Comparative evaluation of the precision of cone-beam computed tomography and surgical intervention in the determination of periodontal bone defects: A clinicoradiographic study.

Unnati Pitale1, Hitesh Mankad1, Rohit Pandey1, Pritish Chandra Pal1, Shikha Dhakad2, Antush Mittal1.   

Abstract

CONTEXT: Major limitations of conventional radiography are overlapping and lack of 3D information. Surgical exposure, though being able to provide accurate information, provides very little time to plan-out the type of periodontal regeneration required during surgery. Cone Beam Computed Tomography (CBCT) has emerged as a feasible tool and found to be accurate. Unfortunately, in-vivo studies are still scarce. AIM: Aim of the present study was to assess the efficacy of CBCT in the detection of periodontal bony defects while determining its quantitative precision in the measurement of alveolar bone height as against the open flap debridement (OFD) procedure which is set as the gold standard. SETTING AND
DESIGN: Present study is a cross-sectional study.
MATERIALS AND METHODS: The present study includes patients with Chronic Periodontitis indicated for periodontal surgeries. Bone defects were measured with the help of CBCT and with William's periodontal probe during surgical intervention and compared. STATISTICAL ANALYSIS: Measurements were compared with Student's t-test; unpaired t-test & correlation were tested with Pearson's correlation coefficient test. P < 0.05 was considered statistically significant.
RESULTS: The mean CBCT & surgical value of palatal/lingual & distal sites of anterior teeth showed statistically significant difference (P = 0.001). All the values for posterior teeth were statistically non-significant.
CONCLUSION: Statistically CBCT & clinical measurement with OFD have similar potential of accuracy to access the bony topography but CBCT provides good accessibility to visualize the sites which are difficult to access during surgical interventions like palatal sites & the distal sites of the posterior teeth. Copyright:
© 2020 Journal of Indian Society of Periodontology.

Entities:  

Keywords:  Alveolar bone height; UNC 15 periodontal probe; chronic periodontitis; cone-beam computed tomography; periodontal bone defects

Year:  2020        PMID: 32189840      PMCID: PMC7069118          DOI: 10.4103/jisp.jisp_118_19

Source DB:  PubMed          Journal:  J Indian Soc Periodontol        ISSN: 0972-124X


INTRODUCTION

Periodontal disease causes attachment loss, alveolar bone loss, and consequent tooth mobility.[12] Radiographs play an important role in diagnosis because they can reveal the amount and type of damage. One major limitation of conventional radiography is the presence of considerable overlapping of the anatomical structures and lack of three-dimensional information.[345] In a plethora of situations, this becomes an actual hindrance in distinction between the buccal and palatal/lingual cortical aspects, making it difficult to obtain an effective evaluation, especially the crater defects and furcation involvements.[6789] This is most commonly observed when periapical, interproximal, and panoramic radiographs are used to evaluate the alveolar bone architecture. Thus, precision of the defect or hidden pathology cannot be ascertained, for which surgical exposure has been the only means to accurately detect the extent of the bone loss, as well as to evaluate the gain in bone height after a definitive periodontal treatment.[10] Surgical exposure, though provides accurate information of the area, gives very little time to plan out for periodontal regeneration required. Recently, cone-beam computed tomography (CBCT) has emerged as a feasible tool in dentistry, providing a lower cost alternative to the conventional computed tomography (CT) with high-quality images and lower radiation exposure to the patients.[1112] For the detection of smallest periodontal osseous defects, CBCT can display the image in all its three dimensions by removing the interfering anatomical structures and making it possible to evaluate each root and surrounding bone.[131415161718] CBCT is found to be as accurate as direct measurements using a periodontal probe and as reliable as intraoral periapical radiograph for the detection of interproximal areas, which are considered to be the highest of the resolution till date. Recent in vitro studies have shown better precision in the evaluation of bony changes associated with periodontal disease using CBCT when compared to the conventional and digital radiography,[1920] but there are scarcity of in vivo studies. Therefore, the aim of the present study was to assess the efficacy of CBCT in detection of periodontal bony defects in chronic periodontitis cases and compare it with the open flap debridement procedure which is the gold standard for the evaluation of bony topography.

MATERIALS AND METHODS

This study was designed as a cross-sectional study and patients were selected from the outpatient department. In inclusion criteria, the individuals between 25 and 55 years of age with moderate-to-severe chronic periodontitis which were indicated for surgery and with no history of systemic disease and tobacco use were included. All the individuals were clinically and radiographically evaluated as a routine diagnostic procedure with the exclusion criteria of pregnant and lactating females. In the present study, 12 patients with 12 teeth (6 anterior and 6 posterior) each were selected, and thus, a total of 144 teeth with chronic periodontitis were selected. For anterior teeth, 12 patients were considered with six teeth and eight sites as shown in Figure 1. The number of sites calculated was 12 patients × 6 teeth × 8 sites = 576. For posterior teeth, 12 patients were considered with six teeth and 12 sites as shown in Figure 2. The number of sites calculated was 12 patients × 6 teeth × 12 sites = 864. Hence, the total sites calculated as: anterior sites (576) + posterior sites (864) = 1440, which are shown in Tables 1 and 2. Presurgically, each patient received an initial full-mouth scaling and root planing, oral hygiene, and plaque control instructions. Once the patients showed plaque proficiency up to the mark, then surgeries were initiated. CBCT images were acquired for the same area of interest in a single 360° rotation around the head of the patient. The standard image-acquisition time was 36 s in which the patient received pulsed radiation. All images were then stored in Digital Imaging Communications in Medicine file format. Following image acquisition, measurements were made using the proprietary NNT software (Version 5.4, cefla dental group, newtom GiANO, Italy) integrated with Newtom-Giano CBCT equipment which has multiple adjustable field of views. The data were acquired with an image size of 400 × 400 × 400, voltage of 90 kV, voxel size of 400 μm, and current of 10 mA. The examiner underwent training for performing the measurements. The CBCT measurements were performed with the same pattern as clinical measurements of sites. In case of horizontal defects, distance between cementoenamel junction (CEJ) and alveolar crest was measured, and in case of vertical defects, the distance between CEJ and the base of the defect was measured. The viewer software accuracy was 0.2 mm. Informed consent was signed by every subject who fulfilled the inclusion criteria and agreed to participate voluntarily, after thorough explanation of the nature, risk, and benefit of the clinical investigations and associated procedure. The University Ethical Committee approved the consent form and experimental protocol (IEC No-IEC/MDCRC/2013–14/0106).
Figure 1

Sites for measurement of anterior teeth

Figure 2

Sites for measurement of posterior teeth

Table 1

Eight sites selected in anterior teeth for evaluation

12345678
BBDBMDMPPDPM
BuccalBuccal/DistalBuccal/MesialDistalMesialPalatalPalatal/DistalPalatal/Mesial
Buccal: Mean of B, BD, BMDistalMesialPalatal: Mean of P, PD, PM

In anterior teeth, 4 mean values were considered for statistical analysis from measurement of 8 sites

Table 2

Twelve sites selected in posterior teeth for evaluation

123456789101112
BBDBMDDBDPMMBMPLLDLM
BuccalBuccal/DistalBuccal/MesialDistalDistal/BuccalDistal/PalatalMesialMesial/BuccalMesial/PalatalLingualLingual/DistalLingual/Mesial
Buccal: Mean of B, BD, BM)Distal: Mean of D, DB, DPMesial: Mean of M, MB, MPLingual: Mean of L, LD, LM

In posterior teeth, 4 mean values were considered for statistical analysis from measurement of 12 sites

Sites for measurement of anterior teeth Sites for measurement of posterior teeth Eight sites selected in anterior teeth for evaluation In anterior teeth, 4 mean values were considered for statistical analysis from measurement of 8 sites Twelve sites selected in posterior teeth for evaluation In posterior teeth, 4 mean values were considered for statistical analysis from measurement of 12 sites

Surgical procedure

During the surgical intervention, periodontal osseous defects were measured with the help of UNC 15 periodontal probe to the nearest millimeter from the CEJ to the deepest level of the bony defect, which is considered as the “gold standard” in literature.[613] Assessment of depth of the bony defect was measured during the surgical procedure in the same region where CBCT measurements were evaluated [Figures 3a, b and 4a-d].
Figure 3

(a) Measurement of bony defects around the mandibular molar teeth with a UNC 15 periodontal probe with surgical intervention; (b) Cone-beam computed tomography images of bony defects around the mandibular molar teeth in coronal, sagittal, axial plane and three-dimensional reconstruction of the area of interest

Figure 4

(a and d) Measurement of bony defects around the mandibular canine teeth with a UNC 15 periodontal probe with surgical intervention; (b) Cone-beam computed tomography images of bony defects around the mandibular canine in coronal, sagittal, axial plane and three-dimensional reconstruction of the area of interest; (c) Cone-beam computed tomography images of bony defects around the mandibular canine

(a) Measurement of bony defects around the mandibular molar teeth with a UNC 15 periodontal probe with surgical intervention; (b) Cone-beam computed tomography images of bony defects around the mandibular molar teeth in coronal, sagittal, axial plane and three-dimensional reconstruction of the area of interest (a and d) Measurement of bony defects around the mandibular canine teeth with a UNC 15 periodontal probe with surgical intervention; (b) Cone-beam computed tomography images of bony defects around the mandibular canine in coronal, sagittal, axial plane and three-dimensional reconstruction of the area of interest; (c) Cone-beam computed tomography images of bony defects around the mandibular canine

Postoperative care

All the patients were instructed to rinse with 10 ml 0.2% chlorhexidine twice daily till 2 weeks postoperatively. Periodontal dressing and sutures were removed postoperatively after 7 days.

Statistical analysis

Clinical and CBCT measurements were compared in each anterior and posterior tooth with the help of Student's paired t-test. If P ≥ 0.05, the difference observed was considered statistically insignificant, and if P ≤ 0.05, it was considered significant. Pearson's correlation coefficient (z) test was used to test correlation. Unpaired t-test was used to compare the data in the same measurement method.

RESULTS

All 144 teeth sites were measured and the mean of each site was calculated which comes as four sites (mean of 8 sites) in anterior and four sites (mean of 12 sites) in posterior as shown in Tables 1 and 2 which is according to Figures 1 and 2.

Descriptive values for anterior teeth

Buccal sites – The mean CBCT value was 4.156 mm and mean surgical value was 4.195 mm. The mean difference was statistically nonsignificant (P = 0.171) [Tables 3-5 and Graph 1]
Table 3

Mean and standard deviation values of anterior teeth measured through cone-beam computed tomography and surgical intervention

Site (anterior)MethodMeanSDSE
BuccalSurgical4.19501.846850.13766
CBCT4.15561.922590.14330
Palatal/lingualSurgical4.18221.934780.14421
CBCT4.04441.988300.14820
MesialSurgical4.19822.109640.27235
CBCT4.15472.123410.27413
DistalSurgical3.52171.317170.17005
CBCT3.36671.352550.17461

SD – Standard deviation; SE – Standard error, CBCT – Cone-beam computed tomography; Surgical – Surgical intervention

Table 5

Pearson’s correlation test for all sites of anterior teeth and the significance status of sites

Site (anterior)MethodCorrelation (Z)P
BuccalSurgical0.980<0.001**
CBCT
Palatal/lingualSurgical0.964<0.001**
CBCT
MesialSurgical0.956<0.001**
CBCT
DistalSurgical0.968<0.001**
CBCT

**Highly statistically significant; P≤0.05 – Considered statistically significant, P≤0.001 – Highly statistically significant. CBCT – Cone-beam computed tomography; P – Probability value, Z – Pearson’s correlation coefficient, Surgical – Surgical intervention

Graph 1

Comparison between mean values of buccal sites of anterior teeth showing better result with surgical intervention than cone-beam computed tomography

Palatal/lingual sites – The mean CBCT value was 4.0444 mm and mean surgical value was 4.1822 mm. The mean difference was statistically significant (P = 0.001) [Tables 3, 4 and Graph 2]
Table 4

The comparison of t values for the anterior teeth by Student’s paired t-test

Site (anterior)MethodtP
BuccalSurgical1.3750.171
CBCT
Palatal/lingualSurgical3.5040.001**
CBCT
MesialSurgical0.2870.775
CBCT
DistalSurgical3.5240.001**
CBCT

**Highly statistically significant; P ≤ 0.05 – Considered statistically significant, P≤ 0.001 – Highly statistically significant. CBCT – Cone-beam computed tomography, P– Probability value, t – Student’s t-test value; Surgical – Surgical intervention

Graph 2

Comparison between mean values of lingual/palatal sites of anterior teeth also showing better result with surgical intervention than cone-beam computed tomography

Mesial sites – The mean CBCT value was 4.1547 mm and mean surgical value was 4.1982 mm. The mean difference was statistically nonsignificant (P = 0.775) [Tables 3, 4 and Graph 3]
Graph 3

Comparison between mean values of mesial sites of anterior teeth showing better result with surgical intervention than cone-beam computed tomography

Distal sites – The mean CBCT value was 3.3667 mm and mean surgical value was 3.5217 mm. The mean difference was statistically significant (P = 0.001) [Tables 3, 4 and Graph 4]. All the measurements were found to be in correlation with each other.
Graph 4

Comparison between mean values for distal sites of anterior teeth showing better result with surgical intervention than cone-beam computed tomography

Comparison between mean values of buccal sites of anterior teeth showing better result with surgical intervention than cone-beam computed tomography Comparison between mean values of lingual/palatal sites of anterior teeth also showing better result with surgical intervention than cone-beam computed tomography Comparison between mean values of mesial sites of anterior teeth showing better result with surgical intervention than cone-beam computed tomography Comparison between mean values for distal sites of anterior teeth showing better result with surgical intervention than cone-beam computed tomography Mean and standard deviation values of anterior teeth measured through cone-beam computed tomography and surgical intervention SD – Standard deviation; SE – Standard error, CBCT – Cone-beam computed tomography; Surgical – Surgical intervention The comparison of t values for the anterior teeth by Student’s paired t-test **Highly statistically significant; P ≤ 0.05 – Considered statistically significant, P≤ 0.001 – Highly statistically significant. CBCT – Cone-beam computed tomography, P– Probability value, t – Student’s t-test value; Surgical – Surgical intervention Pearson’s correlation test for all sites of anterior teeth and the significance status of sites **Highly statistically significant; P≤0.05 – Considered statistically significant, P≤0.001 – Highly statistically significant. CBCT – Cone-beam computed tomography; P – Probability value, Z – Pearson’s correlation coefficient, Surgical – Surgical intervention

Descriptive values for posterior teeth

Buccal sites – The mean CBCT value was 6.0656 mm and mean surgical value was 5.9833 mm. The mean difference was statistically nonsignificant (P = 0.150) [Tables 6-8 and Graph 5]
Table 6

Mean and standard deviation values of posterior teeth through cone-beam computed tomography and surgical intervention

Site (posterior)MethodMeanSDSE
BuccalSurgical5.98331.832880.13657
CBCT6.06561.786520.13316
Palatal/lingualSurgical6.15561.729860.12894
CBCT6.11111.739320.12964
MesialSurgical5.64441.632840.12170
CBCT5.69561.57740.11757
DistalSurgical6.03892.028740.15121
CBCT6.08671.977340.14738

SD – Standard deviation; SE – Standard error; CBCT – Cone-beam computed tomography; Surgical – Surgical intervention

Table 8

Pearson correlation test for all sites of posterior teeth and the significance status of sites

Site (posterior)MethodCorrelation (Z)P
BuccalSurgical0.911<0.001**
CBCT
Palatal/lingualSurgical0.915<0.001**
CBCT
MesialSurgical0.947<0.001**
CBCT
DistalSurgical0.937<0.001**
CBCT

**Highly statistically significant; P≤0.05 – Considered statistically significant, P ≤0.001 – Highly statistically significant. CBCT – Cone-beam computed tomography; P– Probability value, Z – Pearson correlation coefficient, Surgical – Surgical intervention

Graph 5

Comparison between mean values for buccal sites of posterior teeth showing better result with cone-beam computed tomography than surgical intervention

Palatal/lingual sites – The mean CBCT value was 6.1111 mm and mean surgical value was 6.1556 mm. The mean difference was statistically nonsignificant (P = 0.405) [Tables 6, 7 and Graph 6]
Table 7

The t values for the posterior teeth by Student’s paired t-test and significance status of sites byPvalue

Site (posterior)MethodtP
BuccalSurgical1.4440.150
CBCT
Palatal/lingualSurgical0.8350.405
CBCT
MesialSurgical1.3100.192
CBCT
DistalSurgical0.8970.731
CBCT

P≤0.05 – Considered statistically significant. CBCT – Cone-beam computed tomography;P – Probability value; t – Student t-test value; Surgical – Surgical intervention

Graph 6

Comparison between mean values of palatal/lingual sites of posterior teeth showing better result with surgical intervention than cone-beam computed tomography

Mesial sites – The mean CBCT value was 5.6956 mm and mean surgical value was 5.6444 mm. The mean difference was statistically nonsignificant (P = 0.192) [Tables 6, 7 and Graph 7]
Graph 7

Comparison between mean values for mesial sites of posterior teeth showing better result with cone-beam computed tomography than surgical intervention

Distal sites – The mean CBCT value was 6.0867 mm and mean surgical value was 6.0389 mm. The mean difference was statistically nonsignificant (P = 0.731) [Tables 6, 7 and Graph 8].
Graph 8

Comparison between mean values of distal sites of posterior teeth showing better result with cone-beam computed tomography than surgical intervention

Comparison between mean values for buccal sites of posterior teeth showing better result with cone-beam computed tomography than surgical intervention Comparison between mean values of palatal/lingual sites of posterior teeth showing better result with surgical intervention than cone-beam computed tomography Comparison between mean values for mesial sites of posterior teeth showing better result with cone-beam computed tomography than surgical intervention Comparison between mean values of distal sites of posterior teeth showing better result with cone-beam computed tomography than surgical intervention Mean and standard deviation values of posterior teeth through cone-beam computed tomography and surgical intervention SD – Standard deviation; SE – Standard error; CBCT – Cone-beam computed tomography; Surgical – Surgical intervention The t values for the posterior teeth by Student’s paired t-test and significance status of sites byPvalue P≤0.05 – Considered statistically significant. CBCT – Cone-beam computed tomography;P – Probability value; t – Student t-test value; Surgical – Surgical intervention Pearson correlation test for all sites of posterior teeth and the significance status of sites **Highly statistically significant; P≤0.05 – Considered statistically significant, P ≤0.001 – Highly statistically significant. CBCT – Cone-beam computed tomography; P– Probability value, Z – Pearson correlation coefficient, Surgical – Surgical intervention

Anterior teeth

CBCT measurements – The mean difference for buccal and palatal/lingual sites was 0.446 mm which was nonsignificant. The mean difference for mesial and distal sites was 0.829 mm which was nonsignificant (P > 0.05) [Table 9]
Table 9

Unpaired t-test for the surgical and cone-beam computed tomography measurements of all teeth and its significance status

MethodSitetP
Surgical (posterior)Mesial2.3020.043**
Distal
CBCT (posterior)Mesial1.7890.075
Distal
Surgical (posterior)Buccal0.9170.360
Palatal
CBCT (posterior)Buccal0.2450.806
Palatal
Surgical (anterior)Buccal0.7670.444
Palatal
CBCT (anterior)Buccal0.7640.446
Palatal
Surgical (anterior)Mesial0.3550.723
Distal
CBCT (anterior)Mesial0.1280.829
Distal

**Statistically significant; P≤0.05 – Considered statistically significant. CBCT – Cone-beam computed tomography, P – Probability value; t – Unpaired t-test value, Surgical – Surgical intervention

Surgical measurement – The mean difference for buccal and palatal/lingual sites was 0.444 mm which was nonsignificant. The mean difference for mesial and distal sites was 0.723 mm which was nonsignificant (P > 0.05). Unpaired t-test for the surgical and cone-beam computed tomography measurements of all teeth and its significance status **Statistically significant; P≤0.05 – Considered statistically significant. CBCT – Cone-beam computed tomography, P – Probability value; t – Unpaired t-test value, Surgical – Surgical intervention

Posterior teeth

CBCT measurement – The mean difference for buccal and palatal/lingual sites was 0.806 mm which was nonsignificant. The mean difference for mesial and distal sites was 0.446 mm which was nonsignificant (P > 0.05) [Table 9] Surgical measurement – The mean difference for buccal and palatal/lingual teeth was 0.360 mm which was nonsignificant. The mean difference for mesial and distal sites was 0.043 mm which was significant (P ≤ 0.05).

DISCUSSION

CBCT has opened a new vista in maxillofacial imaging facilitating transition of diagnostics from two-dimensional to three-dimensional imaging and having the potential to expand the role of dental imaging from mere diagnosis to image guidance of surgical procedures. Numerous studies have validated the use of CBCT for planning of implants, in diagnosis of embedded teeth, in orthodontics and numerous pathologies;[1921222324252627282930313233343536373839] however, very few have analyzed the need of CBCT in periodontal diagnosis, regeneration, and treatment planning. Recently, in vivo studies have been published concerning periodontal bone height measurements using CBCT. This leaves only a few reports on the accuracy of CBCT in measuring periodontal defects while almost none of them concerning horizontal bone defects.[781536] The main advantages of CBCT include good accessibility and easy handling in addition to a real-size data set.[4041] Although Vandenberghe et al.[19] suggested CBCT to be used only for relatively complex periodontal treatment planning and potential use of dental implants, this imaging modality may provide a new tool for diagnosis of various bony defects, interdisciplinary approach to treatment planning such as in case of periodontal–endodontic lesions and in the detection of dehiscence and fenestration defects before initializing the orthodontic treatments. This present study had given us the clear idea about the precision of CBCT measurement and its status. The analysis of anterior teeth concluded that the mean difference for both the methods made for mesial and buccal surfaces was more precise in surgical intervention than CBCT analysis [Table 3], but it was statistically nonsignificant with P = 0.171 and 0.775, respectively. For palatal/lingual surface (P = 0.001) and distal surface (P = 0.001), the mean difference for both the methods was found statistically significant [Table 4]. The said results were found to be correlated with the results obtained in the study conducted by Mol and Balasundaram,[40] who found less accuracy in the measurement of buccal bone defects in the anterior region of mandible in comparison with the posterior region on images with the New Tom 9000 Scanner. Furthermore, using helical CT, Fuhrmann et al.[8] found that bone defects with wall thickness <0.2 mm were not apparent on tomographic imaging. For posterior teeth, the mean difference of the buccal surface (P = 0.150) and palatal/lingual surface (P = 0.405) was statistically nonsignificant. Mesial and distal surface (P = 0.192 and P = 0.731) was also found statistically nonsignificant. The results obtained were in concordance with the results of the previous studies conducted by de Faria et al.,[7] Grimard et al.,[10] and Feijo et al.[18] The difference in the diagnostic accuracy of CBCT in the anterior and posterior teeth is likely the result of the difference in the morphology of the periodontal bone in each region. The lingual plates are considerably thinner in the anterior region and tapers toward the alveolar crest than in the posterior teeth. A thinner bone plate has low image resolution decreasing the precision of linear measurements in case of anterior teeth. Similar to the present study, Misch et al.[16] showed that CBCT analysis was as accurate as direct measurement using a periodontal probe during surgical intervention. However, within same analytical parameters, when unpaired t-test was applied, it was found that there was a significant difference between the mesial and distal sites in posterior teeth in the measurements made during surgical intervention. This difference can be explained by the fact that the distal site accessibility is difficult while performing periodontal surgeries, especially in case of molars while instrumentation as well as for measurements attempted with the help of standard periodontal probing. On the contrary, CBCT provides good accessibility to visualize the sites which are difficult to access during surgical interventions, clinically including the distal sites of posterior teeth. Similarly, Feijo et al.[18] showed that a clinical measurement gives accuracy up to 1 mm whereas CBCT measurements allowed an accuracy of up to three decimal places. Ideally, 0.5 mm discrepancy between clinical and radiographical estimated bone levels is accepted. Smaller or larger errors in locating the CEJ and crestal levels can, respectively, lead to over- and under-estimation of disease. Because of the limitations of surgical re-entry to evaluate regenerative outcomes, different radiographic techniques such as intraoral radiographs, digital subtraction radiography, and computer-assisted densitometric image analysis have been used as surrogate methods.[42] However, these techniques are difficult to standardize from two different time points and having limitation to assess three-dimensional morphologic changes. Hence, this present study gave us a view that the results of surgical intervention and CBCT give similar value and prove the accuracy and reliability of CBCT analysis. Data from CBCT could be helpful in optimization of surgical treatment plan. Involvement of furcation in case of posteriors can be determined accurately with CBCT. In nutshell, CBCT provides better evaluation of the bony topography, which gives precision in the type and depth of the defect pre- and post-operatively after regeneration. Surgical exposure, though being able to provide accurate information of the area, provides very little time for the surgeon to plan the treatment protocol for suitable regeneration required during surgery.

CONCLUSION

This present study compares CBCT analysis and surgical intervention method and concluded having similar potential of accuracy to access the bony topography in both the measurement methods. However, the diagnostic accuracy of CBCT is lesser for anterior teeth as compared to the posterior teeth due to thinner buccal/lingual cortical plates. The quality of the CBCT image slices is insufficient to resolve the alveolar crest reliably in the anterior region.[41] However, CBCT having better accessibility and precision to visualize any area of interest, which is difficult to access during surgical intervention like distal surface of posterior teeth. The bone density/volume can be detected with supremacy by CBCT. It gives benefit to plan the treatment for better prognosis and its outcome, by avoiding surgical re-entry to evaluate postoperative regeneration precisely and help to improve futuristic approach.

Financial support and sponsorship

Nil.

Conflicts of interest

There are no conflicts of interest.
  39 in total

Review 1.  Is there a role for the use of volumetric cone beam computed tomography in periodontics?

Authors:  A H du Bois; B Kardachi; P M Bartold
Journal:  Aust Dent J       Date:  2012-03       Impact factor: 2.291

2.  Essentials of maxillofacial cone beam computed tomography.

Authors:  William C Scarfe; Allan G Farman; Martin D Levin; David Gane
Journal:  Alpha Omegan       Date:  2010-06

3.  Diagnostic validity (or acuity) of 2D CCD versus 3D CBCT-images for assessing periodontal breakdown.

Authors:  Bart Vandenberghe; Reinhilde Jacobs; Jie Yang
Journal:  Oral Surg Oral Med Oral Pathol Oral Radiol Endod       Date:  2007-07-05

4.  Comparison of different methods of assessing alveolar ridge dimensions prior to dental implant placement.

Authors:  Lung-Cheng Chen; Tord Lundgren; Hadar Hallström; Fabrice Cherel
Journal:  J Periodontol       Date:  2008-03       Impact factor: 6.993

5.  Selection of the most accurate method of conventional radiography for the assessment of periodontal osseous destruction.

Authors:  E A Pepelassi; A Diamanti-Kipioti
Journal:  J Clin Periodontol       Date:  1997-08       Impact factor: 8.728

6.  Accuracy assessment of computer-assisted flapless implant placement in partial edentulism.

Authors:  N Van Assche; D van Steenberghe; M Quirynen; R Jacobs
Journal:  J Clin Periodontol       Date:  2010-04       Impact factor: 8.728

7.  Computerized transverse axial scanning (tomography). 1. Description of system.

Authors:  G N Hounsfield
Journal:  Br J Radiol       Date:  1973-12       Impact factor: 3.039

8.  Detection of periodontal bone loss using cone beam CT and intraoral radiography.

Authors:  K de Faria Vasconcelos; K M Evangelista; C D Rodrigues; C Estrela; T O de Sousa; M A G Silva
Journal:  Dentomaxillofac Radiol       Date:  2012-01       Impact factor: 2.419

9.  Early re-entry procedure. Part II. A five year histologic evaluation.

Authors:  B Langer; D A Gelb; D J Krutchkoff
Journal:  J Periodontol       Date:  1981-03       Impact factor: 6.993

10.  A new method to assess and measure palatal masticatory mucosa by cone-beam computerized tomography.

Authors:  Maurício Barriviera; Wagner Rodrigues Duarte; Alessandro Lourenço Januário; Jorge Faber; Ana Cristina Barreto Bezerra
Journal:  J Clin Periodontol       Date:  2009-07       Impact factor: 8.728