Accuracy and user performance evaluation of a blood glucose monitoring system which wirelessly transmits results to compatible insulin pumps.

Michael Caswell1, Daniel Brown2, Joy Frank1, Jane F Wallace2, Scott Pardo2.   

Abstract

Objective: To perform evaluations of the CONTOUR PLUS LINK 2.4 blood glucose monitoring system (BGMS) assessed according to ISO 15197:2013 criteria.
Methods: Clinical trial registered at ClinicalTrials.gov (NCT01824355). In a laboratory study (Study 1), capillary fingertip blood samples from 100 subjects were evaluated in duplicate, using three test strip lots. In a clinical trial (Study 2), 113 adults with diabetes were enrolled, and BGMS results and Yellow Springs Instruments (YSI) analyzer reference measurements were compared for subject- and trial staff-obtained fingertip blood and subject-obtained palm blood. Subjects completed a questionnaire to evaluate BGMS ease of use.
Results: In Study 1, 100% of combined results (all test strip lots) fulfilled ISO 15197:2013 section 6.3 criteria. In Study 2, 97.7% of subject-obtained fingertip results and 100% of trial staff-obtained fingertip results met ISO 15197:2013 section 8 criteria. Additionally, 93.8% of palm results were within ± 15 mg/dL of mean YSI measurements for glucose concentrations <100 mg/dL or ± 15% for glucose concentrations ≥100 mg/dL. Most subjects found the BGMS easy to use. There were three non-serious, non-device related adverse events.
Conclusion: The BGMS exceeded minimum ISO 15197:2013-specified accuracy criteria in the laboratory and in the hands of lay users with diabetes.

Entities:  

Keywords:  Blood glucose self-monitoring; blood glucose; diabetes mellitus; self-management; wireless technology

Mesh:

Year:  2020        PMID: 32096654     DOI: 10.1080/03007995.2020.1734919

Source DB:  PubMed          Journal:  Curr Med Res Opin        ISSN: 0300-7995            Impact factor:   2.580


  2 in total

1.  Accuracy and User Performance of a New Blood Glucose Monitoring System.

Authors:  Leslie Klaff; Pragathi Shelat; Diana Zondorak; Amy Wayland-Smith; Phil Vernes; James M Richardson
Journal:  J Diabetes Sci Technol       Date:  2020-11-26

Review 2.  Technologies for Diabetes Self-Monitoring: A Scoping Review and Assessment Using the REASSURED Criteria.

Authors:  Jessica Hanae Zafra-Tanaka; David Beran; Beatrice Vetter; Rangarajan Sampath; Antonio Bernabe-Ortiz
Journal:  J Diabetes Sci Technol       Date:  2021-03-09
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.