Comparison of accuracy of cup position using portable navigation versus alignment guide in total hip arthroplasty in supine position.

Masanori Okamoto1, Masashi Kawasaki1, Toshiaki Okura1, Satoshi Ochiai2, Hiroyuki Yokoi1.   

Abstract

BACKGROUND: Accelerometer-based portable navigation systems in supine total hip arthroplasty (THA) have been developed, but there are no reports on the accuracy of cup placement. We aimed to investigate and compare the accuracy of the accelerometer-based portable navigation system versus the acetabular alignment guide placed on the pelvis in THA using the direct anterior approach (DAA). Both devices tracked changes in the pelvic position.
METHODS: In this single-centre, retrospective study, we reviewed 115 hips in 113 patients who underwent primary THA via the DAA using an accelerometer-based portable navigation system in the supine position (portable navigation group) and 106 hips in 101 patients who underwent THA using an acetabular alignment guide (alignment guide group) as controls. Hips were evaluated postoperatively using computed tomography to measure cup orientation. The accuracy of cup orientation was compared between the 2 groups.
RESULTS: Absolute values of inclination error were 3.1° ± 2.2° and 2.9° ± 2.3° (p = 0.708) in the portable navigation and alignment guide groups and those of anteversion error were 2.8° ± 2.3° and 3.7° ± 2.7°, respectively (p = 0.005). The number of cups placed within 10° of error was 98.3% and 96.2% in the portable navigation and alignment guide groups, respectively (p = 0.304). The portable navigation group had significantly more hips (72.2%) placed within a 5° margin of error than did the alignment guide group (56.6%) (p = 0.016).
CONCLUSION: High accuracy in cup placement was achieved using accelerometer-based portable navigation in supine THA. Using a navigation system may contribute to improved long-term outcomes.

Entities:  

Keywords:  Accelerometer; HipAlign; acetabular alignment guide; cup placement; navigation; total hip arthroplasty

Mesh:

Substances:

Year:  2020        PMID: 32126836     DOI: 10.1177/1120700020908788

Source DB:  PubMed          Journal:  Hip Int        ISSN: 1120-7000            Impact factor:   2.135


  6 in total

1.  Accuracy of acetabular cup insertion in an anterolateral supine approach using an accelerometer-based portable navigation system.

Authors:  Masahiro Hasegawa; Yohei Naito; Shine Tone; Hiroki Wakabayashi; Akihiro Sudo
Journal:  J Artif Organs       Date:  2020-09-02       Impact factor: 1.731

2.  Improving Acetabular Component Positioning in Total Hip Arthroplasty: A Cadaveric Study of an Inertial Navigation Tool and a Novel Registration Method.

Authors:  Jobe Shatrov; Daniel Marsden-Jones; Matt Lyons; William L Walter
Journal:  HSS J       Date:  2021-11-02

3.  Image-Less THA Cup Navigation in Clinical Routine Setup: Individual Adjustments, Accuracy, Precision, and Robustness.

Authors:  Corinne A Zurmühle; Benjamin Zickmantel; Matthias Christen; Bernhard Christen; Guoyan Zheng; Joseph M Schwab; Moritz Tannast; Simon D Steppacher
Journal:  Medicina (Kaunas)       Date:  2022-06-20       Impact factor: 2.948

4.  Accuracy of a novel accelerometer-based navigation (Naviswiss) for total hip arthroplasty in the supine position.

Authors:  Masahiro Hasegawa; Yohei Naito; Shine Tone; Akihiro Sudo
Journal:  BMC Musculoskelet Disord       Date:  2022-06-04       Impact factor: 2.562

5.  Accuracy of a portable accelerometer-based navigation system for cup placement and intraoperative leg length measurement in total hip arthroplasty: a cross-sectional study.

Authors:  Hiromasa Tanino; Yasuhiro Nishida; Ryo Mitsutake; Hiroshi Ito
Journal:  BMC Musculoskelet Disord       Date:  2021-03-23       Impact factor: 2.362

6.  Effects of body mass index and range of motion on intraoperative change in pelvic tilt during total hip arthroplasty using the direct anterior approach.

Authors:  Masanori Okamoto; Masashi Kawasaki; Toshiaki Okura; Taisuke Seki; Shiro Imagama
Journal:  BMC Musculoskelet Disord       Date:  2021-03-02       Impact factor: 2.362