Dual-fiber laser Doppler velocimeter and its application to the measurements of coronary blood velocity.

F Kajiya1, O Hiramatsu, Y Ogasawara, K Mito, K Tsujioka.   

Abstract

To obtain a smaller sample volume and a suitable sample position for the measurement of blood velocity, we fabricated a laser Doppler velocimeter (LDV) with a dual-fiber pickup. The two fibers (clad: 62.5 micron and core: 50 micron) were placed side by side. An He-Ne laser was introduced into the blood through one fiber and the backscattered light was collected by the other fiber. The Doppler signal was analyzed by a spectrum analyzer. The spectrum of the Doppler shift frequency showed a sharp peaked pattern for both forward and reverse flows and exhibited an excellent correlation with the known blood velocity. The blood velocity in the poststenotic portion of canine coronary artery was successfully measured by the dual-fiber LDV. These results indicate that the dual-fiber LDV is useful for measuring blood velocity accurately with a small sample volume even in disturbed flow fields.