Inadvertent pacing lead placement in the left ventricle (LV) is an uncommon complication of cardiac pacing therapy, first reported by Stillman and Richards in 1969.1 This can occur when the lead was passed through an atrial-septal defect, a patent foramen ovale (PFO),2,3 via perforation of the intraventricular septum1 or subclavian arterial insertion.4 It is associated with increased risk of thromboembolic events and increased morbidity and mortality.5,6 It can be detected by various cardiac imaging modalities including echocardiography or cardiac computed tomography (CT).5 However, incidental detection of pacing lead in the LV by low-resolution CT with myocardial perfusion imaging (MPI) has not been reported previously.Single photon emission computed tomography (SPECT) myocardial perfusion imaging (MPI) is a well-established technique for evaluation of coronary artery disease (CAD).7 To improve image quality, low-resolution CTs commonly are used for anatomical correction and artifact attenuation during SPECT MPI.8 The low-resolution, unenhanced CT images are considered low-quality, therefore, labelled by manufacturers as “non-diagnostic”. The CT portion of the MPI at many centers is used only for image fusion and attenuation correction, and their images are not reviewed or reported routinely by cardiologists.8This report describes a case of incidental detection of inadvertent placement of pacing lead in the LV by low resolution CT used during cardiac MPI. This case suggested that it is appropriate and necessary to review low-resolution CT images during cardiac MPI for incidental findings of clinically significance.
CASE REPORT
An 83-year-old male had a history of thromboembolic stroke years ago. At that time, he had a full neurological and cardiac workup including transesophageal echocardiography (TEE) which showed a small PFO. Medtronic Reveal LINQ™, a continuous cardiac monitor, was inserted for evaluation of potential cryptogenic causes of his stroke. He was lost to follow-up until he had episodes of near syncope and was evaluated in our cardiology clinic.His Reveal LINQ™ interrogation showed multiple sinus pauses of greater than 4.5 seconds without other cardiac arrhythmias. Because of his symptoms suggestive of sick sinus syndrome, a pacemaker (PM) was inserted successfully using a left subclavian venous approach. An anteroposterior (AP) chest X-ray was performed post-implantation and appeared to show a satisfactory lead position without pneumothorax (Figure 1). He was discharged the following day.
Figure 1
Post-operative portable chest X-ray; arrowheads indicate the ventricular lead.
A pre-discharge check showed normal pacemaker sensing and pacing parameters, and the patient was enrolled for remote PM monitoring. When a 15-beats run of non-sustained ventricular tachycardia (Figure 2) was detected by remote PM monitoring three months after insertion, a pharmacological SPECT/CT MPI was obtained to rule out myocardial ischemia. The MPI showed normal myocardial perfusion and LV systolic function, however, the CT imaging revealed that a pacing lead was inadvertently placed in the LV via intra-atrial septum (Figure 3). Because of this finding, he was admitted to a tertiary hospital where he underwent ventricular lead revision successfully. He suffered no apparent clinical sequelae from the inadvertent pacing lead placement in the LV.
Figure 2
Recordings of intracardiac atrial and ventricular electrogram (EGM) from pacemaker interrogation. It showed non-sustained ventricular tachycardia (middle strip) as evidenced by ventriculo-atrial (VA) dissociation (shorter V-V sensing interval than A-A sensing interval) and different QRS morphology of the intracardiac recordings on the EGM2 during tachycardia as compared to those during normal sinus rhythm (marked as VS). Abbreviations: AS: atrial sensing; EGM1: EGM from atrial lead tip-ring recording; EGM2: EGM from ventricular lead tip-ring sensing; VS: ventricular sensing.
Figure 3
CT image of pacemaker leads in the right atrium (RA), left atrium (LA), and left ventricle (LV). Abbreviation: IAS: intra-atrial septum.
DISCUSSION
Inadvertent pacing lead placement in the LV is a rare complication of cardiac pacing therapy and associated with increased risk of thromboembolic events and other complications.5,6 Therefore, during device implantation, special care should be given to patients with known history of PFO. A few simple and helpful tips that might be considered in such patients are:Intra-operatively, after the ventricular lead is placed in the superior vena cava, a soft-tipped, straight stylet should be employed and the lead advanced to the junction of right atrial (RA) and inferior vena cava junction. At that time, the straight stylet can be replaced with a pre-curved one, then the lead can be withdrawn back carefully until it crosses the tricuspid valve into right ventricle (RV). This will decrease the chance of inadvertent lead crossing of PFO into the left-sided chambers.After the lead is secured in the RV, it is necessary to review the RV lead position carefully by fluoroscopy using both antero-posterior (AP) and left anterior oblique (LAO) views, particularly in LAO view to verify its more anterior RV position.Post operatively, AP and lateral chest x-ray should be performed/ reviewed routinely to confirm appropriate RA/RV leads position. A 12-lead electrocardiogram (ECG), with and without magnet should be performed routinely to confirm appropriate ECG pattern of RV pacing.The incidental findings on low-resolution CT during cardiac MPI are frequent.9,10 Clinically significant findings, however, are relatively infrequent. The majority of cardiac MPIs are reviewed and interpreted by nuclear cardiologists. The increased use of hybrid SPECT/ CT for attenuation correction gives rise to the issue of reviewing and interpreting these CT images during cardiac MPI. Since these low-dose, low-resolution CTs are considered “non-diagnostic”, they are not routinely and readily reviewed by cardiologists.8,11–14 A review of incidental findings of cardiac CT by Earls15 suggested that all cardiac CTs should be reconstructed in the maximal field of view available and images should be reviewed adequately to detect pathological findings. This led to an important discussion regarding the role of cardiologists and radiologists in this issue.16 At this time, there are no uniform or consensus recommendations of reporting incidental findings during cardiac CT imaging. They ranged from no recommendations,11 optional reporting,12 recommended reporting,13 to mandatory reporting.14This report described, for the first time, a case of incidental detection of inadvertent placement of the pacing lead in the LV by low resolution CT used during cardiac MPI. It highlighted the importance of reviewing low-resolution CT imaging during cardiac MPI to detect findings of clinically significance, such as the detection of inadvertent pacing lead placement in the LV. Special attention should be given for device implantation in patients, particularly with known history of PFO to avoid inadvertent lead placement in the LV through PFO.
Authors: Robert C Hendel; James R Corbett; S James Cullom; E Gordon DePuey; Ernest V Garcia; Timothy M Bateman Journal: J Nucl Cardiol Date: 2002 Jan-Feb Impact factor: 5.952
Authors: Matthias R Schulze; Ralf Ostermaier; Yvonne Franke; Klaus Matschke; Martin U Braun; Ruth H Strasser Journal: Circulation Date: 2005-12-20 Impact factor: 29.690
Authors: Vasken Dilsizian; Stephen L Bacharach; Rob S Beanlands; Steven R Bergmann; Dominique Delbeke; Sharmila Dorbala; Robert J Gropler; Juhani Knuuti; Heinrich R Schelbert; Mark I Travin Journal: J Nucl Cardiol Date: 2016-07-08 Impact factor: 5.952
Authors: Peter L Tilkemeier; Jamieson Bourque; Rami Doukky; Rupa Sanghani; Richard L Weinberg Journal: J Nucl Cardiol Date: 2017-09-15 Impact factor: 5.952