrum function of
siMPle on a spectrum assigned as polyvinylchloride using ESM1.csv.In this work, the database for automated analysis[24] and the release version (1.0.0) of siMPle was used to analyze all data sets.
Prior to analysis, all spectra were converted from the original file format
(Agilent:.dmd and PerkinElmer:.fsm) or JCAMP-DX files (Bruker and Thermo Fisher
Scientific) into two siMPle file formats, namely .spe and .wno files, allowing fast
data access (see How to Use siMPle.pdf Supplemental Material file for further
instructions). These file formats are accompanied by an extra file, the
MaschineData.ini, which contains all necessary information for size calculations and
data handling. In all cases, the mosaic structure is either kept or newly generated,
allowing faster data handling. After loading the data and reference spectra, the
spectral fit between the two was calculated by Pearson correlation for the untreated
data, the first derivative and the second derivative, resulting in their correlation
factors r0, r1,
r2, respectively. If not further specified, the
following settings were used: omit CO2 peak (upper wavenumber:
2420 cm−1, lower wavenumber: 2200 cm−1), suppress negative
correlations, and include second-order derivatives.To investigate the performance between Bruker OPUS and siMPle, the calculation times
were measured using a HP KP719AV computer (Intel Core 2 Duo Processor, 8 GB RAM, AMD
Radeon HD 5450 graphic card, extra USB 3.0 Controller card and SANDISK Extreme 64 GB
USB-Stick) which is the same as used in previous studies.[21,24-26,28-34] Further calculations were
performed on a HP Z440 workstation (Intel Xeon E5-2630 v.3 CPU, 64 GB RAM, NVidia
Quadro M2000 graphics card) for all other purposes.The automated analysis pipeline (AAP)[21] identifies the recorded spectra based on the results of the Pearson
correlation factors (r) calculated for the respective untreated
spectra r0 and the first derivatives
r1. Only if maximum values of
r0 and r1 are assigned
to the same polymer entry, then the spectra are counted as identified and the
polymer type added to the list of analyzed pixels together with the
x,y coordinates and the summarized hit quality index (HQI, Eq.
1):This type of data represents a false color image which was then in silico treated by
Image Analysis as described in Primpke et al.[21] by a pixel hole closing mechanism prior to the size determination and
particle quantification. For further calculations, the data thresholds described in
Lorenz et al.[32] were used. To avoid confusion for the reader, we did not apply the second
analysis pipeline[1,22,35] available in siMPle for the scope of this study.The siMPle software allows rapid QA/QC for the assigned polymer hits. During image
analysis, a designated file is generated named “_forqc.csv,” which contains
x,y coordinates of the measured spectra together with the hit
quality, the assigned polymer type, and a reference spectrum identifier. Via the
button “Load Pipeline Results,” the QA/QC process will be started (Fig. 2).
Figure 2.
Quality assurance/quality control for siMPle for results of the image
analysis of sample RefEnv124 via the AWI pipeline allowing a
direct comparison (a) between sample spectrum (orange) and database
spectrum (blue) by clicking on the determined pixel (c). The process
allows checking other spectra by clicking on another database entry as
well as the individual Pearson correlation factors. The heatmap (b)
allows the user to locate the assigned spectra on the map.
Quality assurance/quality control for siMPle for results of the image
analysis of sample RefEnv124 via the AWI pipeline allowing a
direct comparison (a) between sample spectrum (orange) and database
spectrum (blue) by clicking on the determined pixel (c). The process
allows checking other spectra by clicking on another database entry as
well as the individual Pearson correlation factors. The heatmap (b)
allows the user to locate the assigned spectra on the map.Using a designated window, each spectrum can be individually assessed and rated from
a perfect assignment down to a full misassignment with values ranging from 1 to
0.01. Following the approach of Primpke et al.,[21] the spectra were rated with five different values (1.00, 0.75, 0.5, 0.25, and
0.01) ranging from best match to a full misassignment, respectively. To screen the
data of the different instruments and evaluate the successful assignment of polymer
identify to particles, a randomly selected number (n = 100) of
spectra were manually reanalyzed for each instrument data set.For matching single spectra, the score used within siMPle was first described in Liu et al.[22] In short, prior to the quantification of particles, the score
(S) for the identification of the polymer type was calculated using Eq. 2:[22]Each result
(r0,r1,r2)
was squared and multiplied by a weight
(k0,k1,k2,
respectively) for the respective correlation factor, which can be assigned by the
user.
Results and Discussion
The software allows two types of data analyses: first, the matching of single spectra
and second, the analysis of large filter areas. To interpret a single spectrum, the
reference database must first be loaded and then a single spectrum in a defined file
format (Paragraph S1; ESM1.csv, Supplemental Material) must be loaded (Fig. 1). In this example, the
spectrum was assigned to polyvinylchloride (PVC) with a S of 0.7246 using the default options of siMPle
(k0, k1,
k1). By assigning weights, the user can
decide which correlation method should be represented by S, e.g., for comparison with studies using Bruker OPUS[28,30-33] where only the first
derivative was used for single spectrum analysis. For the chosen example, values of
k0 = k2 = 0,
k1 = 1 (first derivative only) were used. In this
case, the S decreased to 0.6689. If the user decides to include all correlation results
(k0 = k1 = k2 = 1),
the S was further decreased to 0.6435. Therefore, it is mandatory to state the
weights k0, k1,
k2 within the material and methods section for
comparison of studies if siMPle is used. In general, this analysis is independent
from the instrumental source of the data. All types of data in the described file
format (see Paragraph S1) can be processed independently from manufacturer and
measurement method. An example of this using preprocessed Raman spectra is shown in
Fig. S1. Future data processing steps for single spectra or lists of spectra will be
introduced in future releases. All releases will be accompanied via a change log and
a living manual on the website.During this process, QA/QC is easily available, because the spectrum with the highest
hit is indicated at the end of the analysis together with a hit list for all other
database entries (Fig. 1).
Together with this, the hit result can be exported for further use.
Analysis of Chemical Imaging Files
For larger data sets, the siMPle software allows a time-efficient loading of data
using a harmonized file system for data storage which is independent from the
original file format. The file formats introduced were optimized for the fast
use within the software. The commonly shared file format by the International
Union of Pure and Applied Chemistry (IUPAC) J-CAMPdx file format was limited to
the import of data, due to long loading times of such text type based files.
Currently, siMPle is able to convert native data from Agilent and PerkinElmer
systems using the file import function while for Bruker and ThermoFisher
Scientific systems extra steps are necessary (see How to Use siMPle.pdf,
Supplemental Material).To validate the performance of siMPle, it was tested against existing reference
data sets from literature. These reference sets consist of materials of known
origin (Ref7P) or from environmental samples (RefEnv1 and RefEnv2). These three
samples were analyzed using siMPle and the results were compared to the
automated analysis via Bruker OPUS. Starting with Ref7P, we started to analyze
the performance on an artificial sample only containing MP (polyethylene (PE),
polyester (PEST), polyamide (PA) and polyurethane (PU)), cellulose, quartz, and
diatomaceous earth. Out of these seven particle types, only diatomaceous earth
could not be detected within the wavenumber range of Anodisc. In comparison to
OPUS, siMPle identified almost twice the number of spectra on the specific
polymer types (Fig. 3).
Especially, cellulose (plant fiber in the database) and polycaprolactone (PCL,
not originally introduced as a material) were affected by factors larger than
four (Fig. 3, Ref7P
9Points and OPUS).
Figure 3.
Assigned polymer types for sample Ref7P[24] using different smoothing factors by siMPle (3 to 13) versus
OPUS with an unknown fixed value for polyethylene (PE), polyamide
(PA), polyester (PEST), acrylates/polyurethanes/varnish (APV),
cellulose, quartz and polycaprolactone (PCL).
Assigned polymer types for sample Ref7P[24] using different smoothing factors by siMPle (3 to 13) versus
OPUS with an unknown fixed value for polyethylene (PE), polyamide
(PA), polyester (PEST), acrylates/polyurethanes/varnish (APV),
cellulose, quartz and polycaprolactone (PCL).This difference was rather striking and the main difference between both kinds of
software was found in the data handling for the calculation of the first
derivative. In the default settings, siMPle adds a nine data points smoothing to
reduce the noise. For Bruker OPUS, it is not documented if smoothing is applied.
To test for a better comparison of the results, a range of this value from 3 to
13 data points was investigated for siMPle (Fig. 3).By screening the number of data points for smoothing, it was found that an
optimal hit was reached with nine data points for most polymer type assignment
(Fig. 3). Only
cellulose kept an increase in assigned polymer hits while PCL reached a constant
level. The data determined by OPUS could not be assigned to a smoothing factor
applied by siMPle. Still, for this particular sample, the high number of
assignments to PCL started using this number of data points for smoothing. To
avoid any misassignment issues, a manual reanalysis on the assigned spectra to
PCL was performed.Through quality assurance, it was found that the spectra assigned to PCL were
caused by a misinterpretation of the measured PEST spectrum. This spectrum has a
high similarity with PCL in transmission, which was not visible by using Bruker
OPUS during cluster performance analysis.[24] In Fig. S2, one of the assigned spectra from Ref7P is plotted against the
assigned reference spectrum and the spectra of the original material. The
original database states this material as a pure PEST, but via an extended
material research, it was found that the material was meanwhile relabeled to
copolyester by the manufacturer. Due to these differences, the material could
not be assigned to the original PEST cluster, as no pure PEST spectrum was
yielded. This issue will be addressed in the future by a database update
including more materials and using siMPle for cluster performance analysis. All
samples in the following were analyzed using the default nine data points
smoothing.The data sets RefEnv1 and RefEnv2 were also analyzed using siMPle and OPUS. The
siMPle analysis required only 2 h for RefEnv1 and 3 h for RefEnv2, which is 12
times faster than the analysis with OPUS using spectral correlation only for raw
and first derivative data. With a look at the polymer composition (see Table S1
for details), it was found that siMPle was more sensitive and identified higher
numbers of polymers and also larger sized polymer particles (Fig. 4) in comparison to
OPUS.
Figure 4.
MP size classes derived from the automated image analysis[21] for the reference data sets (a) RefEnv1[24] and (b) RefEnv2[24] analyzed using Bruker OPUS and siMPle.
MP size classes derived from the automated image analysis[21] for the reference data sets (a) RefEnv1[24] and (b) RefEnv2[24] analyzed using Bruker OPUS and siMPle.Both analyses found a strong trend toward smaller MP sizes. Especially striking
was the higher identification rate for cellulose (plant fiber) during the
analysis with siMPle (see RefEnv2, Table S1). Sample RefEnv2 also showed the
largest differences in the size distribution and showed a better particle
assignment compared to the data derived via OPUS (Fig. S3).Here, it was found that larger particles were identified more accurately with
siMPle in comparison to OPUS, which missed areas of larger particles yielding in
a separation into two particles. Furthermore, the analysis using siMPle improved
closing holes, which is important for morphological analysis of the particles,
see for example, the large PP particle on the rightmost edge of the filter. In
summary, these results show that data determination with siMPle is better suited
for the analysis of imaging data due to transparent data handling and easy data
validation.To test the ability for a harmonized analysis of MP, the performance of siMPle
was assessed for the FT-IR imaging data from instruments from the four mentioned
manufacturers. In this case, the computation of a full spectral analysis based
on Pearson correlation for the untreated data, the first derivative and the
second derivative were performed.[22] The calculation time was determined on the same computer systems applied
for the automated analysis via the OPUS software of the Bruker data set. This
allows a comparison of calculation times with existing studies using
OPUS.[26,28] The determined calculation times are summarized in Table 1 together with
further information on the data sets, such as pixel size on the filter area, the
size of the analyzed area and the number of spectra recorded.
Table I.
Calculation times of the different data sets measured on systems of
four different manufacturers using siMPle.
Data set
Size .spe
Spectra
Pixel size
Calculation time
Calculation performance spectra
Filter area measured
GB
N
µm
s
s
mm
Agilent
9.01
3 211 264
5.5
29 979
107
10 × 10
Bruker
4.14
1 806 336
11.05
16 464
110
14.9 × 14.9
PerkinElmer
0.66
215 296
25
2877
75
11.6 × 11.6
ThermoFisher Scientific
0.25
221 184
20
1129
195
11.5 × 7.2
Calculation times of the different data sets measured on systems of
four different manufacturers using siMPle.As mentioned previously, the calculation time on the Bruker data sets was reduced
considerably when applying siMPle (5 h, in comparison to 48 h using OPUS) which
also included the second derivative (it was omitted during OPUS analysis). When
the calculation time was normalized to the number of analyzed spectra, the
spectra from Thermo Fisher Scientific were correlated twice as fast compared to
the other data sets. The reason is the spectral resolution of 16 cm−1
instead of 8 cm−1. Still, one has to keep in mind that the Bruker
system and the Thermo Fisher Scientific system need an additional transformation
step within their respective software which increases the overall handling and
calculation times independent from siMPle. In Fig. 5, the false color images of the
analyzed samples are shown.
Figure 5.
Overview images of the measured filters using the automated analysis
pipeline for the (a) Agilent, (b) Bruker, (c) PerkinElmer, and (d)
Thermo Fisher Scientific samples. Sample (d) was measured in a
rectangular shape and the area of on the right side was left blank
to avoid irritations.
Overview images of the measured filters using the automated analysis
pipeline for the (a) Agilent, (b) Bruker, (c) PerkinElmer, and (d)
Thermo Fisher Scientific samples. Sample (d) was measured in a
rectangular shape and the area of on the right side was left blank
to avoid irritations.Qualitatively, it can be observed from the images that the samples are similar in
nature, containing a large proportion of natural materials (Fig. 6, gray colors), among which a
number of artificial polymers are successfully identified, irrespective of the
manufacturer of the instrument or the various sampling and extraction methods
employed prior to analysis (see ESM2.xslx, Supplemental Material, for details).
Due to the varying nature of the WWTP sampled from, and the variety of sampling
and extraction methods utilized between samples, commentary on any differences
in enumeration of MP between the samples is beyond the scope of this study.
However, application of the software on these real-world example data sets
demonstrates promising consistency in the proportion of particles identified
which are of synthetic origin (3–25%) and of the major polymer types which are
identified across the samples (see ESM2.xslx). Compared to existing commercial
software solutions, the harmonized analysis via siMPle saves working time and
computational costs. Current computers can run several instances of the software
unattended, allowing the data analysis of up to 16 samples per day compared to
OPUS (two days) for 1.8 million spectra per file. Still, it is possible also to
use low-cost office computers which normally can calculate up to three samples
per day containing 1.8 million spectra. As a minimum requirement, a processor
speed of 3 GHz is advised with 8 GB of RAM. To assess the performance of siMPle
on these data sets, a QA/QC analysis on the overall result was performed (Fig. 6).
Figure 6.
Assignment rates of correct and misidentified spectra for the
different instruments based on manual reanalysis similar to Primpke et al.[21]
Assignment rates of correct and misidentified spectra for the
different instruments based on manual reanalysis similar to Primpke et al.[21]In all cases, correct assignment rates >90% were reached, for three systems
(PerkinElmer, Bruker, and Agilent), and these were even >95% (Fig. 6). Those correct
assignment rates were exceedingly high, independently from instrument and sample
preparation, proving the high potential of siMPle as a harmonized tool for MP
analysis. Still, it is suggested and recommended that each study conducts an own
QA/QC analysis for each sample series for each polymer type identified as
demonstrated, e.g., in Lorenz et al.[32] Further questions, such as a comparison between existing analytical
pipelines, their harmonization, and a full QA/QC analysis will be addressed in a
later detailed study.To conclude, it is noteworthy that the siMPle software is not limited to MP
analysis, and it also allows the analysis of other types of data like the
spectral comparison of nano-FT-IR data[36] or the analysis of single algae species (Fig. S4).Using siMPle, single cells can be selected or specific characteristics can be
highlighted (Fig. S4). Here, the data show a strong Halo effect (Fig. S4b)
mainly caused by interference between the sample and the surface of the
CaF2 window, which is not visible using a heatmap (Fig. S4a). In
the future, heatmaps based on the integration of specific regions will be
introduced to allow even more control over the data. Further, additional
functions are currently planned to be introduced, and new possibilities can be
determined by contacting the authors to explore its application in a broader
scope for future research.
Conclusion
With siMPle, we present a freeware data analysis tool for the harmonized and
systematic analysis of spectroscopic data, with application, for example, in the
identification of MP in the environment. It allows data determination and
interpretation in a transparent and reproducible manner. In addition, it provides a
simpler quality QA/QC compared to existing commercial software tools like Bruker
OPUS and shows an increased identification rate. Furthermore, it allows the analysis
independently from the instrument manufacturer for a single spectrum but also for
large fields generated by imaging techniques. In particular, the field of FT-IR
imaging benefits greatly, as the data calculation time is reduced from several days
to 5 h using this software tool. Compared to other techniques, all spectra are
correlated via three different data treatments with the database yielding
high-quality results for all investigated instrument systems. This new tool improves
the application of FT-IR imaging in monitoring studies for MP, as it is accessible
for most types of spectrometers, free of charge and reduces the human bias during
manual data analysis.Click here for additional data file.Supplemental material, sj-zip-1-asp-10.1177_0003702820917760 for Toward the
Systematic Identification of Microplastics in the Environment: Evaluation of a
New Independent Software Tool (siMPle) for Spectroscopic Analysis by Sebastian
Primpke, Richard K. Cross, Svenja M. Mintenig, Marta Simon, Alvise Vianello,
Gunnar Gerdts and Jes Vollertsen in Applied Spectroscopy
Authors: Claudia Lorenz; Lisa Roscher; Melanie S Meyer; Lars Hildebrandt; Julia Prume; Martin G J Löder; Sebastian Primpke; Gunnar Gerdts Journal: Environ Pollut Date: 2019-06-24 Impact factor: 8.071
Authors: Livia Cabernard; Lisa Roscher; Claudia Lorenz; Gunnar Gerdts; Sebastian Primpke Journal: Environ Sci Technol Date: 2018-11-06 Impact factor: 9.028
Authors: Nanna B Hartmann; Thorsten Hüffer; Richard C Thompson; Martin Hassellöv; Anja Verschoor; Anders E Daugaard; Sinja Rist; Therese Karlsson; Nicole Brennholt; Matthew Cole; Maria P Herrling; Maren C Hess; Natalia P Ivleva; Amy L Lusher; Martin Wagner Journal: Environ Sci Technol Date: 2019-01-17 Impact factor: 9.028
Authors: Sebastian Primpke; Silke H Christiansen; Win Cowger; Hannah De Frond; Ashok Deshpande; Marten Fischer; Erika Holland; Michaela Meyns; Bridget A O'Donnell; Barbara Ossmann; Marco Pittroff; George Sarau; Barbara M Scholz-Böttcher; Kara Wiggin Journal: Appl Spectrosc Date: 2020-04-06 Impact factor: 2.388
Authors: Melanie Bergmann; Vanessa Wirzberger; Thomas Krumpen; Claudia Lorenz; Sebastian Primpke; Mine B Tekman; Gunnar Gerdts Journal: Environ Sci Technol Date: 2017-09-11 Impact factor: 9.028
Authors: Albert A Koelmans; Nur Hazimah Mohamed Nor; Enya Hermsen; Merel Kooi; Svenja M Mintenig; Jennifer De France Journal: Water Res Date: 2019-02-28 Impact factor: 11.236
Authors: Peter S Ross; Stephen Chastain; Ekaterina Vassilenko; Anahita Etemadifar; Sarah Zimmermann; Sarah-Ann Quesnel; Jane Eert; Eric Solomon; Shreyas Patankar; Anna M Posacka; Bill Williams Journal: Nat Commun Date: 2021-01-12 Impact factor: 14.919
Authors: Benedikt Hufnagl; Michael Stibi; Heghnar Martirosyan; Ursula Wilczek; Julia N Möller; Martin G J Löder; Christian Laforsch; Hans Lohninger Journal: Environ Sci Technol Lett Date: 2021-12-09
Authors: Darena Schymanski; Barbara E Oßmann; Nizar Benismail; Kada Boukerma; Gerald Dallmann; Elisabeth von der Esch; Dieter Fischer; Franziska Fischer; Douglas Gilliland; Karl Glas; Thomas Hofmann; Andrea Käppler; Sílvia Lacorte; Julie Marco; Maria El Rakwe; Jana Weisser; Cordula Witzig; Nicole Zumbülte; Natalia P Ivleva Journal: Anal Bioanal Chem Date: 2021-07-20 Impact factor: 4.142