System accuracy evaluation of 18 CE-marked current-generation blood glucose monitoring systems based on EN ISO 15197:2015.

Stefan Pleus1, Annette Baumstark2, Nina Jendrike2, Jochen Mende2, Manuela Link2, Eva Zschornack2, Cornelia Haug2, Guido Freckmann2.   

Abstract

OBJECTIVE: Accuracy of 18 current-generation blood glucose monitoring systems (BGMS) available in Europe was evaluated applying criteria adapted from EN ISO 15197:2015 with one reagent system lot. BGMS were selected based on market research data. RESEARCH DESIGN AND METHODS: The BGMS ABRA, Accu-Chek Guide, AURUM, CareSens Dual, CERA-CHEK 1CODE, ContourNext One, eBsensor, FreeStyle Freedom Lite, GL50 evo, GlucoCheck GOLD, GlucoMen areo 2K, GluNEO, MyStar DoseCoach, OneTouch Verio Flex, Pic GlucoTest, Rightest GM700S, TRUEyou, and WaveSense JAZZ Wireless were tested using capillary blood from 100 different subjects and assessing the percentage of results within ±15 mg/dL (0.83 mmol/L) or 15% of comparison method results for BG concentrations below or above 100 mg/dL (5.55 mmol/L), respectively. In addition, the minimal deviation from comparison method results within which ≥95% of results of the respective BGMS were found was calculated.
RESULTS: In total, 14 BGMS had ≥95% of results within ±15 mg/dL (0.83 mmol/L) or ±15% and 3 BGMS had ≥95% of results within ±10 mg/dL (0.55 mmol/L) or ±10% of the results obtained with the comparison method. The smallest deviation from comparison method results within which ≥95% of results were found was ±7.7 mg/dL (0.43 mmol/L) or ±7.7%; the highest deviation was ±19.7 mg/dL (1.09 mmol/L) or ±19.7%.
CONCLUSIONS: This accuracy evaluation shows that not all CE-labeled BGMS fulfill accuracy requirements of ISO 15197 reliably and that there is considerable variation even among BGMS fulfilling these criteria. This safety-related information should be taken into account by patients and healthcare professionals when making therapy decisions. TRIAL REGISTRATION NUMBER: NCT03737188. © Author(s) (or their employer(s)) 2020. Re-use permitted under CC BY. Published by BMJ.

Entities:  

Keywords:  accuracy; blood glucose monitors; blood glucose self-monitoring; health care research

Year:  2020        PMID: 31958308     DOI: 10.1136/bmjdrc-2019-001067

Source DB:  PubMed          Journal:  BMJ Open Diabetes Res Care        ISSN: 2052-4897


  6 in total

1.  Accuracy and User Performance of a New Blood Glucose Monitoring System.

Authors:  Leslie Klaff; Pragathi Shelat; Diana Zondorak; Amy Wayland-Smith; Phil Vernes; James M Richardson
Journal:  J Diabetes Sci Technol       Date:  2020-11-26

2.  Impact of Two Different Reference Measurement Procedures on Apparent System Accuracy of 18 CE-Marked Current-Generation Blood Glucose Monitoring Systems.

Authors:  Guido Freckmann; Annette Baumstark; Nina Jendrike; Jochen Mende; Sebastian Schauer; Manuela Link; Stefan Pleus; Cornelia Haug
Journal:  J Diabetes Sci Technol       Date:  2020-08-19

3.  Do We Need the Replacement of YSI 2300? A View from the Clinical Laboratory.

Authors:  Loukia Spanou; Konstantinos Makris
Journal:  J Diabetes Sci Technol       Date:  2021-05-17

4.  Mean Absolute Relative Difference of Blood Glucose Monitoring Systems and Relationship to ISO 15197.

Authors:  Guido Freckmann; Jochen Mende; Stefan Pleus; Delia Waldenmaier; Annette Baumstark; Nina Jendrike; Cornelia Haug
Journal:  J Diabetes Sci Technol       Date:  2021-03-24

Review 5.  Glucose biosensors in clinical practice: principles, limits and perspectives of currently used devices.

Authors:  Salvatore Andrea Pullano; Marta Greco; Maria Giovanna Bianco; Daniela Foti; Antonio Brunetti; Antonino S Fiorillo
Journal:  Theranostics       Date:  2022-01-01       Impact factor: 11.600

6.  Self-Monitoring of Blood Glucose as an Integral Part in the Management of People with Type 2 Diabetes Mellitus.

Authors:  Stefan Pleus; Guido Freckmann; Sebastian Schauer; Lutz Heinemann; Ralph Ziegler; Linong Ji; Viswanathan Mohan; Luis Eduardo Calliari; Rolf Hinzmann
Journal:  Diabetes Ther       Date:  2022-04-13       Impact factor: 3.595

  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.