Comparison of image quality between synthetic and patients' electrocardiogram-gated 320-row pediatric cardiac computed tomography.

Eriko Maeda1, Go Shirota2, Eisuke Shibata3, Shuhei Komatsu4, Kenji Ino5, Rumiko Torigoe6, Osamu Abe3.   

Abstract

BACKGROUND: Pediatric cardiac computed tomography (CT) can be acquired without electrode placement by using synthetic electrocardiogram (ECG).
OBJECTIVE: To determine whether the depiction of gross cardiac structures and coronary arteries in 320-row pediatric CT is not inferior when CT is gated with synthetic ECG at 150 beats per minute (bpm), compared to the patients' own ECG.
MATERIALS AND METHODS: Sixty 320-row CT examinations performed in children younger than 3 years old with congenital cardiac anomaly were enrolled in this retrospective study. Thirty examinations were scanned using the children's own ECG for gating and 30 examinations were scanned using synthetic ECG at 150 bpm. The image quality was compared between the two gating modes using a 3-point scale to delineate the following anatomical structures: atrial septum, ventricular septum, right atrium, right ventricle, left atrium, left ventricle, main pulmonary artery, ascending aorta, aortic arch including the patent ductus arteriosus, descending aorta, right coronary artery and left main trunk. Beam-hardening artifacts from contrast enhancement material were evaluated using a 3-point scale, and the overall image quality was evaluated using a 5-point scale.
RESULTS: Synthetic ECG was not inferior to the patients' ECG in depicting each structure, beam-hardening artifact and overall image quality. Average indices were clinically acceptable imaging quality, except for subjective image quality of mid and distal coronary arteries.
CONCLUSION: Pediatric cardiac CT in patients younger than 3 years old can be acquired using synthetic ECG gating, with image quality not inferior to the patients' ECG.

Entities:  

Keywords:  320-row computed tomography; Children; Electrocardiogram gating; Image quality; Sedation; Synthetic electrocardiogram

Year:  2019        PMID: 31853572     DOI: 10.1007/s00247-019-04541-y

Source DB:  PubMed          Journal:  Pediatr Radiol        ISSN: 0301-0449


  16 in total

1.  Measurement of observer agreement.

Authors:  Harold L Kundel; Marcia Polansky
Journal:  Radiology       Date:  2003-06-20       Impact factor: 11.105

2.  Radiation dose from single-heartbeat coronary CT angiography performed with a 320-detector row volume scanner.

Authors:  Andrew J Einstein; Carl D Elliston; Andrew E Arai; Marcus Y Chen; Richard Mather; Gregory D N Pearson; Robert L Delapaz; Edward Nickoloff; Ajoy Dutta; David J Brenner
Journal:  Radiology       Date:  2010-03       Impact factor: 11.105

3.  Estimated risks of radiation-induced fatal cancer from pediatric CT.

Authors:  D Brenner; C Elliston; E Hall; W Berdon
Journal:  AJR Am J Roentgenol       Date:  2001-02       Impact factor: 3.959

4.  Synthetic ECG-gated cardiac computed tomography for in vivo imaging of the temporary total artificial heart.

Authors:  Michael A Winkler; Paul F von Herrmann; Michael A Brooks; Charles W Hoopes; Anil Attili; Vincent L Sorrell
Journal:  Circulation       Date:  2013-01-01       Impact factor: 29.690

5.  Initial experience on the application of 320-row CT angiography with low-dose prospective ECG-triggered in children with congenital heart disease.

Authors:  Tong Zhang; Wei Wang; Zhuren Luo; Dan Wang; Jinquan Bai; Dong Han; Baozhong Shen
Journal:  Int J Cardiovasc Imaging       Date:  2011-12-28       Impact factor: 2.357

6.  Use of 320-detector computed tomographic angiography for infants and young children with congenital heart disease.

Authors:  Faris Al-Mousily; Roger Y Shifrin; Frederick J Fricker; Nicholas Feranec; Nancy S Quinn; Arun Chandran
Journal:  Pediatr Cardiol       Date:  2011-01-06       Impact factor: 1.655

7.  Pediatric 64-MDCT coronary angiography with ECG-modulated tube current: radiation dose and cancer risk.

Authors:  Bingsheng Huang; Martin Wai-Ming Law; Henry Ka-Fung Mak; Stephen Ping-Fai Kwok; Pek-Lan Khong
Journal:  AJR Am J Roentgenol       Date:  2009-08       Impact factor: 3.959

8.  Contrast medium application in pediatric high-pitch cardiovascular CT angiography: manual or power injection?

Authors:  Marc Saake; Michael M Lell; Oliver Rompel; Martin Gloeckler; Matthias May; Achim Eller; Stephan Achenbach; Michael Uder; Wolfgang Wuest
Journal:  J Cardiovasc Comput Tomogr       Date:  2014-05-22

9.  Free-breathing 320-row computed tomographic angiography with low-tube voltage and hybrid iterative reconstruction in infants with complex congenital heart disease.

Authors:  Yuzo Yamasaki; Satoshi Kawanami; Takeshi Kamitani; Koji Sagiyama; Seitaro Shin; Takuya Hino; Kenichiro Yamamura; Hidetake Yabuuchi; Michinobu Nagao; Hiroshi Honda
Journal:  Clin Imaging       Date:  2018-02-15       Impact factor: 1.605

10.  Diagnostic Phase of Calcium Scoring Scan Applied as the Center of Acquisition Window of Coronary Computed Tomography Angiography Improves Image Quality in Minimal Acquisition Window Scan (Target CTA Mode) Using the Second Generation 320-Row CT.

Authors:  Eriko Maeda; Kodai Yamamoto; Shigeaki Kanno; Kenji Ino; Nobuo Tomizawa; Masaaki Akahane; Rumiko Torigoe; Kuni Ohtomo
Journal:  ScientificWorldJournal       Date:  2016-02-10
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.