Thomas R Fanshawe1, Margaret Glogowska1, George Edwards1, Philip J Turner1, Ian Smith2, Rosie Steele2, Caroline Croxson3, Jordan S T Bowen2, Gail N Hayward1.
Abstract
INTRODUCTION: Point of care blood testing to aid diagnosis is becoming increasingly common in acute ambulatory settings and enables timely investigation of a range of diagnostic markers. However, this testing allows scope for errors in the pre-analytical phase, which depends on the operator handling and transferring specimens correctly. The extent and nature of these pre-analytical errors in clinical settings has not been widely reported.Entities:
Mesh:
Year: 2020 PMID: 32012203 PMCID: PMC6996845 DOI: 10.1371/journal.pone.0228687
Source DB: PubMed Journal: PLoS One ISSN: 1932-6203 Impact factor: 3.240
Tests available on POCT platforms in use during study period.
| Device | Cartridge | Qualitative (secondary analysis, 2014–15) | Qualitative (primary study, Jan-Feb 2019) | Quantitative |
|---|---|---|---|---|
| i-STAT | ACT | Jan 2016 –Dec 2018 | ||
| BNP | ||||
| CG4+ | ✓ | ✓ | ||
| CHEM8+ | ✓ | ✓ | ||
| Creatinine (Crea) | ||||
| E3+ | ||||
| Glucose (G) | ||||
| β-hCG | ||||
| PT/INR | ✓ | |||
| Troponin (cTnI) | ✓ | |||
| i-STAT Alinity | CG4+ | Apr 2017 –Dec 2018 | ||
| CHEM8+ | ||||
| Creatinine (Crea) | ||||
| Afinion | CRP | ✓ | Jul–Dec 2018 | |
| HbA1c | ||||
| Lipids |
Total test cartridge use, i-STAT.
| Cartridge | Total workload (%) |
|---|---|
| ACT | 4325 (4.1%) |
| BNP | 1713 (1.6%) |
| CG4+ | 10494 (9.9%) |
| CHEM8+ | 22695 (21%) |
| Creatinine (Crea) | 14260 (14%) |
| E3+ | 136 (0.1%) |
| Glucose (G) | 935 (0.9%) |
| β-hCG | 2489 (2.4%) |
| PT/INR | 2902 (2.7%) |
| Troponin (cTnI) | 45269 (43%) |
Total frequencies and error rates of error code categories, i-STAT, February 2017–December 2018.
| Error category | Frequency (%) | Likely user error | Total frequency (%) |
|---|---|---|---|
| Overfill | 573 (0.7%) | ✓ | 1783 (2.3%) |
| Underfill | 709 (0.9%) | ✓ | |
| Insufficient sample | 207 (0.3%) | ✓ | |
| Insertion error | 8 (0.01%) | ✓ | |
| Immunoassay–poor filling | 286 (0.4%) | ✓ | |
| Battery error | 465 (0.6%) | 3321 (4.2%) | |
| Barcode error | 41 (0.05%) | ||
| Coagulation error | 177 (0.2%) | ||
| Cartridge error | 1284 (1.6%) | ||
| Contact error | 394 (0.5%) | ||
| Analyzer/motor error | 522 (0.7%) | ||
| Immunoassay–analysis fluid | 130 (0.2%) | ||
| Immunoassay–error in QC run | 112 (0.1%) | ||
| Immunoassay–atypical data stream | 159 (0.2%) | ||
| Field length | 37 (0.05%) | ||
| TOTAL | 5104 (6.5%) | 5104 (6.5%) |
Percentages are based on a total denominator of 78,153 tests. Refer to S1 Appendix for error category descriptors.
Fig 1i-STAT error rates in 15 locations with highest workload.
Total numbers of returned results, user errors and non-user errors (upper panel); breakdown of all errors by selected error category (middle panel); breakdown of user errors by selected error category (lower panel). In the upper panel, numbers of errors in some locations are too small to be easily visible.
Fig 2Estimated rate of all i-STAT errors and user errors (as a proportion of total workload) by location.
Each line shows the estimate of the proportion in a particular location (blue square) and 95% confidence interval (horizontal bars). The grey vertical line shows the estimated proportion across all locations.
Fig 3Relationship between i-STAT error rate and total workload.
All errors (upper panel) and user-attributed errors post-February 2017 (lower panel). Each point represents a different location. Locations with fewer than 200 total uses are excluded to reduce the effect of small sample variability.
Total frequencies of error code categories, i-STAT Alinity.
| Error category | Frequency | Likely user error | Total frequency (%) |
|---|---|---|---|
| Cause 3 codes (Cartridge rejected) | 40 | ✓ | 95 (5.2%) |
| Cause 6 codes (Excess blood) | 18 | ✓ | |
| Cause 7 codes (Cartridge rejected) | 14 | ✓ | |
| Cause 8 codes (Insufficient blood) | 16 | ✓ | |
| Cause 9 codes (Sample rejected) | 7 | ✓ | |
| Cause 15 codes (Cartridge rejected) | 4 | 15 (0.8%) | |
| Codes of unspecified cause | 11 | ||
| Process error codes | 58 | 58 (3.1%) | |
| TOTAL | 168 | 168 (9.2%) |
Refer to S1 Appendix for error category descriptors.
Total frequencies of error code categories, Afinion.
| Error code | Frequency | Likely user error | TOTAL |
|---|---|---|---|
| 201 (Insufficient volume) | 35 | ✓ | 50 |
| 203 (Wrong sample) | 6 | ✓ | |
| 208 (Test cartridge previously used) | 9 | ✓ | |
| 101, 102 (Hematocrit error) | 7 | 44 | |
| 213, 214 (Cartridge or analyzer failure) | 6 | ||
| 302 (Analyzer failure) | 31 | ||
| 404 (Operator ID error) | 82 | 85 | |
| MT_16002 (Barcode error) | 3 | ||
| TOTAL | 179 | 179 |
Refer to S1 Appendix for error category descriptors.