Optimizing Settings for Office-Based Endoscopic CO2 Laser Surgery Using an Experimental Vocal Cord Model.

Anouk S Schimberg1, Tim M Klabbers1, David J Wellenstein1, Floris Heutink1, Jimmie Honings1, Ilse van Engen-Van Grunsven2, Rudolf M Verdaasdonk3, Robert P Takes1, Guido B van den Broek1.   

Abstract

OBJECTIVES/HYPOTHESIS: To provide insight in the thermal effects of individual laser settings in target tissues to optimize flexible endoscopic span class="Chemical">pan class="Chemical">CO2span> laser surgery treatment. STUDY
paspan>n class="abstract_title">DESIGN: Experimental laboratory study.
METHODS: Thermal effects of the CO2 laser using a fiber delivery system were visualized using the color Schlieren technique in combination with a polyacrylamide gel tissue model. Variable settings were used for emission mode, power, laser fiber distance, and laser duration, which were evaluated in every possible combination. Collateral thermal expansion and incision depth were measured. To validate the model, the results were compared to histology after CO2 laser irradiation of ex vivo human vocal cords, and the intraclass correlation coefficient was calculated. Thermal damage and incision depth were measured by a blinded pathologist.
RESULTS: Of all parameters studied, duration of laser irradiation had the greatest effect on thermal expansion. Increased distance between laser tip and target tissue resulted in significantly reduced incision depth and increased thermal expansion. Pulsed emission modes led to increased incision depths. The intraclass correlation coefficient for consistency between the model setup and the ex vivo human vocal cords was classified as "fair."
CONCLUSIONS: By using high-intensity pulsed lasers at minimal distance to the target tissue, exposure times and subsequent damage to surrounding tissue can be reduced. If an evaporation technique is used, lower power in continuous wave at a larger distance to the target tissue will lead to superficial but broader thermal effects. The model setup used in this study is a valid model to investigate laser-induced thermal effects in vocal cord tissue. LEVEL OF EVIDENCE: NA Laryngoscope, 130:E680-E685, 2020.
© 2020 The Authors. The Laryngoscope published by Wiley Periodicals, Inc. on behalf of The American Laryngological, Rhinological and Otological Society, Inc.

Entities:  

Keywords:  Flexible endoscopic CO2 laser; thermal effects; vocal cords

Year:  2020        PMID: 32022931      PMCID: PMC7687150          DOI: 10.1002/lary.28518

Source DB:  PubMed          Journal:  Laryngoscope        ISSN: 0023-852X            Impact factor:   3.325


INTRODUCTION

Office‐based flexible endoscopic laser surgery using a span class="Chemical">pan class="Chemical">CO2span> laser for laryngeal or pharyngeal lesions is a procedure that is inpaspan>n class="Chemical">creasingly performed.1 The advent of distal chip‐on‐tip flexible laryngoscopes and the development of flexible instruments (e.g., flexible laser fiber delivery systems and biopsy forceps) has made it possible to perform several laryngeal procedures in an office setting. This development has shortened procedure duration, reduced delay in diagnosis and treatment, lowered costs, and reduced complication risks compared to surgery under general anesthesia.2 Laryngeal laser surgery is one of the procedures traditionally performed under general anesthesia but can now be performed under local anesthesia using a flexible endoscope. Laser irradiation of soft tissue inherently produces thermal effects on surrounding healthy tissue in the form of both reversible (altered tissue that will heal) and irreversible (carbonization and pan class="Disease">necrosis) damage.3 Ideally, lasers used in vocal cord surgery cause minimal collateral thermal damage to surrounding tissues, penetrate only superficially, are able to be used for excision and coagulation, and have a flexible delivery system.4 A recent review reported that, currently, no scientific evidence exists on the optimal laser settings in laryngeal procedures; recommendations are usually provided by the laser manufacturer based on empirical evidence.2 Few studies were published investigating thermal effects of span class="Chemical">pan class="Chemical">CO2span> lasers, either using a fiber5, 6, 7 or an articulated arm delivery system.3, 8, 9, 10, 11, 12, 13 However, none of these studies performed a systematic and complete analysis of the effects of individual paspan>n class="Chemical">CO2 laser settings. Hence, the exact effect of adjusting laser settings on thermal energy produced by the laser, the depth of laser incision, and the thermal damage to surrounding tissue remains unknown. Because office‐based laryngeal laser surgery is a novel technique, and experience in tissue effects is still limited, the aim of this study was to com<span class="Chemical">pare the collateral thermal expansion and incision depth after irradiation using a CO2 laser delivered through a fiber with different laser settings. These results can be used to adjust settings depending on the desired surgical effect. For evaporation, a more superficial and broad thermal effect is required, whereas for cutting, a concentrated beam with minimal collateral thermal damage is appropriate. Experiments were performed on a human tissue model, and thermal effects were visualized using the dynamic Schlieren imaging technique.14 The ability of this imaging technique to predict laser‐induced thermal damage in vocal cord tissue was evaluated by comparing the results with conventional histological evaluation of fresh ex vivo human vocal cords irradiated using the same laser settings.

MATERIALS AND METHODS

Laboratory Laser Setup

A span class="Chemical">pan class="Chemical">CO2span> laser (AcuPulse Duo; Lumenis, Yokneam, Israel) with a wavelength of 10.6 μm was used to perform the experiments. The laser was fitted with a hollow fiber wave guide (500 μm, FiberLase; Lumenis). The laser was tested in various settings: in continuous wave (CW), pulsed (P), and super pulsed (SP) modes, using a power of 4, 6, 8, and 10 W, a laser duration of 1, 2, and 3 seconds, and a distance between laser tip and tissue of 2, 5, 10, and 15 mm. These settings were chosen as they were deemed to be representative of the intraoperative conditions.

Dynamic Color Schlieren Imaging Technique

The model used in this study is based on the color Schlieren technique, which allows for the visualization of a temperature gradient in a transparent medium.14 The gradient seen is the result of changes in refraction index of the medium as the temperature of the medium rises. In this study, a polyacrylamide (PA) phantom gel was used as the transparent medium.14 PA gel was chosen as a model tissue as it has similar thermal, optical, mechanical, and electrical properties compared to human tissue.14 A schematic figure of the model is shown in Figure 1. At one end of the setup, a light‐emitting diode (LED) light source was positioned. The light beam emitted by this LED was concentrated using a pinpoint aperture before traveling through a telephoto lens (>400 mm, infinite focus) creating a parallel beam. The parallel light rays passed through the PA gel placed on a plateau perpendicular to the light beam. The flexible laser fiber was positioned directly above the gel using a weighted stand with a wire holder enabling it to be set at variable distances to the gel. After the light rays passed through the gel, a second, decollimating lens (135 mm) focused the light onto a rainbow Schlieren filter consisting of multicolored concentric rings. The filtered light was captured by a digital single‐lens reflex (DSLR) camera at 50 frames per second with 400‐mm zoom and infinite focus.
Figure 1

Schlieren model and path of light‐emitting diode (LED) light beam. DSLR = digital single‐lens reflex.

span class="Chemical">pan class="Chemical">Schlierenspan> model and paspan>th of light‐emitting diode (LED) light beam. DSLR = digital single‐lens reflex. The experimental model is based on the premise that, if the gel has a constant temperature, and thus a constant refraction index, the light beam will remain <span class="Chemical">parallel and most of the light will be blocked by the dark center of the filter. However, irradiation of the gel leads to a temperature gradient away from the incision, causing a gradient in refractive index of the heated gel. As a result, the <span class="Chemical">parallel light beams passing through the gel are refracted at different angles and are no longer blocked by the dark center of the pan class="Chemical">Schlieren filter. Instead, light beams are deflected through different color rings of the pan class="Chemical">Schlieren filter, resulting in a color gradient in the image captured by the DSLR camera. In this model, a large temperature difference, and thus a large refraction index, results in orange‐ and red‐colored images (outer rings of the filter), whereas small temperature changes produce blue or green images (inner rings of the filter).

Experiments in PA Gel

Real‐time thermal span class="Chemical">pan class="Chemical">Schlierenspan> images of the gels were obtained using the aforementioned settings (Fig. 2). Each combination of settings was measured in triplicate. After 1, 2, and 3 seconds, the recording was interrupted and FIJI imaging software (ImageJ; National Institutes of Health, Bethesda, MD)15 was used to measure the thermal effects caused by the laser beam (i.e., the extent of thermal expaspan>nsion surrounding the incision as well as incision depth). The thermal expansion in the gel was defined as the distance from the edge of the incision to the end of the color gradient measured at a depth of 1 mm. In images with an incision depth of <1 mm but with visible thermal effects at 1 mm beneath the gel surface, thermal expansion was measured as the distance from the center of the color gradient to the end of the gradient in the horizontal axis. The incision depth was measured as the distance between the surface of the gel and the deepest point of the incision. If the incision depth exceeded the limits of image captured by the camera, the incision depth was not recorded, as no accurate measurements could be made for these settings.
Figure 2

Schlieren image during irradiation of the polyacrylamide gel using the AcuPulse CO2 laser. Settings: continuous wave, 10‐mm distance laser tip to gel, 4 W. (A) One second. (B) Two seconds. (C) Three seconds.

span class="Chemical">pan class="Chemical">Schlierenspan> image during irradiation of the paspan>n class="Chemical">polyacrylamide gel using the AcuPulse pan class="Chemical">CO2 laser. Settings: continuous wave, 10‐mm distance laser tip to gel, 4 W. (A) One second. (B) Two seconds. (C) Three seconds.

Experiments in Vocal Cord Tissue

To validate the aforementioned model, experiments were conducted on ex vivo tissue samples using the span class="Chemical">pan class="Chemical">CO2span> laser with a fiber delivery system. Twelve larynges obtained from fresh‐frozen paspan>n class="Species">human cadavers of donors who in life had bequeathed their bodies to the Anatomical Institute at the Radboudumc for academic purposes with a valid handwritten testament, were used to carry out this study. The cadaver specimens were obtained according to the Dutch Body Donation Program for Science and Education (Wet op de lijkbezorging, 1991). The larynges were stored at −5°C to prevent tissue degradation. One day before the experiments took place, larynges were thawed to room temperature. We decided not to warm the larynges to body temperature (37°C), as this led to tissue degeneration. On the day of the experiments, the vocal cords were dissected from the larynges and pinned on a paraffin plate while retaining their natural shape. Subsequently, the vocal cords were irradiated in CW, P, and SP mode, with a power of 6 W and 10 W, a duration of 1 and 3 seconds, and a distance of the laser tip of 2 and 10 mm to the tissue. A total of 24 incisions were made in 12 pairs of vocal cords.

Histology

After irradiation, the incision was stained with black ink, and the tissue was fixed in 4% span class="Chemical">pan class="Chemical">formaldehydespan> solution. After 72 to 96 hours, the tissue was serially dehydrated in paspan>n class="Chemical">ethanol baths. The samples were embedded in paraffin and sectioned into 6‐μm coronally cut microsections. The sections were stained with hematoxylin and eosin. Thermal expansion in each section, considered the distance from the edge of the incision to the end of the thermal damage measured at a depth of 0 mm to 1 mm below the epithelium, was evaluated by a pathologist blinded to the laser settings (Fig. 3). Incision depth was also measured and recorded. A photographic file was compiled with microscopic images.
Figure 3

Hematoxylin and eosin–stained histological vocal cord sample after irradiation (AcuPulse CO2 laser, CW, 5 mm, 6 W, 1 second). The black line denotes the extent of collateral thermal damage.

span class="Chemical">pan class="Chemical">Hematoxylinspan> and paspan>n class="Chemical">eosin–stained histological vocal cord sample after irradiation (AcuPulse pan class="Chemical">CO2 laser, CW, 5 mm, 6 W, 1 second). The black line denotes the extent of collateral thermal damage.

Statistical Analysis

The effect of the four laser settings investigated in all possible combinations in the gel on collateral thermal ex<span class="Chemical">pansion and incision depth was analyzed using multivariable linear regression. Both unstandardized β coefficients (B) and standardized β coefficients (β) were calculated. B coefficients represent the inspan class="Chemical">pan class="Chemical">crease or despan>n class="Chemical">crease in dependent variable (i.e., incision depth in millimeters when an independent variable such laser duration is increased by one unit). β coefficients (minimum value −1 and maximum value +1) were calculated to compare the relative effect of the different independent variables. A β of −1 represents a direct negative correlation, whereas a β of +1 represents a direct positive correlation. To evaluate the ability of our model to correctly predict laser‐induced thermal damage in vocal cord mucosa, the intraclass correlation coefficient (ICC) for both incision depth and thermal expansion was determined between the mean of the results acquired using the model and the results from histological analysis. Both the ICC for absolute agreement and the ICC for consistency between the model setup and histological evaluation were obtained. Incisions through the entire vocal cord were not included in the analysis, as no accurate measurements could be made for these settings. The data analysis was performed using SPSS version 25 (IBM, Armonk, NY) for Windows (Microsoft, Redmond, WA). P values <.05 were considered to be statistically significant.

RESULTS

Laser Parameter Evaluation

A total of 432 experiments were performed (three pulse modes, three laser durations, four wattages, and four distances, with three repetitions per combination of settings). Incision depths exceeding the span class="Chemical">pan class="Chemical">Schlierenspan> image frame captured by the camera were observed in 153 of the experiments, and these were excluded from the analysis. Linear regression analysis (Table 1) demonstrated that prolonged laser duration was strongly related with inspan class="Chemical">pan class="Chemical">crspan>eased thermal expaspan>nsion (B = 0.279, β = 0.848, P < .005). A B of 0.279 represents an increase in thermal expansion of 0.279 mm in the tissue, when laser duration is increased by 1 second. Furthermore, increased laser power and distance between laser tip and gel led to a small but significant increase in thermal expansion (B = 0.020, β = 0.167, P < .005 and B = 0.009, β = 0.170, P < .005, respectively); increasing laser power by 1 W would result in an increase in thermal expansion of 0.02 mm, whereas increasing laser distance by 1 mm would result in an increase of 0.009 mm. Additionally, using the CO2 laser in P and SP modes would result in an increase in thermal expansion by 0.043 and 0.046 mm, respectively, compared to CW mode (B = 0.043, β = 0.076 and B = 0.046 and β = 0.082, P < .005, respectively).
Table 1

Linear Regression Analysis Results for the AcuPulse Laser.

Laser SettingBβ
Thermal expansion (mm)Power (W)0.020* 0.167*
F (5, 426) = 305.8Laser duration (sec)0.279* 0.848*
cp < 0.0005Laser distance (mm)0.009* 0.170*
R 2 = 0.782Pulse0.043* 0.076*
Super pulse0.046* 0.082*
Incision depth (mm)Power (W)0.445* 0.612*
F (5, 273) = 193.2Laser duration (sec)1.031* 0.531*
P < .0005Laser distance (mm)−0.241* −0.722*
R 2 = 0.780Pulse1.728* 0.471*
Super pulse0.793* 0.239*

P < .005.

B = unstandardized β coefficient; β = standardized β coefficient.

Linear Regression Analysis Results for the AcuPulse Laser. P < .005. B = unstandardized β coefficient; β = standardized β coefficient. Effects of different settings on incision depth were evaluated by linear regression analysis. The <span class="Chemical">parameter with the greatest effect on incision depth was laser distance; inspan class="Chemical">pan class="Chemical">creasing the distance between laser tip and surface of the gel by 1 mm would result in a despan>n class="Chemical">crease of incision depth by 0.241 mm (B = −0.241, β = −0.722, P < .005). Increased power and laser duration resulted in significantly deeper incisions (B = 0.445, β = 0.612, P < .005 and B = 1.031, β = 0.531, P < .005, respectively); increasing laser power by 1 W would result in an increase in incision depth of 0.445 mm, whereas increasing laser duration by 1 second would result in an increase of 1.031 mm. P and SP modes would result in an increase in incision depth of 1.728 mm and 0.793 mm, respectively, compared to CW mode (B = 1.728, β = 0.471, P < .005 and B = 0.793, β = 0.239, P < .005, respectively).

Gel Model Evaluation

The incision depth and collateral thermal ex<span class="Chemical">pansion in both the gel model and the vocal cords are plotted in Figure 4. The ICC for absolute agreement and consistency for incision depth were 0.221 and 0.534 (P = .037), respectively. The ICC for absolute agreement and consistency of thermal expansion were 0.058 and 0.462 (P = .010), respectively. The values for consistency of incision depth and thermal expansion were classified as fair, using Cicchetti's classification system for interpreting ICC.16
Figure 4

Scatter plots comparing the incision depth and thermal expansion in gel and vocal cord tissue. The dashed line represents y = x, the solid line represents the trend line through the measurements.

Scatter plots com<span class="Chemical">paring the incision depth and thermal ex<span class="Chemical">pansion in gel and vocal cord tissue. The dashed line represents y = x, the solid line represents the trend line through the measurements.

DISCUSSION

Office‐based, flexible endoscopic laser surgery is a promising procedure for the treatment of laryngeal lesions with many potential benefits as an alternative to surgery under general anesthesia.1, 2, 4, 17, 18, 19, 20 As the use of this technique slowly incrspan>eases, it is essential to have knowledge of the thermal effects that individual laser parameters have on vocal cord tissue. An understanding of these effects can assist surgeons in deciding which laser parameter to adjust to achieve the intended surgical effect. This is the first study to systematically describe the thermal effects of a CO2 laser using a flexible fiber delivery system. By comparing the experimental gel model we used in our study to histological evaluation of ex vivo irradiated vocal cords, we demonstrated that, although there does not seem to be absolute agreement, there is a fair level of agreement for consistency. This suggests the tissue model can be used as an alternative to human or porcine vocal cords to interpret the influence of laser settings on thermal effects observed in target tissues. The systematic experiments performed using this model revealed that of all studied <span class="Chemical">parameters, laser duration had the greatest influence on collateral thermal expansion. Additionally, laser duration was significantly related to incision depth. Our study results further indicate that of all laser parameters, power intensity had the greatest positive effect on incision depth. The effect of increasing power on thermal expansion appeared to be relatively small, but significant. Previously, two studies assessed the thermal effect of flexible CO2 laser delivery systems, albeit in only a small selection of setting combinations, in which a trend of increasing thermal damage was seen when using the laser at higher power.3, 7 For incision depth, Wang et al. did not find a difference by altering power settings,7 whereas Wilder‐Smith et al.3 did find increasing incision depths at higher power. The effects of the other parameters have never been evaluated using a flexible fiber delivery system. By comparing our results of a flexible fiber delivery system with results of previous studies testing span class="Chemical">pan class="Chemical">CO2 lasers with a conventional articulated arm or handpiece, we found compaspan>rable outcomes. Although in previous studies on this topic the extent of the tested laser setting combinations was limited, a correlation between laser power and incision depth had been described earlier.3 Additionally, three studies3, 8, 12 also revealed a positive correlation between laser power and thermal damage. In contrast, two other studies did not find a relation between laser power and thermal damage.10, 11 In a study evaluating the surgical margin after CO2 laser surgery in Tis, T1, and T2 glottic carcinoma, Buchanan et al. even observed an inverse relationship between power and thermal tissue damage.9 Remarkably, none of these studies considered laser duration. Buchanan et al. proposed that this inverse correlation could be an indirect effect; using laser irradiation at higher power intensities led to an increase in cutting speed and therefore shorter procedural time, thereby leading to less thermal damage.9 The results of our study emphasize laser duration is a very important parameter to consider in thermal effects of CO2 lasers. This was the first study to assess the influence of the distance between laser and target tissue on thermal effects. Laser distance is a <span class="Chemical">parameter that is of interest only in flexible laser surgery, as laser light emitted by an articulated arm or handpiece is a <span class="Chemical">parallel beam that delivers energy to target tissue independently of the distance. A strong, negative correlation was observed between the incision depth and the laser distance to the gel. This implies using the laser closer to the target tissue results in significantly deeper incisions. This is important to consider when performing office‐based procedures using a flexible fiber, as the distance to the target tissue can vary with the movements of the larynx. Furthermore, span class="Chemical">pan class="Chemical">CO2span> lasers in (super) pulsed modes led to significantly deeper incisions compaspan>red to irradiation in continuous wave mode, with only a minimal effect on thermal expansion. As laser duration was found to have the most prominent effect on thermal expansion of all laser parameters, it may be beneficial when excising vocal cord lesions to use (super) pulsed modes with high‐power intensities to reduce exposure times and thereby reduce unnecessary thermal damage. Similar advice is given in the study by Yan et al., who recommended using the shortest possible time pulse together with the highest possible power to accomplish a procedure with minimal lateral thermal spread.21 Due to the limited amount of larynges available for the com<span class="Chemical">parison between the model setup and vocal cord tissue, we chose only to investigate the span class="Chemical">pan class="Chemical">CO2 laser, as this is currently the most frequently used laser in our center. Because other lasers, such as the span>n class="Chemical">potassium titanyl phosphate laser, have different wavelengths and absorption characteristics, our results cannot be extrapolated directly to these laser types. Additional research must be done using other lasers to compare the effects observed using this model. One of the limitations of the current study is the large amount of missing data for the gel experiments caused by the depth of the incision exceeding the span class="Chemical">pan class="Chemical">Schlierenspan> image frame, which could not be resolved by adjusting the setup. Furthermore, the number of experiments on paspan>n class="Species">human vocal cords was limited due to scarcity of available larynges and due to the fact that several incisions penetrated the entire vocal cord depth, making it impossible to acquire exact measurements. Had our measurements been complete, ICC values might have been higher. For this study, human fresh frozen vocal cords were used to mimic the clinical setting as accurately as possible. However, the tissue properties of these ex vivo samples that were frozen at −5°C and subsequently thawed might differ from in vivo, well‐perfused vocal cord mucosa with a temperature of 37°C, possibly affecting the results. The water content of the vocal cord samples may also differ, which could influence the absorption rate of the CO2 laser and therefore reduce the ablation rate. Despite these limitations, this study remains the first to systematically investigate laser settings to this extent, and the results can be used to gain understanding into the thermal effects caused by these settings.

CONCLUSION

Laser duration was found to be the leading <span class="Chemical">parameter influencing collateral thermal expansion in the Species">human tissue model when using a CO2 laser with a flexible fiber delivery system. Depending on the desired surgical effect (i.e., evaporation or cutting), laser parameters can be adjusted on the basis of our results. Office‐based laryngeal laser surgery is mainly used for lesions that require evaporation as opposed to excision. The results suggest that this can best be achieved by irradiating at greater distance to the tissue, with low‐power intensities and in CW mode, thereby reducing the incision depth and accomplishing broader thermal effects. In contrast, when the laser is applied for cutting thicker lesions, using high‐power (super) pulsed modes at minimal distance to the target tissue can reduce exposure time and subsequent damage to surrounding tissues. Effects seen in the dynamic color Schlieren setup in combination with the tissue model used in this study demonstrate a fair level of consistency with results in ex vivo vocal cords. This indicates that this model can be used to investigate the influence of individual laser parameters on the thermal effects caused by flexible CO2 laser therapy in vocal cord tissue.
  20 in total

1.  Comparison of KTP, Thulium, and CO2 laser in stapedotomy using specialized visualization techniques: thermal effects.

Authors:  Digna M A Kamalski; Rudolf M Verdaasdonk; Tjeerd de Boorder; Robert Vincent; Franco Trabelzini; Wilko Grolman
Journal:  Eur Arch Otorhinolaryngol       Date:  2013-07-24       Impact factor: 2.503

2.  Imaging techniques for research and education of thermal and mechanical interactions of lasers with biological and model tissues.

Authors:  Rudolf M Verdaasdonk; Christiaan F P van Swol; Matthijs C M Grimbergen; Alex I Rem
Journal:  J Biomed Opt       Date:  2006 Jul-Aug       Impact factor: 3.170

3.  Out-patient flexible carbon dioxide laser surgery for benign laryngopharyngeal pathologies via transnasal flexible laryngo-oesophagoscopy.

Authors:  H Mohammed; L Masterson; R Nassif
Journal:  J Laryngol Otol       Date:  2017-04-20       Impact factor: 1.469

4.  Feasibility of transnasal flexible carbon dioxide laser surgery for laryngopharyngeal lesions.

Authors:  Koji Araki; Masayuki Tomifuji; Kosuke Uno; Hiroshi Suzuki; Yuya Tanaka; Shingo Tanaka; Eiko Kimura; Akihiro Shiotani
Journal:  Auris Nasus Larynx       Date:  2019-02-11       Impact factor: 1.863

5.  Fiber-guided CO2 laser surgery in an animal model.

Authors:  Zhi Wang; Anand K Devaiah; Lining Feng; Urmen Dasai; Gil Shapira; Ori Weisberg; David S Torres; Stanley M Shapshay
Journal:  Photomed Laser Surg       Date:  2006-10       Impact factor: 2.796

6.  Surgical utility of a new carbon dioxide laser fiber: functional and histological study.

Authors:  Anand K Devaiah; Stanley M Shapshay; Urman Desai; Gil Shapira; Ori Weisberg; David S Torres; Zhi Wang
Journal:  Laryngoscope       Date:  2005-08       Impact factor: 3.325

7.  Incision properties and thermal effects of three CO2 lasers in soft tissue.

Authors:  P Wilder-Smith; A M Arrastia; L H Liaw; M Berns
Journal:  Oral Surg Oral Med Oral Pathol Oral Radiol Endod       Date:  1995-06

Review 8.  Office-Based Procedures for the Diagnosis and Treatment of Laryngeal Pathology.

Authors:  David J Wellenstein; Henrieke W Schutte; Robert P Takes; Jimmie Honings; Henri A M Marres; James A Burns; Guido B van den Broek
Journal:  J Voice       Date:  2017-09-19       Impact factor: 2.009

9.  Office-based laryngeal laser surgery: a review of 443 cases using three wavelengths.

Authors:  Jamie A Koufman; Catherine J Rees; William D Frazier; Lauren A Kilpatrick; S Carter Wright; Stacey L Halum; Gregory N Postma
Journal:  Otolaryngol Head Neck Surg       Date:  2007-07       Impact factor: 3.497

10.  Histological ex vivo evaluation of peri-incisional thermal effect created by a new-generation CO2 superpulsed laser.

Authors:  G Palaia; A Del Vecchio; A Impellizzeri; G Tenore; P Visca; F Libotte; C Russo; U Romeo
Journal:  ScientificWorldJournal       Date:  2014-02-25
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.