| Literature DB >> 32161995 |
Victor Schmidbauer1, Gudrun Geisl1, Mariana Cardoso Diogo1, Suren Jengojan1, Vsevolod Perepelov1, Michael Weber1, Katharina Goeral2, Florian Lindenlaub1, Katrin Klebermass-Schrehof2, Angelika Berger2, Daniela Prayer1, Gregor Kasprian3.
Abstract
PURPOSE: The purpose of this study was to assess the diagnostic accuracy of T1-weighted and T2-weighted contrasts generated by the MR data postprocessing software SyMRI (Synthetic MR AB, Linköping, Sweden) for neonatal brain imaging.Entities:
Keywords: Magnetic resonance imaging; Newborn; Sensitivity; Software; Specificity
Year: 2020 PMID: 32161995 PMCID: PMC8211598 DOI: 10.1007/s00062-020-00894-2
Source DB: PubMed Journal: Clin Neuroradiol ISSN: 1869-1439 Impact factor: 3.649
Demographics and clinical characteristics
| Neonates | |
|---|---|
| Characteristics/demography | |
| Male/female | 16/16 |
| Term born/preterm | 8/24 |
| Gestational age (weeks)a | 27 + 5 (23 + 3–41 + 5) |
| Inconspicuousb | |
| Hemorrhageb,c | |
| Expired infarctionb | |
| HIEb | |
| PVLb | |
| Venous vessel malformationb | |
HIE hypoxic-ischemic encephalopathy, PVL periventricular leukomalacia
aData represented as median and range
bData represented as total number
cIncluding hyperacute, acute, subacute, and chronic intraventricular, cortical, subcortical, parenchymal, and subdural hemorrhage
MRI sequences and technical features
| Sequence | Plane | Matrix (slices) | FOV (mm) | Voxel size (mm) | TR (ms) | TE (ms) | AT |
|---|---|---|---|---|---|---|---|
| T1 SE | Axial | 144 × 115 × 30 | 120 × 120 × 90 | 0.83 × 1.05 × 3 | 400 | 15 | 3:07 |
| T2 TSE | Axial | 128 × 113 × 34 | 120 × 120 × 102 | 0.94 × 1.06 × 3 | 3000 | 140 | 1:48 |
| T2 TSE | Coronal | 116 × 103 × 36 | 110 × 110 × 108 | 0.94 × 1.06 × 3 | 3000 | 140 | 1:48 |
| T2 TSE | Sagittal | 128 × 113 × 36 | 120 × 120 × 108 | 0.94 × 1.06 × 3 | 3000 | 140 | 1:48 |
| DWI | Axial | 176 × 170 × 28 | 200 × 200 × 92 | 1.14 × 1.15 × 3 | 4066 | 90 | 1:34 |
| SWI | Axial | 200 × 138 × 90 | 170 × 139 × 90 | 0.85 × 1 × 2 | 51 | 12 | 3:35 |
| T1 (3D) | Sagittal | 160 × 160 × 99 | 120 × 120 × 99 | 0.75 × 0.75 × 2 | 25 | 7.6 | 3:46 |
| MDME | Axial | 224 × 159 × 22 | 200 × 165 × 109 | 0.9 × 1 × 4 | 3309 | 13 | 5:24 |
AT acquisition time, DWI diffusion-weighted imaging, FOV field-of-view, MDME multi-dynamic multi-echo sequence, SE spin echo, SWI susceptibility-weighted imaging, TE echo time, TR repetition time, TSE turbo spin echo
Fig. 1Inconspicuous case: brain of a former preterm infant (24 + 4 weeks). The left column (a, b) shows conventionally acquired T1-weighted (a) and T2-weighted contrasts (b). The right column (c, d) shows T1-weighted (c) and T2-weighted (d) contrasts generated by the MR data post-processing software SyMRI
Fig. 2Selection of pathological cases. The upper row shows conventional MR contrasts (a–d). The bottom row shows SyMRI-generated MR contrasts (e–h). a, e Hypoxic ischemic encephalopathy in a term-born infant (39 + 4 weeks) (T1-weighted contrast). b, f Former preterm neonate (34 + 6 weeks) with multifocal intracranial hemorrhage (T1-weighted contrast). c, g Term-born infant (41 + 5 weeks) with expired infarction in the right caudate nucleus (T2-weighted contrast). d, h Former preterm neonate (32 + 4 weeks) with infarction of the left hemisphere (T2-weighted contrast)
Group allocation
| Group Aa | Group Bb | ||
|---|---|---|---|
| Condition/pathology | Condition/pathology | ||
| Inconspicuousc | Inconspicuousc | ||
| Hemorrhagec | Hemorrhagec | ||
| Expired infarctionc | Expired infarctionc | ||
| HIEc | HIEc | ||
| Venous vessel malformationc | PVLc | ||
HIE hypoxic ischemic encephalopathy, PVL periventricular leukomalacia
aAssessed by raters 1 and 3 on the basis of conventionally acquired T1- and T2-weighted contrasts. Assessed by raters 2 and 4 based on SyMRI-generated T1- and T2-weighted contrasts
bAssessed by raters 1 and 3 on the basis of SyMRI-generated T1- and T2-weighted contrasts. Assessed by raters 2 and 4 based on conventionally acquired T1- and T2-weighted contrasts
cData represented as total number
Fig. 3Allocation of the groups to the corresponding raters
Incorrectly identified cases
| Missed | Rater | Modality | Suspected | Group |
|---|---|---|---|---|
| Inconspicuous | 3a | Conventional MRI | HIE | A |
| Expired infarction | 4a | SyMRI | Inconspicuous | A |
| PVL | 3a | SyMRI | Inconspicuous | B |
| Inconspicuous | 4a | Conventional MRI | HIE | B |
| Expired infarction | 1b | SyMRI | Inconspicuous | B |
| 2b | Conventional MRI | |||
| 3a | SyMRI | |||
| 4a | Conventional MRI |
HIE hypoxic ischemic encephalopathy, PVL periventricular leukomalacia
aNovice rater
bExperienced rater
Fig. 4a Sensitivity and specificity of conventional MRI and SyMRI on the basis of the assessment of experienced raters. b Sensitivity and specificity of conventional MRI and SyMRI on the basis of the assessment of novice raters. c Sensitivity and specificity of conventional MRI and SyMRI on the basis of the assessment of all raters