Finite element analysis of different screw constructs in the treatment of unstable femoral neck fractures.

Jiantao Li1, Menglin Wang2, Jianfeng Zhou3, Hao Zhang4, Lianting Li5.   

Abstract

OBJECTIVES: In the present study, we evaluated the mechanical outcome of different configurations formed by partially threaded screws (PTS) alone or combined screws consisting of PTS and fully threaded screws (FTS) in the treatment of unstable femoral neck fracture.
METHODS: The Pauwels type III unstable femoral fracture and screw models of PTS and FTS were created in 3-matic software and UG-NX software respectively. We assembled the different screw fixation types to the fracture model separately to form the fixation models. We used Abaqus software to perform the finite element analysis.
RESULTS: Our results indicated that the peak von Mises stresses of screws increased when some PTSs changed into FTSs in all groups except for the inferior triangle group. FTS in each group underwent the most stress while PTS underwent a little bit of stress. The combined screws fixation types were less likely to be cut-out and was more stable than PTPs alone fixation strategy. Less yielding regions around the screw tunnels for the superior and inferior triangle configuration fixed by combined screws was indicated. Superior triangle fixation model underwent the largest area of stress concentration around the screw holes after screws removal.
CONCLUSIONS: For unstable femoral neck fractures, superior results were obtained by stabilizing the fracture with superior or inferior triangle configuration consisting of one PTS and two FTSs. If screws removal was taken into account after fracture union, fixation type of inferior triangle configuration should be the recommended choice.
Copyright © 2020. Published by Elsevier Ltd.

Entities:  

Keywords:  Finite element analysis; Three-screw fixation; Unstable femoral neck fracture

Mesh:

Year:  2020        PMID: 32151421     DOI: 10.1016/j.injury.2020.02.075

Source DB:  PubMed          Journal:  Injury        ISSN: 0020-1383            Impact factor:   2.586


  4 in total

1.  The efficacy of dynamic compression locking system vs. dynamic hip screw in the treatment of femoral neck fractures: a comparative study.

Authors:  Jian-Zhong Chang; Ya-Ping Xiao; Ling Li; Ming-Jian Bei
Journal:  BMC Musculoskelet Disord       Date:  2022-07-12       Impact factor: 2.562

2.  Development and validation of a predictive nomogram for postoperative osteonecrosis of the femoral head with cannulated screws fixation.

Authors:  Wanbo Zhu; Kai Xie; Xianzuo Zhang; Jiazhao Yang; Lei Xu; Junchen Zhu; Shiyuan Fang; Chen Zhu
Journal:  Ann Transl Med       Date:  2021-02

Review 3.  Finite Element Analysis of Fracture Fixation.

Authors:  Gregory S Lewis; Dominic Mischler; Hwabok Wee; J Spence Reid; Peter Varga
Journal:  Curr Osteoporos Rep       Date:  2021-06-29       Impact factor: 5.163

4.  Analysis on risk factors for neck shortening after internal fixation for Pauwels II femoral neck fracture in young patients.

Authors:  Fulong Zhao; Lijuan Guo; Xuefei Wang; Yakui Zhang
Journal:  Eur J Med Res       Date:  2021-06-24       Impact factor: 2.175

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.