Meeting the New FDA Standard for Accuracy of Self-Monitoring Blood Glucose Test Systems Intended for Home Use by Lay Users.

Laurence B Katz1, Lorna Stewart2, Danielle King2, Hilary Cameron2.   

Abstract

The OneTouch Verio Reflect blood glucose monitor (BGM) has market clearance in several countries based in part on fulfilling the lay user and system accuracy criteria described in ISO15197:2015. However, the Food and Drug Administration (FDA) does not recognize the accuracy criteria in ISO15197 as a basis for gaining regulatory clearance for these devices. The current study evaluates the BGM using the accuracy guidelines issued by the agency for self-monitoring blood glucose test systems for over-the-counter use. Glucose results were accurate vs comparator over a wide glucose range and met lay user and glucose accuracy criteria at extreme glucose values as described in the FDA guidance.Clinicaltrials.gov NCT03851549.

Entities:  

Keywords:  accuracy; blood glucose; blood glucose meter; blood glucose strips; self-monitoring of blood glucose

Year:  2020        PMID: 32059615      PMCID: PMC7753858          DOI: 10.1177/1932296820906184

Source DB:  PubMed          Journal:  J Diabetes Sci Technol        ISSN: 1932-2968


  5 in total

1.  Type 2 Diabetes in the Real World: The Elusive Nature of Glycemic Control.

Authors:  Steven V Edelman; William H Polonsky
Journal:  Diabetes Care       Date:  2017-08-11       Impact factor: 19.112

2.  A survey of blood glucose monitoring in patients with type 2 diabetes: are recommendations from health care professionals being followed?

Authors:  William H Polonsky; Lawrence Fisher; Danielle Hessler; Steven V Edelman
Journal:  Curr Med Res Opin       Date:  2011-11       Impact factor: 2.580

3.  Hematocrit compensation in electrochemical blood glucose monitoring systems.

Authors:  Maria Teodorczyk; Marco Cardosi; Steven Setford
Journal:  J Diabetes Sci Technol       Date:  2012-05-01

4.  Patient Satisfaction With a New, High Accuracy Blood Glucose Meter That Provides Personalized Guidance, Insight, and Encouragement.

Authors:  Laurence B Katz; Lorna Stewart; Brian Guthrie; Hilary Cameron
Journal:  J Diabetes Sci Technol       Date:  2019-08-02

5.  Novel FAD-dependent glucose dehydrogenase for a dioxygen-insensitive glucose biosensor.

Authors:  Seiya Tsujimura; Shinki Kojima; Kenji Kano; Tokuji Ikeda; Mika Sato; Hirokazu Sanada; Hironori Omura
Journal:  Biosci Biotechnol Biochem       Date:  2006-03       Impact factor: 2.043

  5 in total
  2 in total

1.  Analyse Accuracy of Glucose Monitors without any Dedicated Software Package.

Authors:  Himel Mondal; Shaikat Mondal
Journal:  Indian J Endocrinol Metab       Date:  2022-08-04

2.  Integrated Experimental and Theoretical Studies on an Electrochemical Immunosensor.

Authors:  Neda Rafat; Paul Satoh; Scott Calabrese Barton; Robert Mark Worden
Journal:  Biosensors (Basel)       Date:  2020-10-17
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.