| Literature DB >> 31921883 |
Viktor Holmdahl1,2, Olof Backman1, Ulf Gunnarsson1, Karin Strigård1.
Abstract
Purpose: Parastomal hernia is a common complication of an enterostoma. Current methods of repair have high recurrence rates and are associated with severe complications. Autologous full-thickness skin as reinforcement may reduce the recurrence and complication rates. This study aims to investigates the tensile strength of full-thickness skin; information that is essential if we are to proceed with clinical trials on humans.Entities:
Keywords: biological mesh; composite mesh; full-thickness skin; parastomal hernia; tensile strength
Year: 2019 PMID: 31921883 PMCID: PMC6915078 DOI: 10.3389/fsurg.2019.00069
Source DB: PubMed Journal: Front Surg ISSN: 2296-875X
Figure 1Clamping device with an attached biological collagen matrix sample cut to the dumbbell shape.
Figure 2Clamped skin sample with penetrating suture attached to the dynamometer.
Patient characteristics and individual test results.
| 1 | Hemicolectomy | 70–80 | X | X | X | 2, 2 | 156 | 713 | 34 | − | 0 |
| 2 | Abdominoplasty | 50–60 | − | − | − | 2, 4 | 764 | 3,183 | 76 | 0 | X |
| 3 | Breast reconstruction | 60–70 | − | − | − | 1, 8 | 396 | 2,198 | 83 | X | X |
| 4 | Breast reconstruction | 50–60 | − | − | − | 2 | 1,395 | 6,977 | 66 | − | X |
| 5 | Abdominoplasty | 30–40 | − | − | − | 3,7 | 912 | 2,465 | 72 | X | 0 |
| 6 | Hemicolectomy | 80–90 | X | − | − | 1, 8 | 564 | 3,133 | 64 | X | 0 |
| 7 | Hemicolectomy | 70–80 | X | − | − | 1, 3 | 477 | 3,672 | 65 | X | X |
| 8 | Abdominoplasty | 60–70 | − | − | − | 1, 9 | 714 | 3,759 | 67 | X | X |
| 9 | Whipple operation | 70–80 | X | − | − | 2 | 1,294 | 6,469 | 71 | X | 0 |
| 10 | Abdominoperineal resection | 60–70 | X | − | − | 2 | 644 | 3,219 | 64 | X | 0 |
| 11 | Hemicolectomy + liver resection | 70–70 | X | − | − | 0 | 0 | 0 | 0 | 0 | 0 |
| 12 | Hemicolectomy | 50–60 | X | X | − | 1, 9 | 463 | 2,439 | 71 | X | − |
| 13 | Whipple operation | 80–90 | X | − | − | 1, 4 | 259 | 1,852 | 60 | − | − |
| 14 | Liver resection | 60–70 | X | − | − | 0 | 0 | 0 | 0 | 0 | 0 |
Break not at waist: true value higher. X, yes; −, no; 0, no measurement.
Comparison between the different materials.
| Synthetic | 6 | 40 (5) | 0.0007 | 6 | 30 (5) | 0.0007 |
| Biologic | 13 | 208 (36) | 0.0005 | 7 | 63 (6, 1) | 0.0687 |
| Skin | 12 | 604 (408) | REF | 12 | 67 (8) | REF |
Median (IQR).