Literature DB >> 32098638

Comparative evaluation of the microbicidal activity of low-temperature sterilization technologies to steam sterilization.

William A Rutala1, Maria F Gergen2, Emily E Sickbert-Bennett1,3, David J Weber1,3.   

Abstract

OBJECTIVE: To compare the microbicidal activity of low-temperature sterilization technologies (vaporized hydrogen peroxide [VHP], ethylene oxide [ETO], and hydrogen peroxide gas plasma [HPGP]) to steam sterilization in the presence of salt and serum to simulate inadequate precleaning.
METHODS: Test carriers were inoculated with Pseudomonas aeruginosa, Escherichia coli, Staphylococcus aureus, vancomycin-resistant Enterococcus, Mycobacterium terrae, Bacillus atrophaeus spores, Geobacillus stearothermophilus spores, or Clostridiodes difficile spores in the presence of salt and serum and then subjected to 4 sterilization technologies: steam, ETO, VHP and HPGP.
RESULTS: Steam, ETO, and HPGP sterilization techniques were capable of inactivating the test organisms on stainless steel carriers with a failure rate of 0% (0 of 220), 1.9% (6 of 310), and 1.9% (5 of 270), respectively. The failure rate for VHP was 76.3% (206 of 270).
CONCLUSION: Steam sterilization is the most effective and had the largest margin of safety, followed by ETO and HPGP, but VHP showed much less efficacy.

Entities:  

Mesh:

Substances:

Year:  2020        PMID: 32098638     DOI: 10.1017/ice.2020.2

Source DB:  PubMed          Journal:  Infect Control Hosp Epidemiol        ISSN: 0899-823X            Impact factor:   3.254


  4 in total

1.  Comparative evaluation of four hydrogen peroxide-based systems to decontaminate N95 respirators.

Authors:  Shawn Clark; Herman Ng; Ge Wu; Devika Jain; Garry Bassi; Rita Kandel; Tony Mazzulli
Journal:  Antimicrob Steward Healthc Epidemiol       Date:  2021-08-18

2.  Reply to "Comparative evaluation of the microbicidal activity of low-temperature sterilization technologies to steam sterilization".

Authors:  Randal Eveland
Journal:  Infect Control Hosp Epidemiol       Date:  2020-04-27       Impact factor: 3.254

3.  Monitoring Steam Penetration in Channeled Instruments: An Evidence-Based Worst-Case for Practical Situations.

Authors:  Francesco Tessarolo; Michela Masè; Andrea Visonà; Josephus P C M van Doornmalen Gomez Hoyos
Journal:  Front Med Technol       Date:  2020-11-16

Review 4.  What's new and notable in bacterial spore killing!

Authors:  Peter Setlow; Graham Christie
Journal:  World J Microbiol Biotechnol       Date:  2021-08-05       Impact factor: 3.312

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.