Shuai Sun1,2, Hong Zhang2, Xing-Peng Liu1. 1. Heart Center, Beijing Chao-Yang Hospital, Capital Medical University, Beijing, China. 2. Department of Cardiology, Shan Xi Provincial People's Hospital, Taiyuan City, Shanxi Province, China.
Abstract
Entities:
Keywords:
His bundle pacing; QRS complex; atrioventricular block; concealed conduction; right bundle branch block; ventricular function
His-Purkinje system pacing is divided into His bundle pacing (HBP) and left bundle
branch area pacing (LBBaP).[1,2]
Pacing in the left bundle branch area can produce a right bundle branch block (RBBB)
pattern. When atrioventricular (AV) delay of the pacemaker is prolonged, fusion
waveforms appear between right bundle branch descending excitation and left bundle
area pacing. This situation creates a QRS wave that resembles that of a normal
electrocardiogram (ECG). We present a case with a distinct bundle branch block
pattern. Our patient showed a continuous RBBB pattern during HBP after an operation.
When we prolonged the AV delay to the maximum value after the operation, the patient
still showed RBBB, which may be due to a concealed conduction mechanism.
Case report
A 75-year-old male patient was hospitalized on 7 May 2018 because of a slow heart
rate lasting for 2 days. The patient’s history showed a slow heart rate for 2 days
before admission and a self-measured pulse rate of approximately 40 times/minute,
without dizziness or amaurosis. Thereafter, a dynamic ECG at the local hospital
showed 2: 1 atrioventricular block (AVB) with complete left bundle branch block
(CLBBB). A physical examination after admission showed that the pulse rate was 42
times/minutes and blood pressure was 150/70 mmHg. Routine blood tests did not show
any major abnormalities in blood biochemistry. A chest X-ray showed increased lung
texture, but normal heart shadow size and shape.An ECG showed 2: 1 AVB and CLBBB (Figure 1). An ultrasonic cardiogram indicated that aortic valve
degeneration was accompanied by a small amount of regurgitation and tricuspid valve
regurgitation, the left ventricular ejection fraction (LVEF) was 45%, and left
ventricular end-stage diameter (diastolic/systolic) was 51 mm/39 mm. A negative
troponin test excluded the possibility of AVB caused by acute myocardial infarction.
We considered using HBP or LBBaP to correct AVB and CLBBB because these can
partially correct CLBBB. If correction of CLBBB failed during the operation, we
would have considered upgrading to cardiac resynchronization therapy.
Figure 1.
Preoperative electrocardiogram shows 2: 1 atrioventricular block and complete
left bundle branch block. The black arrows show P waves and the red arrows
show QRS waves. Atrioventricular conduction is 2: 1. The QRS morphology of
V1 is rS and that of V6 is RsR’, which represent left bundle branch
block.
Preoperative electrocardiogram shows 2: 1 atrioventricular block and complete
left bundle branch block. The black arrows show P waves and the red arrows
show QRS waves. Atrioventricular conduction is 2: 1. The QRS morphology of
V1 is rS and that of V6 is RsR’, which represent left bundle branch
block.A Medtronic 3830-69 cm pacing electrode with the assistance of a Medtronic C315 His
sheath (Medtronic, Minneapolis, MN, USA) was used to map the His bundle potential
(unipolar) during the operation. The Carto-3 three-dimensional mapping system
(Johnson & Johnson Company, New Brunswick, NJ, USA) was used to record His
bundle potential. A Medtronic 2290 programmable controller was used to test sensing,
pacing, and impedance parameters, and to record current of injury. The split His
bundle potential was recorded during the operation (Figure 2). Dropped beats correlated with the
missing H2 component, which indicated intrahisian block. During a threshold test,
the QRS complex became narrower and the V1 lead rS wave became an rSR’ wave and
showed a pattern of continuous RBBB. CLBBB was corrected at a threshold of 1.4 V at
0.5 ms (Figure 3). When the
output was adjusted to 5 V at 1.0 ms and 10 V at 1.0 ms, it remained in the RBBB
mode. Because of the patient’s restlessness during the operation and a lack of
dependence on pacing during cessation of pacing, no ventricular backup lead was
implanted (Figures 4 and
5). The pacemaker was
temporarily programmed with a long AV delay. However, the paced QRS complexes
continued to have RBBB morphology (Figure 6).
Figure 2.
Intraoperative intraluminal view shows the His bundle separation potential
(white arrow). When atrioventricular block occurred, the conduction block
was in the His bundle.
Figure 3.
During His bundle pacing, the rS wave of V1 leads became an rSR’ wave, and
showed a pattern of continuous right bundle branch block and correction of
complete left bundle branch block.
Figure 4.
Postoperative pacing electrocardiogram suggests correction of complete left
bundle branch block. With His bundle pacing, the QRS wave in V1 leads is
M-shaped (black arrow) and that in V5 and V6 leads is Rs (red arrow), which
represent right bundle branch block along with disappearance of left bundle
branch block.
Figure 5.
Fluoroscopic projection images. The left image was taken at the 45° left
anterior oblique view and the right image was taken at the 30° right
anterior oblique view.
Figure 6.
When the AV delay was prolonged, the pattern of right bundle branch block
remained.
Intraoperative intraluminal view shows the His bundle separation potential
(white arrow). When atrioventricular block occurred, the conduction block
was in the His bundle.During His bundle pacing, the rS wave of V1 leads became an rSR’ wave, and
showed a pattern of continuous right bundle branch block and correction of
complete left bundle branch block.Postoperative pacing electrocardiogram suggests correction of complete left
bundle branch block. With His bundle pacing, the QRS wave in V1 leads is
M-shaped (black arrow) and that in V5 and V6 leads is Rs (red arrow), which
represent right bundle branch block along with disappearance of left bundle
branch block.Fluoroscopic projection images. The left image was taken at the 45° left
anterior oblique view and the right image was taken at the 30° right
anterior oblique view.When the AV delay was prolonged, the pattern of right bundle branch block
remained.The patient was followed up for 6 months after the operation. The patient was in good
condition and the pacing threshold was stable. The threshold was 1.5 V/0.5 ms and
the LVEF increased from 45% to 71%. The left ventricular end-stage diameter
(diastolic/systolic) was 45 mm/27 mm.The study protocol was approved by the ethics committee of Shanxi Provincial People’s
Hospital. The patient provided written consent for publication.
Discussion
A meta-analysis showed that the HBP threshold was 1.76 V in the acute phase (<3
months) and 1.79 V in the chronic phase.[3] The authors of this meta-analysis also conducted safety assessment of 18
studies and showed that the leads were revised in 26 of 966 patients because of
dislodgement and 20 had an increased pacing threshold. In 2017, the international
consensus of experts on permanent HBP stated that the His bundle capture threshold
should be <2.5 V at 1.0 ms.[4] Therefore, a CLBBB correction threshold of 1.4 V at 0.5 ms was acceptable in
our patient.His bundle fibers are arranged longitudinally and separated by connective tissue.
Because of a lack of side to side intercellular low resistance junctions, His bundle
transmission is dissociated longitudinally. Therefore, the branches of the left and
right bundles are separated within the His bundle.[5-7] The His bundle itself can be
divided into the unbranched part and bifurcated part. The unbranched part is part of
the atrioventricular junction, while the bifurcated part forms the beginning of the
left and right bundle branches. Because of the histological and anatomical
characteristics of the longitudinal separation, bundle branch block may occur in the
His bundle and HBP can correct bundle branch block, similar to our case. In our
case, the patient showed a continuous RBBB pattern during HBP after the operation
rather than permanent damage to the right bundle branch during implantation of the
3830 electrode. When pacing was temporarily stopped during the operation, the
original 2: 1 AVB with CLBBB pattern was restored (Figure 7). Huang et al.[8] reported the first case of LBBaP in 2017. When AV delay was gradually
prolonged, a fusion waveform appeared between right bundle branch descending
excitation and LBBaP, which created a QRS wave that was close to normal. In our
case, the patient had 2: 1 AVB before the operation, and CLBBB was corrected after
the operation, which produced a RBBB morphology. We speculate that prolonging the AV
delay could encourage self-excitation to be transmitted downward through the right
bundle branch and intermittent 2: 1 excitation conducted through the right bundle
branch might also appear. This situation could result in one RBBB and one normal QRS
waveform. However, this did not occur in our patient. When the AV delay was
prolonged to the maximum value after operation, the patient still showed RBBB (Figure 6). This finding may be
due to a concealed conduction mechanism. In 1948, Langendorf[9] first proposed the phenomenon of concealed conduction. Concealed conduction
refers to when excitation reaches a certain area, which is transitioning between
absolute and relative refractory periods, and the excitability of that area is low.
During this time, the 0-phase rise of the action potential in that area is low and
excitation cannot spread to the surrounding area, preventing normal depolarization.
However, because this area has been excited, continuous excitation prevents
completion of the refractory period and the next excitement cannot be transmitted
normally. Therefore, concealed conduction is not really “concealed”, but is an
“incomplete penetrating excitement”. In our case, the patient had 2: 1 AVB with
CLBBB before the operation, and therefore, supraventricular excitation was
transmitted downward through the right bundle branch at 2: 1. The combination of
incomplete conduction block in the right bundle branch and a long refractory period
formed by the previous excitation results in the next excitation conduction block.
In our case, excitation of the left bundle branch due to continuous HBP could have
caused concealed excitation of the right bundle branch. This would then have
affected transmission of excitation through the right bundle branch and resulted in
the pattern of continuous RBBB in HBP (Figure 8).
Figure 7.
When pacing was stopped, 2: 1 atrioventricular block with complete left
bundle branch block occurred again.
Figure 8.
Schematic diagram. (a) Previous excitation generated a long refractory period
through the right bundle branch, preventing the next excitation from being
transmitted. (b) His bundle pacing induced left bundle branch excitation
that could continue to produce concealed excitation of the right bundle
branch. This prevented continuous transmission of upper excitation through
the right bundle branch.
When pacing was stopped, 2: 1 atrioventricular block with complete left
bundle branch block occurred again.Schematic diagram. (a) Previous excitation generated a long refractory period
through the right bundle branch, preventing the next excitation from being
transmitted. (b) His bundle pacing induced left bundle branch excitation
that could continue to produce concealed excitation of the right bundle
branch. This prevented continuous transmission of upper excitation through
the right bundle branch.RBBB can occur at all ages and can be found in a selection of otherwise normal people
in addition to a variety of organic heart diseases. Right ventricular electrical
activity is achieved by slow conduction through the interventricular septum after
left ventricular excitation. Some previous studies have reported that RBBB may also
cause delayed left ventricular activation. The majority of evidence suggests that
RBBB can cause delayed right ventricular electrical excitation and decreased right
ventricular diastolic function.[10] Furthermore, this has little effect on left ventricular electrical excitation
and left ventricular diastolic function. Therefore, the effect of RBBB on left
ventricular function is still controversial. At present, the guidelines do not
include patients with RBBB and a left ventricular ejection fraction >35% as
indications for cardiac resynchronization therapy. Although our patient showed RBBB
after the operation, follow-up several months later showed that left ventricular
function was improved.HBP is a type of physiological pacing. Study of the anatomy and electrophysiological
mechanism of the His system is beneficial in supporting the development of pacing
electrophysiology. This study examined whether prolonging AV delay by LBBaP could
generate a fusion waveform of pacing via the right bundle branch and left bundle
branch. In our case, concealed conduction of the left bundle branch to the right
bundle branch after capture of the left bundle branch was observed after HBP,
resulting in continuous RBBB.
Authors: Francesco Zanon; Kenneth A Ellenbogen; Gopi Dandamudi; Parikshit S Sharma; Weijian Huang; Daniel L Lustgarten; Roderick Tung; Hiroshi Tada; Jayanthi N Koneru; Tracy Bergemann; Dedra H Fagan; John Harrison Hudnall; Pugazhendhi Vijayaraman Journal: Europace Date: 2018-11-01 Impact factor: 5.214