| Literature DB >> 31888556 |
Jae Ho Jung1, Je Hyun Seo2, Min Seung Kang3, Jonghoon Shin4,5,6.
Abstract
BACKGROUND: To compare the glaucoma diagnostic ability of the ganglion cell-inner plexiform layer (GCIPL) thickness depending on the range around the fovea using wide-angle, swept-source optical coherence tomography (SS-OCT).Entities:
Keywords: Ganglion cell-inner plexiform layer; Glaucoma; Wide-angle swept-source OCT
Mesh:
Year: 2019 PMID: 31888556 PMCID: PMC6937720 DOI: 10.1186/s12886-019-1283-y
Source DB: PubMed Journal: BMC Ophthalmol ISSN: 1471-2415 Impact factor: 2.209
Fig. 1a A wide-angle scan with swept-sources optic coherence tomography (SS-OCT; DRI-OCT-1 Atlantis; Topcon, Tokyo, Japan) demonstrating the ganglion cell-inner plexiform layer (GCIPL) thickness for each 1-mm2 grid square of a 12 mm × 9 mm (horizontal × vertical) scan; and these 108 data were collected (green rectangles). We made zone classifications based on the scanned area (centered on the fovea) and the arcuate configuration of the papillomacular bundle. b Zone 1 (narrow area); the widest horizontal × vertical scanned length was 4 × 4 mm, with a total of 12 grid squares and 3 squares per quadrant. c Zone 2 (mid-sized area); the widest horizontal × vertical scanned length was 6 × 6 mm, with a total of 24 squares with 6 per quadrant included. d Zone 3 (wide area); the widest horizontal X vertical scanned length was 8 × 8 mm, and a total of 40 squares with 10 squares per quadrant were included. e Conventional area; 6 × 6 mm annulus area centered on the fovea
Demographic characteristics in the present study
| Parameters | Normal control | Glaucoma group | |
|---|---|---|---|
| Age | 54.22 ± 11.91 | 54.82 ± 10.06 | 0.766 a |
| Sex (Male/Female) | 33/26 | 30/31 | 0.471 b |
| Central corneal thickness, μm | 533.18 ± 41.15 | 525.12 ± 35.05 | 0.514 a |
| Spherical equivalent | - 0.92 ± 2.37 | - 0.49 ± 2.18 | 0.313 a |
| Axial length | 23.93 ± 1.06 | 23.97 ± 1.15 | 0.856 a |
| Average RNFL thickness, μm | 103.90 ± 1.81 | 66.58 ± 3.21 | < 0.001 a |
| Humphrey 24–2 visual field | |||
| MD, dB | - 0.42 ± 0.95 | - 8.14 ± 5.01 | < 0.001 c |
| PSD, dB | 1.56 ± 0.45 | 8.61 ± 4.11 | < 0.001 c |
| VFI, % | 99.18 ± 1.09 | 78.98 ± 16.15 | < 0.001 c |
RNFL Retinal nerve fiber layer, MD mean deviation, PSD pattern standard deviation, VFI visual field index
a Student t-test, b Fisher’s exact test, c Mann-Whitney U test
Comparison of GCIPL thickness between normal an glaucoma using SS-OCT
| Parameters (μm) | Normal | Glaucoma | |
|---|---|---|---|
| In conventional 6 × 6 mm | |||
| Superonasal sector | 72.76 ± 7.33 | 64.90 ± 10.09 | < 0.001 |
| Superior sector | 67.79 ± 6.89 | 59.00 ± 9.31 | < 0.001 |
| Superotemporal sector | 72.42 ± 4.91 | 62.67 ± 8.13 | < 0.001 |
| Inferonasal sector | 69.88 ± 7.47 | 60.78 ± 9.53 | < 0.001 |
| Inferior sector | 64.45 ± 5.84 | 54.32 ± 7.13 | < 0.001 |
| Inferotemporal sector | 71.37 ± 5.20 | 58.98 ± 6.53 | < 0.001 |
| Average | 69.78 ± 5.14 | 60.11 ± 7.07 | < 0.001 |
| Minimum | 63.69 ± 6.10 | 51.90 ± 7.09 | < 0.001 |
| In Zone 1 | |||
| Superonasal sector | 84.18 ± 5.86 | 73.34 ± 13.49 | < 0.001 |
| Superotemporal sector | 78.51 ± 5.25 | 68.03 ± 11.47 | < 0.001 |
| Inferonasal sector | 79.20 ± 6.36 | 64.81 ± 11.04 | < 0.001 |
| Inferotemporal sector | 80.36 ± 5.78 | 61.90 ± 9.83 | < 0.001 |
| Average | 80.56 ± 5.23 | 67.02 ± 9.87 | < 0.001 |
| Minimum | 76.61 ± 5.07 | 59.15 ± 9.70 | < 0.001 |
| In Zone 2 | |||
| Superonasal sector | 68.83 ± 5.02 | 60.89 ± 9.29 | < 0.001 |
| Superotemporal sector | 70.56 ± 4.81 | 61.90 ± 8.01 | < 0.001 |
| Inferonasal sector | 60.42 ± 5.26 | 52.12 ± 6.35 | < 0.001 |
| Inferotemporal sector | 65.18 ± 5.21 | 55.31 ± 4.86 | < 0.001 |
| Average | 66.25 ± 4.35 | 57.56 ± 5.96 | < 0.001 |
| Minimum | 60.22 ± 5.11 | 51.23 ± 6.25 | < 0.001 |
| In Zone 3 | |||
| Superonasal sector | 59.58 ± 4.67 | 52.86 ± 7.04 | < 0.001 |
| Superoteomporal sector | 62.98 ± 4.23 | 56.55 ± 6.28 | < 0.001 |
| Inferonasal sector | 52.11 ± 4.56 | 46.29 ± 5.22 | < 0.001 |
| Inferotemporal sector | 56.78 ± 4.51 | 51.48 ± 4.32 | < 0.001 |
| Average | 57.86 ± 3.82 | 51.79 ± 4.66 | < 0.001 |
| Minimum | 51.95 ± 4.40 | 45.66 ± 5.17 | < 0.001 |
a Student t-test
Univariate correlation of GCIPL thickness in between automated 6 × 6 mm and zone 2
| Parameters | Normal | Glaucoma | Total |
|---|---|---|---|
| Superior sector | 0.805* | 0.886* | 0.901* |
| Inferior sector | 0.746* | 0.879* | 0.899* |
| Average | 0.834* | 0.953* | 0.948* |
| Minimum | 0.580* | 0.791* | 0.836* |
* P < 0.001 by Pearson correlation analysis
Bland-Altman Analysis of GCIPL measurements with each parameter between automated 6 × 6 mm and zone 2
| 95% LoA in Normal | 95% LoA in Glaucoma | |||||
|---|---|---|---|---|---|---|
| Lower | Upper | Width | Lower | Upper | Width | |
| Superior sector | - 7.8 | 5.2 | 13.0 | −8.9 | 7.4 | 16.3 |
| Inferior sector | - 12.7 | 1.2 | 13.9 | - 11.0 | 2.4 | 13.4 |
| Average | - 9.1 | 2.0 | 11.1 | - 6.9 | 1.8 | 8.7 |
| Minimum | - 14.0 | 7.1 | 21.1 | - 8.2 | 6.9 | 15.1 |
Area under the receiver operating characteristic curves of GCIPL parameters between normal and glaucoma
| Parameters | Mean (SD) |
|---|---|
| In conventional 6 × 6 mm | |
| Superonasal sector | 0.737 (0.047) |
| Superior sector | 0.788 (0.043) |
| Superotemporal sector | 0.850 (0.035) |
| Inferonasal sector | 0.801 (0.041) |
| Inferior sector | 0.870 (0.027) |
| Inferotemporal sector | 0.920 (0.027) |
| Average | 0.872 (0.033) |
| Minimum | 0.908 (0.029) |
| In Zone 1 | |
| Superonasal sector | 0.764 (0.044) |
| Superotemporal sector | 0.778 (0.043) |
| Inferonasal sector | 0.871 (0.032) |
| Inferotemporal sector | 0.931 (0.024) |
| Average | 0.894 (0.028) |
| Minimum | 0.945 (0.019) |
| In Zone 2 | |
| Superonasal sector | 0.770 (0.044) |
| Superotemporal sector | 0.822 (0.039) |
| Inferonasal sector | 0.839 (0.036) |
| Inferotemporal sector | 0.918 (0.024) |
| Average | 0.879 (0.031) |
| Minimum | 0.865 (0.032) |
| In Zone 3 | |
| Superonasal sector | 0.777 (0.042) |
| Superotemporal sector | 0.817 (0.039) |
| Inferonasal sector | 0.800 (0.041) |
| Inferotemporal sector | 0.809 (0.040) |
| Average | 0.840 (0.036) |
| Minimum | 0.821 (0.039) |
The parameter of minimum and inferotemporal sector in conventional 6 × 6 mm and zone 1, and inferotemporal sector measurement in zone 2 which have occupied an area of 0.9 or more have shown higher diagnostic value than other GCIPL parameters
Fig. 2Receiver operating characteristic (ROC) curves for the minimum ganglion cell-inner plexiform layer (GCIPL) measurements of zone 1 and conventional area and the GCIPL thickness of the inferotemporal area in zone 1, zone 2, and the conventional area. These curves show that the areas under the receiver operating characteristic (AUROCs) curves were greater than 0.9
Pairwise comparison of area under the receiver operating characteristic of GCIPL thickness in glaucoma group conducted by DeLong et al.’s method
| Difference between areas | Inferotemporal sector of zone 1 | Minimum of Zone 1 |
|---|---|---|
| Inferotemporal sector of zone 2 | 0.011 | 0.024 |
| Average of zone 2 | 0.058* | 0.072* |
| Minimum of zone 2 | 0.023 | 0.036 |
| Inferotemporal sector of conventional 6 × 6 mm | 0.013 | 0.027 |
| Average of conventional 6 × 6 mm | 0.051* | 0.066* |
| Minimum of conventional 6 × 6 mm | 0.066* | 0.079* |
() P value by methods described in DeLong et al, * P value < 0.05