Test-Retest Precision of Myocardial Blood Flow Measurements With 99mTc-Tetrofosmin and Solid-State Detector Single Photon Emission Computed Tomography.

R Glenn Wells1, Ivana Radonjic1, Duncan Clackdoyle1, Jeffrey Do1, Brian Marvin1, Clare Carey1, Robert A deKemp1, Terrence D Ruddy1.   

Abstract

BACKGROUND: Measurement of myocardial blood flow (MBF) with single photon emission computed tomography (SPECT) is feasible using cardiac cameras with solid-state detectors. SPECT MBF has been shown to be accurate when compared with positron emission tomography MBF measured in the same patients. However, the value of a test result applied to an individual patient depends strongly on the precision or repeatability of the test. The purpose of our study is to measure the precision of SPECT MBF measurements using 99mTc-tetrofosmin and a solid-state cardiac camera.
METHODS: SPECT MBF was measured in 30 patients and repeated at a mean interval of 18 days. MBF was evaluated from images with and without attenuation correction based on a separately acquired CT scan. The dynamic images were processed independently by 2 operators using in-house kinetic analysis software that applied a 1-tissue-compartment model. The K1 rate constant was converted to MBF using previously determined extraction fraction corrections. Correction for patient body motion was applied manually.
RESULTS: The average coefficient of variation (COV) in the differences between the 2 MBF measurements was between 28% and 31%. The interobserver COV was between 11% and 15%. Myocardial flow reserve is the ratio of MBF measured at stress and rest, and the COV is correspondingly higher. The COV for the difference in repeated myocardial flow reserve was 33% to 38%, whereas the interobserver COV was 13% to 22%.
CONCLUSIONS: The COV for the difference in SPECT MBF measurements obtained on separate days is 28% to 31%. The corresponding COV for myocardial flow reserve is 33% to 38%.

Entities:  

Keywords:  regional blood flow; tomography, emission-computed, single-photon

Year:  2020        PMID: 32069116     DOI: 10.1161/CIRCIMAGING.119.009769

Source DB:  PubMed          Journal:  Circ Cardiovasc Imaging        ISSN: 1941-9651            Impact factor:   7.792


  4 in total

Review 1.  Role of nuclear cardiology in diagnosis and risk stratification of coronary microvascular disease.

Authors:  Terrence D Ruddy; Anahita Tavoosi; Viviany R Taqueti
Journal:  J Nucl Cardiol       Date:  2022-07-18       Impact factor: 3.872

2.  Accuracy and Reproducibility of Myocardial Blood Flow Quantification by Single Photon Emission Computed Tomography Imaging in Patients With Known or Suspected Coronary Artery Disease.

Authors:  Ana Carolina do A H de Souza; Hendrik J Harms; Laurel Martell; Courtney Bibbo; Meagan Harrington; Kyle Sullivan; Jon Hainer; Sharmila Dorbala; Ron Blankstein; Viviany R Taqueti; Marie Foley Kijewski; Mi-Ae Park; Alejandro Meretta; Christopher Breault; Nathaniel Roth; Alexis Poitrasson-Rivière; Prem Soman; Grant T Gullberg; Marcelo F Di Carli
Journal:  Circ Cardiovasc Imaging       Date:  2022-06-08       Impact factor: 8.589

3.  More evidence for adequate test-retest repeatability of myocardial blood flow quantification with 82Rb PET/CT.

Authors:  Robert A deKemp; Emel Celiker Guler; Terrence D Ruddy
Journal:  J Nucl Cardiol       Date:  2020-06-25       Impact factor: 5.952

Review 4.  Quantitative clinical nuclear cardiology, part 2: Evolving/emerging applications.

Authors:  Piotr J Slomka; Jonathan B Moody; Robert J H Miller; Jennifer M Renaud; Edward P Ficaro; Ernest V Garcia
Journal:  J Nucl Cardiol       Date:  2020-10-16       Impact factor: 5.952

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.