Study of minimally invasive radiofrequency ablation of the ciliary body for the treatment of glaucoma in rabbits.

Baoke Hou1, Fengxiang Wang1, Zi Ye1, Xin Jin1, Yu Fu1, Zhaohui Li1.   

Abstract

The aim of the present study was to explore the safety and effectiveness of radiofrequency ablation (RFA) of the ciliary body for the treatment of glaucoma. A glaucoma model was established in New Zealand white rabbits, which were then treated with RFA of the ciliary body, utilizing an XL‑1‑type RF meter developed by the Chinese PLA General Hospital. After treatment, general ocular investigation, including ocular pressure was carried out, the anterior chamber was imaged via ultrasound biomicroscopy, and the pathological changes were observed via hematoxylin and eosin (H&E) staining. It was determined that the glaucoma model was successfully established in the New Zealand white rabbit by inducing high intraocular pressure (IOP). After RFA treatment, ablation spots were observed but no clear anterior chamber reaction was found. The ablation group showed a steady and continuous decrease of IOP, which was significantly lower than the model group at days 3 and 7 (P<0.05). A sclera pathway was observed in the ablation site 1‑day post‑treatment, which had mostly recovered by day 7. H&E staining demonstrated shedding of the ciliary epithelium, and an unclear boundary between muscle layer and blood vessel at day 1. This had fully recovered by day 14, with clear ciliary layers and well‑arranged muscle structures observed. The present study suggested that treatment with RFA could decrease IOP without substantial side effects in the glaucoma model in the rabbit. Therefore, it could be used as a strategy to control IOP and as a treatment for glaucoma in the clinic.

Entities:  

Mesh:

Year:  2019        PMID: 31894284      PMCID: PMC7003027          DOI: 10.3892/mmr.2019.10906

Source DB:  PubMed          Journal:  Mol Med Rep        ISSN: 1791-2997            Impact factor:   2.952


Introduction

Glaucoma is a progressive optic nerve disorder that is characterized by structural changes of the optic nerve head, and commonly accompanied by the development of visual functional defects (1). The World Health Organization has listed glaucoma as the second leading cause of irreversible blindness (2). According to World Health Organization statistics, ~6,680 million patients suffer from primary glaucoma and 600 million patients are affected by secondary glaucoma, resulting in 670 million patients exhibiting blindness (3). It is estimated by Quigley and Broman (4) that the number of patients with global primary glaucoma will grow to 7,960 million by the year 2020, while blindness caused by glaucoma will increase to 1,120 million. It is widely known that elevated intraocular pressure (IOP) is the most critical risk factor that leads to optic nerve damage in glaucoma (5). Therefore, controlling IOP has been suggested as one of the main strategies to prevent the development of glaucoma from ocular hypertension, eliminate the progression of early glaucoma disease and block the occurrence of progressive visual function damage in the clinic (6–8). However, current clinically employed treatments for reducing IOP are not ideal. The treatment of glaucoma in clinical settings can be divided into drug therapy and surgical procedures (9). There are a wide range of drugs available that reduce IOP, but constant medication is required due to their limited effects. Consequently, a number of patients must resort to surgery after drug treatment (10). Anti-glaucoma surgery can be classified into two types: Procedures that facilitate the outflow of the aqueous humor, including aqueous humor shunts, and trabecular, goniotomy and iridotomy surgeries; and procedures that reduce aqueous humor secretion by destroying cells of the ciliary body, such as cyclocryotherapy and cyclophotocoagulation (11). However, the former group of procedures is flawed, due to scarring problems and their limited long-term effects, while the latter group lacks precise control and can only be utilized for patients with absolute glaucoma who have already lost their vision (12–15). Therefore, there is an urgent need to develop more safe and effective therapeutic approaches for glaucoma. Radiofrequency ablation (RFA) is a new, minimally invasive technique (16). During treatment, microelectrodes are directly punctured into the focal site, and high-power radiofrequency energy is delivered inside through the catheter, making the local temperature high and resulting in irreversible coagulating necrosis (17). RFA has a number of advantages: It offers precise control, it is non-invasive and it demonstrates high effectiveness. Therefore, it has been applied in various departments, including cancer, cardiology, otorhinolaryngology, digestion, orthopedics, obstetrics, stomatology, surgery, physical therapy and dermatology (18). Currently, conductive keratoplasty has also been utilized in the clinic for the treatment of presbyopia, in which the radiofrequency energy is employed to change the shape of the eye's surface (19,20). The successful application of conductive keratoplasty indicates that radiofrequency treatment may be safe and effective for ocular diseases. In the present study, the minimally invasive technique of radiofrequency catheter ablation was used, guided by ultrasound biomicroscopy (UBM), to treat glaucoma. The ciliary body was specifically targeted and ablated, following which the production of the aqueous humor was reduced, which significantly lowered IOP. RFA technology not only controls IOP precisely, reliably, safely and durably, but it is also simple and repeatable, with a relatively low cost (21). Thus, it was hypothesized that minimally invasive RFA may be a highly desirable treatment procedure for glaucoma, from a social and economic standpoint.

Materials and methods

Experimental animals

A total of 30 New Zealand rabbits (age, 4–6 months; weight, 2.0–2.5 kg; sex, 16 female and 14 male) were purchased from the Animal Research Center of the Chinese PLA General Hospital. Eye disease was excluded before research. All rabbits were allowed free access to food and water, and lived in an environment with 12-h light/dark cycles, a temperature of 25°C and ~50% humidity. Animals were randomly divided into 3 groups: The model group (n=10), in which the drug was injected into the anterior chamber to induce glaucoma, but no treatment was applied; the RFA group (n=10), where rabbits were treated with RFA after the induction of glaucoma; and the sham group (n=10), in which saline instead of the drug was injected into the anterior chamber. All protocols were conducted in accordance with the National Institutes of Health Guidelines for the Care and Use of Laboratory Animals (22) and approved by the Institutional Animal Care and Use Committee of the Chinese PLA General Hospital.

Establishment of the glaucoma model in rabbits

Glaucoma was induced by injection of carbomer and dexamethasone solution. Rabbits were anesthetized after intravenous administration of sumianxin (0.1 ml/kg; consists of dihydroetorphine hydrochloride, dimethylaniline thiazole, ethylenediam-inetetraacetic acid and haloperidol) (23) and topical application of 0.5% dicaine (24). Firstly, the corneal limbus of the paranasal side was punctured with a 1-ml syringe needle to form a pinhole. Then, 1 ml of aqueous humor was extracted from the corneal limbus on the temporal side, followed by an injection of 0.2 ml of carbomer solution containing 0.3% carbomer and 0.025% dexamethasone (Sigma-Aldrich; Merck KGaA). After 24 h, IOP was checked and a value of IOP >22 mmHg was used to determine the successful establishment of the glaucoma model (success ratio was 85% in the present study).

RFA treatment for glaucoma

RFA surgery was performed based on an XL-1-type RF exposure system developed by the Chinese PLA General Hospital (with independent intellectual property rights), which can produce ~300–500-µm wide, 1–1.5-mm deep tissue ablation. After local anesthesia (0.5% dicaine, one drop) (24) combined with general anesthesia (sumianxin mixture, 0.1 ml/kg) (23), 5 model eyes were randomly selected, and the sclera was punctured at 1.5 mm from the corneal margin with a depth of 1 mm. This was exposed to a probe that generated a 500-KHz radiofrequency field at specific absorption rate of 0.5–0.8 W/kg. Output power was set at 0.6–1.0 W, duration was 1–2 sec and the number of ablation sites was 8.

IOP measurement

Under local anesthesia, IOP was measured before and on days 1, 3 and 7 post-RFA with a Schiotz tonometer (Rudolf Riester GmbH). Scalp acupuncture connected to the Schiotz tonometer was inserted into the anterior chamber through the margin between the sclera and the cornea. After 5 min of stabilization, the value was read three times and the average value was calculated. The value obtained before ablation surgery was considered as the basal level of IOP.

Investigation of changes to the eyeball and anterior segments with UBM and slit lamp

Investigation with the slit lamp was carried out before surgery and on days 1, 3 and 7 post-surgery. Before investigation, rabbits were anesthetized with 3.5% sodium phenobarbital (35 mg/kg) (25). After the corneal reflex and corneal touch reflection disappeared, the transparency of the cornea, shades of the anterior chamber, the oscillation or opacity of the aqueous humor, and damages to the iris and lens were checked. Subsequently, our self-designed lens-water bath transformation system (patent no. ZL201420207350.0) was utilized to investigate the rabbit eyes with UBM for the purpose of assessing the influence of ablation on the ciliary body. During the UBM test, 0.2 ml 0.1% (w/v) hyaluronic acid sodium drops were administered to the junctival sac with the external optic cup utilized.

Histopathological assay after RFA

On 1, 3, 7 and 14 days after RFA, 2 rabbits on each day were anesthetized with an intravenous injection of 3.5% sodium phenobarbital (35 mg/kg) and sacrificed by ear vein air embolism (10–20 ml air). Following this, the eyeball was removed and 0.1 ml 4% formaldehyde was injected into the vitreous body through the site of the optic nerve. The eye was then soaked in 4% formaldehyde at 25°C for 24 h. The cornea, lens and posterior segment were removed from the fixed eyes. Then, the ciliary body and the sclera were collected, embedded in paraffin and cut into 5-µm sections. They were stained with hematoxylin for 5 min and eosin for 3 min, respectively, to observe the pathological changes in the retinal ganglion cells using light microscopy (magnification, ×400).

Statistical analysis

SPSS 17.0 software (SPSS, Inc.) was used for data analysis. Results of IOP were expressed as mean ± SD. Comparisons of IOP among three groups were performed using one-way ANOVA, followed by LSD post hoc test for multiple comparisons. P<0.05 was considered to be statistically significant.

Results

General observation after RFA

After radiofrequency catheter ablation of the ciliary body, the eyes were examined with UBM and a slit lamp. A melting spot was observed (Fig. 1), but no puncture point penetrating the ciliary body into the eye was shown in the UBM check. Slit-lamp examination results showed no significant anterior chamber reaction.
Figure 1.

Melting spot investigation of the ciliary body in glaucoma model rabbits. Melting spots were observed at (A) 1, (B) 3, (C) 7 and (D) 14 days after radiofrequency ablation.

Changes in IOP after RFA

In comparison with the sham group, rabbits of the glaucoma model showed ~2.5× higher IOP (~33 mmHg). At 1 day after RFA of the ciliary body, IOP was still higher than in the sham group (~15%), but 3 days post-RFA, IOP was decreased by ~40%, and significantly lower than the model group (P=0.047). At day 7, IOP of the RFA group continued to decrease (~25%), although it was not completely restored to the level of the sham group. However, it was significantly lower compared with the model group (P=3.64×10−4; Fig. 2).
Figure 2.

Changes in IOP in the RFA, sham and model groups at 1, 3 and 7 days post-RFA of the ciliary body in glaucoma model rabbits. *P<0.05 vs. sham group; #P<0.05 vs. model group. IOP, intraocular pressure; RFA, radiofrequency ablation.

UBM investigation of anterior segment after RFA

Stable UBM images were obtained by utilizing the objective-bath conversion device coupled with UBM that we developed. Based on this method, the changes of the anterior segments at different time points after ablation were investigated. Fig. 3A shows the anterior segments of the sham group, in which the corneal, anterior chamber angle, iris and the ciliary body could be seen clearly. Fig. 3B-D are days 1, 3 and 7 post-ablation, respectively. At 1 day post-RFA, a straight RFA pathway was displayed in the ciliary body, damages within the ciliary body can be seen, but no penetration of the ciliary body was observed (Fig. 3B). At day 3, ablation-caused damage had decreased, peripheral damage was repaired and the ablation pathway showed a rounded rectangular shape (Fig. 3C). At day 7 post-RFA, the area of damages continued to shrink, displaying an oval shape (Fig. 3D).
Figure 3.

Ultrasound biomicroscopy investigation of the anterior segment after RFA. (A) Anterior segment of the sham group. (B) 1, (C) 3 and (D) 7 days post-RFA. RFA, radiofrequency ablation. Red dotted lines signify the channel of radiofrequency ablation probe into ciliary body at different time points.

H&E staining and pathological changes after RFA

At 1, 3, 7 and 14 days post-RFA of the ciliary body, pathological changes were assessed in H&E stained slices with a light microscope. The sham group displayed a normal chamber angle, the muscle layers of the ciliary body were arranged densely and regularly, and the epithelium was intact (Fig. 4A). At day 1 post-RFA, the sclera fiber was fractured at the site of the ablation pathway and the endothelium of the ciliary body had become disordered, but the ciliary body was not penetrated (Fig. 4B). At 7 days post-RFA, the layer boundaries of the ciliary body had become clear, muscle layer fibers on the ablation pathway were arranged loosely and the scar had healed (Fig. 4C). At 14 days post-RFA, the scleral ablation pathway had almost healed, with only a few cells infiltrated, the ciliary body had healed and the epithelium was integrated (Fig. 4D).
Figure 4.

Hematoxylin and eosin staining and pathological changes after RFA of the ciliary body in rabbits. Images shown at (A) 1, (B) 3, (C) 7 and (D) 14 days post-RFA. RFA, radiofrequency ablation. The white arrows show the channel of radiofrequency ablation probe into ciliary body at different time points. Magnification, ×100.

Discussion

RFA was first introduced by Harvey Cushing and WT Bovie for the clinical treatment of neurological diseases, followed by Huang in 1987 (26) who applied it in the treatment of arrhythmia disease (27). In the 21st century, RFA technologies began to be widely utilized in minimally invasive surgery for the treatment of various diseases (18). Based on the characteristics of glaucoma and the structure of eyes, the aim of the present study was to find out whether RFA technology could also be applicable in the treatment of glaucoma, which has not been reported previously. In the current study, a New Zealand white rabbit model of glaucoma was used to examine the potential of RFA of the ciliary body for the treatment of glaucoma. As expected, the results of the present study showed that region-specific ablation of the ciliary body could reduce the production of aqueous humor, which is the cause of increased ocular pressure, and gradually restore IOP to a normal level over a long period of time. Radiofrequency (RF) techniques have been used to heat biological tissues for a number of years. Laser ablation (LA) is also a thermal ablation technique (28). However, their heating mechanisms are different. In RFA, resistive heating is generated by the agitation of ions, due to the alternating directions of the electric field, while in LA, hyperthermia is produced by increased light absorption and heat conduction (29). To achieve a reduction in IOP, the output power required is 1.5–2 W for 1.5–2 sec, and the temperature of the tissue is set between 70°C and 110°C in LA (30), which seem to be higher than those used for RFA in the present study (output power, 0.6 W; duration, 0.6 sec; temperature, 55–75°C). Thus, LA surgery could induce an inflammatory reaction and lead to several related complications, such as cystoid macular edema, hyphema and persistent hypotony (31). In addition, RFA is completed by moving the electrode under the guide of UBM and thereby precisely acting on the target lesion regions (32), and thermometers can be incorporated into the tips of the electrodes that allow for the continuous monitoring of tissue temperatures (33). These both prevent underlying damage to the adjacent structures. Furthermore, the fiber used in the LA may be difficult to place into the target sites, so the treatment efficiency and safety may be lower than RFA. Therefore, RFA may be more suitable for the treatment of glaucoma. The technique of minimally invasive RFA demonstrated here has a high potential for application in the clinical treatment of glaucoma (34,35). First of all, the rabbit's eye has the power to proliferate and self-repair (36), while its size is comparable to the human eye. The body size of the rabbit is small, which facilitates the use of the slit lamp microscope. These advantages make the rabbit model of glaucoma more suited to study common clinical issues than other animal models, and consequently it is widely used for the preclinical study of glaucoma (37). This is the reason that the ocular hypertension model in rabbits was chosen for the present study. Secondly, invasive RFA technology can not only control IOP precisely, reliably, safely and durably, but it is also easy to operate, it is a repeatable therapy, and it is relatively less expensive, resulting in a high socio-economic value. Since standard wands were used, some scleral damage occurred after ablation. This damage could be seen in the images obtained using ultrasound biomicroscopy, UBM anterior segment imaging and pathological imaging. To this end, a new type of RF wand was designed for ablation of the ciliary body (patent pending). This new wand is loaded with a specially designed insulating sheath that can more effectively protect the tissue around the sclera puncture track, making ablation more restricted to the needle tip of the wand or handle and thus, the ciliary processes (Fig. 5).
Figure 5.

Proposed radiofrequency ablation of the ciliary body using our self-designed ciliary ablation handle.

However, there are some limitations to the present study. Firstly, this was only an animal model study and further clinical studies with short or long-term follow up are also needed to confirm the effectiveness and safety of RFA. Risks may include cataract development, cognitive effects and possible carcinogenic effects due to the frequency used, which is within the range of 30 kHz to 300 GHz. Secondly, a direct comparison between RFA and LA for glaucoma is lacking. The results of the present study indicated that XL-1 meter-based RFA of the ciliary body for controlling ocular pressure in patients with glaucoma is mostly safe and effective, showing only a few anterior chamber reactions and almost no influence on intraocular structures. In further studies, the new handle for the ciliary body ablation, the dose-effect relationship of the ciliary body ablation in the treatment of glaucoma and the application of RFA of the ciliary body for treatment of absolute glaucoma in clinical settings will be investigated.
  33 in total

1.  Epicardial radiofrequency ablation of ventricular myocardium: factors affecting lesion formation and damage to adjacent structures.

Authors:  Guilherme Fenelon; Kleber Ponzi Pereira; Angelo A V de Paola
Journal:  J Interv Card Electrophysiol       Date:  2006-01       Impact factor: 1.900

Review 2.  Endoscopic cyclophotocoagulation for the treatment of glaucoma.

Authors:  Amir Cohen; Sze H Wong; Shriji Patel; James C Tsai
Journal:  Surv Ophthalmol       Date:  2016-10-05       Impact factor: 6.048

3.  Vitrectomy for injury: the effect on intraocular proliferation following perforation of the posterior segment of the rabbit eye.

Authors:  G W Abrams; T M Topping; R Machemer
Journal:  Arch Ophthalmol       Date:  1979-04

Review 4.  Glaucoma as a neurodegenerative disease.

Authors:  Neeru Gupta; Yeni H Yücel
Journal:  Curr Opin Ophthalmol       Date:  2007-03       Impact factor: 3.761

5.  Elevated intraocular pressure decreases response sensitivity of inner retinal neurons in experimental glaucoma mice.

Authors:  Ji-Jie Pang; Benjamin J Frankfort; Ronald L Gross; Samuel M Wu
Journal:  Proc Natl Acad Sci U S A       Date:  2015-02-09       Impact factor: 11.205

6.  The number of people with glaucoma worldwide in 2010 and 2020.

Authors:  H A Quigley; A T Broman
Journal:  Br J Ophthalmol       Date:  2006-03       Impact factor: 4.638

Review 7.  [Glaucoma today: detection and therapeutic progress].

Authors:  Christophe Baudouin; Alexandre Denoyer; William Rostène
Journal:  Biol Aujourdhui       Date:  2013-10-10

8.  Conductive keratoplasty: an approach for the correction of residual hyperopia in post-lasik pseudophakia.

Authors:  Pan-Pan Ye; Wen Xu; He-Sheng Xu; Zhao-Chun Li; Jun-Ting Shi; Feng-Ying He; Ke Yao
Journal:  Int J Ophthalmol       Date:  2012-10-18       Impact factor: 1.779

9.  A combined analysis of four observational studies evaluating the intraocular pressure-lowering ability and tolerability of bimatoprost 0.01% in patients with primary open-angle glaucoma or ocular hypertension.

Authors:  Annemie Stevens; Milko E Iliev; Leo de Jong; Ioana Grobeiu; Anton Hommer
Journal:  Clin Ophthalmol       Date:  2016-04-06

Review 10.  Intracranial Pressure and Its Relationship to Glaucoma: Current Understanding and Future Directions.

Authors:  Uttio Roy Chowdhury; Michael P Fautsch
Journal:  Med Hypothesis Discov Innov Ophthalmol       Date:  2015