COMMENTARY

The polyaxial pedicle screw (PAPS) system is widely used for posterior spinal stabilization. Unlike monoaxial screws, the PAPS has the advantage of moving along several different axes, allowing the surgeon some flexibility in their placement. Nevertheless, pedicle screws are associated with their own unique surgical complications.[1,4] Here, we report a case in which the head of the PAPS accidentally separated/dislocated during posterior fixation for a lumbar burst fracture.

CASE SUMMARY

A 21-year-old male, neurologically intact following a motor vehicle accident, had a L3 burst fracture as demonstrated on the emergent spinal computed tomography-scan [Figure 1a]. He underwent a posterior surgical reduction, decompression, and stabilization using PAPS [Figure 1b]. However, during the right L4 PAPS tightening, the head was suddenly detached from the screw shank [Figure 2]. This caused the surgeon to directly injure the lumbar spine with the screwdriver; fortunately, this did not result in any adverse sequelae. The screw was removed and a second spare screw was finally replaced. The operation ended successfully and the postoperative recovery was uneventful. He was discharged on the 10th postoperative day.
Figure 1:

A burst fracture of L3 as seen on sagittal reconstruction computed tomography scan (bone windows) before operation (a). Postoperative lumbar spinal X-rays showing the emplacement of polyaxial pedicle screws (on L1, L2, L4, and L5), decompressive posterior laminectomy (on L3 and L2) with bilateral connecting rod screws (b).

Figure 2:

Polyaxial pedicle screws before use. These multiaxial screws shave 360° head polar mobility (a). Picture of the dislocated polyaxial screw: The screw head was detached from screw shank (b).

A burst fracture of L3 as seen on sagittal reconstruction computed tomography scan (bone windows) before operation (a). Postoperative lumbar spinal X-rays showing the emplacement of polyaxial pedicle screws (on L1, L2, L4, and L5), decompressive posterior laminectomy (on L3 and L2) with bilateral connecting rod screws (b). Polyaxial pedicle screws before use. These multiaxial screws shave 360° head polar mobility (a). Picture of the dislocated polyaxial screw: The screw head was detached from screw shank (b).

IMPLANT RETURNED TO MANUFACTURER

The implant was sent to the manufacturer for their examination; they found “witness marks” and “deformation” noted on the inner diameter surface below the retaining ring pocket of the head; this suggested the ring may not have been fully seated in the ring screw pocket [Figure 3]. The manufacturer’s conclusion was consistent with a manufacturing error, resulting in the foregoing event.
Figure 3:

The technical analysis of the manufacturer has found witness marks and deformation noted on the inner diameter surface below retaining ring pocket of the head screw (visual review [a] and microscopic review [b]).

The technical analysis of the manufacturer has found witness marks and deformation noted on the inner diameter surface below retaining ring pocket of the head screw (visual review [a] and microscopic review [b]). This case emphasizes the importance of proper checking of the integrity of the surgical implant before its use. In addition, industry has to guarantee the quality of their product. Here, this unpredictable incident was recognized immediately without serious consequences for the patient. One similar case was previously reported; however, there was no clear manufacturer defect.[3] Device component failure should always be considered as a potential risk associated with any spinal system, which may induce neurological/spinal complications and injuries to vessels, nerves, and other organs. Furthermore, incidents and accidents with medical devices must be reported to the National Medicines Agency of each country and the manufacturers to prevent future possible complications.[2,5]
  5 in total

1.  Quality improvement report: Learning from adverse incidents involving medical devices.

Authors:  John Amoore; Paula Ingram
Journal:  BMJ       Date:  2002-08-03

2.  The surgeon and medical devices: adverse incident reporting and off-label use.

Authors:  Nicola Lennard; Melissa Coutinho; Bruce Campbell
Journal:  Ann R Coll Surg Engl       Date:  2013-07       Impact factor: 1.891

3.  Traumatic dislocation of the S1 polyaxial pedicle screw head: a case report.

Authors:  Pieter N B Du Plessis; Bernard P H Lau; Hwee Weng Dennis Hey
Journal:  J Spine Surg       Date:  2017-03

Review 4.  Clinically relevant complications related to pedicle screw placement in thoracolumbar surgery and their management: a literature review of 35,630 pedicle screws.

Authors:  Oliver P Gautschi; Bawarjan Schatlo; Karl Schaller; Enrico Tessitore
Journal:  Neurosurg Focus       Date:  2011-10       Impact factor: 4.047

5.  Pedicle screw malposition following spinal lumbar injury.

Authors:  Ali Akhaddar; El Mehdi Atmane
Journal:  Pan Afr Med J       Date:  2014-04-12