Literature DB >> 32050228

Reasons for explantation, demographics, and material analysis of 200 intraocular lens explants.

Tabitha Neuhann1, Timur M Yildirim, Hyeck-Soo Son, Patrick R Merz, Ramin Khoramnia, Gerd U Auffarth.   

Abstract

PURPOSE: To report demographics, reasons for explantation, and material changes in explanted intraocular lenses (IOLs).
SETTING: The David J. Apple International Laboratory for Ocular Pathology, Department of Ophthalmology, University of Heidelberg, Germany.
DESIGN: Retrospective study laboratory investigation.
METHODS: IOL explants that were sent consecutively to the laboratory were assessed for demographics (patient sex and age), duration of implant, IOL type, model, power, and reason for IOL explantation. In opacified lenses histological staining, scanning electron microscopy, and energy-dispersive X-ray spectroscopy was performed.
RESULTS: The analysis included 200 IOLs that were explanted in median after 5.8 years. The median time the IOL was in the eye was 5.8 years. IOL opacification was the main cause for explantation in 153 (76.5%) cases. Only 27 (13.5%) were explanted due to dislocation. Evaluation of IOL type showed that 167 (83.5%) were made from hydrophilic acrylic material, with 125 (62%) from hydrophilic acrylic material with a hydrophobic surface coating. Analysis of opacities revealed superficial and subsurface deposits of calcium phosphate in most of the opacified lenses (152/153). In total, 22 different manufacturers were represented, with 119 (59.5%) lenses from a single manufacturer.
CONCLUSION: In this cross-sectional study, late IOL calcification was the main reason for IOL explantation. The second most common reason was IOL dislocation. Most explants were lenses from a single manufacturer exchanged due to primary IOL calcification.

Entities:  

Mesh:

Year:  2020        PMID: 32050228     DOI: 10.1016/j.jcrs.2019.08.045

Source DB:  PubMed          Journal:  J Cataract Refract Surg        ISSN: 0886-3350            Impact factor:   3.351


  9 in total

Review 1.  Etiology, pathogenesis, and management of acute intraocular lens opacification: a systematic review.

Authors:  Bita Momenaei; Mohammad Reza Akbari; Seyed Ali Tabatabaei; Mohammad Soleimani; Mahdi Soleymanzadeh; Kaveh Fadakar; Ahmed Alshaheeb; Mahtab Malekpour Khazari; Yasaman Vaseghi; Kasra Cheraqpour
Journal:  Int J Ophthalmol       Date:  2022-07-18       Impact factor: 1.645

2.  Variation in intraocular lens calcification under different environmental conditions in eyes with supplementary sulcus-supported lenses.

Authors:  Timur M Yildirim; Ramin Khoramnia; Sonja K Schickhardt; Donald J Munro; Patrick R Merz; Hyeck-Soo Son; Ingo Lieberwirth; Gerd U Auffarth
Journal:  Am J Ophthalmol Case Rep       Date:  2020-06-24

3.  Visual and Refractive Outcomes Following Exchange of an Opacified Multifocal Intraocular Lens.

Authors:  Stephen A Stewart; Richard N McNeely; Wing C Chan; Jonathan E Moore
Journal:  Clin Ophthalmol       Date:  2022-06-09

4.  Development of a standardized in vitro model to reproduce hydrophilic acrylic intraocular lens calcification.

Authors:  Leoni Britz; Sonja Katrin Schickhardt; Timur Mert Yildirim; Gerd Uwe Auffarth; Ingo Lieberwirth; Ramin Khoramnia
Journal:  Sci Rep       Date:  2022-05-10       Impact factor: 4.996

5.  Analysis of opacification patterns in intraocular lenses (IOL).

Authors:  Marc Mackert; Daniel Rudolf Muth; Efstathios Vounotrypidis; Constanze Deger; David Goldblum; Mehdi Shajari; Pascal Willy Hasler; Siegfried Priglinger; Armin Wolf
Journal:  BMJ Open Ophthalmol       Date:  2021-02-11

6.  Transient Clouding of a Sutureless Scleral Fixated Hydrophilic Intraocular Lens with Spontaneous Resolution: A Case Report and in vitro Experimental Test.

Authors:  Carla Danese; Renzo Calabresi; Paolo Lanzetta
Journal:  Case Rep Ophthalmol       Date:  2021-04-12

7.  Laboratory evaluation of higher-order aberrations and light scattering in explanted opacified intraocular lenses.

Authors:  Grzegorz Łabuz; Timur M Yildirim; Gerd U Auffarth; Hyeck-Soo Son; Ramin Khoramnia
Journal:  Eye Vis (Lond)       Date:  2021-04-22

8.  Comparison Between L-312 Hydrophobic-Hydrophilic Acrylate and US-860 UV Hydrophilic Acrylate IOL Opacification Characteristic.

Authors:  Jin Xie; Jie Sun; Ting Liu; Shilan Mao; Yunhai Dai
Journal:  Front Med (Lausanne)       Date:  2022-04-08

9.  Wave-like calcification on the posterior surface of an acrylic hydrophilic bag-in-the-lens (BIL) implant.

Authors:  Sorcha Ní Dhubhghaill; Caroline Janssen; Diana Carmen Dragnea; Luc Van Os; Jos Rozema; Liliana Werner; Dirk Van Dyck; Marie-José Tassignon
Journal:  Am J Ophthalmol Case Rep       Date:  2022-08-27