| Literature DB >> 31864332 |
Sook-Yee Chong1,2, Lu Shen3, Sandra Frantz4.
Abstract
BACKGROUND: The two-stage revision protocol represents the current gold standard for treating infected total knee replacement implants. Allowing early mobility with weight-bearing between staged procedures will enable early restoration to knee function. So, the mechanical performance of knee spacers is a key issue. Commercially available moulds are often used as they are easy to prepare and produce smoother surfaces of the articulating parts. However, they are costly, and only for single use. A cost-effective alternative is the surgeon-made hand-moulded spacers. In this study, we wanted to determine how the hand-moulded spacers will compare biomechanically with the commercially available COPAL spacers.Entities:
Keywords: Axial displacement; Dynamic knee spacers; Hand-mould spacers; Load; Stiffness
Mesh:
Year: 2019 PMID: 31864332 PMCID: PMC6925492 DOI: 10.1186/s12891-019-2982-5
Source DB: PubMed Journal: BMC Musculoskelet Disord ISSN: 1471-2474 Impact factor: 2.362
Fig. 1The femoral (a, b) and tibial components (c) of the knee spacer fabricated from COPAL knee moulds
Fig. 2X-ray of one specimen with COPAL dynamic spacer before testing
Fig. 3One knee specimen implanted with hand-moulded spacers
Fig. 4X-ray of one specimen with hand-moulded dynamic spacer before testing
Fig. 5One knee specimen mounted at 10 degrees flexion between the upper and lower grips of the MTS machine
Fig. 6First protocol where the knee specimen underwent a single axial loading test
Fig. 7Load-displacement curve with stiffness calculated from linear gradient (red line)
Fig. 8Axial force (top) and displacement (bottom) of 30-1200 N cyclic loading test of one specimen implanted with COPAL spacer. Axial displacement (d) at the 1000th cycle was determined. c is the maximum amount of deviation from the first cycle
Results of specimens that underwent the axial load-to-failure test, and equality of variance
| Mean ± SD | variance | |||
|---|---|---|---|---|
| COPAL ( | hand-moulded (n = 4) | p-value | ||
| Maximum Load (N) | 5202.0 ± 486.9 | 4509.7 ± 1092.6 | 0.32 | 0.36 |
| Stiffness (N/mm) | 1098.3 ± 201.5 | 1008.7 ± 275.4 | 0.64 | 0.56 |
Results of mean displacement d (in mm) of specimens underwent cyclic loading test, and equality of variance
| Mean ± SD | variance | |||
|---|---|---|---|---|
| Cyclic load | COPAL ( | hand-moulded ( | ||
| [30 400] | 0.36 ± 0.14 | 0.36 ± 0.20 | 0.99 | 0.56 |
| [30 600] | 0.63 ± 0.30 | 0.48 ± 0.19 | 0.42 | 0.33 |
| [30 800] | 0.83 ± 0.40 | 0.65 ± 0.30 | 0.48 | 0.67 |
| [30 1000] | 1.01 ± 0.44 | 0.84 ± 0.43 | 0.62 | 0.96 |
| [30 1200] | 1.19 ± 0.57 | 0.89 ± 0.30 | 0.36 | 0.26 |