Automated MRI volumetry as a diagnostic tool for Alzheimer's disease: Validation of icobrain dm.

Hanne Struyfs1, Diana Maria Sima2, Melissa Wittens1, Annemie Ribbens3, Nuno Pedrosa de Barros3, Thanh Vân Phan3, Maria Ines Ferraz Meyer4, Lene Claes3, Ellis Niemantsverdriet5, Sebastiaan Engelborghs6, Wim Van Hecke3, Dirk Smeets3.   

Abstract

Brain volumes computed from magnetic resonance images have potential for assisting with the diagnosis of individual dementia patients, provided that they have low measurement error and high reliability. In this paper we describe and validate icobrain dm, an automatic tool that segments brain structures that are relevant for differential diagnosis of dementia, such as the hippocampi and cerebral lobes. Experiments were conducted in comparison to the widely used FreeSurfer software. The hippocampus segmentations were compared against manual segmentations, with significantly higher Dice coefficients obtained with icobrain dm (25-75th quantiles: 0.86-0.88) than with FreeSurfer (25-75th quantiles: 0.80-0.83). Other brain structures were also compared against manual delineations, with icobrain dm showing lower volumetric errors overall. Test-retest experiments show that the precision of all measurements is higher for icobrain dm than for FreeSurfer except for the parietal cortex volume. Finally, when comparing volumes obtained from Alzheimer's disease patients against age-matched healthy controls, all measures achieved high diagnostic performance levels when discriminating patients from cognitively healthy controls, with the temporal cortex volume measured by icobrain dm reaching the highest diagnostic performance level (area under the receiver operating characteristic curve = 0.99) in this dataset.
Copyright © 2020 The Authors. Published by Elsevier Inc. All rights reserved.

Entities:  

Keywords:  Alzheimer's disease (AD); Brain segmentation software; Dementia; Magnetic resonance imaging (MRI)

Mesh:

Year:  2020        PMID: 32193172      PMCID: PMC7082216          DOI: 10.1016/j.nicl.2020.102243

Source DB:  PubMed          Journal:  Neuroimage Clin        ISSN: 2213-1582            Impact factor:   4.881


Introduction

Structural neuroimaging with magnetic resonance imaging (MRI) (or computed tomography (CT)) plays a key role in the diagnostic work-up of dementia. It allows to rule out structural lesions of the brain that might cause cognitive problems. In addition, structural neuroimaging may contribute to the early and differential diagnosis of the neurodegenerative disease underlying the dementia syndrome (Chui et al., 1992; Roman et al., 1993; Chan et al., 2001; Rosen et al., 2002a, 2002b; Boccardi et al., 2003). Indeed, neurodegenerative disorders that cause dementia are often associated with typical brain atrophy patterns. Alzheimer's disease (AD), for instance, is characterized by medial temporal lobe atrophy, including the hippocampus, and parietal atrophy. Frontotemporal dementia, on the other hand, mainly presents with atrophy of the frontal and (anterior and / or lateral parts of the) temporal lobes. Dementia with Lewy bodies usually does not show specific structural abnormalities, while vascular dementia is mainly characterized by global atrophy and diffuse white matter lesions, lacunes and/or strategic infarcts. As such, global and focal atrophy together with vascular disease are important factors to consider when establishing a differential dementia diagnosis. Gradually, these factors are being included into diagnostic clinical criteria for dementia (McKhann et al., 1984; Roman et al., 1993; Neary et al., 1998; McKhann et al., 2011; McKeith et al., 2017). Besides contributing to differential diagnosis of prevalent dementia, structural neuroimaging may also aid in predicting progression to dementia in subjects who have not reached the dementia stage yet. MRI studies have shown hippocampal atrophy to be associated with increased risk of progression to dementia due to AD (Dubois, 2018). Hippocampal atrophy is included as a biomarker for early AD diagnosis in the revised diagnostic criteria of the National Institute on Aging – Alzheimer Association working group (Albert et al., 2011; Sperling et al., 2011). In order to segment brain regions-of-interest and measure brain atrophy, fully automated processing techniques have been developed. These can be used in large study cohorts, saving both time and costs, and are easily reproducible, as opposed to manual segmentation by neuroanatomical experts or semi-automated measures that still require a priori information on the region-of-interest (Duchesne et al., 2002; Barnes et al., 2008; Kennedy et al., 2009; Dewey et al., 2010; Boccardi et al., 2011; Doring et al., 2011; Bosco et al., 2017). FreeSurfer is a very frequently used automatic tool (Fischl, 2012); depending on hardware, may require a long computation time of up to several tens of hours per scan (http://surfer.nmr.mgh.harvard.edu/). Applying automated measures of brain volumes on individual dementia patients requires a low measurement error and high reliability. For instance, a meta-analysis pointed to an annual atrophy rate of the hippocampus of 4.66% in AD patients compared with 1.41% in controls (Barnes et al., 2009). Hence, the measurement error of the brain volumetric measures should be minimal, in order to draw meaningful conclusions in individual patients. In this study we validate an automated method to measure volumes of the whole brain (WB), total gray matter (GM), frontal, parietal and temporal cortex, hippocampi, and lateral ventricles. In order to evaluate the applicability of the method for brain volume quantification of individual dementia patients, this paper focuses on the accuracy, reliability and diagnostic performance of these volumetric measures.

Materials and methods

Dataset 1.a (accuracy)

Dataset 1.a was acquired from 35 healthy subjects (mean age 34 (±20 SD) years, 67% females,) as part of the OASIS project (http://www.oasis-brains.org). Manual brain segmentations were produced by Neuromorphometrics, Inc. (neuromorphometrics.com) using the brainCOLOR labeling protocol. The data were part of the 2012 MICCAI Multi-Atlas Labeling Challenge, where 15 subjects were used as training and the remaining 20 images were used for testing. Since all 35 manual segmentations were made available, we do not make this distinction and, thus, we report results on all 35 images. The 3D magnetization-prepared rapid gradient-echo (MP-RAGE) T1-weighted MRIs were acquired using a 1.5T Siemens Vision MR scanner, voxel size of 1 × 1 × 1 mm and dimensions up to 256 × 334 × 256 mm.

Dataset 1.b (accuracy)

Dataset 1.b was acquired from 46 subjects of a memory clinic-based research population who participated in a study at the University of Antwerp, Belgium (mean age 72.0 (±7.8 SD) years, 50.0% females, Mini–Mental State Examination (MMSE) score 25.8 ± 3.1). This population consisted of 6 cognitively healthy controls as well as patients with subjective cognitive decline (n = 3), mild cognitive impairment (n = 28) and dementia due to AD (n = 9). Local ethics committees (Hospital Network Antwerp and University of Antwerp / Antwerp University Hospital) approved the study and all patients signed informed consent forms. MR imaging was performed on each subject on a 3T whole body scanner with a 32-channel head coil (Siemens Trio/PrismaFit, Erlangen, Germany). The 3D MP-RAGE (TR/TE = 2200/2.45 ms) was used to obtain 176 axial slices without slice gap and 1.0 mm nominal isotropic resolution (FOV = 192 × 256 mm). An expert (LC) performed bilateral manual hippocampus segmentation on all subjects according to the EADC-ADNI harmonized hippocampus segmentation guidelines (Boccardi et al., 2015). These manual segmentations were further used as ground truth references.

Dataset 2 (reproducibility)

Dataset 2 consisted of 42 cognitively healthy subjects (i.e., having score 0 on the Clinical Dementia Rating scale) who received longitudinal scans up to 10 days apart (mean age 61.4 (±8.6 SD) years, 59.5% females), provided by the publicly available database OASIS-3 (http://www.oasis-brains.org). MR imaging was performed on each subject on a 3T whole body scanner with a 16-channel head coil (Siemens TIM Trio or BioGraph mMR PET-MR, Erlangen, Germany). The baseline and follow-up scans of three subjects were done on the same scanner, while all other 39 subjects had different scanner types for their baseline and follow-up scans. The MP-RAGE protocol of TIM Trio scanner was as follows: TR/TE = 2400/3.16 ms, ±176 axial slices without slice gap and 1.0 mm nominal isotropic resolution (FOV = 256 × 256 mm). The MP-RAGE protocol of BioGraph mMR PET-MR scanner was as follows: TR/TE = 2300/2.95 ms, ±176 axial slices without slice gap and 1.2 mm nominal isotropic resolution (FOV = 256 × 256 mm).

Dataset 3 (diagnostic performance)

Dataset 3 consisted of 46 AD patients (age 71.5 ± 7.2, 60.9% females, Mini–Mental State Examination (MMSE) 19.2 ± 4) and 23 cognitively healthy subjects (age 70.4 ± 7.1, 47.8% females, MMSE 29.4 ± 0.8) of the publicly available MIRIAD database (miriad.drc.ion.ucl.ac.uk). An overview of the MIRIAD demographics, diagnostic procedures, and imaging protocol was published previously (Malone et al., 2013). In brief, AD patients were diagnosed with mild–moderate probable AD according to the NINCDS–ADRDA clinical criteria (McKhann et al., 1984), while the control subjects did not have subjective cognitive complaints, nor evidence of cognitive impairment. All scans were conducted on a 1.5T whole body scanner (GE Medical systems Signa, Milwaukee, Wisconsin, USA). Three-dimensional T1-weighted (T1w) images were acquired with an IR-FSPGR (inversion recovery prepared fast spoiled gradient recalled) sequence, FOV 240 mm, 256 × 256 matrix, 124 1.5 mm coronal partitions, TR/TE = 15/5.4 ms. A summary of the 3 datasets can be found in Table 1.
Table 1

Short overview of datasets used for method validation.

DATA# subjectsAgeCognitive stateSource
Dataset 1.a: accuracy3534 ± 20Healthy controlsMICCAI 2012 challenge neuromorphometrics.com
Dataset 1.b: accuracy4672.0 ± 7.8MMSE: 25.8 ± 3.1University of Antwerp, Belgium
Dataset 2: reproducibility4261.4 ± 8.6Healthy controlsOASIS-3 www.oasis-brains.org
Dataset 3: diagnostic performance4671.5 ± 7.2MMSE: 19.2 ± 4MIRIAD miriad.drc.ion.ucl.ac.uk
2370.4 ± 7.1Healthy controls
Short overview of datasets used for method validation.

MRI analysis

icobrain dm

icobrain dm (version 4.3) is a medical device software that measures relevant volumes of brain structures to assist radiologic assessment of dementia patients. The general icobrain pipeline segments a T1w image into white matter, gray matter and cerebrospinal fluid. When a FLAIR image is available, white matter FLAIR hyper-intensities are also identified and included in the white matter segmentation. The main blocks of the icobrain pipeline have been described previously (Jain et al., 2015); in short, after skull stripping and bias correction, the T1w image is segmented using a probabilistic image intensity model and non-rigidly propagated tissue priors from an MNI atlas (Evans et al., 1992). Lesion segmentation is obtained as intensity outliers to a probabilistic FLAIR image segmentation, and the tissue segmentation is improved iteratively by re-segmenting the lesion-filled T1w image. Volumes are normalized for head size, using the determinant of the affine transformation to MNI atlas as a scaling factor. icobrain dmfurther refines this main tissue segmentation in order to obtain sub-segmentations of cortical gray matter lobes and of the hippocampi. Sub-segmentations of cortical lobes are obtained from the icobrain cortical gray matter segmentation, annotated according to a set of cortical labels available in MNI space (Klein and Tourville, 2012). Initial non-rigid registration (Modat et al., 2010) between the patient's T1w image and the MNI template is used to obtain a first propagation of the cortical labels from atlas space (“CGM labels”) to the patient's T1w image space. This label propagation is further refined through a second non-rigid registration between the skeleton of the patient's binarized cortical gray matter segmentation and the skeleton of the binarized propagated CGM labels. Finally, each cortical gray matter voxel ias assigned as the cortical label matching the closest voxel in the skeleton of the non-rigidly propagated CGM labels. Segmentation of the hippocampi starts from the T1w scans preprocessed by the icobrain pipeline, including bias field correction, brain orientation and skull stripping. After preprocessing, a multi-atlas segmentation approach registers binary anatomical priors (i.e., a set of manually annotated hippocampi corresponding to the guidelines of the EADC-ADNI harmonized protocol - (Boccardi et al., 2015)) for left and right hippocampi to the T1w image space using an affine and a non-rigid image registration algorithm. The propagated segmentations are then combined into one probabilistic segmentation for each hippocampus. This label fusion is based on a local ranking using the locally normalized cross correlation as a similarity metric (Cardoso et al., 2013). Subsequently, the probabilistic segmentation of each hippocampus is used as a prior in an intensity-based 2-step maximum likelihood expectation-maximization algorithm (Cardoso, 2012) to obtain the final hippocampus segmentation. As a post-processing step, voxels mainly considered as CSF by the main tissue segmentation are excluded from the hippocampus segmentation, to keep in line with the EADC-ADNI harmonized protocol, which agreed on excluding internal CSF pools from manual hippocampus segmentation. icobrain dm was executed on a Linux server with 8 CPU cores (Intel Xeon Platinum 8000) and 16 GB RAM, and required between 15 and 30 min per scan to complete.

FreeSurfer

The Freesurfer image analysis suite (version 6.0) is documented and freely available for download online (http://surfer.nmr.mgh.harvard.edu/) and has been thoroughly described elsewhere (Fischl et al., 2002; Fischl, 2012). In this paper, we used the recon-all stream with fully-automated directive -all, in order to reconstruct all brain volumes, including cortical and subcortical parcellations. Since we used very diverse datasets, they were all processed with identical command and default parameters, without optimizing for a specific dataset (e.g., without −3T or -mprage options). Cortical labels corresponding to the frontal, temporal and parietal gray matter regions were grouped in order to obtain volumes of the same three cortical lobe regions as for icobrain dm. When reporting volumes normalized for head size, in order to obtain brain volumes in the same range as icobrain, we performed a scaling of the FreeSurfer volumes using the formula below, where 1985.026 ml is the intracranial volume of the MNI template used in icobrain and ‘Estimated Total Intracranial Volume’ is the total intracranial volume reported by FreeSurfer: FreeSurfer's more recent functionality for segmentation of hippocampal subfields and nuclei of the amygdala (Iglesias et al., 2015) was also applied on the accuracy datasets 1.a and 1.b, from which volumes of the whole left and right hippocampi were extracted. FreeSurfer was executed on a Linux server with 16 CPU cores (Intel Xeon Platinum 8000) and 64GB RAM, and required between 9 and 13 h per scan to complete. Both icobrain and FreeSurfer used only the T1w images as input.

Validation

icobrain dm and FreeSurfer were validated in terms of accuracy, reproducibility and diagnostic performance of all measures. Accuracy of the hippocampal segmentation received special attention, as it was compared against two different approaches implemented in FreeSurfer. Statistical analyses were performed using the integrated development environment for R programming language, RStudio (version 1.0.136) (Team R, 2016). Per experiment, significant differences between icobrain dm and FreeSurfer were evaluated using the nonparametric Wilcoxon signed-rank test, using R package ‘MASS’ (Venables and Ripley, 2002), at significance level 0.01. First, we quantified measurement error of all structures and in particular of the hippocampus segmentation with respect to manual ground truth segmentation (datasets 1.a and 1.b). The measurement error was computed as the (absolute) volume difference between ground truth volume and icobrain dm or FreeSurfer volume. In addition, accuracy of the hippocampal segmentation was assessed by the Dice similarity coefficient (DSC). DSC was used to measure the similarity between the ground truth and the automatic segmentation results separately for left and right hippocampus and for total hippocampal volume for each method. According to (Dill et al., 2015) a DSC of 0.80 can be considered a good accuracy value, since it was measured by previous studies as the average rate of similarity between two manual hippocampus segmentations performed by experienced operators. Subsequently, we assessed reproducibility of all measures on test–retest images from cognitively healthy subjects (dataset 2), based on the absolute volume difference between these pairs of images. Finally, the diagnostic performance of the measures to distinguish AD patients and cognitively healthy subjects was evaluated (dataset 3) by means of a receiver operating characteristic curve (ROC) analysis with DeLong tests at significance level 0.05, using the ‘pROC’ package (Robin et al., 2011).

Results

Accuracy of brain (sub)structures segmentation

Fig. 1 illustrates the accuracy results for the brain segmentation obtained by icobrain dm and FreeSurfer on dataset 1.a (MICCAI 2012 challenge). These results are also summarized in Table 2. It is obvious that several volumes are biased with respect to the ground truth volumes obtained from manual segmentation, and icobrain dm and FreeSurfer typically have the same bias direction (i.e. underestimation for WB, GM and the cortical lobes), with the exception of the hippocampi, where FreeSurfer's default hippocampus segmentation overestimates most of the volumes. On the other hand, FreeSurfer's hippocampal subfield functionality underestimates them. For all measurements, icobrain dm has lower bias and lower absolute error. Moreover, there are fewer outliers.
Fig. 1

Scatter plots illustrating the brain volumes segmentations by icobrain dm and FreeSurfer (including FreeSurfer's hippocampal subfield functionality, denoted “FS subfields”) compared to expert manual segmentation on dataset 1.a.

Table 2

Accuracy of volumes obtained by icobrain dm and FreeSurfer when compared with expert manual segmentation on dataset 1.a (MICCAI 2012 challenge), where volume differences are computed as ground truth segmentation volume minus volume computed automatically by icobrain dm, FreeSurfer or FreeSurfer's hippocampal subfield functionality, “FS subfields”.

volume differences to ground truth
absolute volume differences to ground truth
number of volumetric outliers
P values icobrain dm vs. FreeSurfer
icobrain dmFreeSurfericobrain dmFreeSurfericobrain dmFreeSurfer
Whole brain78.6116.078.6116.026< 0.001
(65.5; 91.3)(97.4; 127.4)(65.5; 91.3)(97.4; 127.4)
Gray matter45.1149.645.5149.6121< 0.001
(31.1; 71.4)(123.7; 170.4)(31.8; 71.4)(124; 170.4)
Frontal lobe14.038.714.138.7213< 0.001
(9.4; 20.8)(34.1; 44.7)(9.6; 20.8)(34.1; 44.7)
Parietal lobe10.027.210.627.213< 0.001
(4.5; 17.6)(18.8; 33.8)(5.3; 17.6)(18.8; 33.8)
Temporal lobe21.742.921.742.9122< 0.001
(14.7; 24.3)(35.8; 48.5)(14.7; 24.3)(35.8; 48.5)
Hippocampus0.3−0.70.30.9123< 0.001
(0.1; 0.6)(−1.0; −0.2)(0.2; 0.6)(0.5; 1.4)
Left hippocampus0.1−0.40.20.51100.01
(−0.1; 0.2)(−0.5; −0.2)(0.1; 0.3)(0.2; 0.7)
Right hippocampus0.2−0.30.20.40130.02
(0.0; 0.3)(−0.6; −0.1)(0.1; 0.3)(0.3; 0.6)
Lateral ventricles0.61.70.92.0040.006
(−0.3; 1.7)(0.6; 3.1)(0.5; 1.7)(0.8; 3.3)

Note: Values in the first 4 columns are median (25–75th quantiles) volume differences or absolute volume difference in ml (not normalised for head size). Volumetric outliers are defined as measurements below (25th percentile - 1.5 interquartile range) or above (75th percentile + 1.5 interquartile range), where these limits are obtained from the volumetric errors of icobrain dm (first column of the table). P values are obtained from Wilcoxon signed-rank tests applied on absolute volume differences for icobrain dm and Freesurfer.

Scatter plots illustrating the brain volumes segmentations by icobrain dm and FreeSurfer (including FreeSurfer's hippocampal subfield functionality, denoted “FS subfields”) compared to expert manual segmentation on dataset 1.a. Accuracy of volumes obtained by icobrain dm and FreeSurfer when compared with expert manual segmentation on dataset 1.a (MICCAI 2012 challenge), where volume differences are computed as ground truth segmentation volume minus volume computed automatically by icobrain dm, FreeSurfer or FreeSurfer's hippocampal subfield functionality, “FS subfields”. Note: Values in the first 4 columns are median (25–75th quantiles) volume differences or absolute volume difference in ml (not normalised for head size). Volumetric outliers are defined as measurements below (25th percentile - 1.5 interquartile range) or above (75th percentile + 1.5 interquartile range), where these limits are obtained from the volumetric errors of icobrain dm (first column of the table). P values are obtained from Wilcoxon signed-rank tests applied on absolute volume differences for icobrain dm and Freesurfer.

Accuracy of hippocampus segmentation

Continuing with the dataset 1.a, we report the DSC for hippocampus segmentations for icobrain dm at 0.8223 (0.8142; 0.8321) (median and interquartile range), while FreeSurfer's default hippocampus segmentation scores a DSC of 0.7988 (0.7867; 0.8158). FreeSurfer's newer hippocampal subfield functionality (Iglesias et al., 2015) scores a slightly lower DSC of 0.7953 (0.7867; 0.8092). Fig. 2 illustrates the accuracy of the hippocampus segmentation obtained by icobrain dm and FreeSurfer (dataset 1.b), with panel A showing the absolute volume difference from ground truth, panel B the DSC, and panel C scatter plots of automated measurements versus manual ground truth. These results are also summarized in Table 3. The median absolute volume difference of icobrain dm was significantly lower than that of FreeSurfer's default stream and FreeSurfer's hippocampal subfield functionality, which is also supported by a significantly higher DSC for icobrain dm compared with FreeSurfer methods. It should be noted that 44/46 subjects had a DSC above 0.80 when segmented by icobrain dm compared with 35/46 subjects for FreeSurfer and 42/46 for FreeSurfer's hippocampal subfield functionality.
Fig. 2

Accuracy of hippocampus segmentation by icobrain dm and FreeSurfer, including FreeSurfer's hippocampal subfield functionality, denoted “FS subfields”, when compared with expert manual segmentation on dataset 1.b. A. Absolute volume difference between manual and automated segmentation. B. Dice similarity coefficient between manual and automated segmentation. C. Scatterplots comparing ground truth volumes to those obtained from icobrain dm and FreeSurfer. Note: p-values are obtained from Wilcoxon signed-rank tests.

Table 3

Accuracy of hippocampus segmentation by icobrain dm and FreeSurfer, including FreeSurfer's hippocampal subfield functionality, denoted “FS subfields”, when compared with expert manual segmentation on dataset 1.b (only hippocampal segmentations), where volume differences are computed as ground truth segmentation volume minus volume computed automatically by icobrain dm, FreeSurfer or “FS subfields” software.

Volume difference, ml, median (25–75 quantiles)Absolute volume difference, ml, median (25–75 quantiles)Dice similarity coefficient, median (25–75 quantiles)Number of volumetric outliers
icobrain dmHippocampus0.190.230.871
(−0.06; 0.46)(0.14; 0.48)(0.86; 0.88)
Left hippocampus0.100.150.871
(−0.05; 0.34)(0.07; 0.35)(0.84; 0.88)
Right hippocampus0.040.130.880
(−0.03; 0.21)(0.04; 0.22)(0.87; 0.89)
FreeSurferHippocampus−0.700.700.8219
(−1.01; −0.36)(0.36; 1.01)(0.80; 0.83)
Left hippocampus−0.380.380.818
(−0.53; −0.16)(0.18; 0.53)(0.79; 0.83)
Right hippocampus−0.320.320.8317
(−0.50; −0.21)(0.21; 0.50)(0.81; 0.84)
FS subfieldsHippocampus0.820.820.823
(0.55; 1.03)(0.61; 1.03)(0.81; 0.83)
Left hippocampus0.470.480.821
(0.24; 0.61)(0.29; 0.61)(0.80; 0.83)
Right hippocampus0.360.360.832
(0.26; 0.45)(0.26; 0.45)(0.82; 0.84)
P values icobrain dm vs. FreeSurferHippocampus<0.0001<0.0001<0.0001
Left hippocampus.00020.0049<0.0001
Right hippocampus<0.0001<0.0001<0.0001
P values icobrain dm vs. FS subfieldsHippocampus<0.0001<0.0001<0.0001
Left hippocampus<0.0001<0.0001<0.0001
Right hippocampus<0.0001<0.0001<0.0001

Note: Hippocampal volumes are not normalized for intracranial volume as the analyses are performed in native space. Manual segmentation volumes ranged from 3.8 ml to 8.6 ml. P values are obtained from Wilcoxon signed-rank tests.

Volumetric outliers are defined as measurements below (25th percentile - 1.5 interquartile range) or above (75th percentile + 1.5 interquartile range), where these limits are obtained from the volumetric errors of icobrain dm.

Accuracy of hippocampus segmentation by icobrain dm and FreeSurfer, including FreeSurfer's hippocampal subfield functionality, denoted “FS subfields”, when compared with expert manual segmentation on dataset 1.b. A. Absolute volume difference between manual and automated segmentation. B. Dice similarity coefficient between manual and automated segmentation. C. Scatterplots comparing ground truth volumes to those obtained from icobrain dm and FreeSurfer. Note: p-values are obtained from Wilcoxon signed-rank tests. Accuracy of hippocampus segmentation by icobrain dm and FreeSurfer, including FreeSurfer's hippocampal subfield functionality, denoted “FS subfields”, when compared with expert manual segmentation on dataset 1.b (only hippocampal segmentations), where volume differences are computed as ground truth segmentation volume minus volume computed automatically by icobrain dm, FreeSurfer or “FS subfields” software. Note: Hippocampal volumes are not normalized for intracranial volume as the analyses are performed in native space. Manual segmentation volumes ranged from 3.8 ml to 8.6 ml. P values are obtained from Wilcoxon signed-rank tests. Volumetric outliers are defined as measurements below (25th percentile - 1.5 interquartile range) or above (75th percentile + 1.5 interquartile range), where these limits are obtained from the volumetric errors of icobrain dm. Fig. 3 shows two illustrations of hippocampus segmentations by icobrain dm and FreeSurfer with high and low DSCs, respectively.
Fig. 3

Illustrations of hippocampus segmentation by an expert (ground truth), icobrain dm, and FreeSurfer from dataset 1.b. The top panel shows segmentations with high Dice similarity coefficient (0.90 for icobrain dm, 0.84 for FreeSurfer and 0.85 FreeSurfer's hippocampal subfield functionality), while segmentations with lower Dice similarity coefficients are presented in the bottom panel (0.79 for icobrain dm, 0.77 for FreeSurfer and 0.75 for FreeSurfer's hippocampal subfield functionality).

Illustrations of hippocampus segmentation by an expert (ground truth), icobrain dm, and FreeSurfer from dataset 1.b. The top panel shows segmentations with high Dice similarity coefficient (0.90 for icobrain dm, 0.84 for FreeSurfer and 0.85 FreeSurfer's hippocampal subfield functionality), while segmentations with lower Dice similarity coefficients are presented in the bottom panel (0.79 for icobrain dm, 0.77 for FreeSurfer and 0.75 for FreeSurfer's hippocampal subfield functionality).

Reproducibility

Fig. 4 illustrates the absolute volume differences between test and retest scans (dataset 2) for all measures. Detailed volume differences are presented in Table 4. The segmentations obtained by icobrain dm systematically tended to have lower volume differences than FreeSurfer, except for parietal lobe volume, with significant differences for whole brain, total gray matter, and hippocampal volumes.
Fig. 4

Reproducibility of segmentations by icobrain dm and FreeSurfer on dataset 2, measured by the absolute volume difference between test-retest segmentations. Note: P values are obtained from Wilcoxon signed-rank tests.

Table 4

Reproducibility of segmentations by icobrain dm and FreeSurfer on dataset 2, measured by the absolute volume difference in millilitres between test and retest quantifications.

icobrain dmFreeSurfer
Whole brain7.9128.43
(3.55–15.05)(14.79–38.15)
Gray matter6.3311.01
(2.62–10.73)(6.81–21.20)
Frontal lobe2.964.80
(0.99–4.56)(1.58–7.57)
Parietal lobe3.602.61
(1.21–5.31)(1.20–5.01)
Temporal lobe1.642.54
(1.07–3.66)(1.28–4.07)
Hippocampus0.1110.330
(0.032–0.232)(0.188–0.444)
Left hippocampus0.0940.161
(0.057–0.176)(0.080–0.228)
Right hippocampus0.1020.174
(0.054–0.175)(0.078–0.256)
Lateral ventricles0.480.69
(0.22–0.83)(0.35–1.17)

Note: Values are median (25–75th quantiles) absolute volume differences in ml (normalised for head size). FreeSurfer's hippocampal segmentations are obtained with the default stream.

Reproducibility of segmentations by icobrain dm and FreeSurfer on dataset 2, measured by the absolute volume difference between test-retest segmentations. Note: P values are obtained from Wilcoxon signed-rank tests. Reproducibility of segmentations by icobrain dm and FreeSurfer on dataset 2, measured by the absolute volume difference in millilitres between test and retest quantifications. Note: Values are median (25–75th quantiles) absolute volume differences in ml (normalised for head size). FreeSurfer's hippocampal segmentations are obtained with the default stream.

Diagnostic performance

As shown in Table 5, all measures from both icobrain dm and FreeSurfer have high area under the curve (AUC) levels to distinguish AD patients from cognitively healthy controls (dataset 3). Temporal lobe volume measured by icobrain dm produced the highest AUC (0.9896), which was significantly higher than the temporal lobe AUC produced by FreeSurfer (0.9565, P = 0.04646).
Table 5

Diagnostic performance to differentiate AD patients from age-matched controls on dataset 3.

icobrain dmFreeSurferP value
Whole brain0.9395 (0.8941–0.9849)0.9414 (0.8964–0.9864).9414
Gray matter0.9386 (0.8955–0.9816)0.9282 (0.8730–0.9834).7313
Frontal lobe0.7963 (0.7055–0.8872)0.8790 (0.8109–0.9472).0767
Parietal lobe0.8601 (0.7848–0.9355)0.8960 (0.8299–0.9621).3242
Temporal lobe0.9896 (0.9770–1.0000)0.9565 (0.9187–0.9944).0465
Hippocampus0.9022 (0.8426–0.9617)0.9168 (0.8631–0.9706).2802
Left hippocampus0.8776 (0.8000–0.9551)0.9055 (0.8400–0.9709).1735
Right hippocampus0.8965 (0.8365–0.9565)0.8885 (0.8253–0.9517).6343
Lateral ventricles0.8899 (0.8180–0.9617)0.8488 (0.7660–0.9315).0013

Note: Values are areas under the receiver operating characteristic curve (95% confidence interval). DeLong tests were used to test whether AUC levels differed significantly between icobrain dm and FreeSurfer. FreeSurfer's hippocampal segmentations are obtained with the default stream.

Diagnostic performance to differentiate AD patients from age-matched controls on dataset 3. Note: Values are areas under the receiver operating characteristic curve (95% confidence interval). DeLong tests were used to test whether AUC levels differed significantly between icobrain dm and FreeSurfer. FreeSurfer's hippocampal segmentations are obtained with the default stream.

Discussion

In this paper, the automated method icobrain dm for measuring brain volumes is presented and compared to the widely used FreeSurfer. In order to assess the use of this method in clinical practice on MRI scans of individual dementia patients, the reliability of the method is evaluated in terms of accuracy, reliability and diagnostic performance of all measures. Results are compared to FreeSurfer, a well-validated and extensively used method for measuring brain volumes in clinical studies and trials. icobrain dm and FreeSurfer results on dataset 1.a demonstrated bias in most volumes compared to manual delineations. A systematic bias is not dangerous as such, because volumes obtained with a certain automated software would typically only be compared with the same software between patient groups or between patients and healthy controls. A reason for bias to manual delineations could be the absence of partial volume effect in the manual ground truth. Both icobrain dm and FreeSurfer compute theirs volumes from probability maps, where the voxels close to the brain contour are partly brain tissue, partly CSF, without sharp edges. Hippocampus segmentation showed however a divergent trend between the 2 automated methods, with FreeSurfer's default stream overestimating most volumes, and icobrain dm slightly underestimating them. On the other hand, FreeSurfer's hippocampal subfields segmentation module (Iglesias et al., 2015), which is currently included in FreeSurfer's development version and thus is not yet the default algorithm, underestimates the considered manual segmentations slightly more than icobrain dm. A recent paper (Ataloglou et al., 2019) reported state-of-the-art hippocampus segmentation results using deep convolutional neural network (CNN) ensembles, reaching a Dice score of 0.88 on the same MICCAI 2012 challenge dataset. However, the authors had to tune their CNN with transfer learning on a training subset of the MICCAI 2012 challenge dataset in order to reach these maximal performance results. Deep learning is increasingly superior to classical brain segmentation approaches, but it is limited by the amount, the diversity and the quality of the data used for training. icobrain dm results on dataset 1.b demonstrated a small measurement error for hippocampus segmentation, with a median absolute volume difference from ground truth of 0.230 ml. The similarity with ground truth was generally high, with a median DSC of 0.87 and 44/46 segmentations with a DSC above 0.80. The accuracy of icobrain dm was significantly higher than that of both the default hippocampal segmentation in FreeSurfer 6.0 recon-all stream and FreeSurfer's hippocampal subfields segmentation module (Iglesias et al., 2015), confirming the same trends observed in dataset 1.a. Bias in hippocampal volumes between automated methods and manual annotations is not surprising, since not all methods and all manual raters use the same definition of the hippocampal borders seen on MRI. The recent EADC-ADNI harmonized protocol (Boccardi et al., 2015), which is used for the multi-atlas approach of icobrain dm, is more clearly defined compared to prior protocols, but it differs from the Center for Morphometric Analysis (CMA) guidelines (Filipek et al., 1994) underlying FreeSurfer's probabilistic atlas used by the default recon-all stream. Other recent studies such as (Schmidt et al., 2018) found that FreeSurfer 6.0 overestimates the hippocampal volume by 20% compared to manual raters, which is explained by the fact that FreeSurfer includes further caudal regions, resulting in larger tails, as well as some voxels between hippocampus and lateral ventricles. On the other hand, the newer FreeSurfer hippocampal subfields segmentation module (Iglesias et al., 2015) is based on a quite different definition of the hippocampal formation at the subregion level, using ultra-high resolution ex vivo MRIs. The total hippocampal volume obtained with this approach underestimates the volumes obtained from manual segmentations in both accuracy datasets considered in this paper. A potential explanation for this bias towards smaller volumes is that the hippocampus subfield atlas was built using elderly subjects, and was based on a detailed ex vivo MRI delineation protocol that cannot be performed on in vivo brain scans. The test-retest error on dataset 2 was lower for icobrain dm for all measures except parietal lobe volume, although these differences were significant only for whole brain, total gray matter, and hippocampal volumes. Regarding hippocampal volume, the average test-retest absolute volume difference of the hippocampus is 0.111 ml, which represents 1.20% of the average icobrain dm hippocampal volume (measured by icobrain dm; test and retest combined). As such, the measurement error is below the average annual hippocampal atrophy rates of 1.41% in healthy individuals (Barnes et al., 2009). For FreeSurfer's hippocampal subfields segmentation, which we explored in the accuracy experiments (Iglesias et al., 2015), Iglesias et al. (2016) reported test-retest reliability of around 2.5% for the whole left and right hippocampus. It should also be noted that test-retest exercises are usually performed with datasets on the same scanner. In this manuscript we evaluated test-retest reliability on different scanner types. This increases variability and is better in line with clinical practice. Finally, when using dataset 3, we found that all measures achieve high diagnostic performance levels when discriminating AD patients from cognitively healthy controls. The temporal lobe volume measured by icobrain dm reached the highest diagnostic performance level (AUC = 0.9896). Although hippocampal atrophy is considered the most disease-specific for Alzheimer's disease, it is not surprising that this structure has slightly lower diagnostic performance compared to the temporal lobe volume, since lower volumes (such as hippocampus) are likely affected by proportionally higher measurement errors. Moreover, not all subjects had severe dementia, as dataset 3 consisted of mild-moderate probable AD. Of note, the frontal lobe produced the lowest diagnostic performance levels, with FreeSurfer showing stronger differences compared to icobrain dm. In fact icobrain dm finds the frontal cortex volumes in this particular dataset as being close to normal values for that age. As this region is the least of all included measures affected in AD (McKhann et al., 2011), this result is in line with our expectations. We conclude that due to its low measurement error, icobrain dm could be of added value to the clinical diagnostic practice of AD patients. In future studies the performance of the measures to diagnose (very) early stages of AD as well as to distinguish between different dementia illnesses should be further investigated.

CRediT authorship contribution statement

Hanne Struyfs: Formal analysis, Investigation, Writing - original draft. Diana Maria Sima: Methodology, Software, Validation, Formal analysis, Writing - original draft, Writing - review & editing, Supervision. Melissa Wittens: Resources, Data curation, Validation, Writing - review & editing. Annemie Ribbens: Methodology, Project administration, Funding acquisition. Nuno Pedrosa de Barros: Methodology, Software, Validation, Writing - review & editing. Thanh Vân Phan: Methodology, Software, Validation, Writing - review & editing. Maria Ines Ferraz Meyer: Resources, Data curation, Software. Lene Claes: Resources, Data curation. Ellis Niemantsverdriet: Resources, Data curation. Sebastiaan Engelborghs: Writing - review & editing, Supervision, Funding acquisition. Wim Van Hecke: Conceptualization, Funding acquisition. Dirk Smeets: Conceptualization, Supervision, Project administration.

Declaration of Competing Interest

The following authors are employed (or have been employed at the time of performing the work relevant for this paper) by icometrix: Hanne Struyfs, Diana M. Sima, Annemie Ribbens, Nuno Pedrosa de Barros, Thanh Vân Phan, Lene Claes, Maria Ines Ferraz Meyer, Wim Van Hecke, Dirk Smeets. Melissa Wittens and Ellis Niemantsverdriet have no competing interests. Sebastiaan Engelborghs has received unrestricted research grants from Janssen Pharmaceutica NV and ADx Neurosciences (paid to institution).
  37 in total

1.  The diagnosis of mild cognitive impairment due to Alzheimer's disease: recommendations from the National Institute on Aging-Alzheimer's Association workgroups on diagnostic guidelines for Alzheimer's disease.

Authors:  Marilyn S Albert; Steven T DeKosky; Dennis Dickson; Bruno Dubois; Howard H Feldman; Nick C Fox; Anthony Gamst; David M Holtzman; William J Jagust; Ronald C Petersen; Peter J Snyder; Maria C Carrillo; Bill Thies; Creighton H Phelps
Journal:  Alzheimers Dement       Date:  2011-04-21       Impact factor: 21.566

Review 2.  Frontotemporal lobar degeneration: a consensus on clinical diagnostic criteria.

Authors:  D Neary; J S Snowden; L Gustafson; U Passant; D Stuss; S Black; M Freedman; A Kertesz; P H Robert; M Albert; K Boone; B L Miller; J Cummings; D F Benson
Journal:  Neurology       Date:  1998-12       Impact factor: 9.910

3.  Anatomical mapping of functional activation in stereotactic coordinate space.

Authors:  A C Evans; S Marrett; P Neelin; L Collins; K Worsley; W Dai; S Milot; E Meyer; D Bub
Journal:  Neuroimage       Date:  1992-08       Impact factor: 6.556

4.  Patterns of temporal lobe atrophy in semantic dementia and Alzheimer's disease.

Authors:  D Chan; N C Fox; R I Scahill; W R Crum; J L Whitwell; G Leschziner; A M Rossor; J M Stevens; L Cipolotti; M N Rossor
Journal:  Ann Neurol       Date:  2001-04       Impact factor: 10.422

5.  Delphi definition of the EADC-ADNI Harmonized Protocol for hippocampal segmentation on magnetic resonance.

Authors:  Marina Boccardi; Martina Bocchetta; Liana G Apostolova; Josephine Barnes; George Bartzokis; Gabriele Corbetta; Charles DeCarli; Leyla deToledo-Morrell; Michael Firbank; Rossana Ganzola; Lotte Gerritsen; Wouter Henneman; Ronald J Killiany; Nikolai Malykhin; Patrizio Pasqualetti; Jens C Pruessner; Alberto Redolfi; Nicolas Robitaille; Hilkka Soininen; Daniele Tolomeo; Lei Wang; Craig Watson; Henrike Wolf; Henri Duvernoy; Simon Duchesne; Clifford R Jack; Giovanni B Frisoni
Journal:  Alzheimers Dement       Date:  2014-08-15       Impact factor: 21.566

6.  Clinical diagnosis of Alzheimer's disease: report of the NINCDS-ADRDA Work Group under the auspices of Department of Health and Human Services Task Force on Alzheimer's Disease.

Authors:  G McKhann; D Drachman; M Folstein; R Katzman; D Price; E M Stadlan
Journal:  Neurology       Date:  1984-07       Impact factor: 9.910

7.  Vascular dementia: diagnostic criteria for research studies. Report of the NINDS-AIREN International Workshop.

Authors:  G C Román; T K Tatemichi; T Erkinjuntti; J L Cummings; J C Masdeu; J H Garcia; L Amaducci; J M Orgogozo; A Brun; A Hofman
Journal:  Neurology       Date:  1993-02       Impact factor: 9.910

Review 8.  Diagnosis and management of dementia with Lewy bodies: Fourth consensus report of the DLB Consortium.

Authors:  Ian G McKeith; Bradley F Boeve; Dennis W Dickson; Glenda Halliday; John-Paul Taylor; Daniel Weintraub; Dag Aarsland; James Galvin; Johannes Attems; Clive G Ballard; Ashley Bayston; Thomas G Beach; Frédéric Blanc; Nicolaas Bohnen; Laura Bonanni; Jose Bras; Patrik Brundin; David Burn; Alice Chen-Plotkin; John E Duda; Omar El-Agnaf; Howard Feldman; Tanis J Ferman; Dominic Ffytche; Hiroshige Fujishiro; Douglas Galasko; Jennifer G Goldman; Stephen N Gomperts; Neill R Graff-Radford; Lawrence S Honig; Alex Iranzo; Kejal Kantarci; Daniel Kaufer; Walter Kukull; Virginia M Y Lee; James B Leverenz; Simon Lewis; Carol Lippa; Angela Lunde; Mario Masellis; Eliezer Masliah; Pamela McLean; Brit Mollenhauer; Thomas J Montine; Emilio Moreno; Etsuro Mori; Melissa Murray; John T O'Brien; Sotoshi Orimo; Ronald B Postuma; Shankar Ramaswamy; Owen A Ross; David P Salmon; Andrew Singleton; Angela Taylor; Alan Thomas; Pietro Tiraboschi; Jon B Toledo; John Q Trojanowski; Debby Tsuang; Zuzana Walker; Masahito Yamada; Kenji Kosaka
Journal:  Neurology       Date:  2017-06-07       Impact factor: 9.910

9.  Automatic segmentation and volumetry of multiple sclerosis brain lesions from MR images.

Authors:  Saurabh Jain; Diana M Sima; Annemie Ribbens; Melissa Cambron; Anke Maertens; Wim Van Hecke; Johan De Mey; Frederik Barkhof; Martijn D Steenwijk; Marita Daams; Frederik Maes; Sabine Van Huffel; Hugo Vrenken; Dirk Smeets
Journal:  Neuroimage Clin       Date:  2015-05-16       Impact factor: 4.881

10.  MIRIAD--Public release of a multiple time point Alzheimer's MR imaging dataset.

Authors:  Ian B Malone; David Cash; Gerard R Ridgway; David G MacManus; Sebastien Ourselin; Nick C Fox; Jonathan M Schott
Journal:  Neuroimage       Date:  2012-12-28       Impact factor: 6.556

View more
  8 in total

1.  Diagnostic performance of hippocampal volumetry in Alzheimer's disease or mild cognitive impairment: a meta-analysis.

Authors:  Ho Young Park; Chong Hyun Suh; Hwon Heo; Woo Hyun Shim; Sang Joon Kim
Journal:  Eur Radiol       Date:  2022-05-04       Impact factor: 7.034

2.  Automated quantitative MRI volumetry reports support diagnostic interpretation in dementia: a multi-rater, clinical accuracy study.

Authors:  Hugh G Pemberton; Olivia Goodkin; Ferran Prados; Ravi K Das; Sjoerd B Vos; James Moggridge; William Coath; Elizabeth Gordon; Ryan Barrett; Anne Schmitt; Hefina Whiteley-Jones; Christian Burd; Mike P Wattjes; Sven Haller; Meike W Vernooij; Lorna Harper; Nick C Fox; Ross W Paterson; Jonathan M Schott; Sotirios Bisdas; Mark White; Sebastien Ourselin; John S Thornton; Tarek A Yousry; M Jorge Cardoso; Frederik Barkhof
Journal:  Eur Radiol       Date:  2021-01-15       Impact factor: 5.315

3.  A Contrast Augmentation Approach to Improve Multi-Scanner Generalization in MRI.

Authors:  Maria Ines Meyer; Ezequiel de la Rosa; Nuno Pedrosa de Barros; Roberto Paolella; Koen Van Leemput; Diana M Sima
Journal:  Front Neurosci       Date:  2021-08-31       Impact factor: 4.677

4.  The role of hippocampal theta oscillations in working memory impairment in multiple sclerosis.

Authors:  Lars Costers; Jeroen Van Schependom; Jorne Laton; Johan Baijot; Martin Sjøgård; Vincent Wens; Xavier De Tiège; Serge Goldman; Miguel D'Haeseleer; Marie Beatrice D'hooghe; Mark Woolrich; Guy Nagels
Journal:  Hum Brain Mapp       Date:  2020-11-28       Impact factor: 5.038

5.  Structural brain dynamics across reading development: A longitudinal MRI study from kindergarten to grade 5.

Authors:  Thanh Van Phan; Diana Sima; Dirk Smeets; Pol Ghesquière; Jan Wouters; Maaike Vandermosten
Journal:  Hum Brain Mapp       Date:  2021-07-01       Impact factor: 5.038

Review 6.  Technical and clinical validation of commercial automated volumetric MRI tools for dementia diagnosis-a systematic review.

Authors:  Hugh G Pemberton; Lara A M Zaki; Olivia Goodkin; Ravi K Das; Rebecca M E Steketee; Frederik Barkhof; Meike W Vernooij
Journal:  Neuroradiology       Date:  2021-09-03       Impact factor: 2.804

Review 7.  The Use, Standardization, and Interpretation of Brain Imaging Data in Clinical Trials of Neurodegenerative Disorders.

Authors:  Adam J Schwarz
Journal:  Neurotherapeutics       Date:  2021-04-12       Impact factor: 7.620

8.  Diagnostic Performance of Automated MRI Volumetry by icobrain dm for Alzheimer's Disease in a Clinical Setting: A REMEMBER Study.

Authors:  Mandy Melissa Jane Wittens; Diana Maria Sima; Ruben Houbrechts; Annemie Ribbens; Ellis Niemantsverdriet; Erik Fransen; Christine Bastin; Florence Benoit; Bruno Bergmans; Jean-Christophe Bier; Peter Paul De Deyn; Olivier Deryck; Bernard Hanseeuw; Adrian Ivanoiu; Jean-Claude Lemper; Eric Mormont; Gaëtane Picard; Ezequiel de la Rosa; Eric Salmon; Kurt Segers; Anne Sieben; Dirk Smeets; Hanne Struyfs; Evert Thiery; Jos Tournoy; Eric Triau; Anne-Marie Vanbinst; Jan Versijpt; Maria Bjerke; Sebastiaan Engelborghs
Journal:  J Alzheimers Dis       Date:  2021       Impact factor: 4.472

  8 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.