Gianluca Scuderi1, Serena Fragiotta1, Luca Scuderi2, Clemente Maria Iodice3, Andrea Perdicchi1.
Abstract
Glaucoma is a group of optic neuropathies characterized by a progressive degeneration of retina ganglion cells (RGCs) and their axons that precedes functional changes detected on the visual field. The macular ganglion cell complex (GCC), available in commercial Fourier-domain optical coherence tomography, allows the quantification of the innermost retinal layers that are potentially involved in the glaucomatous damage, including the retinal nerve fiber (RNFL), ganglion cell and inner plexiform layers. The average GCC thickness and its related parameters represent a reliable biomarker in detecting preperimetric glaucomatous damage. The most accurate GCC parameters are represented by average and inferior GCC thicknesses, and they can be associated with progressive visual field loss. Although the diagnostic accuracy increases with more severe glaucomatous damage and higher signal strength values, it is not affected by increasing axial length, resulting in a more accurate discrimination of glaucomatous damage in myopic eyes with respect to the traditional RNFL thickness. The analysis of the structure-function relationship revealed a good agreement between the loss in retinal sensitivity and GCC thickness. The use of a 10-2° visual field grid, adjusted for the anatomical RGCs displacement, describes more accurately the relationship between RGCs thickness and visual field sensitivity loss.Entities:
Keywords: retinal ganglion cells; retinal nerve fiber layer; spectral-domain optical coherence tomography; standard automated perimetry
Year: 2020 PMID: 32099501 PMCID: PMC6999543 DOI: 10.2147/EB.S226319
Source DB: PubMed Journal: Eye Brain ISSN: 1179-2744
Overview of the Ganglion Cell Layer Imaging Performed by Different Spectral-Domain OCT Devices
| Instrument | Macular Scan Protocol | Macular Thickness Measurements |
|---|---|---|
| RTVue-100 (Optovue Inc., Fremont, CA) | 7×7-mm square area consisting of a horizontal 7-mm line scan (467 A-lines) and 15 vertical 7-mm line scans (400 A-lines) spaced at 0.5-mm intervals, centered 0.75−1 mm temporal to the fovea. | Ganglion cell complex (GCC) |
| Cirrus HD-OCT (Carl Zeiss Meditec, Inc., Dublin, CA) | Macular cube 200 × 200 protocol (6 × 6 mm square grid) containing 200 B-scans (200 A-scan lines) or Macular Cube 512 × 128 protocol (128 B-scans, 512 A-scans per B-scan). The macular thickness data are calculated on 14.13 mm2 elliptical annulus area centered on the fovea. | Ganglion cell analysis (GCA)
Macular RNFL GCL+IPL GCC |
| RS-3000 SD-OCT (Nidek, Gamagori, Aichi, Japan) | 30° × 30° square area (~ 9 × 9 mm square area) consisting of a vertical B scan (512 A-scans) × 128 B-scans horizontally. | Macular GCC
Superior/inferior (S/I) semicircle map 8-sector map or GCchart |
| Spectralis (Heidelberg Engineering, Inc., Heidelberg, Germany) | 30° × 25° volume scan consisting of 61 horizontal B scans (1024 A scans/line) within the central 20°. | Posterior pole asymmetry analysis (PPAA) |
| 3D OCT (Topcon, Inc., Tokyo, Japan) | Macular 3D raster scan composed by 512 × 128 A scans covering a cube of 6 × 6 × 2 mm | Macular RNFL GCL/IPL GCC |
Abbreviations: RNFL, retinal nerve fiber layer; GCL, ganglion cell layer; IPL, inner plexiform layer.
Figure 1Ganglion cell complex (GCC) analysis and segmentation. (A) Thickness map, automatically calculated using RTVue Fourier-domain optical coherence tomography (FD-OCT); (B) Cross-sectional FD-OCT B-scan with segmentation of the GCC. The inset (dotted square) depicts the three anatomical layers included in the GCC analysis: nerve fiber layer (NFL), ganglion cell layer (GCL) and inner plexiform layer (IPL). The traced boundaries for the GCC scan (white arrows) pass from the inner limiting membrane and outer IPL.
Figure 2Quantification of ganglion cell complex (GCC) using Fourier-domain optical coherence tomography (FD-OCT) in the case of severe glaucomatous ganglion cell loss. Three different colorimetric maps were automatically computed by the FD-OCT software (RTVue-100 software; Optovue, Inc., Fremont, CA). (A) Thickness map, displaying diffuse blue and black colors indicating severe and diffuse loss of GCC thickness. The central black spot indicates the fovea where the ganglion cells are missing. (B) Deviation map, illustrating the percentage of thickness loss with respect to the normative database. Black coloring indicates a GCC loss of 50% or greater. (C) Significance map, displaying the significant thickness change variation from normal, where green represents values within the normal range (p-value 5–95%), yellow indicates borderline results (< 5%) and red represents results outside the normal limits (< 1%); (D) Quantitative indexes calculated by the software, including average thickness, hemifield thicknesses (superior, inferior and intra-eye) and the two pattern-based diagnostic indexes (i.e., focal loss volume [FLV] and global loss volume [GLV]); (E) Cross-sectional B-scan displays the segmentation used for GCC analysis.