Studying the Optical 3D Accuracy of Intraoral Scans: An In Vitro Study.

Pokpong Amornvit1, Sasiwimol Sanohkan1, Chaimongkon Peampring1.   

Abstract

There are various scanners available in dental practice with various accuracies. The aim of this study was to compare the 3D capturing accuracy of scans obtained from Trios 3 and Dental Wings scanner. A reference mandibular model was printed from FormLab with reference points in three axes (X, Y, and XY and Z). The printed model was scanned 5 times with 3 scans: normal scan by Trios 3 (Trios 3A), high-resolution scan by Trios 3 (Trios 3B), and normal scan by Dental Wings. After scan, the stereolithography (stl) files were generated. Then, the measurements were made from the computer software using Rhinoceros 3D (Rhino, Robert McNeel & Associates for Windows, Washington DC, USA). The measurements made with digital caliper were taken as control. Statistical analysis was done using one-way ANOVA with post hoc using Sheffe (P < 0.01). Trios 3 presented higher accuracy than Dental Wings and high resolution showed better results. The Dental Wings showed less accuracy at the measurements >50 mm of length and >30 mm in width. There was no significant difference (P > 0.05) of control with the Trios 3A and Trios 3B. Similarly, for the measurements in Z-axis, there was no significant difference of control with each scan (Trios 3A, Trios 3B, and Dental Wings). Accuracy of the scan is affected by the length of the scanning area and scanning pattern. It is less recommended to Dental Wings scan >3-unit prosthesis and that crosses the midline.
Copyright © 2020 Pokpong Amornvit et al.

Entities:  

Mesh:

Year:  2020        PMID: 32148745      PMCID: PMC7042536          DOI: 10.1155/2020/5739312

Source DB:  PubMed          Journal:  J Healthc Eng        ISSN: 2040-2295            Impact factor:   2.682


1. Introduction

There has been massive advancement in digital dentistry in the recent decade, especially since the invention of computer-aided design/ computer-aided manufacturing (CAD/CAM) system, milling systems, rapid and automated prototyping, and three-dimensional (3D) printing of dental biomaterials, and these have revolutionized and created a new modality in dentistry [1]. Currently, CAD/CAM is widely used in virtual occlusal records, full-mouth reconstruction, and orthodontics [2-4]. Moreover, they are extensively used in both the dental laboratory and the dental clinic for the fabrication of various prosthesis, such as inlays, onlays, veneers, crowns, fixed partial dentures, orthodontic aligners, surgical guides, and implant abutments [3]. A 3D scanning is a process that is used to capture the shape of an object using a 3D scanner. After scanning, a 3D file of an object is created which can be edited, and 3D printed. An intraoral scan (IOS) can be based on many different technologies, each with its own limitations, advantages, and costs [5]. Many limitations in the kind of objects that can be digitized are still present. For example, optical technology may encounter many difficulties with shiny, reflective, or transparent objects. Teeth, especially the anterior, play in the esthetics of face [6]. For a successful esthetic dental restoration, a good dental impression is important [7, 8]. With the use of digital dentistry, the intraoral conditions can be transferred digitally and printed. When dental laboratories receive a digital impression, they create a model from the data and either continue with the traditional fabrication procedure or rescan the model and fabricate the prosthesis. The dental technician can do all the design restorative works directly on the computer based on the digital file received. Hence, the digital impression plays an important in the fabrication of all digital works and the accuracy of the impression is very important. There are various 3D scanners technologies, such as image capturing or video capturing type contact scanning or noncontact scanning [9-12]. Contact canners probe the subject through physical touch while the object is in contact with or resting on a precision flat surface plate, ground and polished to a specific maximum of surface roughness. Noncontact scanners emit some kind of radiation or light and detect its reflection or radiation passing through the object in order to probe an object or environment [11]. Nowadays, the noncontact scanning technique is recommended widely. Types of scanning technology: the 3D scanning technologies rely on different physical principles and are explained in following categories [5]: Laser triangulation 3D scanning technology uses either a laser line or a single laser point to scan across an object. Structured light 3D scanning technology uses trigonometric triangulation but not the laser. Photogrammetry 3D scan scanning technology (photography) reconstructs 3D from 2D captures with computer vision and computational geometry algorithms. Contact-based 3D scanning technology is based on contact form of 3D data collection and uses a contact probe. Accuracy comprises precision and trueness (ISO 5725-1) [13]. Precision describes how close repeated measurements are to each other [14]. The higher the precision, the more predictable the measurement is. Trueness describes how far the measurement deviates from the actual dimensions of the measured object. A high trueness delivers a result that is close or equal to the actual dimensions of the measured object. Many factors influence the accuracy of the IOS such as [15-17]: Scanner: ability to record details and its accuracy Operator: scanning principles and span of scanning Scanning area: size of scanning area, arch length, and surface irregularities Intraoral environmental factors: temperature, relative humidity, and illumination The IOS accuracy is enhanced by reducing the span of scanning, and ensuring the scanned surfaces exhibit minimal irregularities [10]. The problem with IOS is that it can be difficult to detect deep margin lines in prepared teeth and/or in case of deep margins or bleeding [18]. In addition, various studies done in evaluating the digital impression highlights several issues such as distortion of the digital models, problems with the intraoral conditions, and lower precision compared to conventional impressions [19, 20]. In addition, digital scanners with high accuracy are currently limited to small measurement fields such as single teeth or quadrants [19, 21]. The aim of this study was to compare the 3D capturing accuracy of scans obtained from Trios 3 and Dental Wings scanner in an in vitro study design.

2. Materials and Methods

A method modified from the American National Standard/ American Dental Association (ANS)/ADA) Standard No. 132 for the scanning accuracy was used in this study [22]. The study consists of fabrication of dental model, scanning, and measurements. The details of the study are shown in (Figure 1).
Figure 1

Details of the study overview.

2.1. Fabrication of Dental Model

A digital mandibular model is made in the computer. Various points were marked on the digital model where the measurements can be measured in three axes (X, Y, and Z) (Figure 2).
Figure 2

Digital dental model with various points marked in the three axes (X, Y, and Z).

The model was printed using FormLab following manufacturing recommendations. From 3 dental models, the best model was selected for this study as shown in Figure 3.
Figure 3

Printed mandibular dental model.

2.2. Scanning

The printed model was scanned 5 times each with 3 Shape Trios 3A: normal scan, 3 Shape Trios 3B: high resolution (3 Shape Trios A/S 2018, Copenhagen, Denmark), and Dental Wings (Dental Wings Inc., Montreal QC, Canada) (Figure 4) according to the manufacturer's recommendation. After the scan, the scanned files were saved as stereolithography (STL) files.
Figure 4

Intraoral scanners (IOS): (a) Trios 3 and (b) Dental Wings.

2.3. Measurements

Then, for the scanned files, the measurements were made from computer software using the Rhinoceros 3D modeling software (Rhino, Robert McNeel & Associates for Windows, Washington, DC, USA). The measurements were done in 3 axes (X, Y, and XY and Z) of various lengths as follows (Table 1 and Figures 5 and 6). The measurements made on the printed model with digital caliper were taken as the control (Figure 5).
Table 1

Measurements in three axes (X, Y, XY, and Z) of various lengths.

X-axis Y-axis XY-axis Z-axis
X1: Mesiodistal width on teeth #21 (2 mm) Y1: 2 mm buccolingual width on teeth #27 (2 mm)AR: Diagonal distance from #12 to #27 (65 mm) Z1: Buccal notch on #17 (2 mm)
X2: Distance from #11 to #21 (10 mm) Y2: Buccolingual width on teeth #26 (10 mm) Z2: Buccal notch on #27 (4 mm)
X3: Distance from #12 to #22 (30 mm) Z3: Buccal notch on #16 (6 mm)
X4: Distance from #13 to #31 (40 mm) Y3: Buccolingual width from #25 to #27 (20 mm)AR: Diagonal distance from #22 to #17 (65 mm) Z4: Buccal notch on #27 (8 mm)
X5: Distance from #14 to #41 (50 mm)
X6: Distance from #16 to #61 (60 mm) Y4: Buccolingual width from #23 to #27 (30 mm)
Figure 5

Reference points on the model and various measurements measured on model in different axes: X-axis (a), Y-axis (b), XY-axis (c), and Z-axis (d).

Figure 6

Reference points on the model and various measurements in the X-axis (length) and Y-axis (length) from software of 1 scan.

In addition, the quality of the scans and capturing details of the both scanners were also evaluated.

2.4. Statistical Analysis

Microsoft Excel 2010 and SPSS version 20 (IBM Company, Chicago, USA) were used for the descriptive statistics and expressed as mean and standard deviation. Multiple comparison was done using one-way ANOVA with post hoc using Sheffe to see the significant difference (P < 0.01) between the control (dental model) and scans.

3. Results

Tables 2–4 shows the descriptive statistics of the scans of various lengths in the three axes (X, Y, XY, and Z).
Table 2

Descriptive statistics of measurements of various groups in the X-axis.

MeasurementsGroupsMeanSD95% CI for meanMinMax
Lower boundUpper bound
X1Control1.9380.0391.8881.9871.902.00
Trios 3A1.8760.0321.8361.9161.851.93
Trios 3B1.8860.0361.8411.9311.841.94
Dental Wings1.8760.0571.8051.9461.831.96
Control9.9380.0229.9119.9649.909.95

X2Trios 3A9.8640.0279.8359.8929.849.90
Trios 3B9.9680.0289.93310.0029.929.99
Dental Wings9.8660.0719.7789.9549.799.97
Control29.8520.0629.7729.92629.7829.93

X3Trios 3A29.7340.02229.70629.76129.7029.76
Trios 3B29.9140.03329.87229.95529.8729.95
Dental Wings29.8060.17229.59230.01929.6030.03
Control39.9100.05439.84239.97839.8239.96

X4Trios 3A39.7160.01139.70239.7339.7039.73
Trios 3B39.6980.24839.38940.00739.3939.93
Dental Wings40.1120.17739.89240.33239.8840.28
Control50.030.03549.98650.07449.9850.07

X5Trios 3A49.860.11849.71350.00649.7550.02
Trios 3B50.210.13650.0450.37949.9850.31
Dental Wings50.8320.42350.30651.35750.4351.55
Caliper60.6160.05260.55260.6860.5460.67
Control60.0680.27159.73260.40459.6660.42

X6Trios 3A60.3680.51459.72861.00759.7760.83
Dental Wings61.9520.37461.48762.41761.5562.41

SD = standard deviation; CI = confidence interval for mean; min = minimum; max = maximum.

Table 3

Descriptive statistics of measurements of various groups in Y-axis and XY-axis.

MeasurementsGroupsMeanSD95% CI for meanMinMax
Lower boundUpper bound
Y1Control1.9320.0081.9211.9421.921.94
Trios 3A1.8560.0351.8121.8991.801.89
Trios 3B1.8540.0771.7581.9491.781.97
Dental Wings1.8280.0171.8051.851.801.85

Y2Control9.8760.0219.859.9019.869.91
Trios 3A9.8060.1499.6199.9929.549.89
Trios 3B9.810.0829.7089.9129.749.93
Dental Wings9.80.0749.7079.8929.719.90

Y3Control19.9340.04719.87519.99219.8920.01
Trios 3A19.710.1219.56119.85919.5619.84
Trios 3B19.810.06219.73319.88619.7519.90
Dental Wings19.6360.51818.99320.27818.8120.02

Y4Control30.1620.33929.74830.57529.8630.70
Trios 3A29.9740.39129.48830.45929.3830.35
Trios 3B30.1860.59329.4530.92129.6031.05
Dental Wings29.7540.42529.22530.28229.1730.37

ARControl65.1260.03765.07965.17365.1065.19
Trios 3A64.8720.11964.72465.01964.7865.07
Trios 3B65.1240.17364.90865.33964.9065.32
Dental Wings66.0440.62465.26866.81965.1766.88

ALControl65.0940.05365.02765.1665.0365.15
Trios 3A65.430.48464.82866.03164.8865.81
Trios 3B65.4080.32665.00265.81364.9765.89
Dental Wings65.9420.13365.77666.10765.8466.14

SD = standard deviation; CI = confidence interval for mean; min = minimum; max = maximum.

Table 4

Descriptive statistics of measurements of various groups in the Z-axis.

MeasurementsGroupsMeanSD95% CI for meanMinMax
Lower boundUpper bound
Z1Control2.0140.0231.9852.0421.992.04
Trios 3A2.0180.0371.9712.0641.962.06
Trios 3B1.9980.0471.9382.0571.962.08
Dental Wings1.9900.0631.9112.0681.882.04

Z2Control3.9860.0113.9724.0013.974.00
Trios 3A3.9880.0723.8984.0773.884.05
Trios 3B3.9520.0733.8614.0423.904.08
Dental Wings3.970.0543.9024.0373.894.02

Z3Control6.150.1375.9796.325.976.32
Trios 3A5.9060.2095.6456.1665.716.19
Trios 3B5.9780.0115.9645.9915.965.99
Dental Wings6.1420.1555.9496.3355.926.31

Z4Control8.0460.047.9958.0968.018.11
Trios 3A7.9680.0857.8618.0747.868.04
Trios 3B7.9780.1367.8088.1477.848.14
Dental Wings7.9960.1587.7998.1927.728.11

SD = standard deviation; CI = confidence interval for mean; min = minimum; max = maximum.

The multiple comparisons between the measurements of dental model and the scan are shown in Tables 5–7. It was seen that there was significant difference (P value <0.01) of the measurements X1, X2, Y1, Y2, Y3, AR, AL, Z1, Z3, and Z4 between the dental model and the scans.
Table 5

Multiple comparison of the various measurements in the X-axis of control with other scan groups (Trios 3A, Trios 3B, and Dental Wings).

MeasurementsComparison groupsMean difference P value
X1ControlTrios 3A0.0620.190
Trios HD0.0520.322
Dental Wings0.0620.190

X2ControlTrios 3A0.0740.082
Trios HD−0.030.727
Dental Wings0.0720.093

X3ControlTrios 3A0.1180.299
Trios HD−0.0620.777
Dental Wings0.0460.893

X4ControlTrios 3A0.1940.309
Trios HD0.2120.240
Dental Wings−0.2020.277

X5ControlTrios 3A0.170.719
Trios HD−0.180.683
Dental Wings−0.8020.001

X6ControlTrios 3A0.5480.143
Trios HD0.2480.737
Dental Wings−1.336<0.001

Significant difference at P value <0.05.

Table 6

Multiple comparison of the various measurements in the Y-axis and the XY-axis of control with other scan groups (Trios 3A, Trios 3B, and Dental Wings).

MeasurementsComparison groupsMean differenceSig.
Y1ControlTrios 3A0.0760.092
Trios 3B0.0780.081
Dental Wings0.1040.015

Y2ControlTrios 3A0.070.711
Trios 3B0.0660.746
Dental Wings0.0760.657

Y3ControlTrios 3A0.2240.637
Trios 3B0.1240.910
Dental Wings0.2980.407

Y4ControlTrios 3A0.1880.929
Trios 3B−0.0241.000
Dental Wings0.4080.567

ARControlTrios 3A0.2540.692
Trios 3B0.0021.000
Dental Wings−0.9180.005

ALControlTrios 3A−0.3360.402
Trios 3B−0.3140.459
Dental Wings−0.8480.004

Significant difference at P value <0.05.

Table 7

Multiple comparisons of the various measurements in the Z-axis of control with other scan groups (Trios 3A, Trios 3B, and Dental Wings).

MeasurementsComparison groupsMean differenceSig.
Z1ControlTrios 3A−0.0040.999
Trios 3B0.0160.957
Dental Wings0.0240.872

Z2ControlTrios 3A−0.0021.000
Trios 3B0.0340.836
Dental Wings0.0160.979

Z3ControlTrios 3A0.2440.119
Trios 3B0.1720.366
Dental Wings0.0081.000

Z4ControlTrios 3A0.0780.766
Trios 3B0.0680.831
Dental Wings0.0500.923

Significant difference at P value <0.05.

For the measurements in the X-axis, X1–X4, there was no significant difference of each scan (Trios 3A, Trios 3B, and Dental Wings) compared to the control as shown in Table 5. But, X5–X6, Dental Wings showed there was significant difference (P value <0.01) from the dental model (control). Hence, Dental Wings showed less accuracy at the measurement length 50 mm and 60 mm. Similarly, for the measurements in the Y-axis, Y2–Y4, there was no significant difference of each scan (Trios 3A, Trios 3B, and Dental Wings) compared to the control as shown in Table 6. But, for Y1, Dental Wings showed there was significant difference (P value = 0.015) from the dental model (control). Furthermore, for the measurements in the XY-axis, AR and AL, there was significant difference (P value <0.005) of Dental Wings compared to the control. But, there was no significant difference (P value >0.05) of control with the Trios 3A and Trios 3B for the AR and AL. Hence, Dental Wings showed less accuracy in the measurements. Similarly, for the measurements in the Z-axis, Z1–Z4, there was no significant difference (P=0.05) of control with each scan (Trios 3A, Trios 3B, and Dental Wings) as shown in Table 7. Regarding the quality and capturing details, Trios 3A showed the best results followed by Trios 3B and Dental Wings.

4. Discussion

Digital impressions reduce the patient discomfort; intraoral scanners (IOS) are time-efficient and simplify clinical procedures for the dentist and the laboratory technician, eliminating plaster models and allowing better communication with the dental technician and with patients. The accuracy of which influences the fit of the restorations, an important factor in the longevity of the final restoration [7, 8]. Renne et al. [23] compared 7 different IOS and they found that the Planscan had the best accuracy (trueness and precision) while the 3Shape Trios was found to be the poorest for sextant scanning. The order of trueness for complete arch scanning was as follows: 3Shape D800 > iTero > 3Shape TRIOS 3 > Carestream 3500 >Planscan > CEREC Omnicam > CEREC Bluecam. The order of precision for complete-arch scanning was as follows: CS3500 > iTero > 3Shape D800 > 3Shape TRIOS 3 >CEREC Omnicam > Planscan > CEREC Bluecam. For the secondary outcome evaluating the effect time has on trueness and precision, the complete-arch scan time was highly correlated with both trueness and precision. They concluded that for complete-arch scanning, the 3 Shape Trios was found to have the best balance of speed and accuracy. Park et al. [15] designed an intraoral environment simulator to assess the accuracy of 2 IOS using the simulator and found no difference due to the intraoral environment. The simulator contributes to the higher accuracy of IOS. Mutwalli et al. [24] studied the trueness and precision of different IOS when scanning fully edentulous arch with multiple implants. They found that there were significant differences between all IOS. For the implant measurements, Trios 3 had the lowest trueness, followed by Trios 3 mono and Itero element. Trios had the lowest precision, followed by Itero element and Trios 3 mono. Regarding the interarch distance measurements, Trios 3 had the lowest trueness, followed by Trios 3 mono and Itero element. Trios 3 had the lowest precision, followed by Itero element and Trios 3 mono. But, in our study, Trios 3 presented higher accuracy than Dental Wings and high resolution showed better results. There can be minor errors in the measurements by the IOS in various steps. While scanning 1 arch, generally, the IOS captures around 1200 images. Errors in scanning may be due to overlapping of the partial images, especially in the anterior region [19, 25]. The occurrence of more errors of digital impression in the anterior regions is due to the less structured tooth surface and steep inclines. The superimposition process leads to the deviation. These errors might be reduced or avoided with further software improvements. In addition, there can be errors in computer processing, which may be due to filter algorithms and calibration errors of the scanner [25]. There can be errors in the x-axis, y-axis, and z-axis, but in our study, there were more errors in the z-axis (depth of scanning). The errors can be avoided by a longitudinal measurement of a calibrated length specimen. The dimension measured can be implemented as follows. In the anterior region, X1 (2 mm) represents a scan body or an onlay, X2 (10 mm) represents 1-unit restoration or prosthesis, X3 (30 mm) represents the 4-unit restoration, X4 (40 mm) represents the 6-unit restoration, X5 (50 mm) represents the 10-unit restoration, and X6 (60 mm) represents the 14-unit restoration or full-arch restoration. X1–X4 represent the dimensions in one quadrant in the same arch (upper or lower). X5 and X6 extend to 2 quadrants in the same arch. Trios 3 allows us to record full-arch and high-resolution scans that are more accurate than normal scans, but there was no significant difference. In addition, Dental Wings allow for maximum 6-unit restorations. In the posterior region, Y1 (2 mm) represents a scan body or an onlay, Y2 (10 mm) represents 1-unit restoration or prosthesis, Y3 (20 mm) represents the 2-unit restoration, and Y4 (30 mm) represents the 4-unit restoration. Similarly, AR and AL (65 mm) represent the full-arch restorations. For posterior measurements, all scanners showed acceptable accuracy. But more than 3-unit restoration showed less accuracy (P value <0.05). If we want to fabricate prosthesis with model less, >3-unit restoration is not recommended. For Dental Wings, the accuracy is less for the restoration that crosses the midline; hence, there is a need to be careful. It is less recommended to scan >3-unit prosthesis.

5. Conclusion

Within the limitations of this study, Trios 3 presented higher accuracy, better quality, and captured more details than Dental Wings, and high resolution showed better results. Accuracy of the scanners is affected by the length of the scanning area and scanning pattern. It is less recommended to use Trios 3 for scanning >3-unit prosthesis (50 mm) and that crosses the midline.
  22 in total

Review 1.  Final impressions: a review of material properties and description of a current technique.

Authors:  Nikolaos Perakis; Urs C Belser; Pascal Magne
Journal:  Int J Periodontics Restorative Dent       Date:  2004-04       Impact factor: 1.840

2.  Digitization of simulated clinical dental impressions: virtual three-dimensional analysis of exactness.

Authors:  Anna S K Persson; Agneta Odén; Matts Andersson; Gunilla Sandborgh-Englund
Journal:  Dent Mater       Date:  2009-03-04       Impact factor: 5.304

Review 3.  Digital dentistry: an overview of recent developments for CAD/CAM generated restorations.

Authors:  F Beuer; J Schweiger; D Edelhoff
Journal:  Br Dent J       Date:  2008-05-10       Impact factor: 1.626

4.  Evaluation of the accuracy of 7 digital scanners: An in vitro analysis based on 3-dimensional comparisons.

Authors:  Walter Renne; Mark Ludlow; John Fryml; Zach Schurch; Anthony Mennito; Ray Kessler; Abigail Lauer
Journal:  J Prosthet Dent       Date:  2016-12-23       Impact factor: 3.426

5.  Comparison of digital intraoral scanners by single-image capture system and full-color movie system.

Authors:  Meguru Yamamoto; Yu Kataoka; Atsufumi Manabe
Journal:  Biomed Mater Eng       Date:  2017       Impact factor: 1.300

6.  Nepalese Esthetic Dental (NED) Proportion in Nepalese Population.

Authors:  D Rokaya; J Kitisubkanchana; A Wonglamsam; P Santiwong; T Srithavaj; M Humagain
Journal:  Kathmandu Univ Med J (KUMJ)       Date:  2015 Jul-Sep

Review 7.  Intraoral scanners in dentistry: a review of the current literature.

Authors:  Francesco Mangano; Andrea Gandolfi; Giuseppe Luongo; Silvia Logozzo
Journal:  BMC Oral Health       Date:  2017-12-12       Impact factor: 2.757

8.  A comparison of the accuracy of intraoral scanners using an intraoral environment simulator.

Authors:  Hye-Nan Park; Young-Jun Lim; Won-Jin Yi; Jung-Suk Han; Seung-Pyo Lee
Journal:  J Adv Prosthodont       Date:  2018-02-12       Impact factor: 1.904

9.  Intraoral Digital Impressions for Virtual Occlusal Records: Section Quantity and Dimensions.

Authors:  Eneko Solaberrieta; Asier Garmendia; Aritza Brizuela; Jose Ramon Otegi; Guillermo Pradies; Andras Szentpétery
Journal:  Biomed Res Int       Date:  2016-01-10       Impact factor: 3.411

10.  Trueness and Precision of Three-Dimensional Digitizing Intraoral Devices.

Authors:  Hussam Mutwalli; Michael Braian; Deyar Mahmood; Christel Larsson
Journal:  Int J Dent       Date:  2018-11-26
View more
  4 in total

1.  Comparison of the accuracy of intraoral digital impression system and conventional impression techniques for multiple implants in the full-arch edentulous mandible.

Authors:  Firas-Abdulameer Farhan; Ali-Jameel-Abdul Sahib; Abdalbseet-Ahmad Fatalla
Journal:  J Clin Exp Dent       Date:  2021-05-01

2.  Three-dimensional analysis of the root canal preparation with Reciproc Blue®, WaveOne Gold® and XP EndoShaper®: a new method in vivo.

Authors:  Javier Caviedes-Bucheli; Nestor Rios-Osorio; Diana Usme; Cristian Jimenez; Adriana Pinzon; Jorge Rincón; María M Azuero-Holguin; Alvaro Zubizarreta-Macho; Jose F Gomez-Sosa; Hugo R Munoz
Journal:  BMC Oral Health       Date:  2021-02-25       Impact factor: 2.757

3.  Comparison of Accuracy of Current Ten Intraoral Scanners.

Authors:  Pokpong Amornvit; Dinesh Rokaya; Sasiwimol Sanohkan
Journal:  Biomed Res Int       Date:  2021-09-13       Impact factor: 3.411

4.  A new digital evaluation protocol applied in a retrospective analysis of periodontal plastic surgery of gingival recessions.

Authors:  Tiago Marques; N M Santos; Joana Fialho; J Montero; A Correia
Journal:  Sci Rep       Date:  2021-10-14       Impact factor: 4.379

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.