Frequency response evaluation of radial artery catheter-manometer systems: sinusoidal frequency analysis versus flush method.

H A Schwid1.   

Abstract

It is well recognized that catheter-manometer systems significantly distort direct radial artery pressure measurements. Sinusoidal frequency analysis and the flush method of assessing the degree of distortion caused by the monitoring system were compared to determine whether these two methods agree in the estimation of natural frequency and damping coefficient. The frequency response of 30 radial artery catheter-manometer systems used for intensive-care unit patients was measured by the flush method and sinusoidal frequency analysis. The monitoring system consisted of a 20-gauge cannula, 150-cm pressure tubing, two plastic stopcocks, a continuous infusion device with fast flush valve, an American Edwards dome, a Hewlett-Packard quartz transducer, and a Hewlett-Packard blood pressure amplifier. Sinusoidal frequency analysis demonstrated second-order underdamped response for all 30 catheter-manometer systems. No secondary resonance peaks were observed up to a frequency of 200 Hz. The measured frequency response demonstrated that the average catheter-manometer system in use in our intensive care unit would cause significant distortion of the radial artery pressure, with the mean natural frequency (fn) of 14.7 +/- 3.7 Hz and the mean damping coefficient (zeta) of 0.24 +/- 0.07. Although the 30 monitoring systems had identical configurations and visible bubbles were carefully removed, a wide range of frequency responses was found (fn = 10.2 to 25.3; zeta = 0.15 to 0.44).(ABSTRACT TRUNCATED AT 250 WORDS)

Entities:  

Mesh:

Year:  1988        PMID: 3210066     DOI: 10.1007/bf01621814

Source DB:  PubMed          Journal:  J Clin Monit        ISSN: 0748-1977


  11 in total

1.  A CRITICAL STUDY OF THE RESPONSE OF MANOMETERS TO FORCED OSCILLATIONS.

Authors:  H M YANOF; A L ROSEN; N M MCDONALD; D A MCDONALD
Journal:  Phys Med Biol       Date:  1963-11       Impact factor: 3.609

2.  Physiologic recording by modern instruments with particular reference to pressure recording.

Authors:  D L FRY
Journal:  Physiol Rev       Date:  1960-10       Impact factor: 37.312

3.  An evaluation of modern pressure recording systems.

Authors:  D L FRY; A J MALLOS; F W NOBLE
Journal:  Circ Res       Date:  1957-01       Impact factor: 17.367

4.  Quantification and simulation of fluid-filled catheter/transducer systems.

Authors:  B C Taylor; D M Ellis; J M Drew
Journal:  Med Instrum       Date:  1986 May-Jun

5.  Direct blood-pressure measurement: gadgets versus progress.

Authors:  E P McCutcheon; J M Evans; R R Stanifer
Journal:  Anesth Analg       Date:  1972 Sep-Oct       Impact factor: 5.108

6.  Intra-arterial pressure measurements. A review and analysis of the methods relevant to anaesthesia and intensive care.

Authors:  V G Hughes; C Prys-Roberts
Journal:  Anaesthesia       Date:  1971-10       Impact factor: 6.955

7.  Evaluation and correction of manometer systems with two degrees of freedom.

Authors:  J Melbin; M Spohr
Journal:  J Appl Physiol       Date:  1969-11       Impact factor: 3.531

8.  Damped and undamped frequency responses of underdamped catheter manometer systems.

Authors:  G G Shapiro; L J Krovetz
Journal:  Am Heart J       Date:  1970-08       Impact factor: 4.749

9.  Square-wave liquid pressure generator for testing blood pressure transducers.

Authors:  J E Levasseur; F C Funk; J L Patterson
Journal:  J Appl Physiol       Date:  1969-09       Impact factor: 3.531

10.  Direct blood pressure measurement--dynamic response requirements.

Authors:  R M Gardner
Journal:  Anesthesiology       Date:  1981-03       Impact factor: 7.892

View more
  5 in total

1.  Dynamic response of a neonatal catheter-manometer system in situ.

Authors:  H van Langen; P Brienesse; K Kopinga; P Wijn
Journal:  J Clin Monit       Date:  1993-11

2.  Prevention of air introduction in catheter-manometer systems for accurate neonatal blood pressure measurement: an in vitro study.

Authors:  H R van Genderingen; M Gevers; W W Hack
Journal:  J Clin Monit       Date:  1994-01

3.  The fast flush test--is the clinical comparison equivalent to its in vitro simulation?

Authors:  B Kleinman; K Frey; R Stevens
Journal:  J Clin Monit Comput       Date:  1998-12       Impact factor: 2.502

4.  Progress in the development of a fluorescent intravascular blood gas system in man.

Authors:  C K Mahutte; C S Sassoon; J R Muro; D R Hansmann; T P Maxwell; W W Miller; M Yafuso
Journal:  J Clin Monit       Date:  1990-04

5.  Intraarterial catheter diameter and dynamic response of arterial pressure monitoring system: a randomized controlled trial.

Authors:  Hyongmin Oh; Suk Hyung Choe; Yoon Jung Kim; Hyun-Kyu Yoon; Hyung-Chul Lee; Hee-Pyoung Park
Journal:  J Clin Monit Comput       Date:  2021-02-01       Impact factor: 2.502

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.