| Literature DB >> 32104107 |
Yoshio Tsunezuka1, Nobuhiro Tanaka1, Hideki Fujimori1.
Abstract
OBJECTIVE: To assess bleeding following transection of the pulmonary artery with powered and manual endoscopic staplers.Entities:
Keywords: bleeding; pulmonary artery; stapler; surgery
Year: 2020 PMID: 32104107 PMCID: PMC7024768 DOI: 10.2147/MDER.S240343
Source DB: PubMed Journal: Med Devices (Auckl) ISSN: 1179-1470
Evaluation of Intraoperative Bleeding After Endoscopic Stapler Use in Pulmonary Artery Transection
| p value <0.0001a | Patients n (%) | Bleeding Casesb n (%) N=24 | Closed Anvil Jaw Space, µm |
|---|---|---|---|
| 82 (34.3) | 0 (0.0) | – | |
| ECHELON ENDOPATH™ 35mm Vascular Reloads (VASECR35) | 82 (34.3) | 0 (0.0) | 0.636 |
| 94 (39.3) | 17 (70.8) | - | |
| Gray cartridge (Endo-GIA™ 45mm Curved Tip Articulating Vascular Reload [EGIA45CTAV]) | 61 (25.5) | 15 (62.5) | 0.106 |
| White cartridge (Endo-GIA-II white cartridge 45) | 4 (1.7) | 2 (8.3) | 0.212 |
| Tan cartridge (Endo-GIA™ 45mm Curved Tip Articulating Vascular/Medium Reload with Tri-Staple™ Technology [EGIA45]) | 29 (12.1) | 0 (0.0) | 0.530 |
| 63 (26.4) | 7 (29.2) | - | |
| Endo-GIA™ Ultra Universal Stapler XL (EGIAUXL) - gray cartridge (Endo-GIA™ 45mm Curved Tip Articulating Vascular Reload [EGIA45CTAV]) | 25 (10.5) | 4 (16.7) | 0.424 |
| Endo-GIA™ Ultra Universal Stapler XL (EGIAUXL) - white cartridge (Endo-GIA-II white cartridge 45) | 8 (3.3) | 2 (8.3) | 0.212 |
| Endo-GIA™ Ultra Universal Stapler XL (EGIAUXL) - tan cartridge (Endo-GIA™ 45mm Curved Tip Articulating Vascular/Medium Reload with Tri-Staple™ Technology [EGIA45CTAVM]) | 17 (7.1) | 0 (0.0) | 0.530 |
| ECHELON FLEX™ ENDOPATH® - white cartridge (Endoscopic Linear Cutter Reloads [ECR45W]) | 13 (5.4) | 1 (4.2) | 0.212 |
Notes: aFisher Exact test. b2 bleeding cases were from thoracotomy, 10 bleeding cases were from video-assisted thoracoscopic surgery. Groups 1 and 2 are powered staplers, Group 3 are manual staplers. ECHELON staplers are manufactured by Ethicon, Endo-GIA™ staplers are manufactured by Medtronic.
Figure 1(A) Comparison of powered endoscopic staplers (B) comparison of jaw angles and (C) histological findings after pulmonary arterial stapling. Histological sections of pulmonary artery tissue adjacent to the area of stapling were stained with Elastica van Gieson stain (X200). In (C) the pulmonary artery section is from stapling with the Endo-GIA™ iDrive™ plus gray cartridge. White arrows indicate the ruptured medial layer.
Abbreviation: PVS, powered vascular stapler.