Ultrasonic dissection versus electrocautery for immediate prosthetic breast reconstruction.

Dongeun Lee1, Bok Ki Jung1, Tai Suk Roh1, Young Seok Kim1.   

Abstract

BACKGROUND: Ultrasonic dissection devices cause less thermal damage to the surrounding tissue than monopolar electrosurgical devices. We compared the effects of using an ultrasonic dissection device or an electrocautery device during prosthetic breast reconstruction on seroma development and short-term postoperative complications.
METHODS: We retrospectively reviewed the medical records of patients who underwent implant-based reconstruction following mastectomy between March 2017 and September 2018. Mastectomy was performed by general surgeons and reconstruction by plastic surgeons. From March 2017 to January 2018, a monopolar electrosurgical device was used, and an ultrasonic dissection device was used thereafter. The other surgical.
METHODS: were the same in both groups.
RESULTS: The incidence of seroma was lower in the ultrasonic dissection device group than in the electrocautery group (11 [17.2%] vs. 18 [31.0%]; P=0.090). The duration of surgery, total drainage volume, duration of drainage, overall complication rate, surgical site infection rate, and flap necrosis rate were comparable between the groups. Multivariate analysis revealed that the risk of seroma development was significantly lower in the ultrasonic dissection device group than in the electrocautery group (odds ratio for electrocautery, 3.252; 95% confidence interval, 1.242-8.516; P=0.016).
CONCLUSIONS: The findings of this study suggest that the incidence of seroma can be reduced slightly by using an ultrasonic dissection device for prosthesis-based breast reconstruction. However, further randomized controlled studies are required to verify our.
RESULTS: and to assess the cost-effectiveness of this technique.

Entities:  

Keywords:  Breast; Breast implants; Electrocoagulation; High-intensity focused ultrasound ablation; Mammaplasty

Year:  2020        PMID: 31964119      PMCID: PMC6976748          DOI: 10.5999/aps.2019.00759

Source DB:  PubMed          Journal:  Arch Plast Surg        ISSN: 2234-6163


INTRODUCTION

Seroma, which is one of the most common complications in prosthetic breast reconstruction, increases the incidence of other morbidities, such as surgical site infection, and the length of hospital stay. Of particular note, in patients with breast cancer, seroma development can lead to delays in the administration of adjuvant therapy. Several hypotheses have been proposed regarding seroma development. Seromas may be the consequence of surgical disruption of both the lymphatics and capillaries [1], injuries of the surrounding lymph vessels due to heat dispersion [2], and the consequence of inflammatory exudates [3]. In our opinion, all three of these hypotheses may be valid to some extent [4]. The causes of seroma development in breast reconstruction are thought to include the presence of a large irregular dead space, lymphatic disruption that occurs after total mastectomy, insertion of a foreign body, chest wall movement, and persistent inflammation in response to the tissue damage that occurs during the surgical process [1,4-6]. Monopolar electrosurgical devices have been frequently used for nearly 50 years. Fulguration, precise vaporization, and coaptation of large vessels are facilitated by using an electrosurgical device instead of a conventional scalpel; thus, bleeding is controlled relatively easily, which is advantageous for securing visibility and reducing operating time [7]. However, the use of a monopolar electrosurgical device is associated with the risk of seroma formation after surgery because it causes thermal damage to the surrounding tissue. Studies have reported that using a monopolar electrosurgical device resulted in an increase in the seroma formation rate and drainage volume compared to a conventional scalpel [8]. Therefore, various advantages and disadvantages are associated with the use of monopolar electrosurgical devices in breast surgery. Recently, ultrasonic dissection devices have been introduced as an alternative. These devices use ultrasonic waves to dissect the tissue, making it possible to perform a more sophisticated dissection with a narrower circumference of collateral thermal damage to the surrounding tissue compared to the damage caused by monopolar electrosurgical devices [9]. Furthermore, bleeding control and dissection can be performed simultaneously, and there is a lack of surgical smoke and reduced charring; consequently, a clearer visual field can be secured. Studies have already reported that using an ultrasonic dissection device instead of an electrosurgical device for radical mastectomy was associated with a significant decrease in the incidence of seroma development [10-12]. To date, no study has investigated the surgical outcomes obtained by using an ultrasonic dissection device in comparison with those obtained by using a monopolar electrosurgical energy device during prosthetic breast reconstruction. Therefore, this study compared the effects of using an ultrasonic dissection device or a monopolar electrosurgical device during prosthesis-based breast reconstruction on seroma formation and short-term postoperative complications.

METHODS

Study design

The study was conducted at the plastic surgery department of our hospital, and the study protocol was approved by the Institutional Review Board of Gangnam Severance Hospital. Patients who underwent total mastectomy at the breast surgery unit and immediate reconstruction with a prosthesis from March 2017 to September 2018 were enrolled. We retrospectively reviewed the patients’ medical records. Patients who underwent bilateral breast reconstruction, those who underwent additional procedures such as contralateral breast manipulations or fat grafting, and those with implants in the prepectoral plane were excluded. We recorded patients’ age, body mass index (BMI), the presence of hypertension and diabetes mellitus, smoking history, breast ptosis grade, mastectomy type (nipple-sparing mastectomy or total mastectomy), axillary operation status, neoadjuvant treatment, breast specimen weight, and reconstruction type (tissue expander or direct-to-implant). Two general surgeons performed the mastectomies, and the reconstructions were performed by a single plastic surgeon. From March 2017 to January 2018, patients underwent surgery with a monopolar electrosurgical device (group 1) and from February 2018 to September 2018, an ultrasonic dissection device was used (group 2).

Surgical technique

After total mastectomy was performed by the breast surgeon, immediate breast reconstruction was performed by the plastic surgeon. First, we performed hemostasis in the whole surgical field using an ultrasonic dissection device (Harmonic; Ethicon, Cincinnati, OH, USA). The origin of the pectoralis major muscle was dissected and elevated using the instrument (Fig. 1). After the muscle was adequately elevated, we made a slit incision on a piece of acellular dermal matrix (ADM), which had been immersed in an antibiotic solution, and fixed the ADM in the new inframammary fold location using Vicryl 3-0 sutures. We then fixed the edges of the pectoralis major muscle with the ADM to create a pocket where the mammary implant or tissue expander would be inserted. A 400-mL Hemovac was inserted in the margin of the breast pocket. We copiously irrigated the breast pocket with a triple antibiotic solution (cefazolin, gentamycin, and metronidazole). After we applied amorphous betadine on the breast skin, we changed into aseptic gloves and inserted the mammary implant (or tissue expander) into the breast pocket. Then, we closed the pocket by suturing the ADM and the pectoralis major muscle. After subcutaneous suturing, subdermal suturing was performed using the buried technique, and skin closure was performed using 5-0 nylon sutures. A mild compressive dressing was applied, and the operation was completed. Similar surgical techniques were used in the monopolar electrosurgical device group and the ultrasonic scalpel group; only the dissection device was different between the groups.
Fig. 1.

Device and intraoperative image of its use

(A) Harmonic Focus, ultrasonic shears, curved tip (Ethicon EndoSurgery, Inc., Cincinnati, OH, USA). (B, C) During the ultrasonic dissection to elevate the pectoralis major muscle, there was no muscle contraction, and hemostasis was straightforward. Furthermore, there was less thermal damage to the nearby tissue.

Outcome variables

The outcome variables were duration of surgery (minutes), drain volume (mL), duration of drainage (days) and postoperative complications (e.g., seroma, hematoma, flap necrosis, or surgical site infection). Seroma was defined as the presence of a fluid collection within 30 days of surgery beneath the skin flaps after drain removal, in sufficient quantity to cause the patient discomfort. Surgical site infections were assessed using the Centers for Disease Control and Prevention criteria, according to which an SSI is defined as an infection of the skin and subcutaneous tissue that occurs within 30 days postoperatively and satisfies at least one of the following criteria: (1) purulent drainage from the superficial incision; (2) organisms (other than Staphylococcus epidermidis) isolated from an aseptically obtained culture of fluid or tissue from the superficial incision; (3) at least one of the given signs or symptoms of infection (i.e., pain or tenderness; localized swelling; redness; heat; superficial incision deliberately opened by surgeon and was culture-positive or not cultured [a culture-negative finding does not meet this criterion]); and (4) diagnosis of a superficial incisional surgical site infection by the surgeon or attending physician. Hematoma was defined as the need to perform aspiration or surgical evacuation due to the collection of blood under the flap. Flap necrosis was defined as the need to perform surgical revision due to a flap wound that occurred after surgery.

Statistical analysis

Statistical analysis was conducted using SPSS version 23.0 (IBM Corp., Armonk, NY, USA). Baseline demographic characteristics (Table 1), treatment information (Table 2), and outcomes (Tables 3, 4) were compared using the Student t-test, chi-square test, and Fisher exact test according to the intervention device. Univariate and multivariate analyses were performed using logistic regression to analyze the effects of the variables in each group. The multivariate analysis was performed using variables identified as affecting seroma development in the univariate analysis (intervention device, age, weight of specimen, BMI, and reconstruction type).
Table 1.

Patient characteristics

VariableGroup 1Group 2P-value
Age (yr)47.6 ± 7.045.1 ± 6.80.052
BMI (kg/m2)23.3 ± 2.822.4 ± 2.80.089
Comorbidity
 DM4 (6.3)1 (1.7)0.368
 HTN1 (1.6)1 (1.7)> 0.999
 Smoking1 (1.6)3 (5.2)0.345
Ptosis grade0.401
 026 (40.6)32 (55.2)
 124 (37.5)17 (29.3)
 29 (14.1)7 (12.1)
 35 (7.8)2 (3.4)

Values are presented as mean±SD or number (%).

Group 1, monopolar electrosurgical device group; group 2, ultrasonic dissection device group; BMI, body mass index; DM, diabetes mellitus; HTN, hypertension.

Table 2.

Treatment information

VariableGroup 1Group 2P-value
Type of mastectomy0.379
 Nipple-sparing mastectomy27 (42.2)30 (51.7)
 Skin-sparing mastectomy14 (21.9)14 (24.1)
 Total mastectomy23 (35.9)14 (24.1)
Axillary operation0.679
 SLNB49 (76.6)42 (72.4)
 ALND15 (23.6)16 (27.6)
Neoadjuvant chemotherapy17 (26.6)12 (20.7)0.525
Specimen weight (g)460.9 ± 203.9356.3 ± 185.90.004
General surgeon0.425
 A42 (65.6)34 (58.6)
 B22 (34.4)24 (41.4)
Reconstruction type0.229
 Tissue expander insertion40 (62.5)30 (51.7)
 Direct-to-implant24 (37.5)28 (48.3)

Values are presented as number (%) or mean±SD.

Group 1, monopolar electrosurgical device group; group 2, ultrasonic dissection device group; SLNB, sentinel lymph node biopsy; ALND, axillary lymph node dissection.

Table 3.

Outcomes of surgery

VariableGroup 1Group 2P-value
Duration of surgery (min)102.7 ± 24.498.2 ± 31.70.439
Drain volume (mL)1,677.7 ± 639.01,762.7 ± 532.10.484
Duration of drainage (day)18.9 ± 4.517.8 ± 3.30.205

Values are presented as mean±SD.

Group 1, monopolar electrosurgical device group; group 2, ultrasonic dissection device group.

Table 4.

Complications of surgery

VariableGroup 1Group 2P-value
Seroma11 (17.2)18 (31.0)0.090
Hematoma05 (8.6)0.022
Surgical site infection2 (3.1)3 (5.2)0.668
Flap necrosis5 (7.8)4 (6.9)1.000
Overall complication rate14 (26.5)16 (39.1)0.274

Values are presented as number (%).

Group 1, monopolar electrosurgical device group; group 2, ultrasonic dissection device group.

RESULTS

The monopolar electrosurgical device group (group 1) comprised 58 breasts, while the ultrasonic dissection device group (group 2) comprised 64 breasts. In a comparison of the baseline demographic characteristics of the two groups, we found that the patients in group 2 were older (group 1 vs. group 2: 47.6± 7.0 years vs. 45.1±6.8 years, P=0.052), and that the average BMI was lower in group 1 (group 1 vs. group 2: 23.3±2.8 kg/m2 vs. 22.4±2.8 kg/m2, P=0.089). However, neither age nor BMI was statistically significant. Hypertension, diabetes mellitus, smoking, and ptosis grade were also similar between the groups (Table 1). In group 2, the specimen weight was significantly greater (group 1 vs. group 2: 460.9±203.9 g vs. 356.3±185.9 g, P=0.004) We surveyed the types of mastectomy, axillary operation status, chemotherapy, and reconstruction type, and these variables were comparable (Table 2). Furthermore, the duration of surgery, drain volume, and duration of drainage were similar in both groups (Table 3). With respect to surgical outcomes, the seroma development rate was relatively high in group 1 (11 [17.2%] vs. 18 [31.0%], P=0.090). Hematoma did not develop in any of the patients in group 2, but five cases of hematoma development were noted in group 1 (5 [8.6%] vs. 0 [0.0%], P=0.022). The duration of surgery, total drainage volume, duration of drainage, overall complication rate, surgical site infection rate, and flap necrosis rate were comparable (Table 4). Univariate analysis was performed to identify the factors affecting the seroma development rate, and we found that the risk of seroma development was higher in group 1 than in group 2 (odds ratio [OR] for electrocautery, 2.168; 95% CI, 0.922–5.099; P=0.076). Multivariate analysis was performed using variables that could clinically affect the occurrence of seroma. Age, BMI, specimen weight, reconstruction type, and intervention type were included in the multivariable analysis; a significantly lower risk of seroma development was observed in the ultrasonic dissection device group (OR for electrocautery, 3.252; 95% CI, 1.242–8.516; P=0.016) (Table 5).
Table 5.

Multivariate analysis of variables affecting the likelihood of seroma development

Adjusted OR (95% CI)P-value
Intervention0.016
 Ultrasonic dissection1 (reference)
 Electrocautery3.252 (1.242–8.516)
Age1.053 (0.986–1.125)0.123
Weight of specimen1.002 (0.999–1.005)0.198
BMI1.025 (0.831–1.265)0.816
General surgeon0.174
 A1 (reference)
 B0.497 (0.181–1.363)
Reconstruction type0.143
 Direct-to-implant1 (reference)
 Tissue expander2.033 (0.787–5.252)

OR, odds ratio; CI, confidence interval; BMI, body mass index.

DISCUSSION

We aimed to compare the effects of using an ultrasonic dissection device and a monopolar electrosurgical device on seroma formation and short-term outcomes in patients undergoing prosthetic breast reconstruction after mastectomy, and found that that the rate of seroma development was lower in the ultrasonic dissection device group (group 2). The incidence of short-term complications was lower in group 2, but this tendency was not statistically significant. There was no statistically significant difference in the operative time between the two groups. The total drainage volume, duration of drainage, surgical site infection rate, and flap necrosis rate were also comparable between the two groups. Ultrasonic dissection devices have been reported to be superior to monopolar electrosurgical devices in many studies. They provide the advantages of better hemostasis with minimal thermal damage, a reduced risk of nerve damage, the use of fewer instruments due to their combined vessel-sealing, tissue-cutting, and tissue-dissecting functionality, and less visual obstruction from smoke [13-16]. Using the principle of breaking hydrogen bonds due to the generation of ultrasonic energy, these devices dissect the tissue and seal the vessels and lymphatics. This method was first used in minimally invasive surgery, including laparoscopic surgery; since then, it has been gradually proven to be superior to traditional dissection techniques, as shown by a decreased incidence of postoperative complications, including a lower rate of seroma development. The superiority of ultrasonic dissection in breast surgery remains controversial, but a report has described good outcomes for a recently developed modified radical mastectomy technique using an ultrasonic dissection device compared to those reported for mastectomy performed using a monopolar electrosurgical device [10]. Despite being a retrospective study, this study was the first to compare the short-term complications of surgery using a monopolar electrocautery device and an ultrasonic dissection device in immediate implant-based breast reconstruction. Cases in which factors other than the tools used in the intervention could have affected surgical outcomes were excluded from this study. In addition, recommendations of ways to reduce the complication rate in prosthetic breast reconstruction, such as using ADM, irrigation of the breast pocket with antibiotic solution, and the surgeon wearing sterile gloves before insertion of the implant, were applied in the surgical procedures presented herein. Seroma formation can be caused by many factors. Currently, the leading causes of seroma formation are thought to be persistent inflammation after mastectomy, thermal damage to tissue, and lymphatic disruption. In this study, the use of an ultrasonic dissection device decreased the seroma formation rate. It seems that when elevating the pectoralis major muscle, there was less thermal damage of the anterior chest wall and muscle, which reduced the formation of seroma due to inflammation. In the hemostasis process, the ultrasonic scalpel is applied only to selective areas, unlike when a monopolar electrosurgical device is used; therefore, tissue damage is markedly reduced in the overall area of the surgical field. In addition to a reduced incidence of seroma development, ultrasonic dissection devices have many other advantages. Charring, which occurs when using a monopolar electrosurgical device for muscle elevation, is not produced with this device. Charring can be a risk factor for prosthesis-related infection because the charred substance is not entirely removed during the irrigation performed before inserting an implant. Therefore, direct contact between the charred substance and the prosthesis is eliminated, thereby reducing the likelihood of infection. The need to administer a muscle relaxant is eliminated because ultrasonic dissection devices do not cause muscle contractions. They also reduce the possibility of anesthesia-related complications, such as anaphylaxis, and prevent an unnecessary prolongation of anesthesia time. There are several limitations of this study. First, the sample size of the two groups was not sufficiently large, which may explain the presence of between-group differences in specimen weight and BMI. However, a sufficiently large sample size would be expected to eliminate such differences between the two groups. Of particular note, a statistically significant difference was noted for specimen weight (group 1 vs. group 2: 460.9±203.9 g vs. 356.3±185.9 g, P=0.004). However, higher-weight specimens are generally associated with increased seroma incidence and drainage volume. Therefore, the multivariate analysis adjusted for the possible effects of differences in specimen weight on the incidence of seroma in the monopolar electrosurgical group. Second, the general surgeons did not use the ultrasonic dissection device in a way consistent with how the plastic surgeon did so. The surgical technique used in mastectomy is expected to have a significant effect on seroma formation after reconstruction. In conclusion, the rate of seroma formation can be slightly reduced by using an ultrasonic dissection device for immediate implant-based breast reconstruction. However, further randomized controlled studies are required to verify our results and to assess the cost-effectiveness of this technique.
  16 in total

Review 1.  Surgical Efficacy Among Laparoscopic Ultrasonic Dissectors: Are We Advancing Safely? A Review of Literature.

Authors:  Rajesh Devassy; Sreelatha Gopalakrishnan; Rudy Leon De Wilde
Journal:  J Obstet Gynaecol India       Date:  2015-09-14

Review 2.  "Current thoughts" in electrosurgery.

Authors:  K Wang; A P Advincula
Journal:  Int J Gynaecol Obstet       Date:  2007-04-23       Impact factor: 3.561

3.  Seroma formation after breast cancer surgery: incidence and predicting factors.

Authors:  P A Woodworth; M F McBoyle; S D Helmer; R L Beamer
Journal:  Am Surg       Date:  2000-05       Impact factor: 0.688

Review 4.  Energy sources for gynecologic laparoscopic surgery: a review of the literature.

Authors:  Kenneth S K Law; Jason A Abbott; Stephen D Lyons
Journal:  Obstet Gynecol Surv       Date:  2014-12       Impact factor: 2.347

Review 5.  Laparoscopic vessel sealing technologies.

Authors:  Stephen D Lyons; Kenneth S K Law
Journal:  J Minim Invasive Gynecol       Date:  2013 May-Jun       Impact factor: 4.137

Review 6.  Pathophysiology of seroma in breast cancer.

Authors:  Katsumasa Kuroi; Kojiro Shimozuma; Tetsuya Taguchi; Hirohisa Imai; Hiroyasu Yamashiro; Shozo Ohsumi; Shinya Saito
Journal:  Breast Cancer       Date:  2005       Impact factor: 4.239

7.  Aetiology of seroma formation in patients undergoing surgery for breast cancer.

Authors:  J A McCaul; A Aslaam; R J Spooner; I Louden; T Cavanagh; A D Purushotham
Journal:  Breast       Date:  2000-06       Impact factor: 4.380

8.  Harmonic focus versus electrocautery in axillary lymph node dissection for breast cancer: a randomized clinical study.

Authors:  Qingqing He; Dayong Zhuang; Luming Zheng; Ziyi Fan; Peng Zhou; Jian Zhu; Zhen Lv; Jixin Chai; Lei Cao
Journal:  Clin Breast Cancer       Date:  2012-10-03       Impact factor: 3.225

9.  A novel technique of harmonic tissue dissection reduces seroma formation after modified radical mastectomy compared to conventional electrocautery: a single-blind randomized controlled trial.

Authors:  Mohammed Faisal; Hamada Fathy; Hamdy Shaban; Sameh T Abuelela; Ahmed Marie; Islam Khaled
Journal:  Patient Saf Surg       Date:  2018-05-17

10.  Collagen sealant patch to reduce lymphatic drainage after lymph node dissection.

Authors:  Gianluca Di Monta; Corrado Caracò; Anna Crispo; Ugo Marone; Nicola Mozzillo
Journal:  World J Surg Oncol       Date:  2012-12-19       Impact factor: 2.754

View more
  3 in total

1.  Risk Factors for Postoperative Complications Following Aesthetic Breast Surgery: A Retrospective Cohort Study of 4973 Patients in China.

Authors:  Lingya Zhang; Jinlong Zheng; Jingpeng Mu; Yunqian Gao; Gehong Li
Journal:  Aesthetic Plast Surg       Date:  2022-08-03       Impact factor: 2.708

2.  A Prospective Randomised Controlled Study Comparing Ultrasonic Dissector with Electrocautery for Axillary Dissection in Patients of Carcinoma Breast.

Authors:  Ananya Deori; Nikhil Gupta; Arun Kumar Gupta; Raghav Yelamanchi; Himanshu Agrawal; C K Durga
Journal:  Malays J Med Sci       Date:  2021-02-24

3.  Comparing Seroma Formation at the Deep Inferior Epigastric Perforator, Transverse Musculocutaneous Gracilis, and Superior Gluteal Artery Perforator Flap Donor Sites after Microsurgical Breast Reconstruction.

Authors:  Alisha Merchant; Nicole E Speck; Michal Michalak; Dirk J Schaefer; Jian Farhadi
Journal:  Arch Plast Surg       Date:  2022-07-30
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.