Optimization of HU threshold for coronary artery calcium scans reconstructed at 0.5-mm slice thickness using iterative reconstruction.

Kuei-Yuan Hou1, Katsumi Tsujioka2, Ching-Ching Yang3.   

Abstract

PURPOSE: This work investigated the simultaneous influence of tube voltage, tube current, body size, and HU threshold on calcium scoring reconstructed at 0.5-mm slice thickness using iterative reconstruction (IR) through multivariate analysis. Regression results were used to optimize the HU threshold to calibrate the resulting Agatston scores to be consistent with those obtained from the conventional protocol.
METHODS: A thorax phantom set simulating three different body sizes was used in this study. A total of 14 coronary artery calcium (CAC) protocols were studied, including 1 conventional protocol reconstructed at 3-mm slice thickness, 1 FBP protocol, and 12 statistical IR protocols (3 kVp values*4 SD values) reconstructed at 0.5-mm slice thickness. Three HU thresholds were applied for calcium identification, including 130, 150, and 170 HU. A multiple linear regression method was used to analyze the impact of kVp, SD, body size, and HU threshold on the Agatston scores of three calcification densities for IR-reconstructed CAC scans acquired with 0.5-mm slice thickness.
RESULTS: Each regression relationship has R2 larger than 0.80, indicating a good fit to the data. Based on the regression models, the HU thresholds as a function of SD estimated to ensure the quantification accuracy of calcium scores for 120-, 100-, and 80-kVp CAC scans reconstructed at 0.5-mm slice thickness using IR for three different body sizes were proposed. Our results indicate that the HU threshold should be adjusted according to the imaging condition, whereas a 130-HU threshold is appropriate for 120-kVp CAC scans acquired with SD = 55 for body size of 24.5 cm.
CONCLUSION: The optimized HU thresholds were proposed for CAC scans reconstructed at 0.5-mm slice thickness using IR. Our study results may provide a potential strategy to improve the reliability of calcium scoring by reducing partial volume effect while keeping radiation dose as low as reasonably achievable.
© 2019 The Authors. Journal of Applied Clinical Medical Physics published by Wiley Periodicals, Inc. on behalf of American Association of Physicists in Medicine.

Entities:  

Keywords:  Agatston score; coronary artery calcification; iterative reconstruction; protocol optimization

Year:  2019        PMID: 31889419      PMCID: PMC7021007          DOI: 10.1002/acm2.12806

Source DB:  PubMed          Journal:  J Appl Clin Med Phys        ISSN: 1526-9914            Impact factor:   2.102


INTRODUCTION

Coronary artery disease (CAD) is one of the leading causes of death around the world. Meanwhile, atherosclerosis is the most common form of vascular disease and constitutes the major cause of death, with 17.5 million global deaths annually.1 The hallmark lesion in atherosclerosis is the atherosclerotic plaque, whereas rupture‐prone atherosclerotic plaques, that is, vulnerable plaques, may cause life‐threatening events like acute coronary syndrome or stroke. In general, a vulnerable plaque is often found to be associated with a thin fibrous cap, a high inflammation burden, a large lipid pool, macroscopic heterogeneity, and so on. Spotty calcification is a morphological characteristic of a vulnerable plaque phenotype.2 Coronary artery calcium (CAC) examination performed on computed tomography (CT) is widely considered as a reliable technique for screening risk of future cardiac events.3 The reference protocol for CAC acquisition has a tube voltage of 120 kVp, and scan data are reconstructed at 3‐mm slice thickness with filtered back‐projection (FBP). The threshold for the determination of coronary calcification is 130 Hounsfield unit (HU). The amount of calcium is typically quantified with the Agatston score, which is a function of the CAC area and the maximal CAC density.4, 5, 6 The total Agatston score is the sum of each CAC lesion for all coronary arteries extending through the z‐axis of the heart.7 A reduced slice thickness reduces partial volume artifacts, possibly increasing reproducibility of CAC scoring and decreasing overestimation of fully calcified plaques. But this would make the score more susceptible to noise. van der Werf et al. reported that high image noise would increase coronary artery scores.8 According to their multivendor phantom study, Agatston score increased by 1–8% at 40% reduced dose and 23–64% at 80% reduced dose. On the other hand, Willemink et al.9 and Li et al.10 reported that high image noise would decrease coronary artery scores. Depending on the vendor, this underestimation can be up to 50%. This study aimed to improve the reliability of calcium scoring by using 0.5‐mm slice thickness to reduce partial volume effect while keeping radiation dose as low as reasonably achievable (ALARA). CT radiation dose reduction is frequently achieved by applying lower kVp11, 12, 13 and tube current modulation (TCM).14, 15, 16 Iterative reconstruction (IR) methods are developed to reduce image noise by performing a recursive search for the best estimate, so they can provide a similar image quality at higher noise levels compared with FBP. Consequently, applying IR techniques is considered to be one of the strategies for reducing radiation dose in CT.17, 18, 19, 20 However, these dose reduction strategies would affect the image performance of CT in terms of contrast, spatial resolution, and noise pattern, and thus may impact the CAC scores. Therefore, this work first investigated the influence of kVp, mA, and HU threshold on calcium scoring for CAC scans reconstructed at 0.5‐mm slice thickness using IR through anthropomorphic phantom studies. Based on these study results, the optimized HU thresholds were then proposed for CAC scans under various imaging conditions to calibrate the resulting Agatston scores to be consistent with those obtained from reference protocol.

METHODS

Anthropomorphic chest phantom

A thorax phantom (Thorax‐CCI, QRM GmbH, Möhrendorf, Germany) which consists of an anthropomorphic phantom body and a calibration insert was used for CAC scans (Fig. 1). The anthropomorphic phantom body contains artificial lungs and a spine insert surrounded by tissue equivalent material. The outer dimensions of the phantom body are 300 × 200 mm2 in transverse plane and 150 mm in height (QRMsmall). Two extension rings were added to the phantom body to enlarge the phantom size in transverse plane to 350 × 250 mm2 (QRMmedium) and 400 × 300 mm2 (QRMlarge). The calibration insert contains nine cylindrical calcifications of three sizes (1, 3, and 5 mm in diameter and height) and three calcium hydroxyapatite (HA) densities (200, 400, and 800 mg/cm3). In addition, the calibration insert contains two large homogeneous inserts made of water equivalent material and HA with density of 200 mg/cm3. A cardiac simulator (Cardiac Trigger, Model: CTM300, Ivy Biomedical Systems, Branford, CT, USA) was used to synthesize ECG signal at mean heart rate of 60 bpm.
Figure 1

Illustration of the anthropomorphic chest phantom (HA: calcium hydroxyapatite; HU: Hounsfield unit).

Illustration of the anthropomorphic chest phantom (HA: calcium hydroxyapatite; HU: Hounsfield unit).

CAC SCANS

All CAC scans were performed on a 320‐detector row CT system (Aquilion One, Canon, Otawara, Japan) using a 0.35 seconds of gantry rotation time and half‐scan reconstruction. In our department, the routine CAC scans are acquired using TCM (SureExposure 3D, Canon, Otawara, Japan) at 120 kVp with 3‐mm slice thickness and reconstructed by FBP with soft tissue kernel (FC12). The TCM parameter value used to specify the desired level of image noise is the SD value, which is the standard deviation of HU in the central region of an image of a homogeneous water phantom. Besides the SD value, the TCM system allows the operator to define the mA range (minimum‐to‐maximum mA) within which the tube current can be modulated. The TCM system is operated with SD = 55 and mA range 40–300 mA in our routine clinical practice. A total of 14 CAC protocols of various acquisition/reconstruction combinations were studied, including 2 protocols with FBP reconstruction and 12 protocols with IR reconstruction. Table 1 demonstrates the scan parameters and the corresponding volume CT dose index (CTDIvol) displayed on the scanner console of the 14 CAC protocols investigated in this study. For the FBP group, CAC scans were acquired using two different slice thickness (3, 0.5 mm) and TCM with SD = 55. The tube voltage applied in the FBP group was 120 kVp. As for the IR group, CAC scans were acquired using three different tube voltages (120, 100, and 80 kVp) and TCM with four different SD values (55, 75, 115, and 155). The slice thickness applied in the IR group was 0.5 mm, and the reconstruction technique was Adaptive Iterative Dose Reduction 3D (AIDR 3D) at the standard level with FC12 kernel. All CAC scans employed a 22‐cm display field of view.
Table 1

The scan parameters, radiation dose and CT number within background ROI (mean ± standard deviation) of the phantom studies.

Reconstruction algorithm & slice thickness

Tube voltage

&

TCM parameters

QRMsmall QRMmedium QRMlarge
Tube current (mA)

CTDIvol

(mGy)

CT number

(HU)

Tube current (mA)

CTDIvol

(mGy)

CT number

(HU)

Tube current (mA)

CTDIvol

(mGy)

CT number

(HU)

FBP, 3 mm120 kVp, SD = 55 (40–300 mA)*401.141.43 ± 27.52501.440.11 ± 41.471504.240.60 ± 40.69
FBP, 0.5 mm120 kVp, SD = 55 (40‐580 mA)*601.741.86 ± 49.171805.040.16 ± 47.0452015.640.58 ± 45.52
AIDR, 0.5 mm120 kVp, SD = 55 (40–580 mA)*601.745.51 ± 25.321704.845.27 ± 24.2651015.348.00 ± 23.97
120 kVp, SD = 75 (40–580 mA)*401.140.60 ± 28.331002.841.64 ± 26.662908.740.69 ± 28.13
120 kVp, SD = 115 (40–580 mA)*401.141.11 ± 27.51451.341.22 ± 32.921303.639.25 ± 34.58
120 kVp, SD = 155 (40–580 mA)*401.141.70 ± 27.61401.140.28 ± 36.38702.039.10 ± 35.47
100 kVp, SD = 55 (40–580 mA)*1001.734.66 ± 24.333003.634.73 ± 23.5858010.638.61 ± 25.38
100 kVp, SD = 75 (40–580 mA)*601.029.66 ± 26.481702.930.54 ± 25.545009.231.87 ± 26.27
100 kVp, SD = 115 (40–580 mA)*400.729.48 ± 30.03801.431.39 ± 30.042303.930.65 ± 30.87
100 kVp, SD = 155 (40–580 mA)*400.729.61 ± 27.55450.830.28 ± 35.731302.229.53 ± 30.63
80 kVp, SD = 55 (40–580 mA)*2302.012.97 ± 26.075805.415.95 ± 29.125805.422.58 ± 38.22
80 kVp, SD = 75 (40–580 mA)*1401.29.79 ± 32.494404.111.88 ± 30.455805.413.60 ± 39.35
80 kVp, SD = 115 (40–580 mA)*700.66.97 ± 38.392101.810.73 ± 37.215805.416.33 ± 36.92
80 kVp, SD = 155 (40–580 mA)*400.38.61 ± 38.861301.113.75 ± 41.063803.313.32 ± 41.04
The scan parameters, radiation dose and CT number within background ROI (mean ± standard deviation) of the phantom studies. Tube voltage & TCM parameters CTDIvol (mGy) CT number (HU) CTDIvol (mGy) CT number (HU) CTDIvol (mGy) CT number (HU)

IMAGE QUALITY ASSESSMENT

The effects of statistical fluctuation were assessed in both low‐ and high‐contrast objects. For low‐contrast object, a background region of interest (ROI) of 35 by 35 pixels was placed at the center of the large homogeneous insert that is made of water equivalent material to calculate the mean and standard deviation of HU within background ROI. As for high‐contrast object, histograms of 5‐mm calcifications with HA density of 200, 400, and 800 mg/cm3 were generated and fitted by a Gaussian distribution to investigate the intensity distribution of calcifications.21 Spatial resolution was quantified by determining the modulation transfer function (MTF) calculated using the edge technique.22 The edge spread function was determined by a square ROI placed over the boundaries (1) at soft tissue/lung interface and (2) between the large homogeneous insert with HA density of 200 mg/cm3 and the surrounding soft tissue. These image quality assessments were performed in MATLAB 7.1 (The Mathworks, Natick, MA, USA).

CAC SCORING

Evaluating calcium score was performed on the postprocessing workstation (Vitrea FX, Vital Images, Minnetonka, USA), using calcium score analysis software (VScore, Vital Images, Minnetonka, USA). The Agatston score was calculated for each calcified lesion by multiplying the area (Ai) of the lesion by a weighting factor (wi), determined by the density within the insert: For the reference protocol, the minimum HU threshold to identify calcium is predefined at 130 HU. A weighting factor of 1 is applied if the maximum CT value is between 130 and 199 HU, a weighting factor of 2 for 200 to 299 HU, a weighting factor of 3 for 300 to 399 HU, and a weighting factor of 4 for values above 400 HU:where CTmax is the maximum CT number in the lesion. Calcified lesions are excluded from the Agatston score if they have an axial area less than 1 mm2 or a CT number less than the predefined CT number threshold, that is, 130 HU at 120 kVp. It has been reported that lower kVp settings would increase the CT number of calcifications, and thus may result in overestimation of Agatston score.23, 24 In this study, the CAC scans were acquired not at 120 kVp but also at 100 and 80 kVp. Hence, three different HU thresholds were applied to detect calcification for each CAC scan listed in Table 1, including 130, 150, and 170 HU.

MULTIVARIATE ANALYSIS

A multiple linear regression method was used to analyze the impact of (1) tube voltage, (2) SD value, (3) body size, and (4) HU threshold on Agatston score (AS) obtained from CAC scans reconstructed at 0.5‐mm slice thickness using IR. The model to explain the dependence relationship was defined as:where B0 to B4 are the regression coefficient (Bi) to be estimated, and body size is the square root of the product of long and short axis of chest phantom, that is, the effective diameter. The coefficient of determination (R2) was calculated to assess the overall strength of the functional regression model. The standard regression coefficient (βi) was calculated to assess the relative importance of each predictor (Xi), that is, βi is standardized Bi. The student’s t ratio (ti), which is the ratio of Bi and its corresponding standard error, was used to evaluate the significance of each predictor. A predictor was considered statistically significant if P < 0.01. A variance inflation factor (VIFi) is computed for each predictor, using the following formula: VIFi = 1/(1‐Ri 2), where Ri 2 is obtained from the multiple regression in which Xi is predicted from all the other independent variables. A maximal VIFi value in excess of 10 was regarded as an indication that multicollinearity may be unduly influencing the least square estimates. The statistical analysis algorithms were implemented in MATLAB 7.1 (The Mathworks, Natick, MA, USA).

RESULTS

Image quality assessment

Table 1 summarizes the CT numbers within background ROI of the 14 CAC protocols of various acquisition/reconstruction combinations. It was found that reducing slice thickness from 3 mm to 0.5 mm in FBP reconstruction results in fourfold increased dose in QRMmedium and QRMlarge to achieve the same noise level (SD = 55). Due to the mA range in TCM, the CAC scan for QRMsmall reconstructed at 3‐mm slice thickness using FBP was overexposed, so the image noise from 3‐mm FBP reconstruction is smaller than that from 0.5‐mm FBP reconstruction. Consequently, the CTDIvol was increased by only 1.55‐fold when changing slice thickness from 3 mm to 0.5 mm in QRMsmall. For CAC scans acquired with 0.5‐mm slice thickness, all IR‐reconstructed images have lower image noise than FBP‐reconstructed images. Figure 2 demonstrates the CT images acquired with four different acquisition/reconstruction combinations for QRMsmall, QRMmedium, and QRMlarge. It was found that the 1‐mm calcifications cannot be detected in some of the IR protocols (high SD value or large body size), so the total Agatston score for multivariate analysis was the sum over 3‐mm and 5‐mm calcifications. Figure 3. shows the histograms of calcifications with HA density of 200, 400, and 800 mg/cm3 in CAC scans acquired with three different acquisition/reconstruction combinations for QRMmedium. A relative flattening of the calcification histogram was observed in CAC scans with higher image noise. Figure 4 compares the MTF values from 120‐kVp CAC scans acquired with four different acquisition/reconstruction combinations for QRMsmall, QRMmedium, and QRMlarge, including (a) FBP, 0.5 mm, 120 kVp, SD = 55; (b) AIDR, 0.5 mm, 120 kVp, SD = 55; (c) AIDR, 0.5 mm, 120 kVp, SD = 75; and (d) AIDR, 0.5 mm, 120 kVp, SD = 115. It was found that increasing image noise or body size may degrade the MTF.
Figure 2

CT images acquired with four different acquisition/reconstruction combinations for QRMsmall (top row), QRMmedium (mid row), and QRMlarge (bottom row): (a) FBP, 0.5 mm, 120 kVp, SD = 55; (b) AIDR, 0.5 mm, 120 kVp, SD = 55; (c) AIDR, 0.5 mm, 120 kVp, SD = 75; and (d) AIDR, 0.5 mm, 120 kVp, SD = 115

Figure 3

Histograms of calcifications with HA density of 200, 400, and 800 mg/cm3 (left to right) in CT scans acquired with three different acquisition/reconstruction combinations for QRMmedium: (a) FBP, 0.5 mm, 120 kVp, SD = 55; (b) AIDR, 0.5 mm, 120 kVp, SD = 75; and (C) AIDR, 0.5 mm, 120 kVp, SD = 115.

Figure 4

Comparison of MTF values from 120‐kVp CAC scans acquired with four different acquisition/reconstruction combinations for QRMsmall (left), QRMmedium (mid), and QRMlarge (right). ROI was placed over the boundaries (a) at soft tissue/lung interface and (b) between the large homogeneous insert with HA density of 200 mg/cm3 and the surrounding soft tissue.

CT images acquired with four different acquisition/reconstruction combinations for QRMsmall (top row), QRMmedium (mid row), and QRMlarge (bottom row): (a) FBP, 0.5 mm, 120 kVp, SD = 55; (b) AIDR, 0.5 mm, 120 kVp, SD = 55; (c) AIDR, 0.5 mm, 120 kVp, SD = 75; and (d) AIDR, 0.5 mm, 120 kVp, SD = 115 Histograms of calcifications with HA density of 200, 400, and 800 mg/cm3 (left to right) in CT scans acquired with three different acquisition/reconstruction combinations for QRMmedium: (a) FBP, 0.5 mm, 120 kVp, SD = 55; (b) AIDR, 0.5 mm, 120 kVp, SD = 75; and (C) AIDR, 0.5 mm, 120 kVp, SD = 115. Comparison of MTF values from 120‐kVp CAC scans acquired with four different acquisition/reconstruction combinations for QRMsmall (left), QRMmedium (mid), and QRMlarge (right). ROI was placed over the boundaries (a) at soft tissue/lung interface and (b) between the large homogeneous insert with HA density of 200 mg/cm3 and the surrounding soft tissue. The box and whisker plots shown in Fig. 5 summarize the impact of tube voltage, SD, body size, and HU threshold on the total Agatston scores from 108 IR measurements (3 kVp values*4 SD values*3 phantom sizes*3 HU thresholds) for calcifications with HA density of 200, 400, and 800 mg/cm3. The red line in each box represents the median of the distribution, whereas the top and bottom of each box represent the 25th and 75th percentile of the distribution, respectively. The whiskers extend to the 99.3% confidence interval (±2.7 sigma). The asterisks are extreme outliers beyond the whiskers. For calcifications with different HA densities, similar tendency was found in [Fig. 5(a) and (d)]. but not in [Figs 5(b) or 5(c)].
Figure 5

Box and whisker diagrams for Agatston score of calcification with HA density of 200, 400, and 800 mg/cm3 with respective to (a) tube voltage, (b) SD, (c) body size, and (d) HU threshold.

Box and whisker diagrams for Agatston score of calcification with HA density of 200, 400, and 800 mg/cm3 with respective to (a) tube voltage, (b) SD, (c) body size, and (d) HU threshold. The results of regression analysis of Eq. (3) for the Agatston score of calcifications with HA density of 200, 400, and 800 mg/cm3 are shown in Table 2. It was found that all independent variables are statistically significant predictors of Agatston score, except the SD value and body size for AS400HA. For AS200HA, 1/kVp is the most significant predictor (β = 0.72), followed by body size (β = −0.43), 1/threshold (β = 0.41), and SD value (β = −0.16). The corresponding results for AS400HA are 1/threshold (β = 0.79), 1/kVp (β = 0.47), SD value (β = 0.04), and body size (β = −0.02). As for AS800HA, the corresponding results are 1/threshold (β = 0.69), SD value (β = 0.40), body size (β = 0.33), and 1/kVp (β = 0.25). High multicollinearity was not observed among independent variables in the three AS models (VIF < 10). The regression equations that expresses the relationship between Agatston score and the predictors for calcifications with HA density of 200, 400, and 800 mg/cm3 are as follows:
Table 2

Statistical analysis results of the regression model for Agatston score of calcification with HA density of 200, 400, and 800 mg/cm3.

 Bβt (p)VIF
200 HA400 HA800 HA200 HA400 HA800 HA200 HA400 HA800 HA200 HA400 HA800 HA
1/kVp13200.565683.613895.720.720.470.25

23.16*

(P < 0.01)

12.45*

(P < 0.01)

5.92*

(P < 0.01)

1.001.001.00
SD−0.130.020.28−0.160.040.40

−5.12*

(P < 0.01)

1.15

(P = 0.23)

9.45*

(P < 0.01)

1.001.001.00
Body size−3.28−0.082.08−0.43−0.020.33

−13.93*

(P < 0.01)

−0.40

(P = 0.68)

7.63*

(P < 0.01)

1.001.001.00
1/threshold17457.6722171.0624606.010.410.790.69

13.26*

(P < 0.01)

21.02*

(P < 0.01)

16.17*

(P < 0.01)

1.001.001.00
Statistical analysis results of the regression model for Agatston score of calcification with HA density of 200, 400, and 800 mg/cm3. 23.16* (P < 0.01) 12.45* (P < 0.01) 5.92* (P < 0.01) −5.12* (P < 0.01) 1.15 (P = 0.23) 9.45* (P < 0.01) −13.93* (P < 0.01) −0.40 (P = 0.68) 7.63* (P < 0.01) 13.26* (P < 0.01) 21.02* (P < 0.01) 16.17* (P < 0.01) The regression model in Eq. (4), (5), (6) yielded an R2 of 0.9005, 0.8531, and 0.8117, respectively. Figure 6 shows the HU thresholds as a function of SD estimated based on the regression models in Eq. (4), (5), (6) for 120‐, 100‐, and 80‐kVp CAC scans reconstructed at 0.5‐mm slice thickness using IR for three different phantom sizes to reach the same Agatston score obtained from routine CAC scans. These results indicate that the HU threshold should be adjusted according to the imaging condition, whereas a 130‐HU threshold is appropriate for 120‐kVp CAC scans acquired with SD = 55 for QRMsmall.
Figure 6

The HU thresholds as a function of SD estimated based on the regression models for (a) 120‐kVp, (b) 100‐kVp, and (c) 80‐kVp CAC scans reconstructed at 0.5‐mm slice thickness using AIDR 3D for three phantom sizes to reach the same Agatston score obtained from routine CAC scans.

The HU thresholds as a function of SD estimated based on the regression models for (a) 120‐kVp, (b) 100‐kVp, and (c) 80‐kVp CAC scans reconstructed at 0.5‐mm slice thickness using AIDR 3D for three phantom sizes to reach the same Agatston score obtained from routine CAC scans.

DISCUSSION

IR reconstruction in CAC scoring

Decades of prognostic data support the value of Agatston score for clinical risk predication. However, Agatston score uses 3‐mm slice thickness and a 130‐HU threshold for detecting calcium in FBP reconstruction. Implementation of the Agatston approach as the gold standard has, in some ways, slows the development and evaluation of alternative, optimized scan parameter.5 A reduced slice thickness may improve the accuracy in evaluating the actual presence of coronary calcium by decreasing the partial volume effects. The primary benefit would be the early detection of merging microcalcification, and a potential increase in total number of calcified lesions detected. However, thin‐slice reconstruction could be associated with higher image noise, which may bias the detection of calcification. IR allows substantial noise reduction in low‐dose CT, but it has been reported that IR techniques may underestimate CAC scores, particularly at low score ranges. Gephard et al.25 demonstrated a reduction in Agatston score when using the Adaptive Statistical Iterative Reconstruction (ASiR; GE Healthcare, Waukesha, USA) algorithm, and the reduction range is associated with the percentage of ASiR, ranged from 6.0% to 22.4% for 0–100% ASiR in Agatston score. Kurata et al. demonstrated a systematic decrease between 8% and 48% in Agatston score when using 10–50% of Sinogram‐affirmed Iterative Reconstruction (SAFIRE; Siemens Medical Solutions, Malvern, USA) blending.26 On the other hand, Rodrigues et al.27 and Blobel et al.23 showed good agreement in calcium scores between FBP and AIDR 3D protocols. Consistency in Agatston score between FBP and AIDR 3D was observed in our results acquired with 120 kVp and SD = 55 for QRMsmall, but not in those acquired with low‐dose protocols or larger phantoms. Overall, these studies indicate that CAC scores differ considerably by acquisition and reconstruction methods.

RELATIONSHIP BETWEEN IMAGE QUALITY AND AGATSTON SCORES

As seen in Fig. 5, the Agatston scores from 108 IR measurements increased as decreasing kVp, which can be explained by the increase in CT number of calcifications at lower tube voltage.23, 24 This phenomenon was observed in all calcification densities and was most obvious in the Agatston score of calcifications with HA density of 200 mg/cm3. With regards to the impact of SD value and body size, it was found that low tube current or high patient attenuation would lead to the decrease in calcium scores for less dense calcifications (eg, 200 mg/cm3), but this trend was reversed in dense calcifications (eg, 800 mg/cm3). According to Fig. 2 and 4, the spatial resolution of CAC scans was degraded by increased SD value or body size, which may be due to the increase in statistical fluctuation and scattered radiation. The degradation of spatial resolution would lead to the increase in partial volume artifacts, thus increasing the number of calcification pixels but with decreased intensity. For calcifications with HA density of 200 mg/cm3, the decrease in intensity due to partial volume effect may lead to CT numbers less than 130 HU in boundary pixels (Fig. 3), and truncation of these pixels would lower the Agatston score. As for calcifications with HA density of 800 mg/cm3, the CT numbers of boundary pixels were also decreased when increasing SD value or body size, but should still be larger than 400 HU (Fig. 3). Besides, the number of calcification pixels was increased due to partial volume effect, thus increasing the Agatston score. Regarding calcifications with HA density of 400 mg/cm3, it is suspected that the influences of low intensity (small wi) and more pixels (large Ai) owing to partial volume effect on Agatston score cancelled each other out, so changing SD value or body size did not cause obvious variation in AS400HA. By definition, reducing CT number threshold would increase the Agatston score, and vice versa.23, 24 This phenomenon was also observed in our results, which was adapted to determine the optimal HU threshold for CAC scans undergoing various imaging conditions to ensure the quantification accuracy of calcium scoring from CAC scans reconstructed at 0.5‐mm slice thickness using IR relative to the conventional protocol.

OPTIMIZED HU THRESHOLDS BASED ON MULTIVARIATE ANALYSIS

In this work, a multiple linear regression method was used to analyze how the tube voltage, SD value, body size, and HU threshold affect the Agatston scores of 3 calcification densities for IR‐reconstructed CAC scans acquired with 0.5‐mm slice thickness simultaneously. Each regression relationship has R2 larger than 0.80, indicating a good fit to the data. Based on the regression models, the HU thresholds as a function of SD estimated to ensure the quantification accuracy of calcium scores for 120‐, 100‐, and 80‐kVp CAC scans reconstructed at 0.5‐mm slice thickness using IR for patients with body size of 24.50 cm (QRMsmall), 29.58 cm (QRMmedium), and 34.64 cm (QRMlarge) were proposed (Fig. 6). Our results demonstrate that a 130‐HU threshold is appropriate for 120‐kVp CAC scans acquired with low SD values, performed on small‐sized patients. However, for calcifications with HA density of 200 mg/cm3, a lower HU threshold should be applied for 120‐kVp CAC scans acquired with higher SD value to prevent underestimation of Agatston score, especially for large‐sized patients. On the other hand, regarding calcifications with HA density of 800 mg/cm3, a higher HU threshold should be used to avoid overestimation of Agatston score. Once dose reduction is achieved by reducing tube voltage, a higher HU threshold should be applied to prevent overestimation of Agatston score, especially for small‐sized patients with less dense calcifications.

LIMITATIONS

Several limitations to this study need to be acknowledged. First, only a single manufacturer's CT system was investigated, so the proposed CAC scans and the corresponding HU threshold cannot be applied to CT systems from different manufacturers. However, the workflow to optimize data acquisition and quantification in this study may be used for AEC systems which modulate tube current according to a predefined image quality index, for example, SD in Canon and noise index (NI) in GE. Second, because it is not possible to evaluate the same individual with different body sizes, this work concerns a phantom study. It is possible that results differ under more extreme conditions, for example, denser calcification or larger body size. Additional studies assessing the optimization workflow for clinical patients on different CT systems will be needed and valuable.

CONCLUSION

Reduced slice thickness may contribute to the early detection of emerging microcalcification and a potential increase in the total number of calcified lesions detection. But there is an inherent tradeoff between slice thickness and image noise. Hence, this work evaluated the simultaneous influence of kVp, SD, body size, and HU threshold on calcium scoring for CAC scans that were reconstructed at 0.5‐mm slice thickness using IR through multivariate analysis. Regression results were used to optimize the HU threshold under various imaging conditions to calibrate the resulting Agatston scores to be consistent with those obtained from reference protocol. Our study results may provide a potential strategy to improve the reliability of calcium scoring by reducing partial volume effect while achieving the ALARA principle.

CONFLICTS OF INTEREST

The authors have no relevant conflicts of interest to disclose.
  27 in total

1.  Prediction of coronary events with electron beam computed tomography.

Authors:  Y Arad; L A Spadaro; K Goodman; D Newstein; A D Guerci
Journal:  J Am Coll Cardiol       Date:  2000-10       Impact factor: 24.094

2.  Development and application of programs to measure modulation transfer function, noise power spectrum and detective quantum efficiency.

Authors:  R Padgett; C J Kotre
Journal:  Radiat Prot Dosimetry       Date:  2006-02-03       Impact factor: 0.972

3.  Coronary calcium scores are systematically underestimated at a large chest size: A multivendor phantom study.

Authors:  Martin J Willemink; Bronislaw Abramiuc; Annemarie M den Harder; Niels R van der Werf; Pim A de Jong; Ricardo P J Budde; Joachim E Wildberger; Rozemarijn Vliegenthart; Tineke P Willems; Marcel J W Greuter; Tim Leiner
Journal:  J Cardiovasc Comput Tomogr       Date:  2015-04-02

4.  Effects of patient size on radiation dose reduction and image quality in low-kVp CT pulmonary angiography performed with reduced IV contrast dose.

Authors:  Aaron Sodickson; Michael Weiss
Journal:  Emerg Radiol       Date:  2012-04-24

Review 5.  Coronary artery calcification screening: estimated radiation dose and cancer risk.

Authors:  Kwang Pyo Kim; Andrew J Einstein; Amy Berrington de González
Journal:  Arch Intern Med       Date:  2009-07-13

6.  Diagnostic accuracy of coronary CT angiography: comparison of filtered back projection and iterative reconstruction with different strengths.

Authors:  Rui Wang; U Joseph Schoepf; Runze Wu; John W Nance; Biao Lv; Hua Yang; Fang Li; Dongxu Lu; Zhaoqi Zhang
Journal:  J Comput Assist Tomogr       Date:  2014 Mar-Apr       Impact factor: 1.826

7.  Contrast agent and radiation dose reduction in abdominal CT by a combination of low tube voltage and advanced image reconstruction algorithms.

Authors:  Nico Buls; Gert Van Gompel; Toon Van Cauteren; Koenraad Nieboer; Inneke Willekens; Guy Verfaillie; Paul Evans; Sven Macholl; Ben Newton; Johan de Mey
Journal:  Eur Radiol       Date:  2014-11-29       Impact factor: 5.315

Review 8.  Low radiation dose in computed tomography: the role of iodine.

Authors:  Andrik J Aschoff; Carlo Catalano; Miles A Kirchin; Martin Krix; Thomas Albrecht
Journal:  Br J Radiol       Date:  2017-05-04       Impact factor: 3.039

Review 9.  Coronary Atherosclerotic Vulnerable Plaque: Current Perspectives.

Authors:  Christodoulos Stefanadis; Christos-Konstantinos Antoniou; Dimitrios Tsiachris; Panagiota Pietri
Journal:  J Am Heart Assoc       Date:  2017-03-17       Impact factor: 5.501

10.  Size-based protocol optimization using automatic tube current modulation and automatic kV selection in computed tomography.

Authors:  Robert D MacDougall; Patricia L Kleinman; Michael J Callahan
Journal:  J Appl Clin Med Phys       Date:  2016-01-08       Impact factor: 2.102