Interlaboratory comparison of nanoparticle size measurements between NMIJ and NIST using two different types of dynamic light scattering instruments.

Kayori Takahashi1, John A Kramar2, Natalia Farkas3, Keiji Takahata1, Ichiko Misumi1, Kentaro Sugawara1, Satoshi Gonda1, Kensei Ehara1.   

Abstract

The question of how to relate particle sizes measured using a fixed-angle dynamic light scattering (DLS) instrument with those measured using a multi-angle DLS instrument is addressed. A series of nearly monodisperse polystyrene latex (PSL) particles with nominal diameters of 100 nm, 70 nm, 50 nm, and 30 nm were measured using two different types of DLS instruments: one owned by the National Metrology Institute of Japan (NMIJ) of the multi-angle type and the other owned by the National Institute of Standards and Technology (NIST) of the fixed-angle type. The mean particle size of the PSL particles was measured using the multi-angle-type instrument at various scattering angles and at various concentrations of particle suspension. These data were used to establish the functional dependence of the measured particle size on the scattering angle and particle concentration through the least-squares fitting method. The established function was then used to predict the mean particle sizes that would have been obtained if the same scattering angle and particle concentrations as those used at NIST had been selected at NMIJ. The mean particle sizes obtained at NIST and at NMIJ agreed quite well for all four PSL particle samples after compensating for the angle and concentration differences. The result of this study clearly demonstrates that consideration for the dependence of measured particle sizes on the scattering angle and particle concentration is crucial in intra-method comparisons of mean particle sizes obtained using DLS.

Entities:  

Keywords:  dynamic light scattering; nanoparticle; particle concentration; scattering angle; size

Year:  2019        PMID: 32116391      PMCID: PMC7047828          DOI: 10.1088/1681-7575/ab3073

Source DB:  PubMed          Journal:  Metrologia        ISSN: 0026-1394            Impact factor:   3.157


  3 in total

1.  Development of a standard method for nanoparticle sizing by using the angular dependence of dynamic light scattering.

Authors:  Kayori Takahashi; Haruhisa Kato; Shinichi Kinugasa
Journal:  Anal Sci       Date:  2011       Impact factor: 2.081

2.  Dispersion characteristics of various metal oxide secondary nanoparticles in culture medium for in vitro toxicology assessment.

Authors:  Haruhisa Kato; Katsuhide Fujita; Masanori Horie; Mie Suzuki; Ayako Nakamura; Shigehisa Endoh; Yasukazu Yoshida; Hitoshi Iwahashi; Kayori Takahashi; Shinichi Kinugasa
Journal:  Toxicol In Vitro       Date:  2009-12-13       Impact factor: 3.500

3.  Reliable size determination of nanoparticles using dynamic light scattering method for in vitro toxicology assessment.

Authors:  Haruhisa Kato; Mie Suzuki; Katsuhide Fujita; Masanori Horie; Shigehisa Endoh; Yasukazu Yoshida; Hitoshi Iwahashi; Kayori Takahashi; Ayako Nakamura; Shinichi Kinugasa
Journal:  Toxicol In Vitro       Date:  2009-05-03       Impact factor: 3.500

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.