Tomoko Yamashita1, Kohei Ozawa2, Kazuyoshi Gamada1. 1. Graduate School of Medical Technology and Health Welfare Sciences, Hiroshima International University: 555-36 Kurose-Gakuendai, Higashihiroshima, Hiroshima 739-2695, Japan. 2. Mirai Links, Japan.
A valid and reliable method of measuring gliding velocity on ultrasonographic images has
not yet been established. When tissue adhesion limits gliding ability, it may contribute to
a limited range of motion (ROM) or joint contracture1,
2), which can be clearly observed
qualitatively on ultrasonography3). Chen et
al.4) manually measured sliding of the
transverse abdominis by tracing the most lateral point of the deep fascia surrounding the
transverse abdominis, and found that intra-rater reliability was 0.95. In a Japanese
journal, Ichikawa et al.5) manually
measured the distance between two arbitrary points on the deep fascia of the vastus
lateralis during quasi-static knee flexion, and reported good intra-rater and inter-rater
reliability (ICC 0.95–0.98) both in the superficial zone (i.e. subcutaneous tissue side) and
the deep zone (i.e. vastus intermedius side) of the deep fascia. Both studies traced only
one point on each ultrasound image. Tissue Velocity Imaging (TVI), an ultrasound-based
technique with Doppler technology used to analyze cardiac function, can be used to measure
the movement of the musculoskeletal system6, 7). However, the measurement error of TVI
ranged from 31% up to 313% when the phantom peak velocity was 0.03 cm/s8). Moreover, the software for TVI was specially created for
measuring cardiac motion and there is no program available to measure muscular movement.
Therefore, there is an urgent need for a more valid and reliable quantification method of
measuring muscle gliding including; (1) defining arbitrary regions of interest (ROI’s) in
the muscle; (2) tracing a greater number of measurement points in the ROI; (3) calculating
the average or representative value; (4) quantifying the movement velocity of each ROI; (5)
quantifying the relative velocity between ROI’s; (6) visualization of the movement; and (7)
comparing longitudinal data by matching the contours of bones to assure the ultrasound image
observations are the same size. A highly specialized software program is required to achieve
these needs.Echolizer (GLAB Co., Inc., Japan) is a commercial software package that uses an
optical-flow alorithm to measure the movement velocity of muscle bundles on ultrasound
images. Using Echolizer software to determine the movement velocity of muscles and other
soft tissues, the clinical outcomes of physical therapy or hydrorelease9, 10) interventions
could be quantified. Measurement errors can be eliminated by averaging the velocity of
several measurement points within the region of interest (ROI). Abnormal values can be
eliminated by including only those within 2.0 standard deviations of the velocity values. It
is not expected that human factors would cause any measurement errors because ROI’s are
manually defined in a reproducible manner and velocity measurements are fully automated.
Determining the optimal data acquisition method as well as the optimal analysis method such
as the appropriate settings for the video frame rate, image qualities such as brightness and
contrast, effects of the size and position of the ROI is expected to further improve the
validity and reliability of Echolizer measurements. No studies to date have validated the
optical-flow algorithm for velocity measurements of skeletal muscles and, therefore, this
study would provide novel knowledge for musculoskeletal research community. The purposes of
this study were to determine: (1) the validity and reliability of Echolizer software in
measuring the velocity of the muscle bundles; and (2) the optimal image quality and the size
and location of ROI’s to optimize the validity.The optical-flow algorithm calculates velocity of an object and present arrows to present
the velocity and orientation of the object. The Ferneback method calculates movements of all
pixels by comparing sequential images11, 12). This method can be used in automatic
driving technology, which has an absolute error in collision time estimation within 0.5
seconds13). The relative error for blood
flow velocity estimation was within 4%14).
The average relative error for tendon motion tracking on ultrasound image sequences using
the optical flow algorithm was less than 12.1%. Williamson et al.15) reported that tracking changes in liver morphology on
ultrasound images using the optical-flow algorithm demonstrated an accuracy (mean ± standard
deviation) of 0.74 ± 1.03 mm. Tabata et al.16) examined the accuracy of tagged magnetic resonance optical-flow
velocity measurement and found that the relative error was less than 1%. Therefore, the
validity and reliability of Echolizer are considered high.The hypothesis of this study was that, in locations where the muscle bundle can be clearly
seen on ultrasound images, the software would demonstrate high validity and reliability in
measuring muscle bundle movement velocity. This software is expected to allow a wider range
of measurements of any moving tissues without involving human errors in either the detection
or tracking of the specific tissue being examined. To confirm or deny this hypothesis, we
performed a validation study using pork meat. If a method of measuring muscle glide with
high validity and reliability can be established, it may contribute to measuring the
effectiveness of physical therapy interventions on soft tissue movement.
MATERIALS AND METHODS
This study is a cross-sectional study. Since the two pieces of pork meat used as materials
were obtained at a local butcher shop, review of the study protocol by the ethical committee
was not required under ethical guidelines for the use of animals in research.Two pieces of pork thigh meat measuring 3 × 20 × 15 cm each were obtained at a local
butcher shop. The pork meat was vacuum-packed and securely sealed using adhesive tape. Each
vacuum-packed bag was sandwiched between two 28 × 20 cm wooden photo frames (wooden frame)
and fixed by screws. The two frames were then stacked on top of each other. The inside of
the bags and the layer between the two bags were filled with ultrasonic gel to ensure the
muscle fibers of the pork meat would be visible on ultrasonography. On the experimental
bench, the two meat bags in wooden frames were placed between two parallel wooden planks
that formed rails for the wooden frames to glide between (Fig. 1a). A 1 kg weight was placed on the top meat to apply pressure that helped maintain
close contact between the two bags of meat. A wooden block was attached to the bench in
front of the lower meat frame in order to stabilize the meat and prevent wobbling during
towing. The upper meat frame was connected by a wire to the universal testing machine (UTM)
(AG-10TE, Shimadzu Corp., Japan). The wire load strength was 20 kg. A 1 kg weight was hung
from the table by a wire connected to the opposite end of the upper meat frame in order to
maintain constant tension between the UTM and the upper meat frame (Fig. 1b). To attach the ultrasonic probe (Linear Probe L11-3, Konica
Minolta Inc., Japan), a rectangular hole was created in the workbench under the lower meat
frame (Fig. 1c). The probe was inserted from the
bottom and fixed using adhesive tape so that contact with the lower side of the meat bag was
maintained. A high-velocity video (EX-FH25, Casio Computer Co., Ltd., Japan) was installed
over the meat frames and an accelerometer was attached to the upper meat frame to monitor
movement of the meat frames. A ruler was placed on the workbench next to the pork so that
the scale of movement could be recorded by the high-velocity camera. Room temperature was
set and maintained at 24 °C. Using the UTM, the upper meat was pulled at a constant velocity
during each session; 33 sessions were performed at 33 different velocities ranging from
0.5 mm/s to 16.5 mm/s at 0.5 mm/s increments. Ultrasonic B-mode using an ultrasound imaging
device (SONIMAGE HS1. Version 1.31, Konica Minolta, Inc., Japan) was used for imaging. A
high-velocity video camera and an accelerometer were used to monitor the movement of the
wooden frames. The experiment was performed by a physical therapist (TY) with five years of
experience in ultrasound imaging. Since the ultrasound probe was fixed in the wooden
rectangle, it was not expected that there would be any effects of technical variations due
to the investigator’s experience.
Fig. 1.
Experimental setting.
a) The specimen (pork meat) was vacuum-packed and contained in a wooden frame. Two
wooden frames were overlaid and the layer between the specimens was filled with
ultrasonic gel. Hooks were attached proximally and distally to the upper wooden frame
and then wires were attached to these hooks. A stopper (A) was attached to the base so
that the lower wooden frame did not move during towing. Two wooden boards (B) were
attached to the base so that the two wooden frames did not rotate or wobble during
movement.
b) The wooden frame containing the specimen was placed on the base, and the meat was
fixed to the base. The upper frame was connected to the universal tester attachment
via a wire. A 1 kg weight was hung distally to maintain constant tension on the wire.
The 1 kg weight was placed on the meat to maintain the approximation of the two
frames. A high speed camera was positioned above the workbench and an acceleration
sensor as well as a ruler were installed on the upper frame. An ultrasonic probe was
positioned in a hole cut into the surface of the workbench in order to record movement
of the specimens during measurements. The upper frame was pulled at 32 different
velocities (ranging from 0.5 to 16.5 mm/s at increments of 0.5 mm/s).
c) A rectangle hole was created in the workbench to firmly hold the probe during
measurements and an adhesive tape was used to stabilize the probe.
Experimental setting.a) The specimen (pork meat) was vacuum-packed and contained in a wooden frame. Two
wooden frames were overlaid and the layer between the specimens was filled with
ultrasonic gel. Hooks were attached proximally and distally to the upper wooden frame
and then wires were attached to these hooks. A stopper (A) was attached to the base so
that the lower wooden frame did not move during towing. Two wooden boards (B) were
attached to the base so that the two wooden frames did not rotate or wobble during
movement.b) The wooden frame containing the specimen was placed on the base, and the meat was
fixed to the base. The upper frame was connected to the universal tester attachment
via a wire. A 1 kg weight was hung distally to maintain constant tension on the wire.
The 1 kg weight was placed on the meat to maintain the approximation of the two
frames. A high speed camera was positioned above the workbench and an acceleration
sensor as well as a ruler were installed on the upper frame. An ultrasonic probe was
positioned in a hole cut into the surface of the workbench in order to record movement
of the specimens during measurements. The upper frame was pulled at 32 different
velocities (ranging from 0.5 to 16.5 mm/s at increments of 0.5 mm/s).c) A rectangle hole was created in the workbench to firmly hold the probe during
measurements and an adhesive tape was used to stabilize the probe.Analyses were performed by the same investigator (TY) using Echolizer software with the
Farneback method and the optical flow algorithm (Fig.
2). The Farneback method measures the movement of an object within grids on the
ultrasound image without using the characteristic point (or edge) within the ROI. First,
ultrasound movie files were imported into Echolizer program. The range of velocity for
analyses was set between 0.0 and 30.0 mm/s for all variables including image quality
(contrast and brightness), ROI size and location, and image frequency. The ROI’s were
defined using the grid overlaying the ultrasound image so that the ROI locations and sizes
could be reproduced. The frequency of the image files to be analyzed were set at either
frame-1 (30 Hz), frame-2 (15 Hz), or frame-3 (10 Hz), where frame-2 skipped one image and
frame-3 skipped two images. Then, an analysis was performed, and mean velocity and mean
orientation of the bundles in each ROI were obtained as a CSV format. Using data from the
ROI located at the center of the ultrasonic image, regression analyses were performed to
determine actual velocity using measured data from each velocity. The frequency settings of
frame-1, 2 and 3 were respectively used for these analyses. The measured velocity was
modified using the regression equation. Six ROI sizes and six locations were selected for
analysis using the movie file obtained at 6.0 mm/s. The 6 sizes ranged from 6.25 to 900
mm2 (2.5 × 2.5 mm, 5 × 5 mm, 10 × 10 mm, 15 × 15 mm, 20 × 20 mm, 30 × 30 mm
(Fig. 3a). The six ROI locations included three in the superficial area and three in deeper
portions of the meat. The location of each ROI was specified. The ROI size was set at 10 ×
10 mm with intervals of 2.5 mm between the ROI’s (Fig.
3b). Variables of the image quality involved 21 brightness and 21 contrast settings
using a commercial software program (BeeCut, Apowersoft) using the movie file at 6.0 mm/s.
The original image quality was 0 for brightness and 0 for contrast and −100 represented the
lowest level of brightness (darkest) and lowest level of contrast (Fig. 4a), while 100 represented the highest level of brightness and highest level of contrast
(Fig. 4b). When the effects of brightness were
analyzed, the contrast remained constant at 0. When the effects of contrast were analyzed,
the brightness remained constant at 0. The ROI size remained constant at 100 mm2
(10 ×10 mm) during both analyses.
Fig. 2.
Graphic interface of the Echolizer software.
Ultrasonographic images are imported as movie files, and up to six regions of
interest (ROI’s) can be defined. Velocity of muscle bundles is demonstrated by arrows,
and the velocity and covariance are quantified and exported as a CSV format.
Fig. 3.
Region of interest (ROI) settings.
a) Six ROI sizes were defined as 6. 25 mm2 (2.5 mm long × 2.5 mm wide),
25 mm2 (5 × 5 mm), 100 mm2 (10 × 10 mm), 225 mm2
(15 × 15 mm), 400 mm2 (20 × 20 mm), 900 mm2 (30 × 30 mm).
b) Six ROI locations with a consistent size (10 × 10 mm) were also defined.
Fig. 4.
Definitions of brightness.
a) The brightness of the original image was set at 0, with the lowest and highest
brightness being −100 and 100, respectively.
b) The contrast of the original image was set at 0, with the lowest and highest
contrast being −100 and 100, respectively.
Graphic interface of the Echolizer software.Ultrasonographic images are imported as movie files, and up to six regions of
interest (ROI’s) can be defined. Velocity of muscle bundles is demonstrated by arrows,
and the velocity and covariance are quantified and exported as a CSV format.Region of interest (ROI) settings.a) Six ROI sizes were defined as 6. 25 mm2 (2.5 mm long × 2.5 mm wide),
25 mm2 (5 × 5 mm), 100 mm2 (10 × 10 mm), 225 mm2
(15 × 15 mm), 400 mm2 (20 × 20 mm), 900 mm2 (30 × 30 mm).b) Six ROI locations with a consistent size (10 × 10 mm) were also defined.Definitions of brightness.a) The brightness of the original image was set at 0, with the lowest and highest
brightness being −100 and 100, respectively.b) The contrast of the original image was set at 0, with the lowest and highest
contrast being −100 and 100, respectively.A 120 Hz high-velocity camera and acceleration data confirmed whether the upper meat was
being pulled at a constant velocity during the measurement and whether the probe remained
still during towing. The movie from the high-velocity video was qualitatively analyzed to
confirm that the movement of the wooden frame was constant. The accelerometer data was
analyzed to confirm that the acceleration remained at 0 mm/s during towing.SPSS statistics ver. 23 (manufactured by IBM SPSS) and G*Power 3.0 were used for
statistical analyses. The significance level was set at α=0.05. Actual velocity and the
measured data were used in the regression analysis to determine the accuracy and its
reliable range in the velocity. Descriptive statistics were used for the effect of image
quality, size and location of the ROI’s, and frequency. A priori power analysis was
performed and the required sample size was 26 with an Effect size (r) of 0.5, an α error
probability of 0.05, and a Power (1-β prob) of 0.8.
RESULTS
The regression equation, the determination coefficient of the actual velocity and the
measured data under three movie frequencies are shown below:Frame-1 (30 Hz):
y=3.802x + 0.8984, RFrame-2
(15 Hz): y=0.977x + 0.5939, RFrame-3 (10 Hz): y=1.150x + 0.071,
RThe data in frame-3 involved a relative error of 12.1% between 2.5 and 16.5 mm/s (Fig. 5a). After regression equation was applied to the measured data, the absolute error was
an average of 0.02 [95%CI: −0.11, 0.14] mm/s (Fig.
5b), and the relative error was 0.2 [−0.8, 1.2] % (Fig. 5c).
Fig. 5.
Regression analysis of the computed velocity as an independent variable and actual
velocity as the dependent variable.
a) The regression equation was y=1.150x−0.071, and R2 was 0.996.
b) After regression analysis was applied to the computed data, the average [95%CI] of
the absolute error was 0.02 [−0.11, 0.14] mm/s in the range between 2.5 to 16.5 mm/s
(area within the dotted line).
c) After regression analysis was applied to the computed data, the average [95%CI] of
the relative error was 0.2 [−0.8, 1.2]% in the range between 2.5 to 16.5 mm/s (area
within the dotted line).
Regression analysis of the computed velocity as an independent variable and actual
velocity as the dependent variable.a) The regression equation was y=1.150x−0.071, and R2 was 0.996.b) After regression analysis was applied to the computed data, the average [95%CI] of
the absolute error was 0.02 [−0.11, 0.14] mm/s in the range between 2.5 to 16.5 mm/s
(area within the dotted line).c) After regression analysis was applied to the computed data, the average [95%CI] of
the relative error was 0.2 [−0.8, 1.2]% in the range between 2.5 to 16.5 mm/s (area
within the dotted line).Effects of the ROI size were analyzed using six sizes ranging from 25 to 900 mm2
on ultrasound images obtained at 6.0 mm/s (Table
1). The median velocity of the six ROI’s was 6.06 (range: 4.02 to 6.13) mm/s.
The minimum error was observed with a ROI of 225mm2, in which the absolute and
relative errors were 0.03 mm/s and 0.51%, respectively. The maximal error was observed with
a ROI of 900 mm2, in which the absolute and relative errors were −1.98 mm/s and
33.0%, respectively.
Table 1.
Absolute and relative errors in different ROI sizes
ROI size (mm2)
Actual velocity (mm/s)
Software calculation (mm/s)
Absolute error (mm/s)
Relative error (%)
6.25 (2.5 mm × 2.5 mm)
6.00
6.09
0.09
1.5
25 (5 mm × 5 mm)
6.00
6.13
0.13
2.1
100 (10 mm × 10 mm)
6.00
6.10
0.10
1.6
225 (15 mm × 15 mm)
6.00
6.03
0.03
0.5
400 (20 mm × 20 mm)
6.00
5.81
−0.19
−3.1
900 (30 mm × 30 mm)
6.00
4.02
−1.98
−33.0
Effects of the ROI location were analyzed using six ROI locations on ultrasound images
obtained at 6.0 mm/s (Table 2). ROI’s A, B and C were located on the superficial area, while ROI’s D, E and
F were on the deeper area. The average velocity of the six ROI’s was 4.07 [2.19, 5.95] mm/s.
The relative errors for the ROI’s B and C were 2.4 and −4.9%, respectively, whereas those
for the ROI’s A, D, E and F ranged from −25.4% to −79.3%. Note that the image quality of ROI
A was not clear enough to show details of the muscle bundles compared with the image quality
of ROI’s B and C.
Table 2.
Absolute and relative errors in different ROI locations
Location
Actual velocity (mm/s)
Software (mm/s)
Absolute error (mm/s)
Relative error (%)
A
6.00
4.47
−1.53
−25.4
B
6.00
6.14
0.14
2.4
C
6.00
5.71
−0.29
−4.9
D
6.00
1.24
−4.76
−79.3
E
6.00
3.34
−2.66
−44.3
F
6.00
3.52
−2.48
−41.3
Effects of the brightness and contrast were examined at 6.0 mm/s using 21 levels of
brightness and contrast, respectively. The median velocity was 5.77 (0.74 to 5.81) mm/s
(Fig. 6, Table 3). The relative error exceeded 10% with a brightness of −50 or below, but
remained within 5% at a brightness of −30 or above. The median velocity was 5.73 (−0.07 to
5.83) mm/s (Fig. 7, Table 4). The relative error exceeded 10% with a contrast of −60 or below, but
remained within 5% with a contrast between −30 and 80.
Fig. 6.
Effects of brightness on the computed velocity.
The actual speed was 6.0 mm/s (horizontal line). The relative error exceeded 10% with
a brightness of −50 or below, but was within 5% with a brightness of −30 or above.
Table 3.
Absolute and relative errors in velocity at different brightness levels
Brightness
Actual velocity (mm/s)
Software (mm/s)
Absolute error (mm/s)
Relative error (%)
−100
6.00
0.74
−5.26
−87.7
−90
6.00
2.16
−3.84
−64.0
−80
6.00
3.70
−2.30
−38.3
−70
6.00
4.49
−1.51
−25.2
−60
6.00
5.03
−0.97
−16.2
−50
6.00
5.29
−0.71
−11.8
−40
6.00
5.65
−0.35
−5.9
−30
6.00
5.78
−0.22
−3.6
−20
6.00
5.80
−0.20
−3.4
−10
6.00
5.81
−0.19
−3.2
0
6.00
5.79
−0.21
−3.6
10
6.00
5.79
−0.21
−3.6
20
6.00
5.78
−0.22
−3.6
30
6.00
5.79
−0.21
−3.4
40
6.00
5.79
−0.21
−3.6
50
6.00
5.78
−0.22
−3.6
60
6.00
5.78
−0.22
−3.6
70
6.00
5.77
−0.23
−3.8
80
6.00
5.76
−0.24
−3.9
90
6.00
5.76
−0.24
−4.0
100
6.00
5.77
−0.23
−3.9
Fig. 7.
Effects of contrast on the computed velocity.
The actual speed was 6.0 mm/s (horizontal line). The relative error exceeded 10% at a
contrast of −60 or below, but was within 5% at a contrast between −30 and 80.
Table 4.
Absolute and relative errors in the velocity in different contrast
levels
contrast
Actual velocity (mm/s)
Software (mm/s)
Absolute error (mm/s)
Relative error (%)
−100
6.00
−0.07
−6.07
−101.2
−90
6.00
0.22
−5.78
−96.3
−80
6.00
2.66
−3.34
−55.7
−70
6.00
4.76
−1.24
−20.6
−60
6.00
5.38
−0.62
−10.4
−50
6.00
5.60
−0.40
−6.7
−40
6.00
5.69
−0.31
−5.2
−30
6.00
5.73
−0.27
−4.5
−20
6.00
5.76
−0.24
−4.1
−10
6.00
5.77
−0.23
−3.9
0
6.00
5.79
−0.21
−3.6
10
6.00
5.81
−0.19
−3.1
20
6.00
5.82
−0.18
−3.0
30
6.00
5.82
−0.18
−3.0
40
6.00
5.83
−0.17
−2.9
50
6.00
5.82
−0.18
−3.0
60
6.00
5.80
−0.20
−3.3
70
6.00
5.78
−0.22
−3.7
80
6.00
5.71
−0.29
−4.9
90
6.00
5.54
−0.46
−7.7
100
6.00
5.47
−0.53
−8.8
Effects of brightness on the computed velocity.The actual speed was 6.0 mm/s (horizontal line). The relative error exceeded 10% with
a brightness of −50 or below, but was within 5% with a brightness of −30 or above.Effects of contrast on the computed velocity.The actual speed was 6.0 mm/s (horizontal line). The relative error exceeded 10% at a
contrast of −60 or below, but was within 5% at a contrast between −30 and 80.
DISCUSSION
The purposes of this study were to determine: (1) the validity and reliability of Echolizer
software in measuring the velocity of muscle bundles; and (2) the optimal image quality as
well as ROI size and location to enhance the validity. The regression equation of y=1.150x +
0.071 with a determinant coefficient of R2=0.996 shows an excellent fit. After
this equation was applied, the relative error was 0.20%. The measurement accuracy was
decreased for deep ROI locations, ROI’s with low quality visualization of the muscle
bundles, large ROI’s located in a deeper zone, images with low brightness and low or high
contrast.Regression analysis of the measured velocity using Echolizer program with the optical-flow
algorithm and block matching method obtained a relative error of 0.2%, which was considered
excellent. Although we utilized echo movies of 30 Hz, the best determinant coefficient of
0.996 was obtained at 10 Hz by skipping two frames in-between. A previous study showed that
the average relative error of the speckle tracking algorithm with the block matching method
was 1.3% in measuring the velocity of the Achilles tendon in pig cadavers17). In measurements using another
optical-flow algorithms, Williamson et al.15) reported that the accuracy (mean ± standard deviation) was 0.74 ±
1.03 mm, and Tabata et al.16) reported a
relative error of less than 1%. The current study showed an excellent accuracy of 0.2% after
correction for detecting a velocity between 2.5 and 16.5 mm/s.The ROI size and location may influence accuracy. A ROI size of 225 mm2 produced
the smallest relative error of 0.5%, whereas a size of 900 mm2 produced the
largest relative error of 33.0%. For sizes below 225 mm2, the greater the ROI
size, the smaller the relative error. This suggests that the greater number of muscle
bundles visible in a larger ROI may be advantageous to achieving a higher accuracy. However,
a ROI size of 400 mm2 or larger demonstrated a lower accuracy because the ROI’s
were located in the deeper zone and the muscle bundles could not be seen clearly. Regarding
the ROI locations, the relative error at locations B and C were 2.4% and −4.9%,
respectively, where the muscle bundles were clearly shown. In locations A, D, E and F, the
muscle bundles were not clearly demonstrated and therefore, showed significantly lower
accuracy with relative errors of −25.4% or greater. ROI locations in the deeper zone
produced greater relative errors due to poor clarity of the muscle bundles. Although
location A was in the superficial zone, the relative error was −25.4% because of the dark
area on the left side of the image. Although the cause of this darkness was not identified,
the poor clarity of the muscle bundles in the ROI caused a large relative error. The
Farneback method used in Echolizer calculates the optical-flow of all pixels in a ROI8, 9.
Therefore, it is reasonable that the clearness of the muscle bundles influences the
measurement accuracy. Accordingly, higher accuracy of Echolizer program would be achieved by
defining the ROI containing greater number of clear muscle bundles.Effects of the brightness and contrast of the ultrasound image were verified. Good results
with a relative error less than 5% were obtained when the brightness was set at −30 or
greater and the contrast was set between −30 and 80. The velocity was underestimated when
the brightness and contrast were lower than −30, at which clarity of the muscle bundles was
visually inferior and the edges were more difficult to recognize. Visual transparency
reduces the accuracy of the analysis18,19,20,21). Therefore, it is highly important that
the original image quality of the ultrasonography should contain clear muscle bundles, which
requires the examiners’ experience in acquiring the ultrasonographic images. The important
message from this result is that the image contrast nor brightness should not be changed to
a large extent since the measurement accuracy is primarily determined by the quality of the
original ultrasonographic images.We made every effort to improve the internal validity of Echolizer measurement. The actual
movement of the material during the testing was monitored by both an acceleration sensor and
a high-speed camera in order to confirm that the specimen was pulled at a uniform velocity.
Therefore, the movement velocity of the material is considered the true value. Sufficient
statistical power for regression analysis was achieved by testing at 33 different
velocities. This is the first study to validate a software program that quantifies muscle
velocity without using anatomical landmarks.There are several limitations. First, the material was neither human nor living. However,
visual observation of pork meat and muscles of living humans were similar in our preliminary
studies. The ultrasound image can detect the fascia such as tendons and intra-muscular
tendons of the muscles and histological similarities between human muscles and pork meat
have been reported. We carefully checked the appearance of the pork meat on the ultrasound
images and found that the muscle fascicles appeared similar to those of humans. Furthermore,
we used the graphic interface to confirm that the software could successfully detect the
fascicles. Therefore, we believe the selected material was appropriate. Second, the material
was not a living animal and this might have reduced the water content within the meat, which
could have negatively impacted the accuracy. Third, the quality of the ultrasonographic
images affects the accuracy of the velocity measurement. Therefore, we should be cautious in
selecting high-quality ultrasonographic images obtained by an experienced researcher or
technician. Fourth, there is an effect of tissue deformation in the longitudinal direction
of the muscle bundles. We used a wooden frame to contain the specimen and longitudinal
muscle elongation was considered minimal. Fifth, the orientation of the probe was set
parallel to the direction of towing in this study, which limited our ability to validate the
software when the direction of muscle movements is not parallel to the orientation of the
probe22).To conclude, the optical-flow algorithm had a relative error of 12.1% in the measurement of
muscle bundle movements. After correction using the regression equation, the relative error
of Echolizer program was within 0.20% between 2.5 and 16.5 mm/s. The validity can be reduced
if the ROI’s are set to include deep tissue or areas with poor visualization of the muscle
bundles, or image brightness and contrast are set too low or too high. Echolizer program
does not require anatomical landmarks on ultrasonic images. Therefore, this study provides
novel information on the quantification of tissue movements using ultrasonic images. Further
studies should confirm its validity in living human tissues and this software is expected to
reveal local gliding properties in vivo, which would allow therapists to
quantify the effectiveness of physical therapy interventions on various soft tissues.
Conflict of interest
Echolizer program was temporarily licensed by GLAB Co., Ltd.