Clive Buckberry1, Nicholas Hoenich2, Detlef Krieter3, Horst-Dieter Lemke4, Marieke Rüth4, John E Milad1. 1. Quanta Dialysis Technologies Ltd, Alcester, Warwickshire, United Kingdom. 2. Newcastle University, Newcastle upon Tyne, United Kingdom. 3. University of Wurzburg, Wurzburg, Germany. 4. EXcorLab GmbH, Industrie Center Obernburg, Obernburg, Germany.
Abstract
BACKGROUND AND OBJECTIVE: The SC+ haemodialysis system developed by Quanta Dialysis Technologies is a small, easy-to-use dialysis system designed to improve patient access to self-care and home haemodialysis. A prototype variant of the standard SC+ device with a modified fluidic management system generating a pulsatile push-pull dialysate flow through the dialyser during use has been developed for evaluation. It was hypothesized that, as a consequence of the pulsatile push-pull flow through the dialyser, the boundary layers at the membrane surface would be disrupted, thereby enhancing solute transport across the membrane, modifying protein fouling and maintaining the surface area available for mass and fluid transport throughout the whole treatment, leading to solute transport (clearance) enhancement compared to normal haemodialysis (HD) operation. METHODS: The pumping action of the SC+ system was modified by altering the sequence and timings of the valves and pumps associated with the flow balancing chambers that push and pull dialysis fluid to and from the dialyser. Using this unique prototype device, solute clearance performance was assessed across a range of molecular weights in two related series of laboratory bench studies. The first measured dialysis fluid moving across the dialyser membrane using ultrasonic flowmeters to establish the validity of the approach; solute clearance was subsequently measured using fluorescently tagged dextran molecules as surrogates for uraemic toxins. The second study used human blood doped with uraemic toxins collected from the spent dialysate of dialysis patients to quantify solute transport. In both, the performance of the SC+ prototype was assessed alongside reference devices operating in HD and pre-dilution haemodiafiltration (HDF) modes. RESULTS: Initial testing with fluorescein-tagged dextran molecules (0.3 kDa, 4 kDa, 10 kDa and 20 kDa) established the validity of the experimental pulsatile push-pull operation in the SC+ system to enhance clearance and demonstrated a 10 to 15% improvement above the current HD mode used in clinic today. The magnitude of the observed enhancement compared favourably with that achieved using pre-dilution HDF with a substitution fluid flow rate of 60 mL/min (equivalent to a substitution volume of 14.4 L in a 4-hour session) with the same dialyser and marker molecules. Additional testing using human blood indicated a comparable performance to pre-dilution HDF; however, in contrast with HDF, which demonstrated a gradual decrease in solute removal, the clearance values using the pulsatile push-pull method on the SC+ system were maintained over the entire duration of treatment. Overall albumin losses were not different. CONCLUSIONS: Results obtained using an experimental pulsatile push-pull dialysis flow configuration with an aqueous blood analogue and human blood ex vivo demonstrate an enhancement of solute transport across the dialyser membrane. The level of enhancement makes this approach comparable with that achieved using pre-dilution HDF with a substitution fluid flow rate of 60 mL/min (equivalent to a substitution volume of 14.4 L in a 4-hour session). The observed enhancement of solute transport is attributed to the disruption of the boundary layers at the fluid-membrane interface which, when used with blood, minimizes protein fouling and maintains the surface area.
BACKGROUND AND OBJECTIVE: The SC+ haemodialysis system developed by Quanta Dialysis Technologies is a small, easy-to-use dialysis system designed to improve patient access to self-care and home haemodialysis. A prototype variant of the standard SC+ device with a modified fluidic management system generating a pulsatile push-pull dialysate flow through the dialyser during use has been developed for evaluation. It was hypothesized that, as a consequence of the pulsatile push-pull flow through the dialyser, the boundary layers at the membrane surface would be disrupted, thereby enhancing solute transport across the membrane, modifying protein fouling and maintaining the surface area available for mass and fluid transport throughout the whole treatment, leading to solute transport (clearance) enhancement compared to normal haemodialysis (HD) operation. METHODS: The pumping action of the SC+ system was modified by altering the sequence and timings of the valves and pumps associated with the flow balancing chambers that push and pull dialysis fluid to and from the dialyser. Using this unique prototype device, solute clearance performance was assessed across a range of molecular weights in two related series of laboratory bench studies. The first measured dialysis fluid moving across the dialyser membrane using ultrasonic flowmeters to establish the validity of the approach; solute clearance was subsequently measured using fluorescently tagged dextran molecules as surrogates for uraemic toxins. The second study used human blood doped with uraemic toxins collected from the spent dialysate of dialysis patients to quantify solute transport. In both, the performance of the SC+ prototype was assessed alongside reference devices operating in HD and pre-dilution haemodiafiltration (HDF) modes. RESULTS: Initial testing with fluorescein-tagged dextran molecules (0.3 kDa, 4 kDa, 10 kDa and 20 kDa) established the validity of the experimental pulsatile push-pull operation in the SC+ system to enhance clearance and demonstrated a 10 to 15% improvement above the current HD mode used in clinic today. The magnitude of the observed enhancement compared favourably with that achieved using pre-dilution HDF with a substitution fluid flow rate of 60 mL/min (equivalent to a substitution volume of 14.4 L in a 4-hour session) with the same dialyser and marker molecules. Additional testing using human blood indicated a comparable performance to pre-dilution HDF; however, in contrast with HDF, which demonstrated a gradual decrease in solute removal, the clearance values using the pulsatile push-pull method on the SC+ system were maintained over the entire duration of treatment. Overall albumin losses were not different. CONCLUSIONS: Results obtained using an experimental pulsatile push-pull dialysis flow configuration with an aqueous blood analogue and human blood ex vivo demonstrate an enhancement of solute transport across the dialyser membrane. The level of enhancement makes this approach comparable with that achieved using pre-dilution HDF with a substitution fluid flow rate of 60 mL/min (equivalent to a substitution volume of 14.4 L in a 4-hour session). The observed enhancement of solute transport is attributed to the disruption of the boundary layers at the fluid-membrane interface which, when used with blood, minimizes protein fouling and maintains the surface area.
In patients with end-stage renal disease kidney function must be replaced by artificial means (i.e. haemodialysis or peritoneal dialysis) or by transplantation in order to sustain life. The most commonly used treatment is haemodialysis, a process that involves the use of an artificial semi-permeable membrane. During haemodialysis, abnormal patient biochemistry is normalized primarily by diffusion; the fluid gained between treatments is removed by a hydrostatic pressure gradient across the dialyser membrane, a process referred to as ultrafiltration.It is estimated that the number of patients receiving renal replacement therapy globally will increase to 4.9 million by 2025 [1]. The majority of patients receiving haemodialysis do so as outpatients in standalone or facility-based dialysis units, typically three times per week for a minimum of four hours each time. The regimen of kidney replacement therapy is associated with poor health outcomes, can be burdensome for patients and their support networks, and is costly for healthcare payers [2].Haemodialysis offered in the home setting (HHD) is a more cost-effective treatment option in the long term [3]. It also provides patients with the ability to dialyse on a flexible schedule more frequently and/or for longer periods. Importantly, when patients are treated in their own home, they have lower rates of dialysis-related complications, hospitalisations and mortality [4-6]. Moreover, HHD also provides patients with quality-of-life improvements [7].Despite these advantages, uptake of HHD has been low. Several factors have been implicated [8], most notably the fear associated with self-managing haemodialysis treatments at home, which can be a significant barrier when deciding on modality type [9,10].Traditionally, HHD has been performed with machines identical to those used in a hospital setting. Such machines are typically large, cumbersome, intimidating and intrusive in the home setting. Although some have been adapted for home use, more recently manufacturers have begun focusing on machines specifically developed for home therapy use and patient operation [11, 12]. Such an approach has necessitated an improved understanding of industrial design, human factors and ergonomics to ensure that the burden of undertaking treatment in the home is reduced for the patient and their care partner [13, 14].It is desirable for haemodialysis machines to not only be suitable for home use, but also to be suitable for use in dialysis facilities, thereby allowing dialysis programmes to balance and optimize clinical resources and to transition patients from one treatment setting to another using a single platform across the continuum of care. It is with this in mind that the QuantaSC+ personal haemodialysis system was developed.The process of haemodialysis favours the removal of low molecular weight solutes. The removal of middle or high molecular weight solutes is limited by the characteristics of the dialysis membrane or artificial kidney. Limited enhancement of small molecule removal can be achieved by modifying diffusive forces through changes in dialyser design [15, 16], but enhancing the removal of middle or larger molecules requires a different approach [17]. Briefly, such an approach involves the combination of diffusion with convection [18]. This combination may be achieved either by internal filtration [19], haemodiafiltration (HDF) [20], or by the use of a new generation of membranes, such as medium cut-off membranes, in a conventional dialysis setting [21-24]. One of the unintended consequences of these approaches is enhanced protein removal [25].The use of on line haemodiafiltration (olHDF), defined as a blood purification therapy combining diffusive and convective solute transport using a high-flux membrane characterized by an ultrafiltration coefficient greater than 20 mL/h/mmHg/m2 and a sieving coefficient (S) for β2-microglobulin of greater than 0.6, in which the fluid balance is maintained by external infusion of a sterile, non-pyrogenic solution into the patient’s blood, is becoming common, with more than 100,000 patients in Europe and Japan being treated by such an approach [20, 26].olHDF offers a number of advantages compared to haemodialysis [27]; however, the technique does not fit well with the simplified use and time flexibility associated with HHD. On the other hand, expanded dialysis using dialysers utilizing medium cut off membranes requires nothing more than a change in dialyser [21-23].To improve HHD adoption rates and procure the associated benefits, Quanta Dialysis Technologies Ltd (Alcester, UK) has adopted a new approach to haemodialysis system design which encourages patients to safely take control of their own treatments within the home setting (Fig 1). Many design elements of the SC+ system are recognisable in other clinically used dialysis machines but in the SC+ there are a number of key differences, namely that all the elements of the dialysis fluid pathway have been placed onto a disposable cartridge (Fig 2).
Fig 1
Quanta SC+ haemodialysis system.
Fig 2
SC+ dialysate cartridge.
In this paper, a potential future development of the SC+ platform that could enhance solute clearance at higher molecular weights, while remaining true to its design intent of simplicity without compromising standards of care is, described and applied.
Operating principle of the SC+ haemodialysis system
During operation of the SC+ haemodialysis system, the dialysis fluid flows through the cartridge in discrete packets (Fig 3). This is achieved by the application of pneumatic pressure and vacuum to manipulate a flexible PVC membrane that, in turn, opens and closes a sequential series of valves and pump cavities or flow balance chambers. The actuation of the membrane at each valve and cavity is independently operated by a uniquely addressable solenoid valve. During normal operation, the two flow balance chambers are operated “in phase”, such that they simultaneously move fluid to and from the dialyser in 22 mL packets, drawing fresh dialysis fluid into the system and expelling used dialysis fluid during each half cycle. In the current CE certified design used clinically, this sequence is hard coded into the operating software.
Fig 3
Phase varied flow balance pump sequencing.
This study investigated an experimental arrangement whereby the pulsed flow into and out of the dialyser was altered by changing the relative phase of both the valves and pumps associated with each balance chamber. Using this approach, the flow can be either rapidly accelerated through the dialyser or forced across the fibre walls from the dialysate to the blood side and back again (Fig 3). Such an approach enhances the distribution of fresh dialysis fluid around the fibres and disrupts the boundary layer surrounding them, thereby enhancing solute transport.This approach builds on the concept of push-pull in haemodialysis, which has demonstrated enhanced solute clearances compared to conventional haemodialysis without any increased loss of albumin [28-31].
Materials and methods
A stepwise approach was used whereby a proof of concept study was performed initially to demonstrate that the software changes to the pump and valve sequencing produced a significant shift in fluid across the dialyser membrane within the time period available for normal operation at a dialysate flow rate, Qd, of 500 mL/min. Then, using a series of representative molecular markers, the alteration of clearance was quantified and used to form an initial understanding of the dynamics of this development, which would steer future refinements. Finally, the clearance characteristics of the dialysis system using a non-HDF-specific dialyser (Leoceed 18H - Asahi Kasei Medical Co. Ltd, Dusseldorf, Germany) were measured ex vivo with human blood against a benchmark machine used in current clinical practice.All donor blood was received as reagents for ex vivo experiments. Therefore, the human donors did not participate directly in the study. Also, each donor gave his informed consent in writing to a detailed description of what was planned with the donated blood prior to the experiments. This document was written in accordance with the General Data Protection Regulation (GDPR) of the EU. Prospective ethics approval for this study was obtained through members of Quanta’s medical advisory board.
Proof of concept
To demonstrate the proof of concept a dual channel ultrasonic transducer system was attached to the dialysis fluid lines leading to and from the dialyser and used to measure the instantaneous fluid velocity between the flow balance chambers and the dialyser. Simultaneously, the fluctuations in the mass moved across the dialyser were determined gravimetrically and correlated with the transducer data. The blood flow rate, Qb, was maintained at 300 mL/min during the studies to represent a patient whose vascular access cannot deliver the high blood flow rates necessary for HDF therapies.
Solute transport or clearance
Two series of experimental studies to quantify solute transport or clearance were performed. The first series of studies was undertaken at Quanta Dialysis Technologies laboratories using fluorescein-labelled dextrans [32]. A second series of studies was performed independently at eXcorLab GmbH (Obernburg, Germany) in accordance with methods detailed in ISO 8637–1:2017 Extracorporeal systems for blood purification—Part 1: Haemodialysers, haemodiafilters, haemofilters and haemoconcentrators. In this series of studies, the solute clearances of the push-pull mode of the SC+ haemodialysis system and the pre-dilution HDF mode of the Nikkiso DBB-05 system were compared, using the Leoceed 18H dialyser in conjunction with aqueous solution and human blood. The pre-dilution mode was chosen as it is known in the art to be more like push-pull [33]. In addition, aqueous clearance measurements were established with the Nikkiso DBB-05 system in haemodialysis mode (HD mode) to validate the experimental methods and compare the data generated with those given by the dialyser manufacturer.
Establishment of solute clearance using fluorescein-labelled dextrans
The focus in these experiments was more on the relative performance of differing modalities than on absolute clearance. The clearance achieved by the novel push-pull mode of the SC+ system was quantified using molecular analogues for uraemic toxins [32,34] (Table 1). Because of their cost effectiveness, their long association with the measurement of solute transport in membranes used for dialysis and their ready availability across a wide range of molecular weights (4 to 70 kDa), fluorescein-tagged dextrans were chosen and added to dialysis fluid to form a solution with a total conductivity of 14 mS/cm. Fluorescein concentrations were measured using an Aquafluor handheld fluorometer (Turner Designs, San Jose, Ca, USA). Diluted stock solutions were used to assess errors due to variations in temperature, cuvette variation/placement and dialysis fluid composition. The lowest level of detection was 0.2 ppb with a dynamic range of 3 orders of magnitude at a wavelength of 515 nm.
Table 1
Key uraemic molecules and the analogue equivalents used.
Uraemic Toxin (Molecular Weight, Da)
Analogue Equivalent (Molecular Weight, Da)
Urea (60)
Sodium chloride (58.8)
Inulin (522)
Fluorescein (330)
Vitamin B12 (1355)
Dextran + Fluorescein (4000)
β2-microglobulin (11000)
Dextran + Fluorescein (10000)
Myoglobin (16700)
Dextran + Fluorescein (20000)
Immunoglobulin LC (28000–56000)
Dextran + Fluorescein (40000)
Before performing the clearance studies, a series of separate studies was performed to confirm that fluorescein-tagged dextran was not adsorbed by either the PVC tubing used within the extracorporeal tube sets of the SC+ device or by the polysulfone fibres of the dialysers used.The suitability of other markers for smaller molecules in the region of 0.5 kDa to 1.5 kDa was also assessed; however, results demonstrated an unacceptable variability due to their solubility, absorbance to fibres in the dialyser and lack of optical sensitivity. In view of this, fluorescein (which has a molecular weight of 0.33 kDa) was used on its own, with concentrations measured as above. For urea, sodium chloride was used as a proxy, with concentrations measured using a temperature-compensated conductivity meter.Two approaches were used, one in which the fluid containing the marker solutes flowed directly to waste (single pass) while in the other the fluid was recirculated. The selection of single pass or recirculation was based on the practicality of swapping between different molecular weights and the time taken to stabilize operating conditions. The former was used with small molecular weights, while the latter was used for middle and large molecular weight solutes [34].A schematic of the flow circuit is shown in Fig 4. Temperature-controlled water baths were used to stabilize the temperature at 37°C. The “patient” reservoir was suspended from scales to monitor the maintenance of flow balance. Flow rates were calibrated before each run. All machine alarms were enabled except the transmembrane and venous pressure alarms. The flow circuit of the SC+ device (omitted for clarity) provided all the pumping and auto-priming with dialysis fluid for both fluid paths either side of the dialyser.
Fig 4
Experimental flow circuit for measurement of clearance using either single pass or recirculation.
All experiments were allowed to stabilize for 30 minutes in the conventional HD mode before opening valves A and/or C to initiate the clearance testing procedure. Depending on the setting of valves A, B and C, the test solution was either recirculated or passed a single time through the dialyser. All samples were collected at valve B and then transferred to cuvettes for stabilization to room temperature and measurement.In the single pass mode, three samples were taken at 10, 15 and 20 minutes, following stabilization for each molecular weight. When the molecular marker was changed, a 10-minute washout period preceded the stabilization period before sampling.In the recirculation mode, single samples were taken for fluorescence measurement at 0, 3, 6 and 10 minutes following the initiation of the experiment. Samples were then taken at 5- or 10-minute intervals up to 60 or 120 minutes, depending on the molecular weight. To correct for changes due to ultrafiltration, the recirculating volume was compensated for fluid loss between sampling times.The aqueous solute clearance (K) expressed in mL/min was calculated using Eq (1) below:
in which Qb is the blood flow rate (mL/min), Quf is the ultrafiltration rate (mL/min) and Cven and Cart are the venous (dialyser outlet) and arterial (dialyser inlet) solute concentrations, respectively. During the experiments, the ultrafiltration rate, Quf, was maintained at zero and observed at all times by gravimetric flow balance measurements in both single and recirculation modes.
Aqueous solute clearance
Aqueous test solutions (pH 7.4 ± 0.1) composed of urea (1500 mg/L, MW 58 Da), sodium chloride (9000 mg/L, MW 58 Da), creatinine (80 mg/L, MW 113 Da), sodium dihydrogenphosphate x 2 H2O (50 mg/L, MW 156 Da), di-sodium hydrogenphosphate x 2 H2O (230 mg/L, MW 178 Da) and inulin (Sigma, Steinheim, # 57614) (125 mg/L, MW 5200 Da) were used.Dialysers were studied during conventional HD (target Qb = 300 mL/min; target Quf = 0 mL/min; target Qd = 500 mL/min) and pre-dilution HDF (target Qb = 300 mL/min; target Quf = 0 mL/min; target Qsub = 60 mL/min; target Qd = 500 mL/min) conditions in vitro at 37°C using the Nikkiso DBB-05 system. For the QuantaSC+ haemodialysis system, the dialysers were studied in the pulsatile push-pull mode at a Qb = 300 mL/min; target Quf = 0 mL/min; programmed effective target Qsub = 60 mL/min at 37°C.Before the introduction of the test solution, each dialyser was rinsed with saline by the dialysis systems in accordance with the manufacturer’s instructions for use. This was followed by the priming of the extracorporeal circuit using 0.9% saline after which the test solution was introduced and approximately 1.5x the extracorporeal volume was discharged to drain before the commencement of each study. The conditions were checked for stability and samples were drawn simultaneously after 20 minutes from both the blood inlet and outlet to the dialyser, from which the urea, creatinine and phosphate concentrations were determined using a Cobas C111 clinical analyser (Roche Diagnostics GmbH, Mannheim). Inulin concentrations were determined after hydrolysis, using a commercially available assay kit (R-Biopharm, Darmstadt, Germany) using a spectrophotometer (UV-1650PC, Shimadzu Deutschland GmbH, Duisburg, Germany). Solute clearances were calculated in accordance with Eq (1).
Plasma water clearance
For these determinations, heparinised (5 U/mL heparin) whole blood donated by two healthy donors was pooled and adjusted at the start of each experiment to reach a haematocrit of 32 ± 3% (actual range: 31.9–32.5%) and a total protein concentration of 60 ± 5 g/L (actual range: 55.4–61.8 g/L).For the Nikkiso DBB-05 system, the study was performed using pre-dilution HDF (target Qb = 300 mL/min; target Quf = 0 mL/min; target Qsub = 60 mL/min; target Qd = 500 mL/min) conditions ex vivo at 37°C. For the QuantaSC+ haemodialysis system, the study was performed using the pulsatile push-pull mode (Qb = 300 mL/min; target Quf = 0 mL/min; programmed effective target Qsub = 60 mL/min; target Qd = 500 mL/min) ex vivo at 37°C.In both series of measurements, the dialysers attached to the two dialysis systems were primed in parallel to permit simultaneous experiments.Following priming with 0.9% saline, human blood was introduced into the circuit. Approximately 550 g of human blood (~400 mL) was used in each of the circuits. Following the introduction of blood into the circuit, conditions were allowed to equilibrate for 28 minutes. Samples were taken at 30, 32, 34 and 120, 122, 124 and 240, 242 and 244 minutes from the dialyser inlet (Cart) and outlet (Cven) and used to calculate the clearance values at 30, 120 and 240 minutes. A series of five paired experiments were performed.Concentrated haemofiltrate containing β2-microglobulin (11.8 kDa) and myoglobin (17 kDa) extracted from the spent haemofiltrate from chronic kidney diseasepatients was added by infusion into the blood circuit at 28, 118 and 238 minutes before samples were drawn as described above.To determine the albumin and total protein loss across the membrane over 240 minutes, the dialysis fluid flowing to the drain was continuously sampled at a rate of 10 mL/min, using a Ismatec IPC pump (Wertheim, Germany) linked to the dialysis fluid outflow from the dialyser.Albumin content was established using laser nephelometry (BN ProSpec, Siemens Dade-Behring, Marburg, Germany) and total protein content determined using a Cobas C111 clinical analyser (Roche Diagnostics GmbH, Mannheim, Germany).Throughout each experiment, the volume removed from the blood compartment for sampling was substituted by an identical volume of saline.Plasma water clearance (KPW) (mL/min) was calculated according to the Eq (2):
where Qb is the blood flow rate at the inlet of the dialyser, Quf is the ultrafiltration rate, Cven and Cart are the venous (dialyser outlet) and arterial (dialyser inlet) solute concentration, respectively, Hct is the haematocrit at the time of sampling and TP is the total protein concentration (g/L) at the same time point.To account for solute shifts from blood cells, the solute partition coefficient (SPC) was assumed as 0 for β2-microglobulin and myoglobin.Haematocrit was established using an ABX Pentra 60 cell counter (Agon Lab AG, Reichenbach/Stuttgart, Germany).The weight of the blood bag was monitored throughout each experiment. An increase in the weight was noted for the SC+ system despite a target Quf = 0 mL/min, indicating the presence of back filtration (the transfer of fluid from the dialysis fluid pathway into the blood pathway). To minimize the impact of back filtration on blood composition (haematocrit and protein concentration), the pulsatile push-pull mode was occasionally interrupted.Data analysis of plasma water clearance and albumin loss were performed by ANOVA, in which blood pool and the dialysis system (Nikkiso DBB-05 and QuantaSC+) were used as covariates, followed by pairwise comparison according to Tukey using a standard statistical package (Minitab release 17, Additive GmbH, Friedrichsdorf, Germany). A probability of p<0.05 was considered significant.
Results
Estimation of fluid transferred during pulsatile push-pull flow using ultrasonic flow measurement
The ultrasonic flow rates in the standard (unmodified) SC+ system with zero phase delay between the opening of the two flow balance chambers in the inlet and outlet pathways at a dialysate flow rate of 500 mL/min, are shown in Fig 5. Fig 6 shows the effect of a phase delay of 300 ms. Using such a delay, the volume moved across the dialyser membrane was 8 L/hr (equivalent to 133 mL/min) in each direction. The small differences observed between the push and pull portions are a result of minor differences in transmembrane pressure applied by the pneumatic pressure and vacuum cycles.
Fig 5
Ultrasonic flow rates into and out of a dialyser with 0 ms delay between flow balance pumps.
Fig 6
Ultrasonic flow rates into and out of a dialyser with a 300 ms delay between flow balance pumps.
With the phase delay, large volumes of fluid could be moved across the membrane. In practice this was limited by the residence time of the bolus within the dialyser. In order to match the Nikkiso DBB-05 system when used in HDF mode, for the experimental studies described the phase delay was reduced to 200 ms (60 mL/min or 3.6 L/hr). The rationale for this reduction was twofold: first, it matched the substitution fluid infusion rate delivered when using the same dialyser in conventional on line HDF mode; second, it optimized the residence time of a bolus of fluid within the dialyser when in the pulsatile push-pull mode.
Solute clearance using fluorescein-labelled dextrans
Having established the principle of pulsatile push-pull flow, a series of studies were performed to quantify the magnitude of enhancements that this approach delivers by using 4, 10 and 20 kDa fluorescein-labelled dextrans. The studies used the recirculating experimental set up shown in Fig 4. The use of recirculation allowed for repeated measurements to be made, mitigating any effect of collecting small samples at points that might not have been in perfect sequence with the push-pull cycle.Data collected for each of the fluorescein-labelled dextrans are shown in Fig 7a–7c where equivalent data established during normal operation of the SC+ system are also shown.
Fig 7
a. Variation of clearance with time using the SC+ for 4 kDa dextran in conventional flow mode and pulsatile push-pull mode. b. Variation of clearance with time using the SC+ for 10 kDa dextran in conventional flow mode and pulsatile push-pull mode. c. Variation of clearance with time using the SC+ for 20 kDa dextran in conventional flow mode and pulsatile push-pull mode.
a. Variation of clearance with time using the SC+ for 4 kDa dextran in conventional flow mode and pulsatile push-pull mode. b. Variation of clearance with time using the SC+ for 10 kDa dextran in conventional flow mode and pulsatile push-pull mode. c. Variation of clearance with time using the SC+ for 20 kDa dextran in conventional flow mode and pulsatile push-pull mode.All points shown are the average of three measurements and are shown with +/- one standard deviation error bars.The relationship between the observed clearances and molecular weight for the SC+ system with and without the push-pull modification using the same dialyser type (Asahi Kasei Leoceed 18H) is shown in Fig 8. The values shown are the mean values established during the individual experiments.
Fig 8
Relationship between solute clearance and molecular weight in conventional flow mode and pulsatile push-pull mode with a phase delay of 200 ms for the Quanta SC+ dialysis system.
Aqueous solute clearances
Aqueous solute clearance data established using urea, creatinine, phosphate and inulin, based on three experiments (N = 3) conducted at eXcorLab for the SC+ system incorporating pulsatile push-pull flow, were compared with those established using the Nikkiso DBB-05 dialysis system in conventional HD and pre-dilution HDF modes and are shown in Table 2, where the manufacturer’s product insert data for the dialyser used are also provided. The clearance measurements gathered were in broad agreement with the values specified by the manufacturer for the Leoceed 18H haemodialyser. The modified QuantaSC+ system incorporating the pulsatile push-pull flow mode yielded slightly lower clearances for urea, creatinine and phosphate compared to the Nikkiso DBB-05 system, when used in both conventional HD and pre-dilution HDF modes. The observed deviation is unlikely to be of clinical importance and was attributed to reduced diffusive forces resulting from dilution. This observation is in accordance with those of Ficheux et al [35]. For inulin, the QuantaSC+ system showed slightly higher clearances compared to the Nikkiso DBB-05 system in both HD and pre-dilution HDF modes.
Table 2
Aqueous solute clearances for low molecular weights.
Clearance (mL/min), mean +/- SD & N = 3
Machine
Mode
Urea
Creatinine
Phosphate
Inulin
SC+
Pulsatile Push-Pull HDF
254±7
239±9
227±9
114±13
Nikkiso DBB-05
HD
276±3
260±5
248±5
106±1b
Nikkiso DBB-05
Pre-dilution HDF
266±7
249±8
236±7
100±16
Manufacturer’s specificationa
HD
274±7
260
247
n/a
a Leoceed 18H Haemodialyser
All data shown are the mean of three experiments except
bmean of two experiments
a Leoceed 18H HaemodialyserAll data shown are the mean of three experiments exceptbmean of two experiments
Plasma water solute clearance and albumin loss
Plasma water solute clearances measured at 30, 120 and 240 minutes are shown in Table 3. Data are presented as mean ± SD based on five experiments with three samples acquired at each time point resulting in N = 15. Overall, the modified SC+ was comparable to the Nikkiso DBB-05 system in pre-dilution HDF mode.
Table 3
Plasma water solute clearances for middle weight molecules.
Sampling time (min)
Clearance (mL/min), mean +/-SD (N = 15)
Machine
Mode
β2-microglobulin
Myoglobin
SC+
Pulsatile Push-Pull HD
30
61±10
35±11
120
59±7
26±15
240
66±5
30±9a
AUC (mL*min)
12846
6098
Nikkiso DBB-05
Pre-dilution HDF
30
65±9
32±11
120
63±10
26±11
240
57±10
18±8a
AUC (mL*min)
12960
5250
a means are statistically significant from each other (p<0.05) at the same time point
a means are statistically significant from each other (p<0.05) at the same time pointComparing the different time points (30, 120 and 240 minutes), the QuantaSC+ shows constant clearances over all time points. Calculation of the area under the curve for clearance for β2-microglobulin and myoglobin indicates that there is equivalence for β2-microglobulin between the two devices. However, the SC+ shows a 16% improvement for myoglobin overall (30–240 minutes) and a 28% increase in the last two hours (120–240 minutes) of the study period.Mean albumin loss data for the whole study period are given in Table 4 based on five experiments. SC+ with the pulsatile push-pull flow configuration showed lower albumin loss compared to the Nikkiso DBB-05 in pre-dilution HDF mode, but the difference was small and not statistically significant.
Table 4
Albumin loss over 240 minutes.
Albumin Loss (g), mean +/-SD (N = 5)
Machine
Mode
SC+
Pulsatile Push-Pull HDF
0.80±0.49
Nikkiso DBB-05
Pre-dilution HDF
0.88±0.40
All data shown are the mean of five experiments.
All data shown are the mean of five experiments.
Discussion
Although HDF offers a number of advantages over conventional haemodialysis, and is accepted in Europe and Japan, it is less commonly used in the United States [36]. Clinical use of HDF requires increased dialysis fluid purity, additional sterile tubing sets for fluid infusion and additional blood side pumps. This increases complexity and is at odds with self-managed treatments in terms of technology simplification or ease of use. Furthermore, solute transport during HDF is associated with membrane fouling, leading to a loss in performance, increased incidence of nuisance alarms and the need for nursing interventions [25]. Such fouling has been shown to lead to a decrease in ultrafiltration coefficient (KUf) of approximately 12% during HD and 16% during post-dilution HDF over a 3-hour period [35]. On the other hand, the push-pull approach does not require external substitution fluid and relies on alternate repetitions of forward and backward filtration across the dialyser membrane during dialysis treatment [37].In the current series of studies a simple adjustment to the controlling software of the SC+ system alters the sequence and timing of the valves and pumps associated with the flow balancing chambers that push and pull the dialysis fluid through the dialyser and introduces a phase shift. Ultrasonic measurements confirmed that volumes up to 8 L/hr could be easily moved across the membrane using such an approach.Experiments with dextran yielded an increase in clearance for all molecular weights tested. In ex vivo comparative testing using the Nikkiso DBB-05 machine operating in pre-dilution HDF mode (Qsub = 60 mL/min), the modified SC+ produced not only an equivalent clearance, but this clearance was maintained over the full duration of the experimental treatment period (240 minutes) and was without any increase in the loss of albumin.During dialysis, when uraemic solutes are removed from the blood through the dialyser membrane with a variable pore size distribution, concentration polarization occurs and a protein gel layer develops on the membrane surface. The concentration polarization may be viewed as an additional mass boundary layer affecting solute transport, while protein gel layer offers additional structural resistance affecting fluid transport [38,39].The improved results observed using the SC+ system with a modified fluidic management system suggest a disruption of the boundary layers at the blood-membrane interface. This minimizes protein fouling and maintains the surface area available for mass and fluid transport throughout the whole treatment period. Further studies have been initiated to validate this hypothesis, by specifically quantifying the mitigation of the protein layer build up during the process of boundary layer disruption.The potential functional and usability advantages of this approach are multiple: no additional sterile tubing or pumps are required to deliver a functional performance that is comparable with that achieved by HDF; there is no change in system complexity as far as the patient or user is concerned as a shift from conventional haemodialysis to pulsed push-pull haemodialysis can be achieved by the push of a button; and, because pressure fluctuations in the pulsed push-pull HD mode are lower than in HDF, there is a reduction of membrane fouling and the incidence of alarms during use is likely to be reduced. Additionally, this technique may be suitable for use over extended treatment periods, such as nocturnal haemodialysis.All these advantages make the system more suitable for self-care or HHD applications. Importantly, this approach is also fully compatible with all current and future dialyser developments, although further work will be required to confirm that the enhanced performance observed is maintained for a range of clinically used haemodialysers.While the approach described is relatively simple in principle, a number of technical and regulatory challenges remain. Of these, some are relatively simple to address, such as the re-tuning of venous and transmembrane pressure alarms. Others, such as maintaining an accurate flow balance, are more challenging. The most challenging aspect is the attainment and maintenance of microbiological quality during each treatment and regulatory approval.During HDF, the fluid infused is produced continuously on line by the passage of a fraction of the dialysis fluid through bacterial and endotoxin retentive filters validated to produce sterile and non-pyrogenic fluids.Whereas the production of on line infusion fluid involves a treatment cascade in which multiple filters may be used [40], the current approach uses the dialyser to ensure that biological contaminants are not transferred from the dialysis fluid to the patient. This approach relies on the fact that, when using ultrapure dialysis fluid, any biological contaminants present in the fluid are retained or adsorbed within the ultrastructure of the membrane wall and do not pass into the blood stream. Clinically used polymer membranes have differing adsorptive capacities [23, 38, 41] and consequently the potential exists for the transfer of endotoxin fragments and other bacterial substances present in the dialysis fluid into the patient’s blood either by convective transfer (back filtration) or by movement down the concentration gradient (back diffusion) [41, 42]. The long-term clinical relevance of this remains an unexplored aspect of haemodialysis therapy.As the membrane contained in the dialyser itself becomes the final barrier to endotoxins and endotoxin fragments in the sterility chain/cascade, detailed risk assessments will be required to ensure that the required sterility assurance level is achieved and regulatory compliance is met. This will form the primary focus for future development of the approach described.These studies are not without limitations. They are preliminary studies to confirm proof of concept and further studies will be required to establish optimal settings of the push-pull cycle and ultimately clinical trials in vivo. The studies were performed at two different sites, and minor differences may have been introduced by the operating settings. The duration of each study was also short, and longer studies will be required to determine stability. Future studies are planned to gather these data.In the approach described, back filtration and ultrafiltration repeat in a relatively short time, and despite a large amount of filtration, the probability that some ultrafiltrate comes directly from dialysate back filtered during a previous phase cannot be excluded. This, in turn, will influence the results through a reduction in solute concentrations caused by dilution and may manifest as an efficiency reduction. Although the in vitro and ex vivo experiments described have shown that alternating back filtration has a positive influence on inhibiting concentration polarization and permeability reduction, further studies, in terms of pulse frequencies and stroke volumes, will be required to optimize settings.
Recorded results and analysis of aqueous clearance.
Mean and standard deviation error of each condition.(XLSX)Click here for additional data file.Mean and standard deviation error of each condition.(XLSX)Click here for additional data file.(XLSX)Click here for additional data file.(XLSX)Click here for additional data file.26 Nov 2019PONE-D-19-16719Enhancement of solute clearance using pulsatile push-pull dialysate flow for the QuantaSC+: a novel clinic-to-home haemodialysis systemPLOS ONEDear Professor Buckberry,Thank you for submitting your manuscript to PLOS ONE. After careful consideration, we feel that it has merit but does not fully meet PLOS ONE’s publication criteria as it currently stands. Therefore, we invite you to submit a revised version of the manuscript that addresses the points raised during the review process.We would appreciate receiving your revised manuscript by Jan 10 2020 11:59PM. When you are ready to submit your revision, log on to https://www.editorialmanager.com/pone/ and select the 'Submissions Needing Revision' folder to locate your manuscript file.If you would like to make changes to your financial disclosure, please include your updated statement in your cover letter.To enhance the reproducibility of your results, we recommend that if applicable you deposit your laboratory protocols in protocols.io, where a protocol can be assigned its own identifier (DOI) such that it can be cited independently in the future. For instructions see: http://journals.plos.org/plosone/s/submission-guidelines#loc-laboratory-protocolsPlease include the following items when submitting your revised manuscript:A rebuttal letter that responds to each point raised by the academic editor and reviewer(s). This letter should be uploaded as separate file and labeled 'Response to Reviewers'.A marked-up copy of your manuscript that highlights changes made to the original version. This file should be uploaded as separate file and labeled 'Revised Manuscript with Track Changes'.An unmarked version of your revised paper without tracked changes. This file should be uploaded as separate file and labeled 'Manuscript'.Please note while forming your response, if your article is accepted, you may have the opportunity to make the peer review history publicly available. The record will include editor decision letters (with reviews) and your responses to reviewer comments. If eligible, we will contact you to opt in or out.We look forward to receiving your revised manuscript.Kind regards,Pasqual Barretti, Ph.D., MDAcademic EditorPLOS ONEJournal Requirements:When submitting your revision, we need you to address these additional requirements.1. Please ensure that your manuscript meets PLOS ONE's style requirements, including those for file naming. The PLOS ONE style templates can be found at http://www.journals.plos.org/plosone/s/file?id=wjVg/PLOSOne_formatting_sample_main_body.pdf and http://www.journals.plos.org/plosone/s/file?id=ba62/PLOSOne_formatting_sample_title_authors_affiliations.pdf2. Please update the ethics statement in the online submission form and the methods section of the manuscript to include a statement that you obtained prospective ethics approval for this study from your medical advisory board3. Please amend either the abstract on the online submission form (via Edit Submission) or the abstract in the manuscript so that they are identical.4. Thank you for stating the following in the Competing Interests section:'The authors have declared that no competing interests exist'We note that one or more of the authors are employed by a commercial company: Quanta Dialysis Technologies Ltd & EXcorLab GmbH.Please provide an amended Funding Statement declaring this commercial affiliation, as well as a statement regarding the Role of Funders in your study. If the funding organization did not play a role in the study design, data collection and analysis, decision to publish, or preparation of the manuscript and only provided financial support in the form of authors' salaries and/or research materials, please review your statements relating to the author contributions, and ensure you have specifically and accurately indicated the role(s) that these authors had in your study. You can update author roles in the Author Contributions section of the online submission form.Please also include the following statement within your amended Funding Statement.“The funder provided support in the form of salaries for authors [insert relevant initials], but did not have any additional role in the study design, data collection and analysis, decision to publish, or preparation of the manuscript. The specific roles of these authors are articulated in the ‘author contributions’ section.”If your commercial affiliation did play a role in your study, please state and explain this role within your updated Funding Statement.2. Please also provide an updated Competing Interests Statement declaring this commercial affiliation along with any other relevant declarations relating to employment, consultancy, patents, products in development, or marketed products, etc.Within your Competing Interests Statement, please confirm that this commercial affiliation does not alter your adherence to all PLOS ONE policies on sharing data and materials by including the following statement: "This does not alter our adherence to PLOS ONE policies on sharing data and materials.” (as detailed online in our guide for authors http://journals.plos.org/plosone/s/competing-interests) . If this adherence statement is not accurate and there are restrictions on sharing of data and/or materials, please state these. Please note that we cannot proceed with consideration of your article until this information has been declared.Please include both an updated Funding Statement and Competing Interests Statement in your cover letter. We will change the online submission form on your behalf.Please know it is PLOS ONE policy for corresponding authors to declare, on behalf of all authors, all potential competing interests for the purposes of transparency. PLOS defines a competing interest as anything that interferes with, or could reasonably be perceived as interfering with, the full and objective presentation, peer review, editorial decision-making, or publication of research or non-research articles submitted to one of the journals. Competing interests can be financial or non-financial, professional, or personal. Competing interests can arise in relationship to an organization or another person. Please follow this link to our website for more details on competing interests: http://journals.plos.org/plosone/s/competing-interestsAdditional Editor Comments (if provided):Authors must fully answer the questions of the reviewers, in particular of the reviewer 1. After my reading of the text my decision is major revision[Note: HTML markup is below. Please do not edit.]Reviewers' comments:Reviewer's Responses to QuestionsComments to the Author1. Is the manuscript technically sound, and do the data support the conclusions?The manuscript must describe a technically sound piece of scientific research with data that supports the conclusions. Experiments must have been conducted rigorously, with appropriate controls, replication, and sample sizes. The conclusions must be drawn appropriately based on the data presented.Reviewer #1: YesReviewer #2: Yes**********2. Has the statistical analysis been performed appropriately and rigorously?Reviewer #1: NoReviewer #2: Yes**********3. Have the authors made all data underlying the findings in their manuscript fully available?The PLOS Data policy requires authors to make all data underlying the findings described in their manuscript fully available without restriction, with rare exception (please refer to the Data Availability Statement in the manuscript PDF file). The data should be provided as part of the manuscript or its supporting information, or deposited to a public repository. For example, in addition to summary statistics, the data points behind means, medians and variance measures should be available. If there are restrictions on publicly sharing data—e.g. participant privacy or use of data from a third party—those must be specified.Reviewer #1: YesReviewer #2: Yes**********4. Is the manuscript presented in an intelligible fashion and written in standard English?PLOS ONE does not copyedit accepted manuscripts, so the language in submitted articles must be clear, correct, and unambiguous. Any typographical or grammatical errors should be corrected at revision, so please note any specific errors here.Reviewer #1: YesReviewer #2: Yes**********5. Review Comments to the AuthorPlease use the space provided to explain your answers to the questions above. You may also include additional comments for the author, including concerns about dual publication, research ethics, or publication ethics. (Please upload your review as an attachment if it exceeds 20,000 characters)Reviewer #1: Although it is known that the Push / pull mode is advantageous for clearances because of anti-fouling, it is highly novel that the Push / pull mode is incorporated into a home hemodialysis machine and put into practical use.・Figs. should be as clear as possible.・Add the number of original data number and standard deviation to the table and Fig.Reviewer #2: Buckberry CH and coworkers are submitting a study assessing the potential benefit of adding convective clearance to the SC+ hemodialysis system developed by Quanta. By modifying the time pressure algorithm of filling/emptying cassette chambers they generate a push-pull like flow. As proof of concept they developed an invitro study showing that new algorithm was able to generate this alternate push-pull flow by US. They moved subsequently to a prototype with bench testing of various solute clearances (fluorescein-tagged dextran of various molecular weight) thereafter with invitro blood simulation (conductivity as surrogate of urea, myoglobin and B2M) and compared the modified SC+ to a predilution HDF model using conventional HD and predilution HDF system (high flux filter and Nikkiso DBS5 machine). In brief, the authors showed that their modified SC+ system ensured push-pull flow increasing instantaneous clearances by 10 to 15% and compared favorably with that achieved using pre-dilution HDF with a substitution fluid flow rate of 60 ml/min with the same dialyzer and marker molecules in blood simulated experiment. Interestingly clearance values using the push-pull method on the SC+ system, were maintained over the duration of treatment.This is an interesting concept based on invitro studies taking benefits of minimal changes and time pressure algorithm modification in an existing innovative hemodialysis device. Now, the study raises several concerns that need to be addressed for clarification and better understanding:1. This is a proof of concept study and not a clinical validation study meaning that further clinical trials in vivo are required to confirm their findings.2. It is not clear why the modified SC+ with push-pull flow was compared with predilution HDF since my understanding of the algorithm modification is correct, it should be better suited to be compared with postdilution HDF. Internal filtration process associated with push-pull flow regime is mimicking postdilution action and not predilution.3. In the study design, it not clear to me why solute clearances of modified SC+ developing push-pull like flow were not compared with standard SC+ algorithm machine. In other words, purely diffusive HD versus added convective component. Such design would have been better understood for showing the added value of the new algorithm.4. Blood based invitro study is confusing and not necessarily correct since some clearance measurements are performed in a single-pass for small molecules and others in a recirculating pass for larger molecules. It would have been preferable to perform them in the same mode based on tank recirculation with log transformation time concentration decline to obtain a true overall clearance based on the slope decline. Furthermore, there is no mass balance calculation from blood and dialysate side to validate clearance calculation.5. Surprisingly, solute clearances obtained in predilution HDF mode with DBS5 machine are lower over the all spectrum of markers than those obtained in pure hemodialysis. This is struggling and should be explained since it is not in line with the known fact that solute clearances are higher in HDF in particular for larger molecular compounds. One can suspect something went wrong with sampling due to predilution mode or calculations or both. In this setting one can expect clearances higher by about 30 ml/min through molecular weight spectrum.6. From a presentation and wording perspective, introduction and discussion are too long and more related to the benefits of home therapy using the SC+ device. This is not the topic of this study. The authors should stay focused on the aim of the study which is to show some benefits on solute clearances by adding a convective component with a modified algorithm. In addition, the authors referred to benefits of using internal transport phenomenons using MCO membrane in term of solute clearances in standard HD. If this is the case, why not using MCO membrane in standard HD, instead of this modified new algorithm intended to increase convective clearances? How does it compare? What are benefits and risks of combining them?**********6. PLOS authors have the option to publish the peer review history of their article (what does this mean?). If published, this will include your full peer review and any attached files.If you choose “no”, your identity will remain anonymous but your review may still be made public.Do you want your identity to be public for this peer review? For information about this choice, including consent withdrawal, please see our Privacy Policy.Reviewer #1: NoReviewer #2: No[NOTE: If reviewer comments were submitted as an attachment file, they will be attached to this email and accessible via the submission site. Please log into your account, locate the manuscript record, and check for the action link "View Attachments". If this link does not appear, there are no attachment files to be viewed.]While revising your submission, please upload your figure files to the Preflight Analysis and Conversion Engine (PACE) digital diagnostic tool, https://pacev2.apexcovantage.com/. PACE helps ensure that figures meet PLOS requirements. To use PACE, you must first register as a user. Registration is free. Then, login and navigate to the UPLOAD tab, where you will find detailed instructions on how to use the tool. If you encounter any issues or have any questions when using PACE, please email us at figures@plos.org. Please note that Supporting Information files do not need this step.15 Jan 2020Response to reviewers PONE-D-19-16719Comments to the Author1. Is the manuscript technically sound, and do the data support the conclusions?The manuscript must describe a technically sound piece of scientific research with data that supports the conclusions. Experiments must have been conducted rigorously, with appropriate controls, replication, and sample sizes. The conclusions must be drawn appropriately based on the data presented.Reviewer #1: YesReviewer #2: YesAuthor: Accepted2. Has the statistical analysis been performed appropriately and rigorously?Reviewer #1: NoReviewer #2: YesAuthor: Reviewer 1 has not given any details that substantiate their statement so I will add detail as I believe to be appropriate. All the data was analysed using either Excel or ‘Minitab’ that are both recognised with for such purposes by the FDA when validated appropriately as they were. In the case of the independent external test house, excorlab, that was contracted for some of the more complex tests they were specifically chosen because of their accreditation for such work under ISO 17025 both experimentally and analytically.3. Have the authors made all data underlying the findings in their manuscript fully available?The PLOS Data policy requires authors to make all data underlying the findings described in their manuscript fully available without restriction, with rare exception (please refer to the Data Availability Statement in the manuscript PDF file). The data should be provided as part of the manuscript or its supporting information, or deposited to a public repository. For example, in addition to summary statistics, the data points behind means, medians and variance measures should be available. If there are restrictions on publicly sharing data—e.g. participant privacy or use of data from a third party—those must be specified.Reviewer #1: YesReviewer #2: YesAuthor: Accepted4. Is the manuscript presented in an intelligible fashion and written in standard English?PLOS ONE does not copyedit accepted manuscripts, so the language in submitted articles must be clear, correct, and unambiguous. Any typographical or grammatical errors should be corrected at revision, so please note any specific errors here.Reviewer #1: YesReviewer #2: YesAuthor: Accepted5. Review Comments to the AuthorPlease use the space provided to explain your answers to the questions above. You may also include additional comments for the author, including concerns about dual publication, research ethics, or publication ethics. (Please upload your review as an attachment if it exceeds 20,000 characters)Reviewer #1: Although it is known that the Push / pull mode is advantageous for clearances because of anti-fouling, it is highly novel that the Push / pull mode is incorporated into a home hemodialysis machine and put into practical use.・Figs. should be as clear as possible.・Add the number of original data number and standard deviation to the table and Fig.Author: Addressed with changes to table and text regarding sample size and SD.Reviewer #2: Buckberry CH and coworkers are submitting a study assessing the potential benefit of adding convective clearance to the SC+ hemodialysis system developed by Quanta. By modifying the time pressure algorithm of filling/emptying cassette chambers they generate a push-pull like flow. As proof of concept they developed an invitro study showing that new algorithm was able to generate this alternate push-pull flow by US. They moved subsequently to a prototype with bench testing of various solute clearances (fluorescein-tagged dextran of various molecular weight) thereafter with invitro blood simulation (conductivity as surrogate of urea, myoglobin and B2M) and compared the modified SC+ to a predilution HDF model using conventional HD and predilution HDF system (high flux filter and Nikkiso DBS5 machine). In brief, the authors showed that their modified SC+ system ensured push-pull flow increasing instantaneous clearances by 10 to 15% and compared favorably with that achieved using pre-dilution HDF with a substitution fluid flow rate of 60 ml/min with the same dialyzer and marker molecules in blood simulated experiment. Interestingly clearance values using the push-pull method on the SC+ system, were maintained over the duration of treatment.This is an interesting concept based on invitro studies taking benefits of minimal changes and time pressure algorithm modification in an existing innovative hemodialysis device. Now, the study raises several concerns that need to be addressed for clarification and better understanding:1. This is a proof of concept study and not a clinical validation study meaning that further clinical trials in vivo are required to confirm their findings.Author: The reviewer is correct and this has now been clarified in the submission2. It is not clear why the modified SC+ with push-pull flow was compared with predilution HDF since my understanding of the algorithm modification is correct, it should be better suited to be compared with postdilution HDF. Internal filtration process associated with push-pull flow regime is mimicking postdilution action and not predilution.Author: The reviewer makes a valid observation. In post dilution the substitution fluid will first pass to the patient before returning to the dialyser, because the fluid is added post the dialyser. In pre-dilution HDF it is added just prior to the dialyser. As a consequence, the dilution ratio is different. In push-pull the substitution fluid is added directly into the dialyser so is much more akin to pre-dilution. This likeness is also explained more fully by Tattersall ‘Online haemodiafiltration: definition, dose quantification and safety revisited’ Nephrol Dial Transplant (2013) 22: 542-550. This has been addressed and referenced in the text.3. In the study design, it not clear to me why solute clearances of modified SC+ developing push-pull like flow were not compared with standard SC+ algorithm machine. In other words, purely diffusive HD versus added convective component. Such design would have been better understood for showing the added value of the new algorithm.Author: Again a valid point made by the reviewer. The reasons are as follows. Firstly, in practice there are many ways to program the push-pull mechanism on the SC+ device unlike the original methods proposed in this area by Shinzato which was a simple alternating movement of fluid. In our case we vary the cycle to be one of push, push, push followed by pul,l pull, pull. When can also vary the volume in each cycle and the phase depending upon how we wish to disrupt the boundary layer around the fibres. We have chosen therefore to just make a simple comparison in order to first introduce the principle that it can be done. Secondly, we cannot also compare the baseline SC+ to any other machine because of the pulsatile nature of its dialysate flow, compared to the steady state flow of all current dialysis machine in haemodialysis. This would have required a significantly more complex suite of experiments and deconvolution in the analytical phase.4. Blood based invitro study is confusing and not necessarily correct since some clearance measurements are performed in a single-pass for small molecules and others in a recirculating pass for larger molecules. It would have been preferable to perform them in the same mode based on tank recirculation with log transformation time concentration decline to obtain a true overall clearance based on the slope decline. Furthermore, there is no mass balance calculation from blood and dialysate side to validate clearance calculation.Author: Good points raised again. We found during the course of method development that single-pass and recirculation methods were complimentary. Broadly speaking markers for low molecular weights were cheap, abundant and could be monitored continuously, so single pass was used. The middle weight molecules tagged with fluorescent material we more expensive and complicated to manage to maintain constant density, in particular, hence the recirculation was more effective with constant stirring. Critically, the samples taken, both volume and timing of have to be sympathetic to the push-pull cycle to account for the dilution effect of the substitution fluid. This was done and accounted for in the flow balance calculations for net fluid removal error as prescribed under IEC 60601-2-16: 2014 in all experiments. The actual values were calculated for in Eqn2 in the Quf term. I have added a comment to this effect the text.5. Surprisingly, solute clearances obtained in predilution HDF mode with DBS5 machine are lower over the all spectrum of markers than those obtained in pure hemodialysis. This is struggling and should be explained since it is not in line with the known fact that solute clearances are higher in HDF in particular for larger molecular compounds. One can suspect something went wrong with sampling due to predilution mode or calculations or both. In this setting one can expect clearances higher by about 30 ml/min through molecular weight spectrum.Author: We have reviewed the analysis as prompted by the reviewers comment which is valid and well made, we had internal debate on these results too. Allow us to make our counterpoints. We can confirm that following an audit at the time we found no errors in the data collection method or analysis that formed the basis of table 2. We observe that the effect is present for both machines without bias. We also point out to the reviewer that for Inulin the SC+ in push-pull mode is greater than the Nikkiso in HD so it is not true for all spectrum markers as the Reviewer suggests. It is our observation that previous studies that compare HD with HDF modalities the dialyser used will vary. HDF therapies typically employ HDF specific dialysers which are 10% larger in surface area particularily in post-dilution HDF. So often you may see Fx80 dialysers compared to Fx800. in addition a higher dialysate flow rate to counteract the dilution is often employed, so rather than 500ml/min you will see 600ml/min or higher. This gives a bias which we have avoided and may account for the reviewers perception.For transparency we included the manufacturers data in HD mode to compare to the experimental data HD data on the Nikkiso and replicated this very well for urea, Creatine and Phosphate again confirming our methodology experimentally and analytically.What we found most interesting is the effect of blood flow rate Qb to all methods. In pre and post HDF the Qsub is added to the Qb and this amplifies the calculated rate of clearance. In push-pull as we have deployed it Qsub is not strictly present in the same way, especially as it is first negative then positive direction of flow, so whilst the calculation is there the physical action is very different. This is also different to how push-pull was employed by Shinzato. What is important in this study and it’s purpose of course, is how do the two systems compare, we therefore chose to apply the equations as they currently stand without bias. We are planning to write further papers that analytically model our embodiment of pulsatile push-pull as we felt at this stage it would detract from this early proof-of-concept.6. From a presentation and wording perspective, introduction and discussion are too long and more related to the benefits of home therapy using the SC+ device. This is not the topic of this study. The authors should stay focused on the aim of the study which is to show some benefits on solute clearances by adding a convective component with a modified algorithm. In addition, the authors referred to benefits of using internal transport phenomenon’s using MCO membrane in term of solute clearances in standard HD. If this is the case, why not using MCO membrane in standard HD, instead of this modified new algorithm intended to increase convective clearances? How does it compare? What are benefits and risks of combining them?Author: Again, good observations. We appreciate the opinion of the reviewer, but we feel it important to explain the context of attempting this early proof of concept prototype. The delivery of haemodialysis to patients is beginning to undergo a seismic change form clinic to home-based therapy and in writing this we felt it important for the reader to know why such a different form of HDF was being attempted, one that requires no additional tubing or modification to the extracorporeal circuit and or requirement for a ‘special’ dialyser in order to maintain simplicity and ease of use.Use of dialysers with added internal transport such as MCO are an equally viable alternative, but currently they are being charged at a premium like HDF. We asked Baxter for samples but were declined. In this concept enhanced convective therapies based upon software programmable machine changes only would democratise HDF for all without any additional materials and be deliverable by prescription over the internet as appropriate through a secure digital health platform. We feel sure that dialysers optimised for increased internal transport that aid the disruption of the laminar boundary layer will be a significant aid to push-pull HDF in the future and would be an interesting suite of experiments to perform in the future.Submitted filename: Response to reviewers PONE-D-19-16719.docxClick here for additional data file.3 Feb 2020Enhancement of solute clearance using pulsatile push-pull dialysate flow for the QuantaSC+: a novel clinic-to-home haemodialysis systemPONE-D-19-16719R1Dear Dr. Buckberry,We are pleased to inform you that your manuscript has been judged scientifically suitable for publication and will be formally accepted for publication once it complies with all outstanding technical requirements.Within one week, you will receive an e-mail containing information on the amendments required prior to publication. When all required modifications have been addressed, you will receive a formal acceptance letter and your manuscript will proceed to our production department and be scheduled for publication.Shortly after the formal acceptance letter is sent, an invoice for payment will follow. To ensure an efficient production and billing process, please log into Editorial Manager at https://www.editorialmanager.com/pone/, click the "Update My Information" link at the top of the page, and update your user information. If you have any billing related questions, please contact our Author Billing department directly at authorbilling@plos.org.If your institution or institutions have a press office, please notify them about your upcoming paper to enable them to help maximize its impact. If they will be preparing press materials for this manuscript, you must inform our press team as soon as possible and no later than 48 hours after receiving the formal acceptance. Your manuscript will remain under strict press embargo until 2 pm Eastern Time on the date of publication. For more information, please contact onepress@plos.org.With kind regards,Pasqual Barretti, Ph.D., MDAcademic EditorPLOS ONEAdditional Editor Comments (optional):I agree with the reviewers. I believe that the authors address all questions and the manuscript has improved a lot, being able to be published.Reviewers' comments:Reviewer's Responses to QuestionsComments to the Author1. If the authors have adequately addressed your comments raised in a previous round of review and you feel that this manuscript is now acceptable for publication, you may indicate that here to bypass the “Comments to the Author” section, enter your conflict of interest statement in the “Confidential to Editor” section, and submit your "Accept" recommendation.Reviewer #1: (No Response)Reviewer #2: All comments have been addressed**********2. Is the manuscript technically sound, and do the data support the conclusions?The manuscript must describe a technically sound piece of scientific research with data that supports the conclusions. Experiments must have been conducted rigorously, with appropriate controls, replication, and sample sizes. The conclusions must be drawn appropriately based on the data presented.Reviewer #1: (No Response)Reviewer #2: Yes**********3. Has the statistical analysis been performed appropriately and rigorously?Reviewer #1: (No Response)Reviewer #2: Yes**********4. Have the authors made all data underlying the findings in their manuscript fully available?The PLOS Data policy requires authors to make all data underlying the findings described in their manuscript fully available without restriction, with rare exception (please refer to the Data Availability Statement in the manuscript PDF file). The data should be provided as part of the manuscript or its supporting information, or deposited to a public repository. For example, in addition to summary statistics, the data points behind means, medians and variance measures should be available. If there are restrictions on publicly sharing data—e.g. participant privacy or use of data from a third party—those must be specified.Reviewer #1: (No Response)Reviewer #2: Yes**********5. Is the manuscript presented in an intelligible fashion and written in standard English?PLOS ONE does not copyedit accepted manuscripts, so the language in submitted articles must be clear, correct, and unambiguous. Any typographical or grammatical errors should be corrected at revision, so please note any specific errors here.Reviewer #1: (No Response)Reviewer #2: Yes**********6. Review Comments to the AuthorPlease use the space provided to explain your answers to the questions above. You may also include additional comments for the author, including concerns about dual publication, research ethics, or publication ethics. (Please upload your review as an attachment if it exceeds 20,000 characters)Reviewer #1: (No Response)Reviewer #2: Thank you for having addressed my concerns, even if sometimes you were not able to provide the precise or adequate answer. It reads better and more scientifically exact now.**********7. PLOS authors have the option to publish the peer review history of their article (what does this mean?). If published, this will include your full peer review and any attached files.If you choose “no”, your identity will remain anonymous but your review may still be made public.Do you want your identity to be public for this peer review? For information about this choice, including consent withdrawal, please see our Privacy Policy.Reviewer #1: NoReviewer #2: No12 Feb 2020PONE-D-19-16719R1Enhancement of solute clearance using pulsatile push-pull dialysate flow for the QuantaSC+: a novel clinic-to-home haemodialysis systemDear Dr. Buckberry:I am pleased to inform you that your manuscript has been deemed suitable for publication in PLOS ONE. Congratulations! Your manuscript is now with our production department.If your institution or institutions have a press office, please notify them about your upcoming paper at this point, to enable them to help maximize its impact. If they will be preparing press materials for this manuscript, please inform our press team within the next 48 hours. Your manuscript will remain under strict press embargo until 2 pm Eastern Time on the date of publication. For more information please contact onepress@plos.org.For any other questions or concerns, please email plosone@plos.org.Thank you for submitting your work to PLOS ONE.With kind regards,PLOS ONE Editorial Office Staffon behalf ofDr. Pasqual BarrettiAcademic EditorPLOS ONE
Authors: Michael Kraus; John Burkart; Rebecca Hegeman; Richard Solomon; Norman Coplon; John Moran Journal: Hemodial Int Date: 2007-10 Impact factor: 1.812
Authors: Jonas K Eriksson; Martin Neovius; Stefan H Jacobson; Carl-Gustaf Elinder; Britta Hylander Journal: BMJ Open Date: 2016-10-07 Impact factor: 2.692
Authors: Alexander H Kirsch; Raphael Lyko; Lars-Göran Nilsson; Werner Beck; Michael Amdahl; Petra Lechner; Andreas Schneider; Christoph Wanner; Alexander R Rosenkranz; Detlef H Krieter Journal: Nephrol Dial Transplant Date: 2017-01-01 Impact factor: 5.992
Authors: Anna Mathew; Jody-Ann McLeggon; Nirav Mehta; Samuel Leung; Valerie Barta; Thomas McGinn; Gihad Nesrallah Journal: Can J Kidney Health Dis Date: 2018-01-10