Lateral pressure equalisation as a principle for designing support surfaces to prevent deep tissue pressure ulcers.

Colin J Boyle1,2, Diagarajen Carpanen1, Thanyani Pandelani1, Claire A Higgins1, Marc A Masen2, Spyros D Masouros1.   

Abstract

When immobile or neuropathic patients are supported by beds or chairs, their soft tissues undergo deformations that can cause pressure ulcers. Current support surfaces that redistribute under-body pressures at vulnerable body sites have not succeeded in reducing pressure ulcer prevalence. Here we show that adding a supporting lateral pressure can counter-act the deformations induced by under-body pressure, and that this 'pressure equalisation' approach is a more effective way to reduce ulcer-inducing deformations than current approaches based on redistributing under-body pressure. A finite element model of the seated pelvis predicts that applying a lateral pressure to the soft tissue reduces peak von Mises stress in the deep tissue by a factor of 2.4 relative to a standard cushion (from 113 kPa to 47 kPa)-a greater effect than that achieved by using a more conformable cushion, which reduced von Mises stress to 75 kPa. Combining both a conformable cushion and lateral pressure reduced peak von Mises stresses to 25 kPa. The ratio of peak lateral pressure to peak under-body pressure was shown to regulate deep tissue stress better than under-body pressure alone. By optimising the magnitude and position of lateral pressure, tissue deformations can be reduced to that induced when suspended in a fluid. Our results explain the lack of efficacy in current support surfaces and suggest a new approach to designing and evaluating support surfaces: ensuring sufficient lateral pressure is applied to counter-act under-body pressure.

Entities:  

Mesh:

Year:  2020        PMID: 31899778      PMCID: PMC6941906          DOI: 10.1371/journal.pone.0227064

Source DB:  PubMed          Journal:  PLoS One        ISSN: 1932-6203            Impact factor:   3.240


1. Introduction

Supporting the body weight of critically ill, immobilised or paraplegic people without causing soft-tissue injury is not an easy task. The loading induced while lying or sitting for prolonged periods can cause damage to skin, adipose tissue and muscle; this damage is known as a pressure ulcer. Pressure ulcers are estimated to affect one in five hospitalised patients in Europe [1], while prevalence in some patient groups are much higher. For example, 85% of spinal cord injury patients develop a pressure ulcer over their lifetime [2] with associated care costs of approximately $1.2 billion annually in the US [3]. A severe form of pressure ulcer develops in subdermal tissue close to bony prominences such as the ischial tuberosity and sacrum of the pelvis [4,5], and is known as a deep tissue injury. Because of the severity of deep tissue injury, preventative strategies have been a major focus in the field. One approach to preventing pressure ulcers is to design support surfaces to reduce pathological pressures, and this has been a major area of research for the past forty years [6]—indeed ‘invalid beds’ have been developed since the 19th century [7]. While support surfaces have become increasingly high-tech, they have yet to outperform high-specification foam mattresses, and their adoption in clinics has not led to a significant reduction in pressure ulcer prevalence [8,9]. While this lack of progress may indicate that we have reached the limit of support surface design, in this paper, we argue that current designs have been based upon a suboptimal design principle—that of under-body pressure re-distribution. The presumption that high surface pressure leads to pressure ulcers, and therefore should be reduced, seems obvious. However, experimental, computational and clinical evidence suggests that high surface pressures do not necessarily cause pressure ulcers. Peak surface pressures (as measured by pressure mapping sensor arrays) cannot identify at-risk patients [10,11]. High-tech mattresses that reduce peak surface pressures have increasingly been adopted in clinical settings yet their impact on pressure ulcer prevalence has been disappointing [8,9]. Furthermore, soft tissue can tolerate extremely high surface pressures under certain circumstances. The soft tissues of a deep-sea diver, for example, are exposed to 100 kPa of surface pressure for every 10 m descended, yet pressure-related injuries to soft tissues are not a common issue in diving [12,13]. Computational studies have helped to explain these observations, with Oomens et al. [14] demonstrating that peak surface pressure has very little impact on internal deformations near bony prominences—regions where deep tissue injuries are likely to occur [14,15]. Since reducing peak surface pressure has failed to protect deep tissue, we sought to determine if there is any way to manipulate the external pressure profile that can protect deep tissue. Deep tissue pressure ulcers develop as a result of several overlapping processes: ischaemia [16], ischaemia-reperfusion injury [17] lymphatic network obstruction [18,19] and direct cell deformation [20]. Each of these processes is triggered by excessive deformation (exacerbated by shear stresses, microclimate, and other risk factors) within soft tissue, and so in some regards, a pressure ulcer could be more aptly named a ‘deformation ulcer’. Redistributing surface pressure (as current devices aim to do) does not necessarily reduce deformations (and hence pressure ulcers) because soft tissue has very different tolerances to the two components of stress (Fig 1A): deviatoric stress (which tends to change the shape of an object) and dilatational stress (which tends to change the volume of an object, but not its shape) [21,22]. Human soft tissues are almost incompressible [23], and so can tolerate high dilatational stress with minimal deformation. In contrast, soft tissues deform readily with deviatoric stress, therefore it is this stress that must be minimised to prevent ulceration. While submersed, a diver experiences nearly uniform pressures on all surfaces; this tends to induce dilatational stress [22]. On the other hand, interaction with very localised surface pressure–such as when sitting on a chair or lying on a mattress–induces large deviatoric stress (and therefore deformations) as the soft tissue bulges and is displaced laterally away from the load (Fig 1B).
Fig 1

Deformations beneath a bony prominence.

The stress in soft tissue has two components, dilatational and deviatoric (a). Soft tissue is much more resistant to dilatational stress than deviatoric stress. Under a bony prominence, the soft tissue is distorted due to the concentrated pressures at the bone and the support (b). Redistributing the surface pressure has some effect on the outer (superficial) region, but not on the deep tissue. We hypothesise that by applying pressure laterally (termed pressure equalisation), bulging is reduced, and the tissue can bear the load in a more dilatational mode.

Deformations beneath a bony prominence.

The stress in soft tissue has two components, dilatational and deviatoric (a). Soft tissue is much more resistant to dilatational stress than deviatoric stress. Under a bony prominence, the soft tissue is distorted due to the concentrated pressures at the bone and the support (b). Redistributing the surface pressure has some effect on the outer (superficial) region, but not on the deep tissue. We hypothesise that by applying pressure laterally (termed pressure equalisation), bulging is reduced, and the tissue can bear the load in a more dilatational mode. One way to prevent excessive deformations may be to restrain the soft tissue from deforming by applying a supporting lateral pressure. In this paper, we test the plausibility of this principle using a computational model of the weight-bearing pelvis in a seated individual. We hypothesise that actively applying pressure laterally to the soft tissue of the pelvis will reduce the deformation at the ischial tuberosity to a greater extent than the commonly applied method of redistributing under-body pressure (Fig 1B). Our rationale for this work is that by making subtle changes to the design objectives used for support surfaces, we may be able to substantially reduce the risk of ulceration for high-risk patients.

2. Methods

First, we adapted a previously-developed finite element model of seated buttocks [14] to test the hypothesis that applying lateral pressure will reduce tissue deformations (section 2.1). The seated position was chosen because the ischial tuberosity is a common site for deep tissue injury [1]. Next, we used the model to determine whether applying lateral pressure or changing the stiffness of a standard cushion has the greatest effect on deep tissue deformations (section 2.2). To ensure that these effects translate to a more realistic setting, we developed a 3D model from MRI scans (section 2.3). Finally, we sought to formalise the relationship between surface pressure and internal deformations into a design principle—equalising under-body pressure with lateral pressure. To do this, we described the interaction of soft tissue and a support surface as a surface pressure boundary condition, which could be manipulated and studied independently of particular cushion design (section 2.4). All finite element input files and analysis protocols are available in an open database (10.6084/m9.figshare.10510787).

2.1 Model of the seated pelvis with lateral pressure application

2.1.1 Geometry and material models. An axisymmetric geometry was used to model the soft tissue surrounding a single ischial tuberosity in a seated individual. The geometry was similar to that used by Oomens et al. [14] but included more of the pelvis soft tissue to allow lateral pressure to be modelled (Fig 2A). The soft tissue was partitioned into fat, muscle and skin to produce similar patterns as found from MRI imaging (Fig 2B).
Fig 2

Finite element models.

(a) An axisymmetric model of the soft tissue surrounding the ischial tuberosity. The model incorporates a rigid bony prominence, muscle, fat and skin layers interacting with a cushion and a pressure equalisation device. Axisymmetry was assumed, which allowed a force-controlled simulation of weight-bearing (W is the load borne by the ischial tuberosity). The pressure equalisation device was modelled as an air-filled chamber with a controllable internal pressure, P. (b) The axisymmetric region modelled is shown superimposed on saggital and coronal MR images of a seated male (top). A 3D model was generated from the MR images to assess 3D deformations (bottom).

Finite element models.

(a) An axisymmetric model of the soft tissue surrounding the ischial tuberosity. The model incorporates a rigid bony prominence, muscle, fat and skin layers interacting with a cushion and a pressure equalisation device. Axisymmetry was assumed, which allowed a force-controlled simulation of weight-bearing (W is the load borne by the ischial tuberosity). The pressure equalisation device was modelled as an air-filled chamber with a controllable internal pressure, P. (b) The axisymmetric region modelled is shown superimposed on saggital and coronal MR images of a seated male (top). A 3D model was generated from the MR images to assess 3D deformations (bottom). There have been many material models of skeletal muscle [24,25], skin [26], and to a lesser extent fat [27]. However, experimentally-based models that quantify all three tissues together are rarer, making it difficult to combine tissues defined from different experimental setups. In this study, we used the material models based on Oomens et al [14] because it enabled direct comparison with that study, and because all three soft tissues were characterized. Each region was assigned an Ogden hyperelastic material model, and parameter values are listed in Table 1. A flat, 76 mm-thick, two-layered seat cushion was modelled with hyperelastic material properties representing a soft cushion (Table 1). An air-filled chamber (the pressure-equalisation device) was introduced to apply a lateral pressure to the soft tissues (Fig 2A). This was shaped to conform to the seated pelvis and was modelled using membrane elements that can resist tensile, but not bending, loads. The chamber wall material was modelled as sufficiently stiff (Young’s Modulus, E = 10 MPa) so as not to allow appreciable changes in length.
Table 1

Parameters for the Ogden material model for each of the materials modelled.

Ogden strain energy density function, , where λ1,2,3 are the principal stretches, μ and a are material constants and U is the strain energy density.

Materialμ (MPa)a (-)
Skin0.0430
Fat0.02510
Muscle0.0455
Stiff cushion0.0810
Medium cushion0.00510
Soft cushion top0.00357
Soft cushion bottom0.00510

Parameters for the Ogden material model for each of the materials modelled.

Ogden strain energy density function, , where λ1,2,3 are the principal stretches, μ and a are material constants and U is the strain energy density. 2.1.2 Boundary conditions. The proportion of body weight borne by the ischial tuberosities while seated varies from 18% to 77% [14]. We estimated the amount of load supported by the pelvis at 400N —representing approximately 50% of the body weight of an 80 kg adult (with each tuberosity bearing 200N) because it is within the range of experimental findings and enables direct comparison with Oomens et al [14]. Symmetry boundary conditions were prescribed to all nodes lying along the z-axis (Fig 2A). The cushion base was constrained in all directions. Two nodes of the pressure equalisation device were constrained in all directions, and a uniform pressure was applied to the inner surface of the chamber to a maximum of 80 kPa. Frictionless contact was assumed between the support surfaces and the skin. While this is a simplification of the real-world scenario, a sensitivity study revealed friction to have a negligible effect on model predictions (S2 Appendix). Normal contact behaviour was enforced using the penalty method with finite-sliding [28]. 2.1.3 Solution approach and output. A mesh sensitivity analysis was performed leading to a final mesh of 13056 linear quadrilateral elements representing the soft tissues. All models were solved as quasi-static, non-linear analyses using the ABAQUS finite element software (v2016, Dassault Systems, France). To analyse and compare models, the von Mises stresses and shear strains in the soft tissue regions were calculated. Von Mises stress () is a scalar representing the deviatoric part of the stress tensor () and defined as , where is the stress tensor and is the Kronecker delta function. Shear strain was calculated as −, where , are the maximum and minimum principal strains, respectively. These were chosen to represent the level of deviatoric stress and strain, respectively. To summarise the stresses and strains in the deep tissue, we sampled 1600 elements within a 30mm radius of the ischial tuberosity. Since elements vary considerably in size throughout the region, we weighed our sampling by element volume (IVOL output from ABAQUS). For example, an element with a volume 2V is twice as likely to be sampled as an element with volume V. Weighting the results like this ensures that analyses are independent of mesh density, which varies throughout the model. Peak stress was defined as the 95th percentile of the stress data to avoid extreme outliers that may be sensitive to boundary conditions. Effects sizes (in the mean and peak values) between models were estimated by calculating bootstrapped 95% confidence intervals [29]. We also analysed deep tissue stress along a path through the soft tissue directly beneath the ischial tuberosity, and along the surface of the ischial tuberosity.

2.2 Comparing lateral pressure application to changing cushion stiffness

Cushion stiffness is a design variable commonly used to create a more conformable cushion, and this was used as an intervention to compare to adding lateral pressure. The model described above was adapted to model load-bearing on three different cushion designs–a stiff, medium, and soft variety. The stiff and medium cushions were homogeneous, 38 mm thick cushions, while the most compliant (softest) was produced by defining a 76 mm thick two-layer cushion as in the previous section. The bottom layer of this cushion had the medium-stiffness cushion properties, and a softer material was assigned to the 38 mm top layer (Table 1). These cushions were based on those analysed by Oomens et al. [14], which were in turn calibrated to model materials commonly used in wheelchair cushions. For each of the three cushion simulations, load-bearing to 200N was established as in section 2.1. Pressure in the pressure equalisation device was then increased incrementally to a maximum of 80 kPa. The stresses and strains induced when lateral pressure is applied were computed.

2.3 Assessing deformations in 3D

Modelling the pelvis using a 2D axisymmetric model as above requires simplification of the bone and soft tissue geometries. To ensure that the beneficial effect of adding lateral pressure translates to 3D environments, we developed a 3D model that can more accurately capture the geometry of a seated pelvis. The MRI data of a male subject (age 30) was used to generate the 3D geometry (Fig 2B) including skin, fat, muscle and bone. Data usage was approved by the Imperial College Research Ethics committee under ethical approval number ICHTB HTA licence: 12275 and REC Wales approval: 12/WA/0196. To aid comparison of results between models, material properties for each layer were assigned to be consistent with the 2D model. The nodes representing the outer surface of the pelvic bones were constrained to move in the z-direction. Central symmetry was assumed to reduce model size by half. The soft tissue was meshed using 669,995 linear tetrahedral elements. A body force of 200N was applied to the bone nodes. This represents a full body weight of 80 kg, with the assumption that 50% of this travels through the pelvis of a seated individual (as was assumed with the axisymmetric model and is based on Oomens et al. [14], see section 2.1.2). Maintaining this level of loading in the 3D model aids comparison between that and the axisymmetric model. The lateral supports were modelled as air-filled cavities, with a thin outer membrane of material (as in the 2D model). To model the seated load case, the body force was ramped incrementally over the first analysis step. Next, the lateral supports were displaced towards the pelvis while the internal pressure within the support was fixed at 1 kPa. Finally, the pressure within the lateral support was increased to 10 kPa incrementally, and the von Mises stresses and maximum shear strains were calculated over ten increments. This model was solved using Abaqus/Explicit, with mass scaling applied to ensure kinetic energy was less than 1% of internal strain energy.

2.4 Determining the relationship between surface pressure and deep tissue mechanics

We studied how the shape and magnitude of the surface pressure distribution affected internal tissue stress to understand how these stresses can be minimised. The specific cushions and lateral support used in the previous sections were removed and replaced with a surface pressure boundary condition. For our axisymmetric model, pressure is a function of the angle from vertical, P(θ). The surface pressure was constrained to ensure that the model was in static equilibrium: the sum of the vertical forces due to surface pressure is equal to body weight (W), (∮P(θ)cosθdS = W, where dS represents an infinitesimal surface element), and the sum of the horizontal forces is zero (∮P(θ)sinθdS = 0). The surface pressure on the buttocks when seated on a flat cushion follows a characteristic distribution [30]—there is a pressure peak beneath each ischial tuberosity which gradually reduces to zero towards the periphery of the contact area (S1 Appendix). This contact pressure was modelled as a Gaussian distribution (see S1 Appendix). The spread of the pressure peak (α) was varied between 0.2 and 0.35, which represent cushions with stiffness values beyond the range of those tested in section 2.2. We then modelled an externally applied lateral pressure by defining a second Gaussian term. This was controlled by its spread (β), its magnitude (P), and its location (θ0). β and θ0 were fixed (0.4 and π/4 respectively) and P/P (the ratio of lateral pressure relative to under-body pressure) was varied from 0% to 75%. This led to 16 parameter combinations, described in Table 2. Each pressure field was then applied as a boundary condition to the soft-tissue finite element model, and peak stresses and strains were calculated.
Table 2

Parameters governing the spread of the under-body pressure (α) and the magnitude of lateral pressure.

Lateral pressure was defined relative to the peak under-body pressure (P/P).

Pressure re-distribution (α)Lateral Pressure (PL/PV)
0.20
0.250
0.30
0.350
0.20.25
0.250.25
0.30.25
0.350.25
0.20.50
0.250.50
0.30.50
0.350.50
0.20.75
0.250.75
0.30.75
0.350.75

Parameters governing the spread of the under-body pressure (α) and the magnitude of lateral pressure.

Lateral pressure was defined relative to the peak under-body pressure (P/P). 2.4.1 Optimising the surface pressure distribution. We considered the pressure distribution of a body suspended in a fluid as an ideal support scenario [22], as it results in minimal deviatoric stress relative to dilatational stress (see ). We then optimised the location () and relative magnitude of the lateral pressure (/) to minimise the difference between () and the distribution when suspended in a fluid, while ensuring that the full body weight (200N) was supported (see ).

3. Results

3.1 Applying lateral pressure reduces soft tissue deformations

We first set out to determine the effect of adding lateral pressure to a person seated on a standard support surface. We used a finite element model of the pelvis to simulate weight bearing while sitting on a soft cushion. Firstly, we simulated weight bearing without lateral support. In agreement with other studies [14,22,31], the model predicts significant stress concentrations under the ischial tuberosity (Fig 3A), with peak von Mises stresses of 58 kPa produced in the muscle. Then, when lateral pressure is applied, the peak von Mises stress beneath the bony prominence drops to 18 kPa, in support of our hypothesis. Contour plots show the stress is more evenly distributed in the soft tissues (Fig 3A), suggesting that more of the soft tissue is being recruited in transferring the load.
Fig 3

Analysis of load-bearing when seated on a soft cushion.

In the absence of lateral pressure, the model predicts high von Mises stresses under the ischial tuberosity (a). With the introduction of lateral pressure (44 kPa chamber pressure), the region of high stress shrinks dramatically. Histograms of stresses and strains in the muscle tissue within a radius of 30 mm from the ischial tuberosity (b) indicate that von Mises stresses and shear strains are reduced. Analysis of the stress along path ABC (c) show a drop in von Mises stress and shear strain at the bony prominence, and throughout the muscle tissue. Shear strain and von Mises stress are also reduced in the skin and fat layers.

Analysis of load-bearing when seated on a soft cushion.

In the absence of lateral pressure, the model predicts high von Mises stresses under the ischial tuberosity (a). With the introduction of lateral pressure (44 kPa chamber pressure), the region of high stress shrinks dramatically. Histograms of stresses and strains in the muscle tissue within a radius of 30 mm from the ischial tuberosity (b) indicate that von Mises stresses and shear strains are reduced. Analysis of the stress along path ABC (c) show a drop in von Mises stress and shear strain at the bony prominence, and throughout the muscle tissue. Shear strain and von Mises stress are also reduced in the skin and fat layers. The volume of muscle tissue around the ischial tuberosity exposed to high von Mises stress (> 20 kPa) is reduced from 58% to 4% with lateral pressure application (Fig 3B). The volume of muscle tissue exposed to high shear strains (> 0.2) fell from 20% to 0%. Adding lateral pressure reduces the mean von Mises stress by 64% (95% CI 61% to 68%), while mean shear strains are reduced by 42% (95% CI 41% to 43%) when lateral pressure is applied (Fig 3B). Plots of stress along a path through the soft tissue show that von Mises stress is reduced in all tissues under the ischial tuberosity when lateral pressure is applied (Fig 3C), with stress reduction being most pronounced in the muscle (66% in muscle, 43% in fat and 49% in skin). These results show that adding lateral pressure can reduce deep tissue deviatoric stress and deformation.

3.2 Applying lateral pressure reduces deformations to a greater extent than changing cushion stiffness

Having established that applying lateral pressure reduces deep tissue stress and deformation, we next assessed the effect of this intervention compared to a common device design consideration—changing cushion stiffness (Fig 4). Cushion stiffness is usually manipulated to reduce peak pressures by maximising the contact area with the soft tissue. Indeed, the soft cushion provides approximately 3.5 times more contact area than the stiff cushion (72 cm2 in the stiff cushion, 177 cm2 in the medium cushion and 255 cm2 in the soft cushion), indicating that we are capturing a broad range of support surface stiffnesses. This area increase could be expected to achieve a similar reduction in deep tissue deformation. However, von Mises stresses (Fig 4A) and shear strains (Fig 4B) in the deep tissue are relatively less affected—with the change from a stiff to a soft cushion we see a reduction by a factor of 1.4 in peak von Mises stress (Fig 4C). Without lateral pressure, all cushions induce a peak von Mises stress > 50 kPa. We find that introducing lateral pressure reduces the peak von Mises stresses observed with each cushion by a factor of 2.4 on average (1.9 for the stiff cushion, 2.1 for the medium cushion and 3.2 for the soft cushion; Fig 4C). The lowest peak deep tissue stresses are observed when the soft cushion is combined with lateral pressure, which reduces the peak von Mises stress to 18 kPa, suggesting a synergistic effect of combining lateral pressure with a soft cushion (Fig 4C).
Fig 4

Applying lateral pressure is more effective than changing cushion stiffness.

While the contact area varies substantially with cushion stiffness, the pattern of internal stress remains similar (a)—stress is concentrated at the bony prominence. Shear strains in the fat and skin are lower when a softer cushion is used (b), but strains within the muscle remain high for all cushions. These strains are reduced when lateral pressure is introduced. All three cushions benefit from the introduction of lateral pressure, with a soft cushion and lateral pressure providing the lowest von Mises stresses (c) [Violin plots show mean and 95th percentile values, stress difference plot shows the peak difference relative to a stiff cushion only with 95% confidence intervals]. As lateral pressure is gradually increased, the von Mises stress decreases until an optimum pressure is reached (d); beyond this pressure, von Mises stresses begin to increase again. While the magnitude of the optimum lateral pressure is different for each cushion, the ratio of lateral to vertical pressure is between 0.63 and 0.79 for all cushions tested.

Applying lateral pressure is more effective than changing cushion stiffness.

While the contact area varies substantially with cushion stiffness, the pattern of internal stress remains similar (a)—stress is concentrated at the bony prominence. Shear strains in the fat and skin are lower when a softer cushion is used (b), but strains within the muscle remain high for all cushions. These strains are reduced when lateral pressure is introduced. All three cushions benefit from the introduction of lateral pressure, with a soft cushion and lateral pressure providing the lowest von Mises stresses (c) [Violin plots show mean and 95th percentile values, stress difference plot shows the peak difference relative to a stiff cushion only with 95% confidence intervals]. As lateral pressure is gradually increased, the von Mises stress decreases until an optimum pressure is reached (d); beyond this pressure, von Mises stresses begin to increase again. While the magnitude of the optimum lateral pressure is different for each cushion, the ratio of lateral to vertical pressure is between 0.63 and 0.79 for all cushions tested. We noticed that there is an optimum magnitude of lateral pressure which is different for each cushion (38.5 kPa, 37.9 kPa and 12.2 kPa for stiff, medium and soft cushions, respectively); however, the ratio of lateral to under-body pressure is consistently between 0.6 and 0.8 (Fig 4D). This suggests that balancing under-body and lateral pressures is more important for the reduction of deep tissue deviatoric stress than reducing peak under-body pressures.

3.3 Lateral pressure reduces stresses in a 3D model

Upon addition of lateral pressure, stresses were reduced at the ischial tuberosities in a similar way to the 2D model (Fig 5A). The volume of soft tissue exposed to high von Mises stress greatly reduced when lateral pressure was applied (Fig 5B). Adding lateral pressure reduced peak von Mises stress in the muscle by a factor of 2.5 when a stiff cushion was used, 2.6 with a medium-stiffness cushion and 2.4 with a soft cushion (Fig 5C). With optimal lateral pressure applied, the stresses at the greater trochanter and bones of the hemipelvis reached no more than 22% of the load at the ischial tuberosity. These results demonstrate that the effects found in 2D are representative of the 3D environment. They also show that gentle lateral pressure can be applied without compromising the tissue at the femur or sacrum. It should be noted that the lateral device modelled here was a very simple design, and no optimisation of its shape was performed. By contouring the lateral pressure device or other optimisations, further efficacy may be possible.
Fig 5

A 3D model of the seated pelvis under load.

(a) Results for the stiff cushion shown with and without lateral pressure applied. Coronal and transverse sections are shown to indicate von Mises stresses both at the ischial tuberosities and the greater trochanter. (b) The volume of soft tissue exposed to high stresses (>32kPa) is shown in relation to the whole pelvis. The whole pelvis is made transparent to help visualise the location of high stresses (beneath the ischial tuberosity) (c) Change in peak von Mises stress throughout the soft tissue of the pelvis (surrounding both the ischium and the femur) as lateral pressure is increased.

A 3D model of the seated pelvis under load.

(a) Results for the stiff cushion shown with and without lateral pressure applied. Coronal and transverse sections are shown to indicate von Mises stresses both at the ischial tuberosities and the greater trochanter. (b) The volume of soft tissue exposed to high stresses (>32kPa) is shown in relation to the whole pelvis. The whole pelvis is made transparent to help visualise the location of high stresses (beneath the ischial tuberosity) (c) Change in peak von Mises stress throughout the soft tissue of the pelvis (surrounding both the ischium and the femur) as lateral pressure is increased.

3.4 Surface pressure equalisation is necessary to protect deep tissue from deformation

Having found that deep tissue deformations were minimised for each cushion when under-body and lateral pressure were in a specific ratio (0.6–0.8), we sought to test whether this ratio could form the basis of a design principle. We removed the cushion from the model and replaced it with a surface pressure boundary condition that could be manipulated independently of cushion design. We varied both the ratio of lateral to under-body pressure (P/P), and the spread of underbody pressure (α) and measured peak von Mises stress for all combinations (Table 2). As in the previous analyses, in the absence of lateral pressure, redistributing under-body pressure (increasing the spread, α, of the pressure peak), reduces peak von Mises stresses at the ischial tuberosity, but even substantial re-distribution fails to reduce the stress below 100 kPa (112 kPa is observed when α = 0.35; Fig 6A). In contrast, inducing a lateral to under-body pressure ratio of 0.25 reduces peak von Mises stress from 180 kPa to 67 kPa. The presence of lateral pressure appears to reduce the effect of redistributing under-body pressure (Fig 6A), suggesting that when lateral pressure is employed, it becomes the most important factor in reducing deformations. These results indicate that controlling the ratio of lateral to under-body pressure (P/P) is necessary to achieve low deep-tissue stress.
Fig 6

Surface pressure analysis.

Redistributing under-body pressure (P) reduces peak von Mises stresses when no lateral pressure is applied (a), but peak stresses remain above 100 kPa. Counter-acting that pressure with a lateral pressure (P) reduces peak stresses to a greater extent. When the magnitude and angle of lateral pressure is optimised, the deep tissue von Mises stresses approach that of suspension in a fluid (b; arrows illustrate pressure intensity). Path plots of von Mises stress show that lateral pressure can induce a similar stress profile at the bony prominence to that when suspended in a fluid (c).

Surface pressure analysis.

Redistributing under-body pressure (P) reduces peak von Mises stresses when no lateral pressure is applied (a), but peak stresses remain above 100 kPa. Counter-acting that pressure with a lateral pressure (P) reduces peak stresses to a greater extent. When the magnitude and angle of lateral pressure is optimised, the deep tissue von Mises stresses approach that of suspension in a fluid (b; arrows illustrate pressure intensity). Path plots of von Mises stress show that lateral pressure can induce a similar stress profile at the bony prominence to that when suspended in a fluid (c). To understand why lateral pressure may be critical, we studied how this ratio affects the shape of the pressure distribution when compared to two extreme scenarios: the pressure distribution while sitting on a stiff cushion (a high-deformation scenario), and that when suspended in a fluid (a low-deformation scenario). The shape of these distributions is markedly different (Fig 1A in S1 Appendix), with a sharp peak of pressure beneath the ischial tuberosity when seated on a cushion, versus a smooth, even pressure distribution when submersed. A parametric study showed that adding lateral pressure best mimicked the pressure profile of suspension in a fluid (S1 Appendix). We then optimised the magnitude and angle of the lateral pressure to best mimic suspension in a fluid. Contour plots show that optimising this lateral pressure (to P/P = 0.71 and θ0 =61.5°) can mimic the internal stresses experienced while suspended in a fluid (Fig 6B). In addition, with these parameters, von Mises stresses at the ischial tuberosity were either equal to or less than those induced when suspended in a fluid (Fig 6C). In summary, not only can applying lateral pressure reduce deep tissue von Mises stress and deformation, we have found that an optimal magnitude and location of lateral pressure can mimic the environment induced when suspended in a fluid.

4. Discussion

The goal of reducing peak surface pressures at vulnerable body sites has underpinned the design of almost all medical support surfaces to date. Meanwhile, studies have consistently concluded that peak surface pressures do not accurately predict internal tissue mechanics [11,14], nor are they effective in predicting patients at risk of pressure ulcers [10]. In this study, we have shown that ensuring under-body and lateral pressures are balanced—a principle we call pressure equalisation—is more effective at reducing deep tissue deformations than reducing peak under-body pressure. We postulate that devices designed to maintain a prescribed ratio of lateral pressure to under-body pressure will reduce the risk of pressure ulcer formation in the soft tissue of immobile patients. The shift in emphasis from pressure re-distribution to pressure equalisation has implications for support surface design (Fig 7). The synthesised results of multiple clinical trials [8,9] suggest that any well-designed mattress is better than a standard hospital bed, but none are particularly successful at reducing pressure ulcer risk. Pressure redistributing devices (either passive or active, Fig 7A and 7B respectively) may protect against superficial ulcers, while having little effect on deep tissue injuries [14]. Our results show that devices must be capable of providing sufficient lateral support to counter-act the deformations induced by under-body pressure. Immersion/encapsulation-based devices such as water beds [32] aim to increase the contact area between the soft tissue and the support surface; however, while the contact area may increase, the horizontal pressures at the periphery of the contact area are usually minimal (Fig 7C), as pressure is primarily in reaction to gravitational body force. As this force acts in the vertical direction, it would be insufficient to equalise the under-body pressures and prevent bulging. Devices that aim to directly minimise shape change have also been developed [33], but their dependence on vertical translations of pistons means that they are not suited to regulating lateral pressure. We believe that support surface technology may yet reduce pressure ulcer prevalence if they are redirected to achieving pressure equalisation.
Fig 7

Pressure equalisation and its effects on device design.

For surfaces designed to reduce peak pressure passively (a), applying a lateral pressure device helps to avoid lateral bulging (top, showing current devices, bottom showing improved design). Active devices based on individually controlled air cells (b) could be improved by surrounding the soft tissue and changing the control software to aim for equalised pressure, rather than reduce peak pressure. Encapsulation devices achieve large contact areas, but the lateral pressures exerted may be limited (c). These could be improved by active compression or smart materials. Pressure mapping systems (d) currently identify pressure peaks as undesirable. If they could measure pressure around the surface, then they could be re-purposed to measure the level of pressure equalisation.

Pressure equalisation and its effects on device design.

For surfaces designed to reduce peak pressure passively (a), applying a lateral pressure device helps to avoid lateral bulging (top, showing current devices, bottom showing improved design). Active devices based on individually controlled air cells (b) could be improved by surrounding the soft tissue and changing the control software to aim for equalised pressure, rather than reduce peak pressure. Encapsulation devices achieve large contact areas, but the lateral pressures exerted may be limited (c). These could be improved by active compression or smart materials. Pressure mapping systems (d) currently identify pressure peaks as undesirable. If they could measure pressure around the surface, then they could be re-purposed to measure the level of pressure equalisation. The pressure equalisation principle has implications for pressure measurement as a diagnostic tool and as a method of evaluating support surfaces. Surfaces incorporating arrays of pressure sensors have been suggested as early-warning systems for ulceration [34,35], and are frequently used to evaluate new support surfaces [36-38]. Using this technology, devices can be readily differentiated based on the peak pressures they produce. However, when these devices are then compared through clinical outcomes, the differences between them vanish [8,9], and so the current predictive power of pressure measurement is limited. If surface pressure could be measured all around the soft tissues (Fig 7D), then the level of pressure equalisation may be a more predictive tool. Then, a measure of the ratio of lateral to under-body pressure (P/P) could be used to determine ulcer risk, and as a control signal for active devices. The pressure gradient, defined as the spatial change in pressure from the point of peak pressure, has been proposed as an alternative to peak pressure for predicting soft tissue damage [39]. At first glance, pressure equalisation may seem to be equivalent to using pressure gradient as an ulceration indicator. However, the pressure gradient does not account for the direction of pressure, and so a body could be loaded with a low pressure gradient yet have little or no pressure equalisation, because lateral pressures are not considered. In other words, pressure gradient is a local variable, as is peak pressure, whereas pressure equalisation is a measure of the quality of the body support as a whole. From a clinical perspective, devices that can equalise under-body pressure with lateral pressure may be a vital tool in reducing the incidences of pressure ulcers—both hospital-acquired ulcers and those that are acquired in community care settings. The ability to apply sufficient lateral pressure will, however, need to be balanced with other equally important design considerations. For example, wheelchair users must not be exposed to intermittent high lateral pressures as they move in the chair. This inflexibility is one reason why form-fitted cushions are not a solution to preventing tissue distortion. In contrast, a successful device must be flexible enough to provide a well-distributed and well-controlled lateral pressure regardless of patient movement. Other design considerations include regulating the temperature and humidity at the skin surface, as well as ease of installation, use and cleaning. While this work has focused on the biomechanics of lateral pressure in general, the practical application of this principle will be more complex and require significant innovation in device design. The reductions in deep tissue stress and strain possible through surface pressure equalisation could be sufficient to reduce pressure ulcer risk. The safe magnitude of deformation (and even the most appropriate measure) is not yet fully accounted for [40] and it is likely to be patient, environment, and tissue-specific. In this study, we have used two measures of deep tissue mechanics—von Mises stress and shear strain. These measures aim to capture the deformations likely to lead to capillary and lymphatic vessel restriction, and cell deformation, which contribute to pressure ulcer onset. Experiments using rat muscle under compression [41] indicated that stresses greater than 32 kPa induced damage, with this threshold dropping to 9 kPa over prolonged loading, while work quantifying deformations in a similar model [42] indicated that damage occurred above a shear strain of 0.3. Our results indicate that redistributing under-body pressure would not protect soft tissue from these levels of deformation, but that applying lateral pressure could. The 2D finite element model used here simplified the anatomical structure of the pelvis in a similar way to previous studies [14,22,31]. These idealisations allowed us to focus on the general case of a bony prominence transferring load through soft tissue to a support surface, and enabled the comparisons and analyses described here. The 3D model used here were important to support the conclusions drawn from 2D analyses, but there are also limitations to this model. While we have included more anatomical complexity including thighs, femurs and pelvis, more biofidelic models have been proposed [43-45]. For example, we chose tissue mechanical properties in line with Oomens et al. [14], but there are several published models of soft tissue mechanical properties that vary in complexity [24-27]. This makes conclusions based on absolute stress values difficult. A further complexity not accounted for in our model is the posture and secondary supports (arm rests, for example) of the patient, which may affect the loading boundary conditions. For these reasons, we have focused on the relative effects of interventions on stresses and strains, thus making the conclusions robust against the chosen material models and boundary conditions. Using more biofidelic approaches will be a key step in applying the current results in the clinic. In particular, models generated from high-risk patients as opposed to healthy volunteers will be crucial. Physical validation will need to come from measurements of internal tissue deformations, for example through load-bearing MRI [46,47]. Our results suggest a novel method for creating a safer mechanical environment. We hypothesize that the reduced deformations created by lateral pressure equalization will translate into better deep tissue blood perfusion and lower risk of deformation-induced cell damage. A key next step will be to test this hypothesis by measuring the physiological response of the soft tissues of seated patients, for example through measuring transcutaneous gas tension [34]. In conclusion, a change in focus from redistributing under-body pressure to equalising it with lateral pressure will lead to new innovations and improvements to patient care, resulting in a reduction of pressure ulcer prevalence in immobile patients.

Analysis of pressure distributions.

(PDF) Click here for additional data file.

Testing model assumptions.

(PDF) Click here for additional data file. 19 Nov 2019 PONE-D-19-27579 Lateral pressure equalisation as a principle for designing support surfaces to prevent deep tissue pressure ulcers PLOS ONE Dear Dr Boyle, Thank you for submitting your manuscript to PLOS ONE. After careful consideration, we feel that it has merit but does not fully meet PLOS ONE’s publication criteria as it currently stands. Therefore, we invite you to submit a revised version of the manuscript that addresses the points raised during the review process. We would appreciate receiving your revised manuscript by Jan 03 2020 11:59PM. When you are ready to submit your revision, log on to https://www.editorialmanager.com/pone/ and select the 'Submissions Needing Revision' folder to locate your manuscript file. If you would like to make changes to your financial disclosure, please include your updated statement in your cover letter. To enhance the reproducibility of your results, we recommend that if applicable you deposit your laboratory protocols in protocols.io, where a protocol can be assigned its own identifier (DOI) such that it can be cited independently in the future. For instructions see: http://journals.plos.org/plosone/s/submission-guidelines#loc-laboratory-protocols Please include the following items when submitting your revised manuscript: A rebuttal letter that responds to each point raised by the academic editor and reviewer(s). This letter should be uploaded as separate file and labeled 'Response to Reviewers'. A marked-up copy of your manuscript that highlights changes made to the original version. This file should be uploaded as separate file and labeled 'Revised Manuscript with Track Changes'. An unmarked version of your revised paper without tracked changes. This file should be uploaded as separate file and labeled 'Manuscript'. Please note while forming your response, if your article is accepted, you may have the opportunity to make the peer review history publicly available. The record will include editor decision letters (with reviews) and your responses to reviewer comments. If eligible, we will contact you to opt in or out. We look forward to receiving your revised manuscript. Kind regards, Pedro H. Oliveira, Ph.D. Academic Editor PLOS ONE Journal requirements: 1. When submitting your revision, we need you to address these additional requirements. Please ensure that your manuscript meets PLOS ONE's style requirements, including those for file naming. The PLOS ONE style templates can be found at http://www.journals.plos.org/plosone/s/file?id=wjVg/PLOSOne_formatting_sample_main_body.pdf and http://www.journals.plos.org/plosone/s/file?id=ba62/PLOSOne_formatting_sample_title_authors_affiliations.pdf 2. Thank you for providing the following Funding Statement: "This project was partly funded by an Imperial College London and Imperial Innovations Ltd. Proof of Concept funding to CB and SM and EPSRC funding to CH, MM and SM (EP/N026845/1). The funders had no role in study design, data collection and analysis, decision to publish, or preparation of the manuscript.". We note that one or more of the authors is affiliated with the funding organization, indicating the funder may have had some role in the design, data collection, analysis or preparation of your manuscript for publication; in other words, the funder played an indirect role through the participation of the co-authors. If the funding organization did not play a role in the study design, data collection and analysis, decision to publish, or preparation of the manuscript and only provided financial support in the form of authors' salaries and/or research materials, please review your statements relating to the author contributions, and ensure you have specifically and accurately indicated the role(s) that these authors had in your study in the Author Contributions section of the online submission form. Please make any necessary amendments directly within this section of the online submission form.  Please also update your Funding Statement to include the following statement: “The funder provided support in the form of salaries for authors [insert relevant initials], but did not have any additional role in the study design, data collection and analysis, decision to publish, or preparation of the manuscript. The specific roles of these authors are articulated in the ‘author contributions’ section.” If the funding organization did have an additional role, please state and explain that role within your Funding Statement. Please also provide an updated Competing Interests Statement declaring this commercial affiliation along with any other relevant declarations relating to employment, consultancy, patents, products in development, or marketed products, etc. Within your Competing Interests Statement, please confirm that this commercial affiliation does not alter your adherence to all PLOS ONE policies on sharing data and materials by including the following statement: "This does not alter our adherence to  PLOS ONE policies on sharing data and materials.” (as detailed online in our guide for authors http://journals.plos.org/plosone/s/competing-interests). If this adherence statement is not accurate and  there are restrictions on sharing of data and/or materials, please state these. Please note that we cannot proceed with consideration of your article until this information has been declared. 3. We note that you have a patent relating to material pertinent to this article. Please provide an amended statement of Competing Interests to declare this patent (with details including name and number), along with any other relevant declarations relating to employment, consultancy, patents, products in development or modified products etc. Please confirm that this does not alter your adherence to all PLOS ONE policies on sharing data and materials, as detailed online in our guide for authors http://journals.plos.org/plosone/s/competing-interests by including the following statement: "This does not alter our adherence to  PLOS ONE policies on sharing data and materials.” If there are restrictions on sharing of data and/or materials, please state these. Please note that we cannot proceed with consideration of your article until this information has been declared. * This information should be included in your cover letter; we will change the online submission form on your behalf. Please know it is PLOS ONE policy for corresponding authors to declare, on behalf of all authors, all potential competing interests for the purposes of transparency. PLOS defines a competing interest as anything that interferes with, or could reasonably be perceived as interfering with, the full and objective presentation, peer review, editorial decision-making, or publication of research or non-research articles submitted to one of the journals. Competing interests can be financial or non-financial, professional, or personal. Competing interests can arise in relationship to an organization or another person. Please follow this link to our website for more details on competing interests: http://journals.plos.org/plosone/s/competing-interests [Note: HTML markup is below. Please do not edit.] Reviewers' comments: Reviewer's Responses to Questions Comments to the Author 1. Is the manuscript technically sound, and do the data support the conclusions? The manuscript must describe a technically sound piece of scientific research with data that supports the conclusions. Experiments must have been conducted rigorously, with appropriate controls, replication, and sample sizes. The conclusions must be drawn appropriately based on the data presented. Reviewer #1: Yes Reviewer #2: Yes ********** 2. Has the statistical analysis been performed appropriately and rigorously? Reviewer #1: N/A Reviewer #2: Yes ********** 3. Have the authors made all data underlying the findings in their manuscript fully available? The PLOS Data policy requires authors to make all data underlying the findings described in their manuscript fully available without restriction, with rare exception (please refer to the Data Availability Statement in the manuscript PDF file). The data should be provided as part of the manuscript or its supporting information, or deposited to a public repository. For example, in addition to summary statistics, the data points behind means, medians and variance measures should be available. If there are restrictions on publicly sharing data—e.g. participant privacy or use of data from a third party—those must be specified. Reviewer #1: Yes Reviewer #2: Yes ********** 4. Is the manuscript presented in an intelligible fashion and written in standard English? PLOS ONE does not copyedit accepted manuscripts, so the language in submitted articles must be clear, correct, and unambiguous. Any typographical or grammatical errors should be corrected at revision, so please note any specific errors here. Reviewer #1: Yes Reviewer #2: Yes ********** 5. Review Comments to the Author Please use the space provided to explain your answers to the questions above. You may also include additional comments for the author, including concerns about dual publication, research ethics, or publication ethics. (Please upload your review as an attachment if it exceeds 20,000 characters) Reviewer #1: The paper deals with a basic research regarding the opportunity or not to use lateral pressure to be applied to the patient forced to maintain a sitting position for long periods in order to reduce the pressure at the interface with the pillow and consequently the risk of ulcerations. The idea is good and its applicability must be verified. The preliminary study is well written and the methods used are appropriate. Small improvements are required: - always define all the parameters introduced (for example in table 1 and in the equations) - in paragraph 2.1.2 it is well known that the body mass of the torso + head + upper limbs is equal to 2/3 of the whole body mass not 1/2 - in paragraph 2.1.2 next to the figure 80 the unit of measurement is missing - in brackets, after the softwares used, it is necessary to mention the company that distributes them and their location - in paragraph 2.3 justify the choices made on the imposed loads. Reviewer #2: This study sought to present a novel approach to prevent ulcers due to deep tissue pressure. In particular the authors stated that adding a supporting lateral pressure (they called “pressure equalization”) to the classical under-body pressure redistribution approach could provide a more effective solution to reduce deep tissue deformation that can lead to ulcers development. The work is based on Finite Element (FE) Model, which is able to predict the stresses induced on the deep tissue; the authors used this model to assess the differences introduced by the lateral pressure equalization. In particular, they found that managing the ratio of peak lateral pressure to peak under-body pressure and the position of the lateral supports could be used to regulate deep tissue stresses. The authors stated that the overall deformation can be decreased at the induced when the body is suspended in a fluid. The approach can be used to design and assess novel solutions related to supporting devices. General Comments The main working hypothesis at the basis of this paper is clearly reported as far as the main objective. The introduction to the approach is clearly reported and the modelling phase well documented. Although the methodology is not that innovative, the application to is indeed original. I guess the audience of PLOS ONE will appreciate this work. The structure of the article seems to be precise (Abstract, Introduction, Methods [with subheadings], Results, Discussion). Modelling and data analysis seem to be clearly reported and coherent with the work objectives. Several minor concerns have been already reported to the authors. The use of the English language seems to be correct. The references to previous works seem to be precise and up-to-date. Specific Comments Abstract In general this section is ok. Please, could you report also the magnitude of the deformation related to deep tissue and the differences between under-body pressure redistribution and lateral pressure equalization? Introduction Really well written. You well reported the main clinical problem, the state-of-the-art related to the main provided solutions, your hypothesis and the rationale behind your work. Really appreciated Figure 1. Methods • Page 8. Model properties. Oomens’ work is a modelling approach dating back to 2003. Is there any more recent work identifying soft tissues properties and modelling? • Page 11 – 2.1.2 Line 6. I guess that the unit of measurements is missing. • Please provide more hypothesis on the choice of not introducing any friction between the support surface and the skin. Is this condition the real one? • Page 12. Please provide more information about the “adjusting” of data accounting for volume variation. It is not that clear. • Page 16. Table 2. Please provide information about “[0.5ex]”, present in the first line. Results Very well reported both graphically and in the main text. Discussion In general, this section is ok. Since you reported also the most clinical perspective, what about pressure equalization on vessels? Could you speculate on any main issue? ********** 6. PLOS authors have the option to publish the peer review history of their article (what does this mean?). If published, this will include your full peer review and any attached files. If you choose “no”, your identity will remain anonymous but your review may still be made public. Do you want your identity to be public for this peer review? For information about this choice, including consent withdrawal, please see our Privacy Policy. Reviewer #1: No Reviewer #2: Yes: Nicola Francesco Lopomo [NOTE: If reviewer comments were submitted as an attachment file, they will be attached to this email and accessible via the submission site. Please log into your account, locate the manuscript record, and check for the action link "View Attachments". If this link does not appear, there are no attachment files to be viewed.] While revising your submission, please upload your figure files to the Preflight Analysis and Conversion Engine (PACE) digital diagnostic tool, https://pacev2.apexcovantage.com/. PACE helps ensure that figures meet PLOS requirements. To use PACE, you must first register as a user. Registration is free. Then, login and navigate to the UPLOAD tab, where you will find detailed instructions on how to use the tool. If you encounter any issues or have any questions when using PACE, please email us at figures@plos.org. Please note that Supporting Information files do not need this step. 21 Nov 2019 We wish to thank the reviewers for their valuable critique of our manuscript. Pease see below for our responses to each comment. Reviewer #1: The paper deals with a basic research regarding the opportunity or not to use lateral pressure to be applied to the patient forced to maintain a sitting position for long periods in order to reduce the pressure at the interface with the pillow and consequently the risk of ulcerations. The idea is good and its applicability must be verified. The preliminary study is well written and the methods used are appropriate. Small improvements are required: - always define all the parameters introduced (for example in table 1 and in the equations) We have reviewed the manuscript to ensure all parameters are now defined at the point of first instance. - in paragraph 2.1.2 it is well known that the body mass of the torso + head + upper limbs is equal to 2/3 of the whole body mass not ½ We agree that this 50% value does not represent all upper-body weight. Oomens et al [14] discuss loading boundary conditions in pelvis models. Using literature on experimental pressure-reading, they describe that weight borne by the IT during sitting varies from 18% to 77% of body weight. This variation is due to patient posture, including use of arm and thigh support, and inclination of back support. We chose 50% as a value within the experimental range, while also enabling direct comparison with that previous modelling work [14]. We also argue that because we look at relative differences (between support surfaces), rather than draw conclusions based on absolute values, the choice of load magnitude does not affect our conclusions. Changes to text: Section 2.1.2 pg 8: The proportion of body weight borne by the ischial tuberosities while seated varies from 18% to 77% [14]. We estimated the amount of load supported by the pelvis at 400N — representing approximately 50% of the body weight of an 80 kg adult (with each tuberosity bearing 200N) because it is within the range of experimental findings and enables direct comparison with Oomens et al [14]. Discussion pg 25: A further complexity not accounted for in our model is the posture and secondary supports (arm rests, for example) of the patient, which may affect the loading boundary conditions. For these reasons, we have focused on the relative effects of interventions on stresses and strains, thus making the conclusions robust against the chosen material models and boundary conditions. Using more biofidelic approaches will be a key step in applying the current results in the clinic. - in paragraph 2.1.2 next to the figure 80 the unit of measurement is missing Thank you, amended with kPa as units - in brackets, after the softwares used, it is necessary to mention the company that distributes them and their location Amended as suggested - in paragraph 2.3 justify the choices made on the imposed loads. Our objective for developing this 3D model was to ensure that the conclusions drawn from the axisymmetric model translated to a 3D environment. To achieve this objective, we modelled the same loading regime as the axisymmetric model. We accept, however, that a more biofidelic model – based on realistic load cases including body freedom-of-movement and whole-body simulation – will be required as a next step before translation to the clinic. Changes to text (the alterations to section 2.1.2 are also relevant here): Section 2.3 pg 11: A body force of 200N was applied to the bone nodes. This represents a full body weight of 80 kg, with the assumption that 50% of this travels through the pelvis of a seated individual (as was assumed with the axisymmetric model and is based on Oomens et al. [14], see section 2.1.2). Maintaining this level of loading in the 3D model aids comparison between that and the axisymmetric model. Reviewer #2: This study sought to present a novel approach to prevent ulcers due to deep tissue pressure. In particular the authors stated that adding a supporting lateral pressure (they called “pressure equalization”) to the classical under-body pressure redistribution approach could provide a more effective solution to reduce deep tissue deformation that can lead to ulcers development. The work is based on Finite Element (FE) Model, which is able to predict the stresses induced on the deep tissue; the authors used this model to assess the differences introduced by the lateral pressure equalization. In particular, they found that managing the ratio of peak lateral pressure to peak under-body pressure and the position of the lateral supports could be used to regulate deep tissue stresses. The authors stated that the overall deformation can be decreased at the induced when the body is suspended in a fluid. The approach can be used to design and assess novel solutions related to supporting devices. General Comments The main working hypothesis at the basis of this paper is clearly reported as far as the main objective. The introduction to the approach is clearly reported and the modelling phase well documented. Although the methodology is not that innovative, the application to is indeed original. I guess the audience of PLOS ONE will appreciate this work. The structure of the article seems to be precise (Abstract, Introduction, Methods [with subheadings], Results, Discussion). Modelling and data analysis seem to be clearly reported and coherent with the work objectives. Several minor concerns have been already reported to the authors. The use of the English language seems to be correct. The references to previous works seem to be precise and up-to-date. Specific Comments Abstract In general this section is ok. Please, could you report also the magnitude of the deformation related to deep tissue and the differences between under-body pressure redistribution and lateral pressure equalization? Thank you for this suggestion, we have updated the abstract to include magnitudes as well as fold-change. Changes to text: Abstract: A finite element model of the seated pelvis predicts that applying a lateral pressure to the soft tissue reduces peak von Mises stress in the deep tissue by a factor of 2.4relative to a standard cushion (from 113 kPa to 47 kPA) — a greater effect than that achieved by using a more conformable cushion, which reduced von Mises stress to 75 kPa. Combining both a conformable cushion and lateral pressure reduced peak von Mises stresses to 25 kPa. Introduction Really well written. You well reported the main clinical problem, the state-of-the-art related to the main provided solutions, your hypothesis and the rationale behind your work. Really appreciated Figure 1. Methods • Page 8. Model properties. Oomens’ work is a modelling approach dating back to 2003. Is there any more recent work identifying soft tissues properties and modelling? Thank you for this comment. We accept that the material properties chose may not be the definitive material models for each of the soft tissues. For example, more complex models that incorporate transverse isotropy of muscle/skin tissue or viscoelasticity may yield more accurate predictions of absolute stress/strain magnitudes. However, our study focuses on relative differences between groups (on different cushions, and with/without lateral pressure). We argue that this focus on relative differences reduces the sensitivity of our conclusions to specific material models, particularly because our chosen models incorporate the two most important aspects of soft tissue mechanics (large-deformations and material non-linearity). We also note that using these models enabled direct comparison to the study most closely related to our work (Oomens et al [14]). We also understand that incorporating more accurate material models of soft tissues will be a crucial part of creating more biofidelic models, and that this is a next step on the way to validating our pressure-equalisation approach for the clinic. Changes to text: Methods, page 7: There have been many material models of skeletal muscle [24,25], skin [26], and to a lesser extent fat [27]. However, experimentally-based models that quantify all three tissues together are rarer, making it difficult to combine tissues defined from different experimental setups. In this study, we used the material models based on Oomens et al [14] because it enabled direct comparison with that study, and because all three soft tissues were characterized. Each region was assigned an Ogden hyperelastic material model, and parameter values are listed in Table 1. Discussion, page 25: For example, we chose tissue mechanical properties in line with Oomens et al. [14], but there are several published models of soft tissue mechanical properties that vary in complexity [24-27]. This makes conclusions based on absolute stress values difficult. A further complexity not accounted for in our model is the posture and secondary supports (arm rests, for example) of the patient, which may affect the loading boundary conditions. For these reasons, we have focused on the relative effects of interventions on stresses and strains, thus making the conclusions robust against the chosen material models and boundary conditions. Using more biofidelic approaches will be a key step in applying the current results in the clinic. • Page 11 – 2.1.2 Line 6. I guess that the unit of measurements is missing. Amended to include kPa as unit. • Please provide more hypothesis on the choice of not introducing any friction between the support surface and the skin. Is this condition the real one? We used frictionless contact as a simplification of the real-world situation. To assess the impact of this simplification, we ran a sensitivity study where the effect of including friction on peak stresses and strains was quantified (see S2 Appendix). We found that in our load case, friction had negligible effect on peak stresses. However, we accept that in future studies that aim to simulate more biofidelic load cases, modelling friction effects will become increasingly important. In particular, a model that captured patient posture dynamically would require more accurate contact characterisation. We also note that our primary focus in this study was the deep tissue (the source of the most dangerous pressure ulcers), while friction tends to induce superficial pressure ulcers. Changes to text: Methods page 9: While this is a simplification of the real-world scenario, a sensitivity study revealed friction to have a negligible effect on model predicitons (S2 Appendix). • Page 12. Please provide more information about the “adjusting” of data accounting for volume variation. It is not that clear. Thank you for this comment. We developed an analysis approach to ensure that our conclusions were insensitive to the mesh density. Our aim was to summarise the stress/strain in a tissue. One approach is to simply probe one point near the region of interest. However, this is sensitive to the choice of point. Another approach is to summarise all the points within a region of interest (mean, median, 95 percentile, etc). However, this approach is affected by mesh density, so locations with many small elements contribute more to the summary statistic than locations with fewer large elements. To overcome this limitation, we took a weighted sample of all elements in the region of interest. The sample was weighted by the volume of the element, so an element with double the volume is twice as likely to be sampled. Changes to text: Page 10: To summarise the stresses and strains in the deep tissue, we sampled 1600 elements within a 30mm radius of the ischial tuberosity. Since elements vary considerably in size throughout the region, we weighed our sampling by element volume (IVOL output from ABAQUS). For example, an element with a volume 2V is twice as likely to be sampled as an element with volume V. Weighting the results like this ensures that analyses are independent of mesh density, which varies throughout the model. • Page 16. Table 2. Please provide information about “[0.5ex]”, present in the first line. Thank you for spotting this, it is an artefact of a conversion from Latex to MS Word and has been deleted. Results Very well reported both graphically and in the main text. Discussion In general, this section is ok. Since you reported also the most clinical perspective, what about pressure equalization on vessels? Could you speculate on any main issue? If we understand your question correctly, you would like us to discuss how our mechanical results translate into improvements to blood perfusion, thereby keeping tissue oxygenated and uninjured. This is a great point, and we have added a paragraph to the discussion. Changes to text: Discussion, Page 27: Our results suggest a novel method for creating a safer mechanical environment. We hypothesize that the reduced deformations created by lateral pressure equalization will translate into better deep tissue blood perfusion and lower risk of deformation-induced cell damage. A key next step will be to test this hypothesis by measuring the physiological response of the soft tissues of seated patients, for example through measuring transcutaneous gas tension [34]. Submitted filename: responsetoReviewsR1.docx Click here for additional data file. 12 Dec 2019 Lateral pressure equalisation as a principle for designing support surfaces to prevent deep tissue pressure ulcers PONE-D-19-27579R1 Dear Dr. Boyle, We are pleased to inform you that your manuscript has been judged scientifically suitable for publication and will be formally accepted for publication once it complies with all outstanding technical requirements. Within one week, you will receive an e-mail containing information on the amendments required prior to publication. When all required modifications have been addressed, you will receive a formal acceptance letter and your manuscript will proceed to our production department and be scheduled for publication. Shortly after the formal acceptance letter is sent, an invoice for payment will follow. To ensure an efficient production and billing process, please log into Editorial Manager at https://www.editorialmanager.com/pone/, click the "Update My Information" link at the top of the page, and update your user information. If you have any billing related questions, please contact our Author Billing department directly at authorbilling@plos.org. If your institution or institutions have a press office, please notify them about your upcoming paper to enable them to help maximize its impact. If they will be preparing press materials for this manuscript, you must inform our press team as soon as possible and no later than 48 hours after receiving the formal acceptance. Your manuscript will remain under strict press embargo until 2 pm Eastern Time on the date of publication. For more information, please contact onepress@plos.org. With kind regards, Pedro H. Oliveira, Ph.D. Academic Editor PLOS ONE Comments to the Author 1. If the authors have adequately addressed your comments raised in a previous round of review and you feel that this manuscript is now acceptable for publication, you may indicate that here to bypass the “Comments to the Author” section, enter your conflict of interest statement in the “Confidential to Editor” section, and submit your "Accept" recommendation. Reviewer #1: (No Response) Reviewer #2: All comments have been addressed ********** 2. Is the manuscript technically sound, and do the data support the conclusions? The manuscript must describe a technically sound piece of scientific research with data that supports the conclusions. Experiments must have been conducted rigorously, with appropriate controls, replication, and sample sizes. The conclusions must be drawn appropriately based on the data presented. Reviewer #1: (No Response) Reviewer #2: Yes ********** 3. Has the statistical analysis been performed appropriately and rigorously? Reviewer #1: (No Response) Reviewer #2: Yes ********** 4. Have the authors made all data underlying the findings in their manuscript fully available? The PLOS Data policy requires authors to make all data underlying the findings described in their manuscript fully available without restriction, with rare exception (please refer to the Data Availability Statement in the manuscript PDF file). The data should be provided as part of the manuscript or its supporting information, or deposited to a public repository. For example, in addition to summary statistics, the data points behind means, medians and variance measures should be available. If there are restrictions on publicly sharing data—e.g. participant privacy or use of data from a third party—those must be specified. Reviewer #1: (No Response) Reviewer #2: Yes ********** 5. Is the manuscript presented in an intelligible fashion and written in standard English? PLOS ONE does not copyedit accepted manuscripts, so the language in submitted articles must be clear, correct, and unambiguous. Any typographical or grammatical errors should be corrected at revision, so please note any specific errors here. Reviewer #1: (No Response) Reviewer #2: Yes ********** 6. Review Comments to the Author Please use the space provided to explain your answers to the questions above. You may also include additional comments for the author, including concerns about dual publication, research ethics, or publication ethics. (Please upload your review as an attachment if it exceeds 20,000 characters) Reviewer #1: (No Response) Reviewer #2: The authors answered all the issues arisen during the first review. All the comments have been fully addressed. ********** 7. PLOS authors have the option to publish the peer review history of their article (what does this mean?). If published, this will include your full peer review and any attached files. If you choose “no”, your identity will remain anonymous but your review may still be made public. Do you want your identity to be public for this peer review? For information about this choice, including consent withdrawal, please see our Privacy Policy. Reviewer #1: No Reviewer #2: Yes: Nicola Francesco Lopomo 13 Dec 2019 PONE-D-19-27579R1 Lateral pressure equalisation as a principle for designing support surfaces to prevent deep tissue pressure ulcers Dear Dr. Boyle: I am pleased to inform you that your manuscript has been deemed suitable for publication in PLOS ONE. Congratulations! Your manuscript is now with our production department. If your institution or institutions have a press office, please notify them about your upcoming paper at this point, to enable them to help maximize its impact. If they will be preparing press materials for this manuscript, please inform our press team within the next 48 hours. Your manuscript will remain under strict press embargo until 2 pm Eastern Time on the date of publication. For more information please contact onepress@plos.org. For any other questions or concerns, please email plosone@plos.org. Thank you for submitting your work to PLOS ONE. With kind regards, PLOS ONE Editorial Office Staff on behalf of Dr. Pedro H. Oliveira Academic Editor PLOS ONE
  39 in total

1.  Etiology of decubitus ulcers.

Authors:  M KOSIAK
Journal:  Arch Phys Med Rehabil       Date:  1961-01       Impact factor: 3.966

2.  Use of weight-bearing MRI for evaluating wheelchair cushions based on internal soft-tissue deformations under ischial tuberosities.

Authors:  Nogah Shabshin; Gil Zoizner; Amir Herman; Vlad Ougortsin; Amit Gefen
Journal:  J Rehabil Res Dev       Date:  2010

Review 3.  Missing links in pressure ulcer research--an interdisciplinary overview.

Authors:  Christian Gammelgaard Olesen; Mark de Zee; John Rasmussen
Journal:  J Appl Physiol (1985)       Date:  2010-03-18

4.  A validated model of passive muscle in compression.

Authors:  M Van Loocke; C G Lyons; C K Simms
Journal:  J Biomech       Date:  2005-11-28       Impact factor: 2.712

5.  Tissue-interface pressures on three different support-surfaces for trauma patients.

Authors:  B P J A Keller; P H W Lubbert; E Keller; L P H Leenen
Journal:  Injury       Date:  2005-08       Impact factor: 2.586

6.  Full-body interface pressure testing as a method for performance evaluation of clinical support surfaces.

Authors:  F Shelton; R Barnett; E Meyer
Journal:  Appl Ergon       Date:  1998-12       Impact factor: 3.661

Review 7.  Seat cushion design for elderly wheelchair users based on minimization of soft tissue deformation using stiffness and pressure measurements.

Authors:  D M Brienza; P E Karg; C E Brubaker
Journal:  IEEE Trans Rehabil Eng       Date:  1996-12

8.  The aetiology of pressure sores.

Authors:  J Bridel
Journal:  J Wound Care       Date:  1993-07-02       Impact factor: 2.072

Review 9.  A systematic review of risk factors for the development and recurrence of pressure ulcers in people with spinal cord injuries.

Authors:  J Marin; J Nixon; C Gorecki
Journal:  Spinal Cord       Date:  2013-04-16       Impact factor: 2.772

Review 10.  Support surfaces for pressure ulcer prevention.

Authors:  Elizabeth McInnes; Asmara Jammali-Blasi; Sally E M Bell-Syer; Jo C Dumville; Victoria Middleton; Nicky Cullum
Journal:  Cochrane Database Syst Rev       Date:  2015-09-03
View more
  2 in total

1.  Different temporal weight-bearing tendencies of persons with right and left hemiplegia while sitting in a wheelchair.

Authors:  Hunhee Kim; Taekyeong Lee; Kang Hee Cho; Gwang Moon Eom; Junghwa Hong
Journal:  PLoS One       Date:  2022-01-27       Impact factor: 3.240

2.  Subcutaneous Fat Thickness Remarkably Influences Contact Pressure and Load Distribution of Buttock in Seated Posture.

Authors:  Kuan Wang; Yufang Chen; Shangjun Huang; Lejun Wang; Wenxin Niu
Journal:  J Healthc Eng       Date:  2021-11-30       Impact factor: 2.682

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.