When immobile or neuropathic patients are supported by beds or chairs, their soft tissues undergo deformations that can cause pressure ulcers. Current support surfaces that redistribute under-body pressures at vulnerable body sites have not succeeded in reducing pressure ulcer prevalence. Here we show that adding a supporting lateral pressure can counter-act the deformations induced by under-body pressure, and that this 'pressure equalisation' approach is a more effective way to reduce ulcer-inducing deformations than current approaches based on redistributing under-body pressure. A finite element model of the seated pelvis predicts that applying a lateral pressure to the soft tissue reduces peak von Mises stress in the deep tissue by a factor of 2.4 relative to a standard cushion (from 113 kPa to 47 kPa)-a greater effect than that achieved by using a more conformable cushion, which reduced von Mises stress to 75 kPa. Combining both a conformable cushion and lateral pressure reduced peak von Mises stresses to 25 kPa. The ratio of peak lateral pressure to peak under-body pressure was shown to regulate deep tissue stress better than under-body pressure alone. By optimising the magnitude and position of lateral pressure, tissue deformations can be reduced to that induced when suspended in a fluid. Our results explain the lack of efficacy in current support surfaces and suggest a new approach to designing and evaluating support surfaces: ensuring sufficient lateral pressure is applied to counter-act under-body pressure.
When immobile or neuropathicpatients are supported by beds or chairs, their soft tissues undergo deformations that can cause pressure ulcers. Current support surfaces that redistribute under-body pressures at vulnerable body sites have not succeeded in reducing pressure ulcer prevalence. Here we show that adding a supporting lateral pressure can counter-act the deformations induced by under-body pressure, and that this 'pressure equalisation' approach is a more effective way to reduce ulcer-inducing deformations than current approaches based on redistributing under-body pressure. A finite element model of the seated pelvis predicts that applying a lateral pressure to the soft tissue reduces peak von Mises stress in the deep tissue by a factor of 2.4 relative to a standard cushion (from 113 kPa to 47 kPa)-a greater effect than that achieved by using a more conformable cushion, which reduced von Mises stress to 75 kPa. Combining both a conformable cushion and lateral pressure reduced peak von Mises stresses to 25 kPa. The ratio of peak lateral pressure to peak under-body pressure was shown to regulate deep tissue stress better than under-body pressure alone. By optimising the magnitude and position of lateral pressure, tissue deformations can be reduced to that induced when suspended in a fluid. Our results explain the lack of efficacy in current support surfaces and suggest a new approach to designing and evaluating support surfaces: ensuring sufficient lateral pressure is applied to counter-act under-body pressure.
Supporting the body weight of critically ill, immobilised or paraplegic people
without causing soft-tissue injury is not an easy task. The loading induced while
lying or sitting for prolonged periods can cause damage to skin, adipose tissue and
muscle; this damage is known as a pressure ulcer. Pressure ulcers are estimated to
affect one in five hospitalised patients in Europe [1], while prevalence in some patient groups are
much higher. For example, 85% of spinal cord injurypatients develop a pressure
ulcer over their lifetime [2]
with associated care costs of approximately $1.2 billion annually in the US [3]. A severe form of pressure
ulcer develops in subdermal tissue close to bony prominences such as the ischial
tuberosity and sacrum of the pelvis [4,5], and is
known as a deep tissue injury. Because of the severity of deep tissue injury,
preventative strategies have been a major focus in the field.One approach to preventing pressure ulcers is to design support surfaces to reduce
pathological pressures, and this has been a major area of research for the past
forty years [6]—indeed
‘invalid beds’ have been developed since the 19th century [7]. While support surfaces have
become increasingly high-tech, they have yet to outperform high-specification foam
mattresses, and their adoption in clinics has not led to a significant reduction in
pressure ulcer prevalence [8,9]. While this
lack of progress may indicate that we have reached the limit of support surface
design, in this paper, we argue that current designs have been based upon a
suboptimal design principle—that of under-body pressure re-distribution.The presumption that high surface pressure leads to pressure ulcers, and therefore
should be reduced, seems obvious. However, experimental, computational and clinical
evidence suggests that high surface pressures do not necessarily cause pressure
ulcers. Peak surface pressures (as measured by pressure mapping sensor arrays)
cannot identify at-risk patients [10,11]. High-tech
mattresses that reduce peak surface pressures have increasingly been adopted in
clinical settings yet their impact on pressure ulcer prevalence has been
disappointing [8,9]. Furthermore, soft tissue
can tolerate extremely high surface pressures under certain
circumstances. The soft tissues of a deep-sea diver, for example, are exposed to 100
kPa of surface pressure for every 10 m descended, yet pressure-related injuries to
soft tissues are not a common issue in diving [12,13]. Computational studies have helped to
explain these observations, with Oomens et al. [14] demonstrating that peak surface pressure
has very little impact on internal deformations near bony prominences—regions where
deep tissue injuries are likely to occur [14,15]. Since reducing peak surface pressure has
failed to protect deep tissue, we sought to determine if there is any way to
manipulate the external pressure profile that can protect deep tissue.Deep tissue pressure ulcers develop as a result of several overlapping processes:
ischaemia [16],
ischaemia-reperfusion injury [17] lymphatic network obstruction [18,19] and direct cell deformation [20]. Each of these processes is
triggered by excessive deformation (exacerbated by shear stresses, microclimate, and
other risk factors) within soft tissue, and so in some regards, a pressure ulcer
could be more aptly named a ‘deformation ulcer’. Redistributing surface pressure (as
current devices aim to do) does not necessarily reduce deformations (and hence
pressure ulcers) because soft tissue has very different tolerances to the two
components of stress (Fig 1A):
deviatoric stress (which tends to change the shape of an object) and dilatational
stress (which tends to change the volume of an object, but not its shape) [21,22]. Human soft tissues are almost
incompressible [23], and so
can tolerate high dilatational stress with minimal deformation. In contrast, soft
tissues deform readily with deviatoric stress, therefore it is this stress that must
be minimised to prevent ulceration. While submersed, a diver experiences nearly
uniform pressures on all surfaces; this tends to induce dilatational stress [22]. On the other hand,
interaction with very localised surface pressure–such as when sitting on a chair or
lying on a mattress–induces large deviatoric stress (and therefore deformations) as
the soft tissue bulges and is displaced laterally away from the load (Fig 1B).
Fig 1
Deformations beneath a bony prominence.
The stress in soft tissue has two components, dilatational and deviatoric
(a). Soft tissue is much more resistant to dilatational stress than
deviatoric stress. Under a bony prominence, the soft tissue is distorted due
to the concentrated pressures at the bone and the support (b).
Redistributing the surface pressure has some effect on the outer
(superficial) region, but not on the deep tissue. We hypothesise that by
applying pressure laterally (termed pressure equalisation), bulging is
reduced, and the tissue can bear the load in a more dilatational mode.
Deformations beneath a bony prominence.
The stress in soft tissue has two components, dilatational and deviatoric
(a). Soft tissue is much more resistant to dilatational stress than
deviatoric stress. Under a bony prominence, the soft tissue is distorted due
to the concentrated pressures at the bone and the support (b).
Redistributing the surface pressure has some effect on the outer
(superficial) region, but not on the deep tissue. We hypothesise that by
applying pressure laterally (termed pressure equalisation), bulging is
reduced, and the tissue can bear the load in a more dilatational mode.One way to prevent excessive deformations may be to restrain the soft tissue from
deforming by applying a supporting lateral pressure. In this paper, we test the
plausibility of this principle using a computational model of the weight-bearing
pelvis in a seated individual. We hypothesise that actively applying pressure
laterally to the soft tissue of the pelvis will reduce the deformation at the
ischial tuberosity to a greater extent than the commonly applied method of
redistributing under-body pressure (Fig
1B). Our rationale for this work is that by making subtle changes to the
design objectives used for support surfaces, we may be able to substantially reduce
the risk of ulceration for high-risk patients.
2. Methods
First, we adapted a previously-developed finite element model of seated buttocks
[14] to test the
hypothesis that applying lateral pressure will reduce tissue deformations (section
2.1). The seated position was chosen because the ischial tuberosity is a common site
for deep tissue injury [1].
Next, we used the model to determine whether applying lateral pressure or changing
the stiffness of a standard cushion has the greatest effect on deep tissue
deformations (section 2.2). To ensure that these effects translate to a more
realistic setting, we developed a 3D model from MRI scans (section 2.3). Finally, we
sought to formalise the relationship between surface pressure and internal
deformations into a design principle—equalising under-body pressure with lateral
pressure. To do this, we described the interaction of soft tissue and a support
surface as a surface pressure boundary condition, which could be manipulated and
studied independently of particular cushion design (section 2.4). All finite element
input files and analysis protocols are available in an open database
(10.6084/m9.figshare.10510787).
2.1 Model of the seated pelvis with lateral pressure application
2.1.1 Geometry and material models. An axisymmetric geometry was used to model
the soft tissue surrounding a single ischial tuberosity in a seated individual.
The geometry was similar to that used by Oomens et al. [14] but included more of the pelvis soft
tissue to allow lateral pressure to be modelled (Fig 2A). The soft tissue was partitioned into
fat, muscle and skin to produce similar patterns as found from MRI imaging
(Fig 2B).
Fig 2
Finite element models.
(a) An axisymmetric model of the soft tissue surrounding the ischial
tuberosity. The model incorporates a rigid bony prominence, muscle, fat
and skin layers interacting with a cushion and a pressure equalisation
device. Axisymmetry was assumed, which allowed a force-controlled
simulation of weight-bearing (W is the load borne by the ischial
tuberosity). The pressure equalisation device was modelled as an
air-filled chamber with a controllable internal pressure, P. (b) The
axisymmetric region modelled is shown superimposed on saggital and
coronal MR images of a seated male (top). A 3D model was generated from
the MR images to assess 3D deformations (bottom).
Finite element models.
(a) An axisymmetric model of the soft tissue surrounding the ischial
tuberosity. The model incorporates a rigid bony prominence, muscle, fat
and skin layers interacting with a cushion and a pressure equalisation
device. Axisymmetry was assumed, which allowed a force-controlled
simulation of weight-bearing (W is the load borne by the ischial
tuberosity). The pressure equalisation device was modelled as an
air-filled chamber with a controllable internal pressure, P. (b) The
axisymmetric region modelled is shown superimposed on saggital and
coronal MR images of a seated male (top). A 3D model was generated from
the MR images to assess 3D deformations (bottom).There have been many material models of skeletal muscle [24,25], skin [26], and to a lesser extent fat [27]. However,
experimentally-based models that quantify all three tissues together are rarer,
making it difficult to combine tissues defined from different experimental
setups. In this study, we used the material models based on Oomens et al [14] because it enabled
direct comparison with that study, and because all three soft tissues were
characterized. Each region was assigned an Ogden hyperelastic material model,
and parameter values are listed in Table 1. A flat, 76 mm-thick, two-layered
seat cushion was modelled with hyperelastic material properties representing a
soft cushion (Table 1). An
air-filled chamber (the pressure-equalisation device) was introduced to apply a
lateral pressure to the soft tissues (Fig 2A). This was shaped to conform to the
seated pelvis and was modelled using membrane elements that can resist tensile,
but not bending, loads. The chamber wall material was modelled as sufficiently
stiff (Young’s Modulus, E = 10 MPa) so as not to allow appreciable changes in
length.
Table 1
Parameters for the Ogden material model for each of the materials
modelled.
Ogden strain energy density function, , where
λ1,2,3 are the principal stretches,
μ and a are material constants and
U is the strain energy density.
Material
μ (MPa)
a (-)
Skin
0.04
30
Fat
0.025
10
Muscle
0.045
5
Stiff cushion
0.08
10
Medium cushion
0.005
10
Soft cushion top
0.0035
7
Soft cushion bottom
0.005
10
Parameters for the Ogden material model for each of the materials
modelled.
Ogden strain energy density function, , where
λ1,2,3 are the principal stretches,
μ and a are material constants and
U is the strain energy density.2.1.2 Boundary conditions. The proportion of body weight borne by
the ischial tuberosities while seated varies from 18% to 77% [14]. We
estimated the amount of load supported by the pelvis at 400N —representing
approximately 50% of the body weight of an 80 kg adult (with each tuberosity
bearing 200N) because it is within the range of experimental findings and
enables direct comparison with Oomens et al [14]. Symmetry boundary conditions were
prescribed to all nodes lying along the z-axis (Fig 2A). The cushion base was constrained in
all directions. Two nodes of the pressure equalisation device were constrained
in all directions, and a uniform pressure was applied to the inner surface of
the chamber to a maximum of 80 kPa. Frictionless contact was assumed between the
support surfaces and the skin. While this is a simplification of the real-world
scenario, a sensitivity study revealed friction to have a negligible effect on
model predictions (S2 Appendix). Normal contact behaviour was
enforced using the penalty method with finite-sliding [28].2.1.3 Solution approach and output. A mesh sensitivity analysis was performed
leading to a final mesh of 13056 linear quadrilateral elements representing the
soft tissues. All models were solved as quasi-static, non-linear analyses using
the ABAQUS finite element software (v2016, Dassault Systems, France). To analyse
and compare models, the von Mises stresses and shear strains in the soft tissue
regions were calculated. Von Mises stress () is a
scalar representing the deviatoric part of the stress tensor
()
and defined as , where is
the stress tensor and
is the Kronecker delta function. Shear strain was calculated as
−,
where ,
are the maximum
and minimum principal strains, respectively. These were chosen to represent the
level of deviatoric stress and strain, respectively.To summarise the stresses and strains in the deep tissue, we sampled 1600
elements within a 30mm radius of the ischial tuberosity. Since elements vary
considerably in size throughout the region, we weighed our sampling by element
volume (IVOL output from ABAQUS). For example, an element with a volume 2V is
twice as likely to be sampled as an element with volume V. Weighting the results
like this ensures that analyses are independent of mesh density, which varies
throughout the model.Peak stress was defined as the 95th percentile of the stress data to avoid
extreme outliers that may be sensitive to boundary conditions. Effects sizes (in
the mean and peak values) between models were estimated by calculating
bootstrapped 95% confidence intervals [29]. We also analysed deep tissue stress
along a path through the soft tissue directly beneath the ischial tuberosity,
and along the surface of the ischial tuberosity.
2.2 Comparing lateral pressure application to changing cushion
stiffness
Cushion stiffness is a design variable commonly used to create a more conformable
cushion, and this was used as an intervention to compare to adding lateral
pressure. The model described above was adapted to model load-bearing on three
different cushion designs–a stiff, medium, and soft variety. The stiff and
medium cushions were homogeneous, 38 mm thick cushions, while the most compliant
(softest) was produced by defining a 76 mm thick two-layer cushion as in the
previous section. The bottom layer of this cushion had the medium-stiffness
cushion properties, and a softer material was assigned to the 38 mm top layer
(Table 1). These
cushions were based on those analysed by Oomens et al. [14], which were in turn calibrated to model
materials commonly used in wheelchair cushions.For each of the three cushion simulations, load-bearing to 200N was established
as in section 2.1. Pressure in the pressure equalisation device was then
increased incrementally to a maximum of 80 kPa. The stresses and strains induced
when lateral pressure is applied were computed.
2.3 Assessing deformations in 3D
Modelling the pelvis using a 2D axisymmetric model as above requires
simplification of the bone and soft tissue geometries. To ensure that the
beneficial effect of adding lateral pressure translates to 3D environments, we
developed a 3D model that can more accurately capture the geometry of a seated
pelvis. The MRI data of a male subject (age 30) was used to generate the 3D
geometry (Fig 2B) including
skin, fat, muscle and bone. Data usage was approved by the Imperial College
Research Ethics committee under ethical approval number ICHTB HTA licence: 12275
and REC Wales approval: 12/WA/0196. To aid comparison of results between models,
material properties for each layer were assigned to be consistent with the 2D
model. The nodes representing the outer surface of the pelvic bones were
constrained to move in the z-direction. Central symmetry was assumed to reduce
model size by half. The soft tissue was meshed using 669,995 linear tetrahedral
elements. A body force of 200N was applied to the bone nodes. This represents a
full body weight of 80 kg, with the assumption that 50% of this travels through
the pelvis of a seated individual (as was assumed with the axisymmetric model
and is based on Oomens et al. [14], see section 2.1.2). Maintaining this level of loading in the 3D
model aids comparison between that and the axisymmetric model. The lateral
supports were modelled as air-filled cavities, with a thin outer membrane of
material (as in the 2D model). To model the seated load case, the body force was
ramped incrementally over the first analysis step. Next, the lateral supports
were displaced towards the pelvis while the internal pressure within the support
was fixed at 1 kPa. Finally, the pressure within the lateral support was
increased to 10 kPa incrementally, and the von Mises stresses and maximum shear
strains were calculated over ten increments. This model was solved using
Abaqus/Explicit, with mass scaling applied to ensure kinetic energy was less
than 1% of internal strain energy.
2.4 Determining the relationship between surface pressure and deep tissue
mechanics
We studied how the shape and magnitude of the surface pressure distribution
affected internal tissue stress to understand how these stresses can be
minimised. The specific cushions and lateral support used in the previous
sections were removed and replaced with a surface pressure boundary condition.
For our axisymmetric model, pressure is a function of the angle from vertical,
P(θ). The surface pressure was constrained
to ensure that the model was in static equilibrium: the sum of the vertical
forces due to surface pressure is equal to body weight (W),
(∮P(θ)cosθdS =
W, where dS represents an infinitesimal
surface element), and the sum of the horizontal forces is zero
(∮P(θ)sinθdS = 0).The surface pressure on the buttocks when seated on a flat cushion follows a
characteristic distribution [30]—there is a pressure peak beneath each ischial tuberosity which
gradually reduces to zero towards the periphery of the contact area (S1
Appendix). This contact pressure was modelled as a Gaussian distribution
(see S1
Appendix). The spread of the pressure peak (α) was
varied between 0.2 and 0.35, which represent cushions with stiffness values
beyond the range of those tested in section 2.2. We then modelled an externally
applied lateral pressure by defining a second Gaussian term. This was controlled
by its spread (β), its magnitude
(P), and its location
(θ0). β and
θ0 were fixed (0.4 and π/4
respectively) and
P/P
(the ratio of lateral pressure relative to under-body pressure) was varied from
0% to 75%. This led to 16 parameter combinations, described in Table 2. Each pressure field
was then applied as a boundary condition to the soft-tissue finite element
model, and peak stresses and strains were calculated.
Table 2
Parameters governing the spread of the under-body pressure
(α) and the magnitude of lateral pressure.
Lateral pressure was defined relative to the peak under-body pressure
(P/P).
Pressure re-distribution
(α)
Lateral Pressure
(PL/PV)
0.2
0
0.25
0
0.3
0
0.35
0
0.2
0.25
0.25
0.25
0.3
0.25
0.35
0.25
0.2
0.50
0.25
0.50
0.3
0.50
0.35
0.50
0.2
0.75
0.25
0.75
0.3
0.75
0.35
0.75
Parameters governing the spread of the under-body pressure
(α) and the magnitude of lateral pressure.
Lateral pressure was defined relative to the peak under-body pressure
(P/P).2.4.1 Optimising the surface pressure distribution. We considered the pressure
distribution of a body suspended in a fluid as an ideal support scenario [22], as it results in
minimal deviatoric stress relative to dilatational stress (see ). We then optimised the location
() and relative
magnitude of the lateral pressure
(/)
to minimise the difference between
() and the
distribution when suspended in a fluid, while ensuring that the full body weight
(200N) was supported (see ).
We first set out to determine the effect of adding lateral pressure to a person
seated on a standard support surface. We used a finite element model of the
pelvis to simulate weight bearing while sitting on a soft cushion. Firstly, we
simulated weight bearing without lateral support. In agreement with other
studies [14,22,31], the model predicts significant stress
concentrations under the ischial tuberosity (Fig 3A), with peak von Mises stresses of 58
kPa produced in the muscle. Then, when lateral pressure is applied, the peak von
Mises stress beneath the bony prominence drops to 18 kPa, in support of our
hypothesis. Contour plots show the stress is more evenly distributed in the soft
tissues (Fig 3A), suggesting
that more of the soft tissue is being recruited in transferring the load.
Fig 3
Analysis of load-bearing when seated on a soft cushion.
In the absence of lateral pressure, the model predicts high von Mises
stresses under the ischial tuberosity (a). With the introduction of
lateral pressure (44 kPa chamber pressure), the region of high stress
shrinks dramatically. Histograms of stresses and strains in the muscle
tissue within a radius of 30 mm from the ischial tuberosity (b) indicate
that von Mises stresses and shear strains are reduced. Analysis of the
stress along path ABC (c) show a drop in von Mises stress and shear
strain at the bony prominence, and throughout the muscle tissue. Shear
strain and von Mises stress are also reduced in the skin and fat
layers.
Analysis of load-bearing when seated on a soft cushion.
In the absence of lateral pressure, the model predicts high von Mises
stresses under the ischial tuberosity (a). With the introduction of
lateral pressure (44 kPa chamber pressure), the region of high stress
shrinks dramatically. Histograms of stresses and strains in the muscle
tissue within a radius of 30 mm from the ischial tuberosity (b) indicate
that von Mises stresses and shear strains are reduced. Analysis of the
stress along path ABC (c) show a drop in von Mises stress and shear
strain at the bony prominence, and throughout the muscle tissue. Shear
strain and von Mises stress are also reduced in the skin and fat
layers.The volume of muscle tissue around the ischial tuberosity exposed to high von
Mises stress (> 20 kPa) is reduced from 58% to 4% with lateral pressure
application (Fig 3B). The
volume of muscle tissue exposed to high shear strains (> 0.2) fell from 20%
to 0%. Adding lateral pressure reduces the mean von Mises stress by 64% (95% CI
61% to 68%), while mean shear strains are reduced by 42% (95% CI 41% to 43%)
when lateral pressure is applied (Fig 3B).Plots of stress along a path through the soft tissue show that von Mises stress
is reduced in all tissues under the ischial tuberosity when lateral pressure is
applied (Fig 3C), with
stress reduction being most pronounced in the muscle (66% in muscle, 43% in fat
and 49% in skin). These results show that adding lateral pressure can reduce
deep tissue deviatoric stress and deformation.
3.2 Applying lateral pressure reduces deformations to a greater extent than
changing cushion stiffness
Having established that applying lateral pressure reduces deep tissue stress and
deformation, we next assessed the effect of this intervention compared to a
common device design consideration—changing cushion stiffness (Fig 4). Cushion stiffness is
usually manipulated to reduce peak pressures by maximising the contact area with
the soft tissue. Indeed, the soft cushion provides approximately 3.5 times more
contact area than the stiff cushion (72 cm2 in the stiff cushion, 177
cm2 in the medium cushion and 255 cm2 in the soft
cushion), indicating that we are capturing a broad range of support surface
stiffnesses. This area increase could be expected to achieve a similar reduction
in deep tissue deformation. However, von Mises stresses (Fig 4A) and shear strains (Fig 4B) in the deep tissue are
relatively less affected—with the change from a stiff to a soft cushion we see a
reduction by a factor of 1.4 in peak von Mises stress (Fig 4C). Without lateral pressure, all
cushions induce a peak von Mises stress > 50 kPa. We find that introducing
lateral pressure reduces the peak von Mises stresses observed with each cushion
by a factor of 2.4 on average (1.9 for the stiff cushion, 2.1 for the medium
cushion and 3.2 for the soft cushion; Fig 4C). The lowest peak deep tissue stresses
are observed when the soft cushion is combined with lateral pressure, which
reduces the peak von Mises stress to 18 kPa, suggesting a synergistic effect of
combining lateral pressure with a soft cushion (Fig 4C).
Fig 4
Applying lateral pressure is more effective than changing cushion
stiffness.
While the contact area varies substantially with cushion stiffness, the
pattern of internal stress remains similar (a)—stress is concentrated at
the bony prominence. Shear strains in the fat and skin are lower when a
softer cushion is used (b), but strains within the muscle remain high
for all cushions. These strains are reduced when lateral pressure is
introduced. All three cushions benefit from the introduction of lateral
pressure, with a soft cushion and lateral pressure providing the lowest
von Mises stresses (c) [Violin plots show mean and 95th percentile
values, stress difference plot shows the peak difference relative to a
stiff cushion only with 95% confidence intervals]. As lateral pressure
is gradually increased, the von Mises stress decreases until an optimum
pressure is reached (d); beyond this pressure, von Mises stresses begin
to increase again. While the magnitude of the optimum lateral pressure
is different for each cushion, the ratio of lateral to vertical pressure
is between 0.63 and 0.79 for all cushions tested.
Applying lateral pressure is more effective than changing cushion
stiffness.
While the contact area varies substantially with cushion stiffness, the
pattern of internal stress remains similar (a)—stress is concentrated at
the bony prominence. Shear strains in the fat and skin are lower when a
softer cushion is used (b), but strains within the muscle remain high
for all cushions. These strains are reduced when lateral pressure is
introduced. All three cushions benefit from the introduction of lateral
pressure, with a soft cushion and lateral pressure providing the lowest
von Mises stresses (c) [Violin plots show mean and 95th percentile
values, stress difference plot shows the peak difference relative to a
stiff cushion only with 95% confidence intervals]. As lateral pressure
is gradually increased, the von Mises stress decreases until an optimum
pressure is reached (d); beyond this pressure, von Mises stresses begin
to increase again. While the magnitude of the optimum lateral pressure
is different for each cushion, the ratio of lateral to vertical pressure
is between 0.63 and 0.79 for all cushions tested.We noticed that there is an optimum magnitude of lateral pressure which is
different for each cushion (38.5 kPa, 37.9 kPa and 12.2 kPa for stiff, medium
and soft cushions, respectively); however, the ratio of lateral to under-body
pressure is consistently between 0.6 and 0.8 (Fig 4D). This suggests that balancing
under-body and lateral pressures is more important for the reduction of deep
tissue deviatoric stress than reducing peak under-body pressures.
3.3 Lateral pressure reduces stresses in a 3D model
Upon addition of lateral pressure, stresses were reduced at the ischial
tuberosities in a similar way to the 2D model (Fig 5A). The volume of soft tissue exposed to
high von Mises stress greatly reduced when lateral pressure was applied (Fig 5B). Adding lateral
pressure reduced peak von Mises stress in the muscle by a factor of 2.5 when a
stiff cushion was used, 2.6 with a medium-stiffness cushion and 2.4 with a soft
cushion (Fig 5C). With
optimal lateral pressure applied, the stresses at the greater trochanter and
bones of the hemipelvis reached no more than 22% of the load at the ischial
tuberosity. These results demonstrate that the effects found in 2D are
representative of the 3D environment. They also show that gentle lateral
pressure can be applied without compromising the tissue at the femur or sacrum.
It should be noted that the lateral device modelled here was a very simple
design, and no optimisation of its shape was performed. By contouring the
lateral pressure device or other optimisations, further efficacy may be
possible.
Fig 5
A 3D model of the seated pelvis under load.
(a) Results for the stiff cushion shown with and without lateral pressure
applied. Coronal and transverse sections are shown to indicate von Mises
stresses both at the ischial tuberosities and the greater trochanter.
(b) The volume of soft tissue exposed to high stresses (>32kPa) is
shown in relation to the whole pelvis. The whole pelvis is made
transparent to help visualise the location of high stresses (beneath the
ischial tuberosity) (c) Change in peak von Mises stress throughout the
soft tissue of the pelvis (surrounding both the ischium and the femur)
as lateral pressure is increased.
A 3D model of the seated pelvis under load.
(a) Results for the stiff cushion shown with and without lateral pressure
applied. Coronal and transverse sections are shown to indicate von Mises
stresses both at the ischial tuberosities and the greater trochanter.
(b) The volume of soft tissue exposed to high stresses (>32kPa) is
shown in relation to the whole pelvis. The whole pelvis is made
transparent to help visualise the location of high stresses (beneath the
ischial tuberosity) (c) Change in peak von Mises stress throughout the
soft tissue of the pelvis (surrounding both the ischium and the femur)
as lateral pressure is increased.
3.4 Surface pressure equalisation is necessary to protect deep tissue from
deformation
Having found that deep tissue deformations were minimised for each cushion when
under-body and lateral pressure were in a specific ratio (0.6–0.8), we sought to
test whether this ratio could form the basis of a design principle. We removed
the cushion from the model and replaced it with a surface pressure boundary
condition that could be manipulated independently of cushion design. We varied
both the ratio of lateral to under-body pressure
(P/P),
and the spread of underbody pressure (α) and measured peak von
Mises stress for all combinations (Table 2).As in the previous analyses, in the absence of lateral pressure, redistributing
under-body pressure (increasing the spread, α, of the pressure
peak), reduces peak von Mises stresses at the ischial tuberosity, but even
substantial re-distribution fails to reduce the stress below 100 kPa (112 kPa is
observed when α = 0.35; Fig 6A). In contrast, inducing a lateral to
under-body pressure ratio of 0.25 reduces peak von Mises stress from 180 kPa to
67 kPa. The presence of lateral pressure appears to reduce the effect of
redistributing under-body pressure (Fig 6A), suggesting that when lateral pressure is employed, it
becomes the most important factor in reducing deformations. These results
indicate that controlling the ratio of lateral to under-body pressure
(P/P)
is necessary to achieve low deep-tissue stress.
Fig 6
Surface pressure analysis.
Redistributing under-body pressure
(P) reduces peak
von Mises stresses when no lateral pressure is applied (a), but peak
stresses remain above 100 kPa. Counter-acting that pressure with a
lateral pressure (P)
reduces peak stresses to a greater extent. When the magnitude and angle
of lateral pressure is optimised, the deep tissue von Mises stresses
approach that of suspension in a fluid (b; arrows illustrate pressure
intensity). Path plots of von Mises stress show that lateral pressure
can induce a similar stress profile at the bony prominence to that when
suspended in a fluid (c).
Surface pressure analysis.
Redistributing under-body pressure
(P) reduces peak
von Mises stresses when no lateral pressure is applied (a), but peak
stresses remain above 100 kPa. Counter-acting that pressure with a
lateral pressure (P)
reduces peak stresses to a greater extent. When the magnitude and angle
of lateral pressure is optimised, the deep tissue von Mises stresses
approach that of suspension in a fluid (b; arrows illustrate pressure
intensity). Path plots of von Mises stress show that lateral pressure
can induce a similar stress profile at the bony prominence to that when
suspended in a fluid (c).To understand why lateral pressure may be critical, we studied how this ratio
affects the shape of the pressure distribution when compared to two extreme
scenarios: the pressure distribution while sitting on a stiff cushion (a
high-deformation scenario), and that when suspended in a fluid (a
low-deformation scenario). The shape of these distributions is markedly
different (Fig 1A in S1 Appendix), with a sharp peak of pressure
beneath the ischial tuberosity when seated on a cushion, versus a smooth, even
pressure distribution when submersed. A parametric study showed that adding
lateral pressure best mimicked the pressure profile of suspension in a fluid
(S1
Appendix).We then optimised the magnitude and angle of the lateral pressure to best mimic
suspension in a fluid. Contour plots show that optimising this lateral pressure
(to
P/P
= 0.71 and θ0 =61.5°) can mimic the internal
stresses experienced while suspended in a fluid (Fig 6B). In addition, with these parameters,
von Mises stresses at the ischial tuberosity were either equal to or less than
those induced when suspended in a fluid (Fig 6C).In summary, not only can applying lateral pressure reduce deep tissue von Mises
stress and deformation, we have found that an optimal magnitude and location of
lateral pressure can mimic the environment induced when suspended in a
fluid.
4. Discussion
The goal of reducing peak surface pressures at vulnerable body sites has underpinned
the design of almost all medical support surfaces to date. Meanwhile, studies have
consistently concluded that peak surface pressures do not accurately predict
internal tissue mechanics [11,14], nor are
they effective in predicting patients at risk of pressure ulcers [10]. In this study, we have
shown that ensuring under-body and lateral pressures are balanced—a principle we
call pressure equalisation—is more effective at reducing deep tissue deformations
than reducing peak under-body pressure. We postulate that devices designed to
maintain a prescribed ratio of lateral pressure to under-body pressure will reduce
the risk of pressure ulcer formation in the soft tissue of immobile patients.The shift in emphasis from pressure re-distribution to pressure equalisation has
implications for support surface design (Fig 7). The synthesised results of multiple
clinical trials [8,9] suggest that any
well-designed mattress is better than a standard hospital bed, but none are
particularly successful at reducing pressure ulcer risk. Pressure redistributing
devices (either passive or active, Fig
7A and 7B respectively) may protect against superficial ulcers, while
having little effect on deep tissue injuries [14]. Our results show that devices must be
capable of providing sufficient lateral support to counter-act the deformations
induced by under-body pressure. Immersion/encapsulation-based devices such as water
beds [32] aim to increase the
contact area between the soft tissue and the support surface; however, while the
contact area may increase, the horizontal pressures at the periphery of the contact
area are usually minimal (Fig
7C), as pressure is primarily in reaction to gravitational body force. As
this force acts in the vertical direction, it would be insufficient to equalise the
under-body pressures and prevent bulging. Devices that aim to directly minimise
shape change have also been developed [33], but their dependence on vertical
translations of pistons means that they are not suited to regulating lateral
pressure. We believe that support surface technology may yet reduce pressure ulcer
prevalence if they are redirected to achieving pressure equalisation.
Fig 7
Pressure equalisation and its effects on device design.
For surfaces designed to reduce peak pressure passively (a), applying a
lateral pressure device helps to avoid lateral bulging (top, showing current
devices, bottom showing improved design). Active devices based on
individually controlled air cells (b) could be improved by surrounding the
soft tissue and changing the control software to aim for equalised pressure,
rather than reduce peak pressure. Encapsulation devices achieve large
contact areas, but the lateral pressures exerted may be limited (c). These
could be improved by active compression or smart materials. Pressure mapping
systems (d) currently identify pressure peaks as undesirable. If they could
measure pressure around the surface, then they could be re-purposed to
measure the level of pressure equalisation.
Pressure equalisation and its effects on device design.
For surfaces designed to reduce peak pressure passively (a), applying a
lateral pressure device helps to avoid lateral bulging (top, showing current
devices, bottom showing improved design). Active devices based on
individually controlled air cells (b) could be improved by surrounding the
soft tissue and changing the control software to aim for equalised pressure,
rather than reduce peak pressure. Encapsulation devices achieve large
contact areas, but the lateral pressures exerted may be limited (c). These
could be improved by active compression or smart materials. Pressure mapping
systems (d) currently identify pressure peaks as undesirable. If they could
measure pressure around the surface, then they could be re-purposed to
measure the level of pressure equalisation.The pressure equalisation principle has implications for pressure measurement as a
diagnostic tool and as a method of evaluating support surfaces. Surfaces
incorporating arrays of pressure sensors have been suggested as early-warning
systems for ulceration [34,35], and are
frequently used to evaluate new support surfaces [36-38]. Using this technology, devices can be
readily differentiated based on the peak pressures they produce. However, when these
devices are then compared through clinical outcomes, the differences between them
vanish [8,9], and so the current
predictive power of pressure measurement is limited. If surface pressure could be
measured all around the soft tissues (Fig 7D), then the level of pressure equalisation may be a more
predictive tool. Then, a measure of the ratio of lateral to under-body pressure
(P/P)
could be used to determine ulcer risk, and as a control signal for active
devices.The pressure gradient, defined as the spatial change in pressure from the point of
peak pressure, has been proposed as an alternative to peak pressure for predicting
soft tissue damage [39]. At
first glance, pressure equalisation may seem to be equivalent to using pressure
gradient as an ulceration indicator. However, the pressure gradient does not account
for the direction of pressure, and so a body could be loaded with a low pressure
gradient yet have little or no pressure equalisation, because lateral pressures are
not considered. In other words, pressure gradient is a local variable, as is peak
pressure, whereas pressure equalisation is a measure of the quality of the body
support as a whole.From a clinical perspective, devices that can equalise under-body pressure with
lateral pressure may be a vital tool in reducing the incidences of pressure
ulcers—both hospital-acquired ulcers and those that are acquired in community care
settings. The ability to apply sufficient lateral pressure will, however, need to be
balanced with other equally important design considerations. For example, wheelchair
users must not be exposed to intermittent high lateral pressures as they move in the
chair. This inflexibility is one reason why form-fitted cushions are not a solution
to preventing tissue distortion. In contrast, a successful device must be flexible
enough to provide a well-distributed and well-controlled lateral pressure regardless
of patient movement. Other design considerations include regulating the temperature
and humidity at the skin surface, as well as ease of installation, use and cleaning.
While this work has focused on the biomechanics of lateral pressure in general, the
practical application of this principle will be more complex and require significant
innovation in device design.The reductions in deep tissue stress and strain possible through surface pressure
equalisation could be sufficient to reduce pressure ulcer risk. The safe magnitude
of deformation (and even the most appropriate measure) is not yet fully accounted
for [40] and it is likely to
be patient, environment, and tissue-specific. In this study, we have used two
measures of deep tissue mechanics—von Mises stress and shear strain. These measures
aim to capture the deformations likely to lead to capillary and lymphatic vessel
restriction, and cell deformation, which contribute to pressure ulcer onset.
Experiments using rat muscle under compression [41] indicated that stresses greater than 32 kPa
induced damage, with this threshold dropping to 9 kPa over prolonged loading, while
work quantifying deformations in a similar model [42] indicated that damage occurred above a
shear strain of 0.3. Our results indicate that redistributing under-body pressure
would not protect soft tissue from these levels of deformation, but that applying
lateral pressure could.The 2D finite element model used here simplified the anatomical structure of the
pelvis in a similar way to previous studies [14,22,31]. These idealisations allowed us to focus on
the general case of a bony prominence transferring load through soft tissue to a
support surface, and enabled the comparisons and analyses described here. The 3D
model used here were important to support the conclusions drawn from 2D analyses,
but there are also limitations to this model. While we have included more anatomical
complexity including thighs, femurs and pelvis, more biofidelic models have been
proposed [43-45]. For example, we chose
tissue mechanical properties in line with Oomens et al. [14], but there are several published models of
soft tissue mechanical properties that vary in complexity [24-27]. This makes conclusions based on absolute
stress values difficult. A further complexity not accounted for in our model is the
posture and secondary supports (arm rests, for example) of the patient, which may
affect the loading boundary conditions. For these reasons, we have focused on the
relative effects of interventions on stresses and strains, thus making the
conclusions robust against the chosen material models and boundary conditions. Using
more biofidelic approaches will be a key step in applying the current results in the
clinic. In particular, models generated from high-risk patients as opposed to
healthy volunteers will be crucial. Physical validation will need to come from
measurements of internal tissue deformations, for example through load-bearing MRI
[46,47].Our results suggest a novel method for creating a safer mechanical environment. We
hypothesize that the reduced deformations created by lateral pressure equalization
will translate into better deep tissue blood perfusion and lower risk of
deformation-induced cell damage. A key next step will be to test this hypothesis by
measuring the physiological response of the soft tissues of seated patients, for
example through measuring transcutaneous gas tension [34].In conclusion, a change in focus from redistributing under-body pressure to
equalising it with lateral pressure will lead to new innovations and improvements to
patient care, resulting in a reduction of pressure ulcer prevalence in immobile
patients.
Analysis of pressure distributions.
(PDF)Click here for additional data file.
Testing model assumptions.
(PDF)Click here for additional data file.19 Nov 2019PONE-D-19-27579Lateral pressure equalisation as a principle for designing support surfaces to
prevent deep tissue pressure ulcersPLOS ONEDear Dr Boyle,Thank you for submitting your manuscript to PLOS ONE. After careful consideration, we
feel that it has merit but does not fully meet PLOS ONE’s publication criteria as it
currently stands. Therefore, we invite you to submit a revised version of the
manuscript that addresses the points raised during the review process.We would appreciate receiving your revised manuscript by Jan 03 2020 11:59PM. When
you are ready to submit your revision, log on to https://www.editorialmanager.com/pone/ and select the 'Submissions
Needing Revision' folder to locate your manuscript file.If you would like to make changes to your financial disclosure, please include your
updated statement in your cover letter.To enhance the reproducibility of your results, we recommend that if applicable you
deposit your laboratory protocols in protocols.io, where a protocol can be assigned
its own identifier (DOI) such that it can be cited independently in the future. For
instructions see: http://journals.plos.org/plosone/s/submission-guidelines#loc-laboratory-protocolsPlease include the following items when submitting your revised manuscript:A rebuttal letter that responds to each point raised by the academic
editor and reviewer(s). This letter should be uploaded as separate file
and labeled 'Response to Reviewers'.A marked-up copy of your manuscript that highlights changes made to the
original version. This file should be uploaded as separate file and
labeled 'Revised Manuscript with Track Changes'.An unmarked version of your revised paper without tracked changes. This
file should be uploaded as separate file and labeled 'Manuscript'.Please note while forming your response, if your article is accepted, you may have
the opportunity to make the peer review history publicly available. The record will
include editor decision letters (with reviews) and your responses to reviewer
comments. If eligible, we will contact you to opt in or out.We look forward to receiving your revised manuscript.Kind regards,Pedro H. Oliveira, Ph.D.Academic EditorPLOS ONEJournal requirements:1. When submitting your revision, we need you to address these additional
requirements.Please ensure that your manuscript meets PLOS ONE's style requirements, including
those for file naming. The PLOS ONE style templates can be found athttp://www.journals.plos.org/plosone/s/file?id=wjVg/PLOSOne_formatting_sample_main_body.pdf
and http://www.journals.plos.org/plosone/s/file?id=ba62/PLOSOne_formatting_sample_title_authors_affiliations.pdf2. Thank you for providing the following Funding Statement:"This project was partly funded by an Imperial College London and Imperial
Innovations Ltd. Proof of Concept funding to CB and SM and EPSRC funding to CH, MM
and SM (EP/N026845/1). The funders had no role in study design, data collection and
analysis, decision to publish, or preparation of the manuscript.".We note that one or more of the authors is affiliated with the funding organization,
indicating the funder may have had some role in the design, data collection,
analysis or preparation of your manuscript for publication; in other words, the
funder played an indirect role through the participation of the co-authors.If the funding organization did not play a role in the study design, data collection
and analysis, decision to publish, or preparation of the manuscript and only
provided financial support in the form of authors' salaries and/or research
materials, please review your statements relating to the author contributions, and
ensure you have specifically and accurately indicated the role(s) that these authors
had in your study in the Author Contributions section of the online submission form.
Please make any necessary amendments directly within this section of the online
submission form. Please also update your Funding Statement to include the following
statement: “The funder provided support in the form of salaries for authors [insert
relevant initials], but did not have any additional role in the study design, data
collection and analysis, decision to publish, or preparation of the manuscript. The
specific roles of these authors are articulated in the ‘author contributions’
section.”If the funding organization did have an additional role, please state and explain
that role within your Funding Statement.Please also provide an updated Competing Interests Statement declaring this
commercial affiliation along with any other relevant declarations relating to
employment, consultancy, patents, products in development, or marketed products,
etc.Within your Competing Interests Statement, please confirm that this commercial
affiliation does not alter your adherence to all PLOS ONE policies on sharing data
and materials by including the following statement: "This does not alter our
adherence to PLOS ONE policies on sharing data and materials.” (as detailed online
in our guide for authors http://journals.plos.org/plosone/s/competing-interests). If this
adherence statement is not accurate and there are restrictions on sharing of data
and/or materials, please state these. Please note that we cannot proceed with
consideration of your article until this information has been declared.3. We note that you have a patent relating to material pertinent to this article.
Please provide an amended statement of Competing Interests to declare this patent
(with details including name and number), along with any other relevant declarations
relating to employment, consultancy, patents, products in development or modified
products etc. Please confirm that this does not alter your adherence to all PLOS ONE
policies on sharing data and materials, as detailed online in our guide for authors
http://journals.plos.org/plosone/s/competing-interests by including
the following statement: "This does not alter our adherence to PLOS ONE policies on
sharing data and materials.” If there are restrictions on sharing of data and/or
materials, please state these. Please note that we cannot proceed with consideration
of your article until this information has been declared.* This information should be included in your cover letter; we will change the online
submission form on your behalf.Please know it is PLOS ONE policy for corresponding authors to declare, on behalf of
all authors, all potential competing interests for the purposes of transparency.
PLOS defines a competing interest as anything that interferes with, or could
reasonably be perceived as interfering with, the full and objective presentation,
peer review, editorial decision-making, or publication of research or non-research
articles submitted to one of the journals. Competing interests can be financial or
non-financial, professional, or personal. Competing interests can arise in
relationship to an organization or another person. Please follow this link to our
website for more details on competing interests: http://journals.plos.org/plosone/s/competing-interests[Note: HTML markup is below. Please do not edit.]Reviewers' comments:Reviewer's Responses to QuestionsComments to the Author1. Is the manuscript technically sound, and do the data support the conclusions?The manuscript must describe a technically sound piece of scientific research with
data that supports the conclusions. Experiments must have been conducted rigorously,
with appropriate controls, replication, and sample sizes. The conclusions must be
drawn appropriately based on the data presented.Reviewer #1: YesReviewer #2: Yes**********2. Has the statistical analysis been performed
appropriately and rigorously?Reviewer #1: N/AReviewer #2: Yes**********3. Have the authors made all data underlying the
findings in their manuscript fully available?The PLOS Data policy requires authors to make all
data underlying the findings described in their manuscript fully available without
restriction, with rare exception (please refer to the Data Availability Statement in
the manuscript PDF file). The data should be provided as part of the manuscript or
its supporting information, or deposited to a public repository. For example, in
addition to summary statistics, the data points behind means, medians and variance
measures should be available. If there are restrictions on publicly sharing
data—e.g. participant privacy or use of data from a third party—those must be
specified.Reviewer #1: YesReviewer #2: Yes**********4. Is the manuscript presented in an intelligible
fashion and written in standard English?PLOS ONE does not copyedit accepted manuscripts, so the language in submitted
articles must be clear, correct, and unambiguous. Any typographical or grammatical
errors should be corrected at revision, so please note any specific errors
here.Reviewer #1: YesReviewer #2: Yes**********5. Review Comments to the AuthorPlease use the space provided to explain your answers to the questions above. You may
also include additional comments for the author, including concerns about dual
publication, research ethics, or publication ethics. (Please upload your review as
an attachment if it exceeds 20,000 characters)Reviewer #1: The paper deals with a basic research regarding the opportunity or not
to use lateral pressure to be applied to the patient forced to maintain a sitting
position for long periods in order to reduce the pressure at the interface with the
pillow and consequently the risk of ulcerations.The idea is good and its applicability must be verified. The preliminary study is
well written and the methods used are appropriate.Small improvements are required:- always define all the parameters introduced (for example in table 1 and in the
equations)- in paragraph 2.1.2 it is well known that the body mass of the torso + head + upper
limbs is equal to 2/3 of the whole body mass not 1/2- in paragraph 2.1.2 next to the figure 80 the unit of measurement is missing- in brackets, after the softwares used, it is necessary to mention the company that
distributes them and their location- in paragraph 2.3 justify the choices made on the imposed loads.Reviewer #2: This study sought to present a novel approach to prevent ulcers due to
deep tissue pressure. In particular the authors stated that adding a supporting
lateral pressure (they called “pressure equalization”) to the classical under-body
pressure redistribution approach could provide a more effective solution to reduce
deep tissue deformation that can lead to ulcers development. The work is based on
Finite Element (FE) Model, which is able to predict the stresses induced on the deep
tissue; the authors used this model to assess the differences introduced by the
lateral pressure equalization. In particular, they found that managing the ratio of
peak lateral pressure to peak under-body pressure and the position of the lateral
supports could be used to regulate deep tissue stresses. The authors stated that the
overall deformation can be decreased at the induced when the body is suspended in a
fluid. The approach can be used to design and assess novel solutions related to
supporting devices.General CommentsThe main working hypothesis at the basis of this paper is clearly reported as far as
the main objective. The introduction to the approach is clearly reported and the
modelling phase well documented. Although the methodology is not that innovative,
the application to is indeed original. I guess the audience of PLOS ONE will
appreciate this work.The structure of the article seems to be precise (Abstract, Introduction, Methods
[with subheadings], Results, Discussion).Modelling and data analysis seem to be clearly reported and coherent with the work
objectives. Several minor concerns have been already reported to the authors.The use of the English language seems to be correct.The references to previous works seem to be precise and up-to-date.Specific CommentsAbstractIn general this section is ok. Please, could you report also the magnitude of the
deformation related to deep tissue and the differences between under-body pressure
redistribution and lateral pressure equalization?IntroductionReally well written. You well reported the main clinical problem, the
state-of-the-art related to the main provided solutions, your hypothesis and the
rationale behind your work. Really appreciated Figure 1.Methods• Page 8. Model properties. Oomens’ work is a modelling approach dating back to 2003.
Is there any more recent work identifying soft tissues properties and modelling?• Page 11 – 2.1.2 Line 6. I guess that the unit of measurements is missing.• Please provide more hypothesis on the choice of not introducing any friction
between the support surface and the skin. Is this condition the real one?• Page 12. Please provide more information about the “adjusting” of data accounting
for volume variation. It is not that clear.• Page 16. Table 2. Please provide information about “[0.5ex]”, present in the first
line.ResultsVery well reported both graphically and in the main text.DiscussionIn general, this section is ok. Since you reported also the most clinical
perspective, what about pressure equalization on vessels? Could you speculate on any
main issue?**********6. PLOS authors have the option to publish the peer
review history of their article (what does this mean?). If published, this will
include your full peer review and any attached files.If you choose “no”, your identity will remain anonymous but your review may still be
made public.Do you want your identity to be public for this peer review? For
information about this choice, including consent withdrawal, please see our
Privacy Policy.Reviewer #1: NoReviewer #2: Yes: Nicola Francesco Lopomo[NOTE: If reviewer comments were submitted as an attachment file, they will be
attached to this email and accessible via the submission site. Please log into your
account, locate the manuscript record, and check for the action link "View
Attachments". If this link does not appear, there are no attachment files to be
viewed.]While revising your submission, please upload your figure files to the Preflight
Analysis and Conversion Engine (PACE) digital diagnostic tool, https://pacev2.apexcovantage.com/. PACE helps
ensure that figures meet PLOS requirements. To use PACE, you must first register as
a user. Registration is free. Then, login and navigate to the UPLOAD tab, where you
will find detailed instructions on how to use the tool. If you encounter any issues
or have any questions when using PACE, please email us at figures@plos.org. Please note that Supporting Information files do not
need this step.21 Nov 2019We wish to thank the reviewers for their valuable critique of our manuscript. Pease
see below for our responses to each comment.Reviewer #1: The paper deals with a basic research regarding the opportunity or not
to use lateral pressure to be applied to the patient forced to maintain a sitting
position for long periods in order to reduce the pressure at the interface with the
pillow and consequently the risk of ulcerations.The idea is good and its applicability must be verified. The preliminary study is
well written and the methods used are appropriate.Small improvements are required:- always define all the parameters introduced (for example in table 1 and in the
equations)We have reviewed the manuscript to ensure all parameters are now defined at the point
of first instance.- in paragraph 2.1.2 it is well known that the body mass of the torso + head + upper
limbs is equal to 2/3 of the whole body mass not ½We agree that this 50% value does not represent all upper-body weight. Oomens et al
[14] discuss loading boundary conditions in pelvis models. Using literature on
experimental pressure-reading, they describe that weight borne by the IT during
sitting varies from 18% to 77% of body weight. This variation is due to patient
posture, including use of arm and thigh support, and inclination of back support. We
chose 50% as a value within the experimental range, while also enabling direct
comparison with that previous modelling work [14].We also argue that because we look at relative differences (between support
surfaces), rather than draw conclusions based on absolute values, the choice of load
magnitude does not affect our conclusions.Changes to text:Section 2.1.2 pg 8:The proportion of body weight borne by the ischial tuberosities while seated varies
from 18% to 77% [14]. We estimated the amount of load supported by the pelvis at
400N — representing approximately 50% of the body weight of an 80 kg adult (with
each tuberosity bearing 200N) because it is within the range of experimental
findings and enables direct comparison with Oomens et al [14].Discussion pg 25:A further complexity not accounted for in our model is the posture and secondary
supports (arm rests, for example) of the patient, which may affect the loading
boundary conditions. For these reasons, we have focused on the relative effects of
interventions on stresses and strains, thus making the conclusions robust against
the chosen material models and boundary conditions. Using more biofidelic approaches
will be a key step in applying the current results in the clinic.- in paragraph 2.1.2 next to the figure 80 the unit of measurement is missingThank you, amended with kPa as units- in brackets, after the softwares used, it is necessary to mention the company that
distributes them and their locationAmended as suggested- in paragraph 2.3 justify the choices made on the imposed loads.Our objective for developing this 3D model was to ensure that the conclusions drawn
from the axisymmetric model translated to a 3D environment. To achieve this
objective, we modelled the same loading regime as the axisymmetric model. We accept,
however, that a more biofidelic model – based on realistic load cases including body
freedom-of-movement and whole-body simulation – will be required as a next step
before translation to the clinic.Changes to text (the alterations to section 2.1.2 are also relevant here):Section 2.3 pg 11:A body force of 200N was applied to the bone nodes. This represents a full body
weight of 80 kg, with the assumption that 50% of this travels through the pelvis of
a seated individual (as was assumed with the axisymmetric model and is based on
Oomens et al. [14], see section 2.1.2). Maintaining this level of loading in the 3D
model aids comparison between that and the axisymmetric model.Reviewer #2: This study sought to present a novel approach to prevent ulcers due to
deep tissue pressure. In particular the authors stated that adding a supporting
lateral pressure (they called “pressure equalization”) to the classical under-body
pressure redistribution approach could provide a more effective solution to reduce
deep tissue deformation that can lead to ulcers development. The work is based on
Finite Element (FE) Model, which is able to predict the stresses induced on the deep
tissue; the authors used this model to assess the differences introduced by the
lateral pressure equalization. In particular, they found that managing the ratio of
peak lateral pressure to peak under-body pressure and the position of the lateral
supports could be used to regulate deep tissue stresses. The authors stated that the
overall deformation can be decreased at the induced when the body is suspended in a
fluid. The approach can be used to design and assess novel solutions related to
supporting devices.General CommentsThe main working hypothesis at the basis of this paper is clearly reported as far as
the main objective. The introduction to the approach is clearly reported and the
modelling phase well documented. Although the methodology is not that innovative,
the application to is indeed original. I guess the audience of PLOS ONE will
appreciate this work.The structure of the article seems to be precise (Abstract, Introduction, Methods
[with subheadings], Results, Discussion).Modelling and data analysis seem to be clearly reported and coherent with the work
objectives. Several minor concerns have been already reported to the authors.The use of the English language seems to be correct.The references to previous works seem to be precise and up-to-date.Specific CommentsAbstractIn general this section is ok. Please, could you report also the magnitude of the
deformation related to deep tissue and the differences between under-body pressure
redistribution and lateral pressure equalization?Thank you for this suggestion, we have updated the abstract to include magnitudes as
well as fold-change.Changes to text:Abstract:A finite element model of the seated pelvis predicts that applying a lateral pressure
to the soft tissue reduces peak von Mises stress in the deep tissue by a factor of
2.4relative to a standard cushion (from 113 kPa to 47 kPA) — a greater effect than
that achieved by using a more conformable cushion, which reduced von Mises stress to
75 kPa. Combining both a conformable cushion and lateral pressure reduced peak von
Mises stresses to 25 kPa.IntroductionReally well written. You well reported the main clinical problem, the
state-of-the-art related to the main provided solutions, your hypothesis and the
rationale behind your work. Really appreciated Figure 1.Methods• Page 8. Model properties. Oomens’ work is a modelling approach dating back to 2003.
Is there any more recent work identifying soft tissues properties and modelling?Thank you for this comment. We accept that the material properties chose may not be
the definitive material models for each of the soft tissues. For example, more
complex models that incorporate transverse isotropy of muscle/skin tissue or
viscoelasticity may yield more accurate predictions of absolute stress/strain
magnitudes. However, our study focuses on relative differences between groups (on
different cushions, and with/without lateral pressure).We argue that this focus on relative differences reduces the sensitivity of our
conclusions to specific material models, particularly because our chosen models
incorporate the two most important aspects of soft tissue mechanics
(large-deformations and material non-linearity). We also note that using these
models enabled direct comparison to the study most closely related to our work
(Oomens et al [14]).We also understand that incorporating more accurate material models of soft tissues
will be a crucial part of creating more biofidelic models, and that this is a next
step on the way to validating our pressure-equalisation approach for the clinic.Changes to text:Methods, page 7:There have been many material models of skeletal muscle [24,25], skin [26], and to a
lesser extent fat [27]. However, experimentally-based models that quantify all three
tissues together are rarer, making it difficult to combine tissues defined from
different experimental setups. In this study, we used the material models based on
Oomens et al [14] because it enabled direct comparison with that study, and because
all three soft tissues were characterized. Each region was assigned an Ogden
hyperelastic material model, and parameter values are listed in Table 1.Discussion, page 25:For example, we chose tissue mechanical properties in line with Oomens et al. [14],
but there are several published models of soft tissue mechanical properties that
vary in complexity [24-27]. This makes conclusions based on absolute stress values
difficult. A further complexity not accounted for in our model is the posture and
secondary supports (arm rests, for example) of the patient, which may affect the
loading boundary conditions. For these reasons, we have focused on the relative
effects of interventions on stresses and strains, thus making the conclusions robust
against the chosen material models and boundary conditions. Using more biofidelic
approaches will be a key step in applying the current results in the clinic.• Page 11 – 2.1.2 Line 6. I guess that the unit of measurements is missing.Amended to include kPa as unit.• Please provide more hypothesis on the choice of not introducing any friction
between the support surface and the skin. Is this condition the real one?We used frictionless contact as a simplification of the real-world situation. To
assess the impact of this simplification, we ran a sensitivity study where the
effect of including friction on peak stresses and strains was quantified (see S2
Appendix). We found that in our load case, friction had negligible effect on peak
stresses. However, we accept that in future studies that aim to simulate more
biofidelic load cases, modelling friction effects will become increasingly
important. In particular, a model that captured patient posture dynamically would
require more accurate contact characterisation.We also note that our primary focus in this study was the deep tissue (the source of
the most dangerous pressure ulcers), while friction tends to induce superficial
pressure ulcers.Changes to text:Methods page 9:While this is a simplification of the real-world scenario, a sensitivity study
revealed friction to have a negligible effect on model predicitons (S2
Appendix).• Page 12. Please provide more information about the “adjusting” of data accounting
for volume variation. It is not that clear.Thank you for this comment. We developed an analysis approach to ensure that our
conclusions were insensitive to the mesh density. Our aim was to summarise the
stress/strain in a tissue. One approach is to simply probe one point near the region
of interest. However, this is sensitive to the choice of point. Another approach is
to summarise all the points within a region of interest (mean, median, 95
percentile, etc). However, this approach is affected by mesh density, so locations
with many small elements contribute more to the summary statistic than locations
with fewer large elements. To overcome this limitation, we took a weighted sample of
all elements in the region of interest. The sample was weighted by the volume of the
element, so an element with double the volume is twice as likely to be sampled.Changes to text:Page 10:To summarise the stresses and strains in the deep tissue, we sampled 1600 elements
within a 30mm radius of the ischial tuberosity. Since elements vary considerably in
size throughout the region, we weighed our sampling by element volume (IVOL output
from ABAQUS). For example, an element with a volume 2V is twice as likely to be
sampled as an element with volume V. Weighting the results like this ensures that
analyses are independent of mesh density, which varies throughout the model.• Page 16. Table 2. Please provide information about “[0.5ex]”, present in the first
line.Thank you for spotting this, it is an artefact of a conversion from Latex to MS Word
and has been deleted.ResultsVery well reported both graphically and in the main text.DiscussionIn general, this section is ok. Since you reported also the most clinical
perspective, what about pressure equalization on vessels? Could you speculate on any
main issue?If we understand your question correctly, you would like us to discuss how our
mechanical results translate into improvements to blood perfusion, thereby keeping
tissue oxygenated and uninjured. This is a great point, and we have added a
paragraph to the discussion.Changes to text:Discussion, Page 27:Our results suggest a novel method for creating a safer mechanical environment. We
hypothesize that the reduced deformations created by lateral pressure equalization
will translate into better deep tissue blood perfusion and lower risk of
deformation-induced cell damage. A key next step will be to test this hypothesis by
measuring the physiological response of the soft tissues of seated patients, for
example through measuring transcutaneous gas tension [34].Submitted filename: responsetoReviewsR1.docxClick here for additional data file.12 Dec 2019Lateral pressure equalisation as a principle for designing support surfaces to
prevent deep tissue pressure ulcersPONE-D-19-27579R1Dear Dr. Boyle,We are pleased to inform you that your manuscript has been judged scientifically
suitable for publication and will be formally accepted for publication once it
complies with all outstanding technical requirements.Within one week, you will receive an e-mail containing information on the amendments
required prior to publication. When all required modifications have been addressed,
you will receive a formal acceptance letter and your manuscript will proceed to our
production department and be scheduled for publication.Shortly after the formal acceptance letter is sent, an invoice for payment will
follow. To ensure an efficient production and billing process, please log into
Editorial Manager at https://www.editorialmanager.com/pone/, click the "Update My
Information" link at the top of the page, and update your user information. If you
have any billing related questions, please contact our Author Billing department
directly at authorbilling@plos.org.If your institution or institutions have a press office, please notify them about
your upcoming paper to enable them to help maximize its impact. If they will be
preparing press materials for this manuscript, you must inform our press team as
soon as possible and no later than 48 hours after receiving the formal acceptance.
Your manuscript will remain under strict press embargo until 2 pm Eastern Time on
the date of publication. For more information, please contact onepress@plos.org.With kind regards,Pedro H. Oliveira, Ph.D.Academic EditorPLOS ONEComments to the Author1. If the authors have adequately addressed your comments raised in a previous round
of review and you feel that this manuscript is now acceptable for publication, you
may indicate that here to bypass the “Comments to the Author” section, enter your
conflict of interest statement in the “Confidential to Editor” section, and submit
your "Accept" recommendation.Reviewer #1: (No Response)Reviewer #2: All comments have been addressed**********2. Is the manuscript technically sound, and do the data
support the conclusions?The manuscript must describe a technically sound piece of scientific research with
data that supports the conclusions. Experiments must have been conducted rigorously,
with appropriate controls, replication, and sample sizes. The conclusions must be
drawn appropriately based on the data presented.Reviewer #1: (No Response)Reviewer #2: Yes**********3. Has the statistical analysis been performed
appropriately and rigorously?Reviewer #1: (No Response)Reviewer #2: Yes**********4. Have the authors made all data underlying the
findings in their manuscript fully available?The PLOS Data policy requires authors to make all
data underlying the findings described in their manuscript fully available without
restriction, with rare exception (please refer to the Data Availability Statement in
the manuscript PDF file). The data should be provided as part of the manuscript or
its supporting information, or deposited to a public repository. For example, in
addition to summary statistics, the data points behind means, medians and variance
measures should be available. If there are restrictions on publicly sharing
data—e.g. participant privacy or use of data from a third party—those must be
specified.Reviewer #1: (No Response)Reviewer #2: Yes**********5. Is the manuscript presented in an intelligible
fashion and written in standard English?PLOS ONE does not copyedit accepted manuscripts, so the language in submitted
articles must be clear, correct, and unambiguous. Any typographical or grammatical
errors should be corrected at revision, so please note any specific errors
here.Reviewer #1: (No Response)Reviewer #2: Yes**********6. Review Comments to the AuthorPlease use the space provided to explain your answers to the questions above. You may
also include additional comments for the author, including concerns about dual
publication, research ethics, or publication ethics. (Please upload your review as
an attachment if it exceeds 20,000 characters)Reviewer #1: (No Response)Reviewer #2: The authors answered all the issues arisen during the first review. All
the comments have been fully addressed.**********7. PLOS authors have the option to publish the peer
review history of their article (what does this mean?). If published, this will
include your full peer review and any attached files.If you choose “no”, your identity will remain anonymous but your review may still be
made public.Do you want your identity to be public for this peer review? For
information about this choice, including consent withdrawal, please see our
Privacy Policy.Reviewer #1: NoReviewer #2: Yes: Nicola Francesco Lopomo13 Dec 2019PONE-D-19-27579R1Lateral pressure equalisation as a principle for designing support surfaces to
prevent deep tissue pressure ulcersDear Dr. Boyle:I am pleased to inform you that your manuscript has been deemed suitable for
publication in PLOS ONE. Congratulations! Your manuscript is now with our production
department.If your institution or institutions have a press office, please notify them about
your upcoming paper at this point, to enable them to help maximize its impact. If
they will be preparing press materials for this manuscript, please inform our press
team within the next 48 hours. Your manuscript will remain under strict press
embargo until 2 pm Eastern Time on the date of publication. For more information
please contact onepress@plos.org.For any other questions or concerns, please email plosone@plos.org.Thank you for submitting your work to PLOS ONE.With kind regards,PLOS ONE Editorial Office Staffon behalf ofDr. Pedro H. OliveiraAcademic EditorPLOS ONE
Authors: Elizabeth McInnes; Asmara Jammali-Blasi; Sally E M Bell-Syer; Jo C Dumville; Victoria Middleton; Nicky Cullum Journal: Cochrane Database Syst Rev Date: 2015-09-03