Daryl L Roberts1, Frank Chambers2, Mark Copley3, Jolyon P Mitchell4.
Abstract
Background: Determination of aerosol aerodynamic particle size distributions (APSD) from dry-powder inhalers (DPIs), following quality control procedures in the pharmacopeial compendia, requires that the flow through the measurement apparatus, comprising induction port, optional pre-separator, and cascade impactor, starts from zero on actuation of the inhaler, using a solenoid valve to apply vacuum to the apparatus exit. The target flow rate, governed by the inhaler resistance, is reached some time afterward. Understanding the behavior of the DPI design-specific flow rate-rise time curve can provide information about the kinetics of the initial powder dispersion in the inhaler and subsequent transport through the APSD measurement equipment. Accurate and precise measures of the internal volume of each component of this apparatus are required to enable reliable relationships to be established between this parameter and those defining the flow rate-rise time curve.Entities:
Keywords: cascade impactor; induction port; internal volume; pre-separator
Mesh:
Substances:
Year: 2020 PMID: 32101074 PMCID: PMC7407003 DOI: 10.1089/jamp.2019.1590
Source DB: PubMed Journal: J Aerosol Med Pulm Drug Deliv ISSN: 1941-2711 Impact factor: 2.849
FIG. 1.(a, b) Arrangement of syringe and pressure transducer for measuring changes in absolute pressure.
Internal Volumes of Components and Groups of Components Related to Testing with the Next-Generation Impactor for Inhaler Aerosol Aerodynamic Particle Size Distribution
| Component or grouping | Internal volume (mL; current study) | Internal volume (mL; 2005 study) | |
|---|---|---|---|
| By individual apparatus | USP/PhEur Induction Port alone | 67 ± 1 | 85 |
| NGI™ alone | 1123 ± 6 | 1160 | |
| NGI with pre-separator | 1855 ± 21 | 1940[ | |
| Groups calculated from individual apparatus measurements | NGI with induction port | 1172 ± 6 | 1245 |
| NGI with induction port and pre-separator | 1904 ± 22 | 2025 |
NGI and its pre-separator reported separately by Copley et al.(
NGI, Next-Generation Impactor; PhEur, European Pharmacopoeia; USP, United States Pharmacopeia.
Internal Volumes of Components and Groups of Components Related to Testing with the Andersen Cascade Impactor for Inhaler Aerosol Aerodynamic Particle Size Distribution
| Measurement approach | Component or grouping | Internal volume (mL; current study) | Internal volume (mL; 2005 study) |
|---|---|---|---|
| By individual apparatus | USP/PhEur induction port alone | 67 ± 1 | 85 |
| ACI alone; 28.3-L/min configuration (inlet cone, stages 0–7, final filter) | 840 ± 8 | 890 | |
| ACI alone; 60-L/min configuration (inlet cone, stages −1 to 6, final filter) | 842 ± 27 | Not studied | |
| ACI alone; 90-L/min configuration (inlet cone, stages −2 to 5, final filter) | 845 ± 26 | Not studied | |
| ACI; 28.3-L/min configuration (stages 0–7, final filter) with pre-separator (28.3-L/min version) | 1013 ± 7 | 1070[ | |
| ACI; 60-L/min configuration (stages −1 to 6, final filter) with pre-separator (60-L/min version) | 1007 ± 5 | Not studied | |
| ACI; 90-L/min configuration (stages −2 to 5, final filter) with pre-separator (90-L/min version) | 1053 ± 23 | Not studied | |
| Groups calculated from individual apparatus measurements | ACI with induction port | 910 ± 8, 27, 26[ | 975 |
| ACI with induction port and pre-separator | 1077 ± 4[ | 1155[ | |
| 1120 ± 23[ |
28.3-L/min ACI and 28.3-L/min pre-separator reported separately by Copley et al.(
Mean value for the three configurations; uncertainties remain for each configuration.
28.3- and 60-L/min configurations.
28.3-L/min configuration.
Ninety-liter per minute configuration.
ACI, Andersen cascade impactor.
APPENDIX FIG. A2.Cross-section of induction port made to fit the ACI (L) and the NGI™ (R)—The internal dimensions of segments 1–6 are identical in both varieties of the induction port; all dimensions shown are given in the USP (independent variables). NGI, Next-Generation Impactor; USP, United States Pharmacopeia.
Comparison of V0 Values Calculated for Next-Generation Impactor Equipment
| Impactor configuration | Replicate | Analysis method | ||||
|---|---|---|---|---|---|---|
| Nonlinear | Linear, D | Used in 2005, E | ||||
| A | B | C | ||||
| NGI™ alone | 1 | 1124.000 | 1123.912 | 1123.902 | 1165.361 | 1113.789 |
| 2 | 1122.897 | 1122.886 | 1122.885 | 1164.297 | 1122.333 | |
| 3 | 1122.897 | 1122.886 | 1122.885 | 1164.297 | 1122.333 | |
| NGI with pre-separator | 1 | 1849.583 | 1849.457 | 1849.395 | 1890.986 | 1816.373 |
| 2 | 1846.844 | 1846.776 | 1846.766 | 1888.244 | 1838.061 | |
| 3 | 1867.518 | 1867.531 | 1867.517 | 1909.000 | 1853.587 | |
NGI, Next-Generation Impactor.
Comparison of V0 Values Calculated for Andersen Cascade Impactor Configurations
| Impactor configuration | Replicate | Analysis method | ||||
|---|---|---|---|---|---|---|
| Nonlinear | Linear, D | Used in 2005, E | ||||
| A | B | C | ||||
| 28.3-L/min ACI alone | 1 | 838.422 | 838.369 | 838.366 | 879.777 | 834.517 |
| 2 | 832.870 | 833.044 | 833.027 | 874.345 | 844.473 | |
| 3 | 838.741 | 838.659 | 838.652 | 880.081 | 832.079 | |
| 28.3-L/min ACI with pre-separator[ | 1 | 1010.475 | 1010.494 | 1010.479 | 1051.890 | 1007.886 |
| 2 | 998.824 | 998.880 | 998.874 | 1040.255 | 1003.798 | |
| 3 | 1004.006 | 1003.991 | 1003.991 | 1045.395 | 1002.883 | |
| 60-L/min ACI alone | 1 | 835.681 | 835.349 | 835.145 | 876.902 | 817.921 |
| 2 | 840.689 | 840.968 | 840.918 | 882.212 | 862.121 | |
| 3 | 837.625 | 837.947 | 837.876 | 879.170 | 863.688 | |
| 60-L/min ACI with 60-L/min pre-separator | 1 | 1005.683 | 1005.732 | 1005.731 | 1047.118 | 1008.143 |
| 2 | 999.475 | 999.476 | 999.475 | 1040.882 | 997.310 | |
| 3 | 1003.302 | 1003.368 | 1003.362 | 1044.744 | 1006.859 | |
| 90-L/min ACI Alone | 1 | 835.584 | 835.924 | 835.850 | 877.138 | 862.070 |
| 2 | 839.283 | 839.632 | 839.552 | 880.843 | 867.621 | |
| 3 | 838.608 | 839.066 | 838.926 | 880.223 | 873.045 | |
| 90-L/min ACI with 90-L/min pre-separator | 1 | 1039.429 | 1039.677 | 1039.628 | 1080.955 | 1061.956 |
| 2 | 1038.473 | 1038.791 | 1038.707 | 1080.034 | 1069.831 | |
| 3 | 1041.254 | 1041.519 | 1041.471 | 1082.800 | 1064.615 | |
Data set B, only.
ACI, Andersen cascade impactor.
Dimensions and Volumes of Segments of United States Pharmacopeia/European Pharmacopoeia Induction Port
| Segment | Shape type | Larger base (mm) | Smaller base (mm) | Total included angle (degrees) | Segment thickness (mm) | Volume (mL) |
|---|---|---|---|---|---|---|
| 1 | Cone segment | 22.154 | 18.000 | 10.398 | ||
| 2- | Cone segment | 22.154 | 18.025 | 6.005 | ||
| 3 | Cylinder | 51.475 | 14.595 | |||
| 4 | Cylinder | 5.387 | ||||
| 5 | Cylinder | 58.924 | 16.707 | |||
| 6 | Cone segment | 36.576 | 14.256 | |||
Entries in italics are values given in the USP,( the independent variables; values in normal font are dependent variables, calculated from the independent variables and simple geometric principles, some of which are explained in the text.
USP, United States Pharmacopeia.