There are various scanners available in dental practice with various accuracies. The aim of this study was to compare the 3D capturing accuracy of scans obtained from Trios 3 and Dental Wings scanner. A reference mandibular model was printed from FormLab with reference points in three axes (X, Y, and XY and Z). The printed model was scanned 5 times with 3 scans: normal scan by Trios 3 (Trios 3A), high-resolution scan by Trios 3 (Trios 3B), and normal scan by Dental Wings. After scan, the stereolithography (stl) files were generated. Then, the measurements were made from the computer software using Rhinoceros 3D (Rhino, Robert McNeel & Associates for Windows, Washington DC, USA). The measurements made with digital caliper were taken as control. Statistical analysis was done using one-way ANOVA with post hoc using Sheffe (P < 0.01). Trios 3 presented higher accuracy than Dental Wings and high resolution showed better results. The Dental Wings showed less accuracy at the measurements >50 mm of length and >30 mm in width. There was no significant difference (P > 0.05) of control with the Trios 3A and Trios 3B. Similarly, for the measurements in Z-axis, there was no significant difference of control with each scan (Trios 3A, Trios 3B, and Dental Wings). Accuracy of the scan is affected by the length of the scanning area and scanning pattern. It is less recommended to Dental Wings scan >3-unit prosthesis and that crosses the midline.Entities:
Mesh:
Year: 2020 PMID: 32148745 PMCID: PMC7042536 DOI: 10.1155/2020/5739312
Source DB: PubMed Journal: J Healthc Eng ISSN: 2040-2295 Impact factor: 2.682
Figure 1Details of the study overview.
Figure 2Digital dental model with various points marked in the three axes (X, Y, and Z).
Figure 3Printed mandibular dental model.
Figure 4Intraoral scanners (IOS): (a) Trios 3 and (b) Dental Wings.
Measurements in three axes (X, Y, XY, and Z) of various lengths.
|
|
|
|
|
|---|---|---|---|
|
|
| AR: Diagonal distance from #12 to #27 (65 mm) |
|
|
|
|
| |
|
|
| ||
|
|
| AR: Diagonal distance from #22 to #17 (65 mm) |
|
|
| |||
|
|
| ||
Figure 5Reference points on the model and various measurements measured on model in different axes: X-axis (a), Y-axis (b), XY-axis (c), and Z-axis (d).
Figure 6Reference points on the model and various measurements in the X-axis (length) and Y-axis (length) from software of 1 scan.
Descriptive statistics of measurements of various groups in the X-axis.
| Measurements | Groups | Mean | SD | 95% CI for mean | Min | Max | |
|---|---|---|---|---|---|---|---|
| Lower bound | Upper bound | ||||||
|
| Control | 1.938 | 0.039 | 1.888 | 1.987 | 1.90 | 2.00 |
| Trios 3A | 1.876 | 0.032 | 1.836 | 1.916 | 1.85 | 1.93 | |
| Trios 3B | 1.886 | 0.036 | 1.841 | 1.931 | 1.84 | 1.94 | |
| Dental Wings | 1.876 | 0.057 | 1.805 | 1.946 | 1.83 | 1.96 | |
| Control | 9.938 | 0.022 | 9.911 | 9.964 | 9.90 | 9.95 | |
|
| |||||||
|
| Trios 3A | 9.864 | 0.027 | 9.835 | 9.892 | 9.84 | 9.90 |
| Trios 3B | 9.968 | 0.028 | 9.933 | 10.002 | 9.92 | 9.99 | |
| Dental Wings | 9.866 | 0.071 | 9.778 | 9.954 | 9.79 | 9.97 | |
| Control | 29.852 | 0.06 | 29.77 | 29.926 | 29.78 | 29.93 | |
|
| |||||||
|
| Trios 3A | 29.734 | 0.022 | 29.706 | 29.761 | 29.70 | 29.76 |
| Trios 3B | 29.914 | 0.033 | 29.872 | 29.955 | 29.87 | 29.95 | |
| Dental Wings | 29.806 | 0.172 | 29.592 | 30.019 | 29.60 | 30.03 | |
| Control | 39.910 | 0.054 | 39.842 | 39.978 | 39.82 | 39.96 | |
|
| |||||||
|
| Trios 3A | 39.716 | 0.011 | 39.702 | 39.73 | 39.70 | 39.73 |
| Trios 3B | 39.698 | 0.248 | 39.389 | 40.007 | 39.39 | 39.93 | |
| Dental Wings | 40.112 | 0.177 | 39.892 | 40.332 | 39.88 | 40.28 | |
| Control | 50.03 | 0.035 | 49.986 | 50.074 | 49.98 | 50.07 | |
|
| |||||||
|
| Trios 3A | 49.86 | 0.118 | 49.713 | 50.006 | 49.75 | 50.02 |
| Trios 3B | 50.21 | 0.136 | 50.04 | 50.379 | 49.98 | 50.31 | |
| Dental Wings | 50.832 | 0.423 | 50.306 | 51.357 | 50.43 | 51.55 | |
| Caliper | 60.616 | 0.052 | 60.552 | 60.68 | 60.54 | 60.67 | |
| Control | 60.068 | 0.271 | 59.732 | 60.404 | 59.66 | 60.42 | |
|
| |||||||
|
| Trios 3A | 60.368 | 0.514 | 59.728 | 61.007 | 59.77 | 60.83 |
| Dental Wings | 61.952 | 0.374 | 61.487 | 62.417 | 61.55 | 62.41 | |
SD = standard deviation; CI = confidence interval for mean; min = minimum; max = maximum.
Descriptive statistics of measurements of various groups in Y-axis and XY-axis.
| Measurements | Groups | Mean | SD | 95% CI for mean | Min | Max | |
|---|---|---|---|---|---|---|---|
| Lower bound | Upper bound | ||||||
|
| Control | 1.932 | 0.008 | 1.921 | 1.942 | 1.92 | 1.94 |
| Trios 3A | 1.856 | 0.035 | 1.812 | 1.899 | 1.80 | 1.89 | |
| Trios 3B | 1.854 | 0.077 | 1.758 | 1.949 | 1.78 | 1.97 | |
| Dental Wings | 1.828 | 0.017 | 1.805 | 1.85 | 1.80 | 1.85 | |
|
| |||||||
|
| Control | 9.876 | 0.021 | 9.85 | 9.901 | 9.86 | 9.91 |
| Trios 3A | 9.806 | 0.149 | 9.619 | 9.992 | 9.54 | 9.89 | |
| Trios 3B | 9.81 | 0.082 | 9.708 | 9.912 | 9.74 | 9.93 | |
| Dental Wings | 9.8 | 0.074 | 9.707 | 9.892 | 9.71 | 9.90 | |
|
| |||||||
|
| Control | 19.934 | 0.047 | 19.875 | 19.992 | 19.89 | 20.01 |
| Trios 3A | 19.71 | 0.12 | 19.561 | 19.859 | 19.56 | 19.84 | |
| Trios 3B | 19.81 | 0.062 | 19.733 | 19.886 | 19.75 | 19.90 | |
| Dental Wings | 19.636 | 0.518 | 18.993 | 20.278 | 18.81 | 20.02 | |
|
| |||||||
|
| Control | 30.162 | 0.339 | 29.748 | 30.575 | 29.86 | 30.70 |
| Trios 3A | 29.974 | 0.391 | 29.488 | 30.459 | 29.38 | 30.35 | |
| Trios 3B | 30.186 | 0.593 | 29.45 | 30.921 | 29.60 | 31.05 | |
| Dental Wings | 29.754 | 0.425 | 29.225 | 30.282 | 29.17 | 30.37 | |
|
| |||||||
| AR | Control | 65.126 | 0.037 | 65.079 | 65.173 | 65.10 | 65.19 |
| Trios 3A | 64.872 | 0.119 | 64.724 | 65.019 | 64.78 | 65.07 | |
| Trios 3B | 65.124 | 0.173 | 64.908 | 65.339 | 64.90 | 65.32 | |
| Dental Wings | 66.044 | 0.624 | 65.268 | 66.819 | 65.17 | 66.88 | |
|
| |||||||
| AL | Control | 65.094 | 0.053 | 65.027 | 65.16 | 65.03 | 65.15 |
| Trios 3A | 65.43 | 0.484 | 64.828 | 66.031 | 64.88 | 65.81 | |
| Trios 3B | 65.408 | 0.326 | 65.002 | 65.813 | 64.97 | 65.89 | |
| Dental Wings | 65.942 | 0.133 | 65.776 | 66.107 | 65.84 | 66.14 | |
SD = standard deviation; CI = confidence interval for mean; min = minimum; max = maximum.
Descriptive statistics of measurements of various groups in the Z-axis.
| Measurements | Groups | Mean | SD | 95% CI for mean | Min | Max | |
|---|---|---|---|---|---|---|---|
| Lower bound | Upper bound | ||||||
|
| Control | 2.014 | 0.023 | 1.985 | 2.042 | 1.99 | 2.04 |
| Trios 3A | 2.018 | 0.037 | 1.971 | 2.064 | 1.96 | 2.06 | |
| Trios 3B | 1.998 | 0.047 | 1.938 | 2.057 | 1.96 | 2.08 | |
| Dental Wings | 1.990 | 0.063 | 1.911 | 2.068 | 1.88 | 2.04 | |
|
| |||||||
|
| Control | 3.986 | 0.011 | 3.972 | 4.001 | 3.97 | 4.00 |
| Trios 3A | 3.988 | 0.072 | 3.898 | 4.077 | 3.88 | 4.05 | |
| Trios 3B | 3.952 | 0.073 | 3.861 | 4.042 | 3.90 | 4.08 | |
| Dental Wings | 3.97 | 0.054 | 3.902 | 4.037 | 3.89 | 4.02 | |
|
| |||||||
|
| Control | 6.15 | 0.137 | 5.979 | 6.32 | 5.97 | 6.32 |
| Trios 3A | 5.906 | 0.209 | 5.645 | 6.166 | 5.71 | 6.19 | |
| Trios 3B | 5.978 | 0.011 | 5.964 | 5.991 | 5.96 | 5.99 | |
| Dental Wings | 6.142 | 0.155 | 5.949 | 6.335 | 5.92 | 6.31 | |
|
| |||||||
|
| Control | 8.046 | 0.04 | 7.995 | 8.096 | 8.01 | 8.11 |
| Trios 3A | 7.968 | 0.085 | 7.861 | 8.074 | 7.86 | 8.04 | |
| Trios 3B | 7.978 | 0.136 | 7.808 | 8.147 | 7.84 | 8.14 | |
| Dental Wings | 7.996 | 0.158 | 7.799 | 8.192 | 7.72 | 8.11 | |
SD = standard deviation; CI = confidence interval for mean; min = minimum; max = maximum.
Multiple comparison of the various measurements in the X-axis of control with other scan groups (Trios 3A, Trios 3B, and Dental Wings).
| Measurements | Comparison groups | Mean difference |
| |
|---|---|---|---|---|
|
| Control | Trios 3A | 0.062 | 0.190 |
| Trios HD | 0.052 | 0.322 | ||
| Dental Wings | 0.062 | 0.190 | ||
|
| ||||
|
| Control | Trios 3A | 0.074 | 0.082 |
| Trios HD | −0.03 | 0.727 | ||
| Dental Wings | 0.072 | 0.093 | ||
|
| ||||
|
| Control | Trios 3A | 0.118 | 0.299 |
| Trios HD | −0.062 | 0.777 | ||
| Dental Wings | 0.046 | 0.893 | ||
|
| ||||
|
| Control | Trios 3A | 0.194 | 0.309 |
| Trios HD | 0.212 | 0.240 | ||
| Dental Wings | −0.202 | 0.277 | ||
|
| ||||
|
| Control | Trios 3A | 0.17 | 0.719 |
| Trios HD | −0.18 | 0.683 | ||
| Dental Wings | −0.802 | 0.001 | ||
|
| ||||
|
| Control | Trios 3A | 0.548 | 0.143 |
| Trios HD | 0.248 | 0.737 | ||
| Dental Wings | −1.336 | <0.001 | ||
Significant difference at P value <0.05.
Multiple comparison of the various measurements in the Y-axis and the XY-axis of control with other scan groups (Trios 3A, Trios 3B, and Dental Wings).
| Measurements | Comparison groups | Mean difference | Sig. | |
|---|---|---|---|---|
|
| Control | Trios 3A | 0.076 | 0.092 |
| Trios 3B | 0.078 | 0.081 | ||
| Dental Wings | 0.104 | 0.015 | ||
|
| ||||
|
| Control | Trios 3A | 0.07 | 0.711 |
| Trios 3B | 0.066 | 0.746 | ||
| Dental Wings | 0.076 | 0.657 | ||
|
| ||||
|
| Control | Trios 3A | 0.224 | 0.637 |
| Trios 3B | 0.124 | 0.910 | ||
| Dental Wings | 0.298 | 0.407 | ||
|
| ||||
|
| Control | Trios 3A | 0.188 | 0.929 |
| Trios 3B | −0.024 | 1.000 | ||
| Dental Wings | 0.408 | 0.567 | ||
|
| ||||
| AR | Control | Trios 3A | 0.254 | 0.692 |
| Trios 3B | 0.002 | 1.000 | ||
| Dental Wings | −0.918 | 0.005 | ||
|
| ||||
| AL | Control | Trios 3A | −0.336 | 0.402 |
| Trios 3B | −0.314 | 0.459 | ||
| Dental Wings | −0.848 | 0.004 | ||
Significant difference at P value <0.05.
Multiple comparisons of the various measurements in the Z-axis of control with other scan groups (Trios 3A, Trios 3B, and Dental Wings).
| Measurements | Comparison groups | Mean difference | Sig. | |
|---|---|---|---|---|
|
| Control | Trios 3A | −0.004 | 0.999 |
| Trios 3B | 0.016 | 0.957 | ||
| Dental Wings | 0.024 | 0.872 | ||
|
| ||||
|
| Control | Trios 3A | −0.002 | 1.000 |
| Trios 3B | 0.034 | 0.836 | ||
| Dental Wings | 0.016 | 0.979 | ||
|
| ||||
|
| Control | Trios 3A | 0.244 | 0.119 |
| Trios 3B | 0.172 | 0.366 | ||
| Dental Wings | 0.008 | 1.000 | ||
|
| ||||
|
| Control | Trios 3A | 0.078 | 0.766 |
| Trios 3B | 0.068 | 0.831 | ||
| Dental Wings | 0.050 | 0.923 | ||
Significant difference at P value <0.05.