Literature DB >> 32120923

Design of Multiplex Lateral Flow Tests: A Case Study for Simultaneous Detection of Three Antibiotics.

Anastasiya V Bartosh1, Dmitriy V Sotnikov1, Olga D Hendrickson1, Anatoly V Zherdev1, Boris B Dzantiev1.   

Abstract

The presented study is focused on the impact of binding zone location on immunochromatographic test strips on the analytical parameters of multiplex lateral flow assays. Due to non-equilibrium conditions for such assays the duration of immune reactions influences significantly the analytical parameters, and the integration of several analytes into one multiplex strip may cause an essential decrease of sensitivity. To choose the best location for binding zones, we have tested reactants for immunochromatographic assays of lincomycin, chloramphenicol, and tetracycline. The influence of the distance to the binding zones on the intensity of coloration and limit of detection (LOD) was rather different. Basing on the data obtained, the best order of binding zones was chosen. In comparison with non-optimal location the LODs were 5-10 fold improved. The final assay provides LODs 0.4, 0.4 and 1.0 ng/mL for lincomycin, chloramphenicol, and tetracycline, respectively. The proposed approach can be applied for multiplexed assays of other analytes.

Entities:  

Keywords:  antibiotics; immunochromatography; multiparametric assay; non-equilibrium interactions; rapid tests

Year:  2020        PMID: 32120923     DOI: 10.3390/bios10030017

Source DB:  PubMed          Journal:  Biosensors (Basel)        ISSN: 2079-6374


  5 in total

Review 1.  Technical considerations to development of serological tests for SARS-CoV-2.

Authors:  Emilie Ernst; Patricia Wolfe; Corrine Stahura; Katie A Edwards
Journal:  Talanta       Date:  2020-11-10       Impact factor: 6.057

2.  A Comparative Study of Approaches to Improve the Sensitivity of Lateral Flow Immunoassay of the Antibiotic Lincomycin.

Authors:  Kseniya V Serebrennikova; Olga D Hendrickson; Elena A Zvereva; Demid S Popravko; Anatoly V Zherdev; Chuanlai Xu; Boris B Dzantiev
Journal:  Biosensors (Basel)       Date:  2020-12-03

3.  Evaluation of Host Serum Protein Biomarkers of Tuberculosis in sub-Saharan Africa.

Authors:  Thomas C Morris; Clive J Hoggart; Novel N Chegou; Martin Kidd; Tolu Oni; Rene Goliath; Katalin A Wilkinson; Hazel M Dockrell; Lifted Sichali; Louis Banda; Amelia C Crampin; Neil French; Gerhard Walzl; Michael Levin; Robert J Wilkinson; Melissa S Hamilton
Journal:  Front Immunol       Date:  2021-02-25       Impact factor: 7.561

4.  Europium Fluorescent Nanoparticles-Based Multiplex Lateral Flow Immunoassay for Simultaneous Detection of Three Antibiotic Families Residue.

Authors:  Yaping Wang; Biao Ma; Miaomiao Liu; Erjing Chen; Ying Xu; Mingzhou Zhang
Journal:  Front Chem       Date:  2021-12-20       Impact factor: 5.221

5.  Switching from Multiplex to Multimodal Colorimetric Lateral Flow Immunosensor.

Authors:  Simone Cavalera; Fabio Di Nardo; Luca Forte; Francesca Marinoni; Matteo Chiarello; Claudio Baggiani; Laura Anfossi
Journal:  Sensors (Basel)       Date:  2020-11-18       Impact factor: 3.576