Shuwei Zhang1, Gongzi Zhang1, Ye Peng1, Xiang Wang1, Peifu Tang1, Lihai Zhang2.
Abstract
BACKGROUND: It is difficult for the surgeon to measure pelvic displacement in the closed reduction operation for unstable pelvic fracture. We therefore developed a pelvic deformity measurement software program based on standardized radiographs. The objectives of the present study were to evaluate the inter-observer reliability of the program for measuring specific fracture types on preoperative pelvic films and to assess the validity of the measurement software program by comparing it with a gold standard.Entities:
Keywords: 3D reconstruction; Measurement software program; Method and formulas; Pelvic closed reduction; Pelvic fractures
Mesh:
Year: 2020 PMID: 32005205 PMCID: PMC6995216 DOI: 10.1186/s13018-020-1558-2
Source DB: PubMed Journal: J Orthop Surg Res ISSN: 1749-799X Impact factor: 2.359
Fig. 1The specific coordinate system of the pelvis in three dimensions
The representative anatomic landmarks selected for the measurement of the displacement of the hemipelvis
| Radiographs | Representative anatomic landmarks |
|---|---|
| AP view | Anterior superior iliac spine |
| Ischial tuberosity | |
| Outlet view | Superior point of iliac wing |
| Ischial tuberosity | |
| ASIS | |
| Center of sacral endplate | |
| Inlet view | ASIS |
| Anterior SI joint (iliac side) | |
| Center of sacral endplate |
ASIS anterior superior iliac spine
Fig. 2Measurement results in the anteroposterior view
Fig. 3Measurement results in the outlet view
Fig. 4Measurement results in the inlet view
Inter-observer analysis for overall displacement
| Measurement method | Kappa value | 95% CI |
|---|---|---|
| Software program | 0.957 | 0.929–0.984 |
| CT coupled with 3D reconstruction | 0.958 | 0.930–0.987 |
| Hand measurement | 0.853 | 0.823–0.883 |
CT computed tomography, 3D three-dimensional, CI confidence interval
Inter-observer analysis for translational and rotational displacement
| Measurement method | Kappa value | 95% CI | |
|---|---|---|---|
| Translational | Software program | 0.974 | 0.926–1.023 |
| CT coupled with 3D reconstruction | 0.964 | 0.915–1.013 | |
| Hand measurement | 0.894 | 0.844–0.945 | |
| Rotational | Software program | 0.936 | 0.890–0.982 |
| CT coupled with 3D reconstruction | 0.949 | 0.902–0.996 | |
| Hand measurement | 0.802 | 0.752–0.853 |
CT computed tomography, 3D three-dimensional, CI confidence interval
Validity analysis for hemipelvic displacement
| Measurement method | Kappa value | 95% CI |
|---|---|---|
| Software program | 0.939 | 0.905–0.973 |
| CT coupled with 3D reconstruction | 0.969 | 0.934–1.003 |
| Hand measurement | 0.858 | 0.823–0.894 |
CT computed tomography, 3D three-dimensional, CI confidence interval