Pleural fluid glucose testing using a finger stick glucometer: a novel bedside test.

Tony F Abdo1, Himanshu Bhardwaj1, Muhammad K Ishaq1, Jean I Keddissi1, Houssein A Youness1.   

Abstract

BACKGROUND: Pleural fluid glucose (PFG) has diagnostic and therapeutic implications for the management of pleural effusion. The literature examining point-of-care testing of PFG is limited, and no studies exist for the bedside measurement of PFG using a glucometer (B-PFG). In this study, we compared the accuracy of B-PFG measurement to standard in-lab measurement (Lab-PFG).
METHODS: Patients undergoing thoracentesis or thoracostomy were enrolled. PFG was measured at the bedside with a finger stick blood glucometer (ACCU-CHEK® Inform II, Roche) and in the laboratory. Two consecutive measurements were taken using the glucometer, and their mean was compared to the glucose concentration measured in the laboratory. Pearson correlation coefficient and Bland-Altman Plot analysis were used to compare the two measurements.
RESULTS: Sixty patients were included. Mean age was 64.1 years. Forty-nine patients had exudative effusions (41% malignant, 26% parapneumonic, and 33% others). There was a significant correlation between the B-PFG and the Lab-PFG (r=0.98, 95% CI of 0.97 to 0.99; P<0.0001). There was good agreement between the B-PFG and the Lab-PFG with a mean difference of 14.8 mg/dL [95% limit of agreement (LOA) of -2.2 to 31.8 mg/dL]. This agreement was even better at glucose values less than 80 mg/dL.
CONCLUSIONS: PFG measured at the bedside with a glucometer closely correlates with the laboratory measurement. Further studies are needed prior to incorporating this test in clinical practice. 2019 Journal of Thoracic Disease. All rights reserved.

Entities:  

Keywords:  Pleural effusion; glucometer; pleural fluid glucose (PFG); thoracentesis

Year:  2019        PMID: 31903280      PMCID: PMC6940255          DOI: 10.21037/jtd.2019.09.04

Source DB:  PubMed          Journal:  J Thorac Dis        ISSN: 2072-1439            Impact factor:   2.895


  19 in total

1.  BTS guidelines for the management of pleural infection.

Authors:  C W H Davies; F V Gleeson; R J O Davies
Journal:  Thorax       Date:  2003-05       Impact factor: 9.139

2.  Investigation of a unilateral pleural effusion in adults: British Thoracic Society Pleural Disease Guideline 2010.

Authors:  Clare Hooper; Y C Gary Lee; Nick Maskell
Journal:  Thorax       Date:  2010-08       Impact factor: 9.139

3.  Clinically important factors influencing the diagnostic measurement of pleural fluid pH and glucose.

Authors:  Najib M Rahman; Eleanor K Mishra; Helen E Davies; Robert J O Davies; Y C Gary Lee
Journal:  Am J Respir Crit Care Med       Date:  2008-06-12       Impact factor: 21.405

Review 4.  State of the art. The pleura.

Authors:  S A Sahn
Journal:  Am Rev Respir Dis       Date:  1988-07

5.  Low glucose and pH levels in malignant pleural effusions. Diagnostic significance and prognostic value in respect to pleurodesis.

Authors:  F Rodríguez-Panadero; J López Mejías
Journal:  Am Rev Respir Dis       Date:  1989-03

6.  Pleural effusions: the diagnostic separation of transudates and exudates.

Authors:  R W Light; M I Macgregor; P C Luchsinger; W C Ball
Journal:  Ann Intern Med       Date:  1972-10       Impact factor: 25.391

7.  Statistical methods for assessing agreement between two methods of clinical measurement.

Authors:  J M Bland; D G Altman
Journal:  Lancet       Date:  1986-02-08       Impact factor: 79.321

Review 8.  Parapneumonic effusion and empyema.

Authors:  H Hamm; R W Light
Journal:  Eur Respir J       Date:  1997-05       Impact factor: 16.671

9.  Measuring pleural fluid pH: high correlation of a handheld unit to a traditional tabletop blood gas analyzer.

Authors:  G L Kohn; W D Hardie
Journal:  Chest       Date:  2000-12       Impact factor: 9.410

Review 10.  Do we measure pleural fluid pH correctly?

Authors:  Benjamin Putnam; Azeem Elahi; Mark R Bowling
Journal:  Curr Opin Pulm Med       Date:  2013-07       Impact factor: 3.155

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.