Literature DB >> 31863740

Mindray SF-Cube technology: An effective way for correcting platelet count in individuals with EDTA dependent pseudo thrombocytopenia.

Jiankai Deng1, Yaoming Chen1, Shihong Zhang1, Laisheng Li1, Qiong Shi1, Min Liu2, Xuegao Yu3.   

Abstract

Several strategies are applied to determine the precise platelet count in individuals with ethylenediaminetetraacetic acid dependent pseudo thrombocytopenia (EDTA-PTCP) caused by in vitro aggregation of platelets in daily laboratory practice. None of them proves optimal for routine purposes. Thus, Mindray has developed the SF-Cube technology coupled with the CDR mode in the Mindray hematology analyzer to overcome the problem of EDTA-PTCP. With Mindray SF-Cube technology, platelet aggregates dissociate effectively and platelets are correctly counted in the CDR mode without pre-analytical management. In our studies, the EDTA-PTCP blood samples when analyzed with the CDR mode of Mindray BC-6800 plus analyzer, yield a markedly higher platelet count compared to those obtained with PLT-I on Sysmex XN-9000 hematology analyzer. We conclude that in patients with known or suspected EDTA-PTCP Mindray SF-Cube technology is a straightforward and effective way of determining the platelet count in EDTA-anticoagulated blood.
Copyright © 2019 Elsevier B.V. All rights reserved.

Entities:  

Keywords:  Automated hematology analyzer; EDTA; Mindray SF-Cube technology; Platelet aggregation; Pseudo thrombocytopenia

Mesh:

Substances:

Year:  2019        PMID: 31863740     DOI: 10.1016/j.cca.2019.12.012

Source DB:  PubMed          Journal:  Clin Chim Acta        ISSN: 0009-8981            Impact factor:   3.786


  3 in total

Review 1.  Pseudothrombocytopenia-A Review on Causes, Occurrence and Clinical Implications.

Authors:  Benjamin Lardinois; Julien Favresse; Bernard Chatelain; Giuseppe Lippi; François Mullier
Journal:  J Clin Med       Date:  2021-02-04       Impact factor: 4.241

2.  Performance evaluation of optical platelet counting of BC-6000Plus automated hematology analyzer.

Authors:  Xuehua Hu; Yuqun Tang; Mingyi Li; Chunyun Fu; Zuoxin Deng; Wenting Tang; Huideng Zhou; Yan Chen; Yingfeng Li
Journal:  Ann Transl Med       Date:  2021-03

3.  Performance of Platelet Counting in Thrombocytopenic Samples: Comparison between Mindray BC-6800Plus and Sysmex XN-9000.

Authors:  Hanah Kim; Mina Hur; Gun-Hyuk Lee; Seung-Wan Kim; Hee-Won Moon; Yeo-Min Yun
Journal:  Diagnostics (Basel)       Date:  2021-12-29
  3 in total