Cardiovascular implantable electronic device lead removal in a resource-constrained setting: A single-center experience from India.

Sakshi Parasrampuria1, Sirish Chandra Srinath Patloori2, Reka Karuppusami3, David Chase1, John Roshan1.   

Abstract

BACKGROUND: Data from large-volume centers in developed countries, using dedicated tools, show a high success rate with a good safety record for the percutaneous lead removal procedure. However, there are constraints to replicate the results in a resource-poor setting and there is limited data from India.
METHODS: We retrospectively analyzed lead removal procedures performed in our institution from 2008 to 2019.
RESULTS: Seventy-five patients underwent percutaneous removal of 138 leads. Of these, 44 procedures and 80 leads qualified as extraction with a median dwell time of 52.1 (IQR 28.2-117.2) months. Overall, 33/44 (75.0%) procedures were successful and 65/80 (81.2%) leads were successfully extracted. Manual traction was successful in the extraction of 44/57 (77.2%) leads. All leads implanted less than 2.7 years could be removed with manual traction alone. Specialized tools were used in 23 leads and 21 (91.3%) of those could be successfully extracted. Inability to use dedicated tools was an independent predictor of procedural failure (adjusted OR 14.0; 95% CI 1.8-110.2; p-value 0.012). Right-sided implant (adjusted OR 12.6; 95% CI 1.3-119.5; p-value 0.027) was also independently associated with failure. There was 1 death (1.3%) and minor complications occurred in 6 (8.0%) patients.
CONCLUSIONS: In a resource-limited setting, percutaneous lead extraction of predominantly pacemaker leads by manual traction methods achieved success in extracting about three-fourths of the leads. Inability to use specialized tools was the main factor limiting success. The complication rate was low.
Copyright © 2019 Indian Heart Rhythm Society. Production and hosting by Elsevier B.V. All rights reserved.

Entities:  

Keywords:  Cardiovascular implantable electronic devices; Lead explantation; Lead extraction; Percutaneous lead removal

Year:  2019        PMID: 31843556      PMCID: PMC6994293          DOI: 10.1016/j.ipej.2019.12.002

Source DB:  PubMed          Journal:  Indian Pacing Electrophysiol J        ISSN: 0972-6292


Introduction

The growing number of cardiovascular implantable electronic device (CIED) implantations has seen a concomitant increase in the indications for their removal. Percutaneous lead extraction, as opposed to surgical removal, has become the standard of care for the management of infected or dysfunctional CIED systems. In developed countries, large-volume centers with specialized lead extraction programs have evolved. Data from these centers, using dedicated tools, show a high success rate with a good safety record for the procedure [1,2]. Despite all the technical advances, percutaneous lead extraction remains a challenging and risky procedure. Data about lead management strategies from resource-constrained settings is limited. Given the prohibitive cost associated with these dedicated tools, reserving them for selected cases may be reasonable in countries like India. We report a single-center experience of non-laser assisted percutaneous lead removal procedures.

Methods

Study population

We retrospectively analyzed 75 lead removal procedures performed in our institution from 2008 to 2019. Patient data, device details, and follow-up data were collected from electronic medical records. Procedure details were collected from the detailed procedural notes written by the operator.

Lead removal procedure

All patients had pre-procedure blood group typing and cross-matching done. When feasible, antiplatelet drugs and anticoagulants were stopped before the procedure. In cases where they could not be stopped safely, antiplatelets were continued through the procedure. Anticoagulants were stopped 2–3 days before the procedure and bridged with heparin or the procedure was done once the international normalized ratio (INR) was less than 2.0 without using heparin. Pacing-dependent patients who were not expected to receive immediate reimplantation were placed on a transjugular ventricular active-fixation lead connected to an external pulse generator. The choice to operate in the operating room (OR) or electrophysiology (EP) lab was at the discretion of the explanting physician. Regardless of the place of the procedure, thoracic-surgeons were always available on standby. Local anesthesia was used for leads with short dwell time while general anesthesia was used to extract longer dwelling leads. A stepwise approach was used as described previously [[3], [4], [5], [6], [7]]. To summarize, after unscrewing and separating the device from the leads, manual lead traction with the aid of an implantation stylet was performed. If traction from the pocket resulted in a lead fracture or in case of a preexisting free-floating lead, a snare typically used for endovascular procedures (Goose Neck™ snare, Amplatz, Plymouth, Minnesota, USA) was inserted through a 6F catheter via the femoral route and lead retrieval was attempted. This snare, though not specifically designed for lead extraction, was used as an extension of manual traction for removal of retained lead fragments. When available and not constrained by cost, the following types of dedicated extraction tools were used in various combinations according to the choice of the operator - (a) lead locking stylets (Liberator ®, Cook Medical, Bloomington, Indiana, USA, or LLD®, Spectranetics (a Philips company, San Diego, California, USA), (b) snaring device (Needle’s Eye Snare®, Cook Medical), (c) telescoping sheaths (SlightRail™, Spectranetics), and (d) rotating mechanical dilator sheaths (Evolution®, Cook Medical, or TightRail™, Spectranetics). Several procedural caveats apply while performing the extraction with dedicated tools. A skin incision directly over the site of lead’s entrance into the vascular system and dissecting as close as possible to that site helped transmit the coaxial traction. The LLD was passed till the lead tip or to distal-most point possible to achieve effective traction. When manual traction with the lead stylet or LLD did not help, more advanced tools were used. It was made sure that the traction on the LLD was coaxial and controlled so as to ensure that the LLD did not retract. Continuous repeated application of the rotational cutting sheath mechanism was seldom required and waiting between applications while delivering forward counter pressure was more effective.

Reimplantation

When infection was the indication for CIED removal and there was no evidence of bacteremia, reimplantation was at the discretion of the physician depending on the adequacy of wound healing and urgency of reimplantation. In the case of endocarditis, at least 2 weeks of antibiotic therapy was administered and the blood cultures were negative before reimplantation. In infected cases, implantation of a new CIED system was always advised. However, if the patient could not afford a new device and the old device had adequate battery life, it was reused. The leads were never reused. Explanted pulse generators were screened for any significant external damage and interrogated. Devices were washed with a detergent solution to remove particulate debris. They were packed as a double barrier package using commercially available sterilization pouches (Tyvek®, DuPont, Midland, Michigan, USA) and sterilized at least twice with either ethylene oxide (prior to 2017; EO-FCT, Andersen sterilizers, Haw River, North Carolina, USA) or low temperature hydrogen peroxide plasma sterilization (from 2017; STERRAD®, ASP (a Johnson & Johnson company),Irvine, California, USA). The last sterilization was done not more than 24 h before reimplantation. A separate consent explicitly stating the implantation of a refurbished device and the potentially higher risks, was taken.

Definitions

The terms related to lead removal have been defined previously in the expert consensus statements on lead extraction [[7], [8], [9]]. In brief, a procedure was called ‘lead explantation’ when all the removed leads were implanted for less than one year and no special tools (other than implantation stylets) were used. A procedure was defined as ‘lead extraction’ when at least one lead was implanted for more than one year or the procedure required the assistance of specialized equipment designed for lead removal. ‘Complete success’ refers to the removal of all targeted leads and lead material without permanently disabling complications/procedure-related death. ‘Clinical success’ was the removal of all targeted leads with retention of a small portion of the lead (<4 cm) that does not negatively impact the outcome goals. The procedure was considered a ‘failure’ when neither complete nor clinical success could be achieved or a permanently disabling complication/procedure-related death ensued. In addition, for the purpose of this study, we defined ‘manual lead removal’ as removal of leads with traction or using snares not typically designed for lead extraction.

Statistical analysis

Continuous variables were expressed as mean ± standard deviation and in case of skewed distribution, median with interquartile range (IQR) was used. Categorical variables were expressed as frequency and percentage. Means were compared with Student’s t-test and categorical variables were compared with the Chi-square test. Non-parametric tests (Mann-Whitney U test and Fisher’s exact test) were used when a variable had skewed distribution. Multivariable analysis was done using binary logistic regression by selecting variables with p-value < 0.200 on univariate analysis or if the variable was thought to be clinically relevant. Penalized logistic regression was used when the odds ratio could not be estimated due to empty cells or cells with low frequency, because of failure of the maximum likelihood estimate to converge. The point estimates were reported as odds ratio (OR) and 95% confidence interval (CI). A p-value of <0.05 was considered statistically significant. Statistical analysis was done with SPSS® software (Ver. 21.0, IBM, USA) and SAS® (Ver. 9.2, SAS Institute, USA).

Results

Patient characteristics

A total of 86 patients required CIED removal during the study period. Eleven patients underwent direct surgical removal due to other indications for surgery or the treating physician considered percutaneous removal very risky. Most of the surgical referrals were before the ready availability of extraction tools. The remaining 75 patients underwent percutaneous removal of 138 leads (average 1.8 leads per procedure). Majority of the leads (63.0%) were implanted elsewhere and referred to our institution for extraction. The most common reason for device removal was infection in 68/75 (90.7%) patients. The baseline patient characteristics are summarized in Table .1 and the infection characteristics in Table 2. Most infections were confined to the pocket with only 4/68 (5.9%) having endocarditis. Cultures were negative in 29/68 (42.6%) of the cases. Coagulase-negative Staphylococcus aureus (CONS) was the most common organism (22.0%), followed by Pseudomonas aeruginosa (10.3%), Staphylococcus aureus (7.4%), other gram-negative bacilli (7.4%), atypical mycobacteria (2.9%), Candida species (2.9%), and others (4.4%).
Table 1

Baseline characteristics.

Number of patients75
Number of leads138
Age, yrs62.6 ± 13.4
Male55 (73.3)
Device type
 Pacemaker66 (88.0)
 ICD3 (4.0)
 CRT - P3 (4.0)
 CRT - D3 (4.0)
Indication for implantation
 Sinus node dysfunction21 (28.0)
 AV block42 (56.0)
 LV dysfunction/heart failure5 (6.7)
 Others7 (9.3)
Comorbid conditions
 Diabetes24 (32.0)
 Hypertension36 (48.0)
 Ischemic heart disease9 (12.0)
 Chronic kidney disease5 (6.7)
 Left ventricular dysfunction10 (13.3)
 Prosthetic valves2 (2.7)
 Prior CABG2 (2.7)
Medications
 Any antithrombotic drugs18 (24.0)
 Antiplatelets17 (22.7)
 Anticoagulants3 (4.0)
Last procedure
 Initial implant60 (80.0)
 Device change/revision/upgrade15 (20.0)
Reason for device removal
 Infection/erosion68 (90.7)
 Lead dysfunction6 (8.0)
 Device upgrade1 (1.3)

Note: Values are mean ± standard deviation or frequency (%).

Table 2

Infection characteristics (n = 68).

Type of infection
 Pocket infection/erosion64 (94.1)
 Infective endocarditis4 (5.9)
Organisms
 Coagulase negative staphylococcus aureus15 (22.0)
 Staphylococcus aureus5 (7.4)
 Enterococcus sps.2 (2.9)
 Gram-negative enteric bacilli5 (7.4)
 Pseudomonas aeruginosa7 (10.3)
 Atypical mycobacteria2 (2.9)
 Candida sps.2 (2.9)
 Others1 (1.5)
Culture-negative29 (42.6)

Note: Values are frequency (%).

Baseline characteristics. Note: Values are mean ± standard deviation or frequency (%). Infection characteristics (n = 68). Note: Values are frequency (%).

Lead and procedure characteristics

Removal of 138 leads was attempted. Six of the 138 (4.3%) were ICD leads and 68 (49.3%) were passive fixation leads. Seven leads (5.1%) were cut during earlier attempts at extraction before referral to our institution. The median dwell time was 22.1 (IQR 4.5–61.3) months. Eighty leads and 44 procedures qualified as extraction. Seventy of the 75 procedures (93.3%) were performed in the EP lab. The lead and procedure characteristics are summarized in Table .3.
Table 3

Lead and procedure characteristics.

Total leads removed (n = 138)Explantation (n = 58)Extraction (n = 80)p- valuea
Lead type
 Atrial pacing leads50 (36.2)23 (39.7)27 (33.7)
 Ventricular pacing leads75 (54.3)32 (55.2)43 (53.7)
 VDD pacing leads1 (0.7)0 (0)1 (1.2)
 Coronary sinus leads6 (4.3)2 (3.4)4 (5.0)
 Single coil ICD leads2 (1.4)0 (0)2 (2.5)
 Dual coil ICD leads4 (2.9)1 (1.7)3 (3.8)
Lead fixation
 Active70 (50.7)37 (63.8)33 (41.2)0.009
 Passive68 (49.3)21 (36.2)47 (58.8)
Unipolar leads3 (2.2)1 (1.7)2 (2.5)0.758
Vascular access for the leads
 Subclavian/axillary125 (90.6)52 (89.7)73 (91.2)0.752
 Cephalic13 (9.4)6 (10.3)7 (8.8)
Cut leads7 (5.1)0 (0)7 (8.8)0.021
Side of the implant
 Left side79 (57.2)37 (63.8)42 (52.5)0.186
 Right side59 (42.8)21 (36.2)38 (47.5)
Dwell time, months22.1 (4.5–61.3)3.9 (2.9–8.6)52.1 (28.2–117.2)
Place of explantation
 Electrophysiology lab128 (92.8)55 (94.8)73 (91.2)0.424
 Operating room10 (7.2)3 (5.2)7 (8.8)
Technique of explantation
 Manual traction only115 (83.3)58 (100.0)57 (71.2)
 Traction tools (Lead locking device and/or snare)9 (6.5)9 (11.2)
 Cutting sheaths14 (10.1)14 (17.5)
Successful removal
 Success123 (89.1)58 (100.0)65 (81.2)<0.001
 Total removal121 (87.7)58 (100.0)63 (78.8)
 Partial removal (<4 cm remnant left)2 (1.4)0 (0)2 (2.5)

Note: Values are frequency (%) or median (25th percentile - 75th percentile).

– Not applicable or not analyzed.

Comparison between explantation and extraction groups.

Lead and procedure characteristics. Note: Values are frequency (%) or median (25th percentile - 75th percentile). – Not applicable or not analyzed. Comparison between explantation and extraction groups.

Outcomes

Overall, 64/75 (85.3%) lead removal procedures were successful and 123/138 (89.1%) leads were removed. Analysis was done to 80 leads and 44 procedures that qualified as ‘lead extractions’. Complete success was achieved in 33/44 (75.0%) extraction procedures. When analyzed according to leads, 65/80 (81.2%) could be extracted successfully. Manual traction (including use of a goose-neck snare in 3 cases) was successful in 16/25 (64.0%) of the extraction procedures. On lead wise analysis, manual traction alone was successful in extracting 44/57 (77.2%) leads. All leads with a dwell time less than 33 months could be removed with manual traction alone. Dedicated tools were used for extraction of 23 leads - traction tools (LLD and/or dedicated snares) in 9 and cutting sheaths (telescopic and/or rotational cutting sheaths) in 14 leads. When procedures with the use of dedicated tools were analyzed, 17/19 (89.5%) of the procedures were successful and 21/23 (91.3%) of the leads could be successfully extracted. Multivariable analysis was performed only for pacing leads, as the ICD leads were few as compared to pacing leads and technical challenges of ICD implantation are different. Multivariable analysis was performed for predicting procedural failure by adding the following variables to the model - the inability to use dedicated tools, presence of passive fixation leads, right-sided implants, and dwell-time of the oldest lead (Table .4). After adjustment for other variables (Table .4), inability to use dedicated tools was the only significant predictor of procedural failure (adjusted OR 14.00; 95% CI 1.78–110.25; p-value 0.012).
Table 4

Predictors of failure of pacemaker extraction procedures (n = 39)a.

Failure (n = 11)Success (n = 28)p-valueUnivariate OR (95% CI)p-valueAdjusted ORc (95% CI)p-valuec
Age, years61.1 ± 18.065.8 ± 11.60.338
Male sex9 (81.9)21 (75.0)1.000
Comorbid conditions
 Diabetes mellitus2 (18.2)12 (42.9)0.266
 Hypertension6 (54.5)14 (50.0)0.798
 Chronic kidney disease0 (0)2 (7.1)0.918
 Ischemic heart disease1 (9.1)4 (14.3)1.000
 Prior open heart surgery1 (9.1)1 (3.6)0.490
 Left ventricular dysfunctionb0 (0)5 (17.9)0.644
Procedure characteristics
 Procedure in the EP lab11 (100.0)25 (89.3)0.644
 Use of manual traction only (unable to use special tools)9 (81.8)14 (50.0)0.0864.50 (0.82–24.68)b0.083b14.00 (1.78–110.25)0.012
Device characteristics
 Presence of passive leadsb11 (100.0)18 (64.3)0.05913.00 (0.60–280.53)0.1016.57 (0.22–199.54)0.280
 Presence of cut lead3 (27.3)3 (10.7)0.323
 Right-sided implants10 (90.9)13 (46.4)0.01411.54 (1.30–102.65)0.0287.05 (0.74–67.20)0.090
 Dwell time of the oldest lead, months117.2 (80.3–173.8)41.4 (21.9–99.8)0.0031.02 (1.00–1.03)0.0221.01 (1.00–1.02)0.210

Note: Values are mean ± standard deviation, frequency (%) or median (25th percentile - 75th percentile).

– Not applicable or not included in the multivariable model.

ICD leads were not included in multivariable analysis.

Penalized simple logistic regression was performed.

Penalized multiple logistic regression was performed.

Predictors of failure of pacemaker extraction procedures (n = 39)a. Note: Values are mean ± standard deviation, frequency (%) or median (25th percentile - 75th percentile). – Not applicable or not included in the multivariable model. ICD leads were not included in multivariable analysis. Penalized simple logistic regression was performed. Penalized multiple logistic regression was performed. Lead wise univariate analysis for the 75 pacing leads is summarized in Table .5. Multivariable analysis for predicting failure of lead extraction was performed by adding the following variables to the model - the inability to use dedicated tools, passive fixation leads, right-sided implants, and dwell-time. On multivariable analysis (Table .5), right-sided implants (adjusted OR 12.65; 95% CI 1.34–119.49; p-value 0.027), use of manual traction (adjusted OR 17.89; 95% CI 1.82–176.07; p-value 0.013), and dwell-time (adjusted OR 1.02; 95% CI 1.00–1.03; p-value 0.024) were significant.
Table 5

Lead characteristics predicting failure of extraction (n = 75)a.

Failure (n = 15)Success (n = 60)p-valueUnivariate OR (95% CI)p-valueAdjusted OR (95% CI)p-value
Venous access site
 Subclavian/axillary13 (86.7)55 (91.7)0.622
 Cephalic2 (13.3)5 (8.3)
Use of manual traction only (unable to use special tools)13(86.7)42(70.0)0.1922.79 (0.57–13.63)0.20617.89 (1.82–176.07)0.013
Passive fixation leads14 (93.3)33 (55.0)0.00611.46 (1.42–92.75)0.0227.40 (0.72–75.52)0.091
Right-sided implants14 (93.3)24 (40.0)<0.00121.00 (2.59–170.37)0.00412.65(1.34–119.49)0.027
Dwell time, months117.2 (80.3–162.6)46.9 (26.3–93.1)<0.0011.02 (1.01–1.03)0.0021.02(1.00–1.03)0.024

Note: Values are mean ± standard deviation, frequency (%) or median (25th percentile - 75th percentile).

– Not applicable or not included in multivariable analysis.

ICD leads were not included in multivariable analysis.

Lead characteristics predicting failure of extraction (n = 75)a. Note: Values are mean ± standard deviation, frequency (%) or median (25th percentile - 75th percentile). – Not applicable or not included in multivariable analysis. ICD leads were not included in multivariable analysis.

Complications

Complications occurred in 7 (9.3%) patients. Six of them were minor complications not resulting in death or permanent disability - 4 patients had bleeding requiring a blood transfusion, 1 patient developed sepsis post-procedure and recovered with antibiotic therapy, and 1 patient developed deep vein thrombosis of the distal lower limb veins. No deaths occurred during the procedure, but one patient died due to massive pulmonary embolism 1 week after failed extraction. Twenty-seven patients (36.0%) did not have long-term (>1month) follow-up. Of the remaining, none of the patients had a complication after a median follow-up of 21.4 months (IQR 7.1 to 53.4). Fifty-five patients underwent reimplantation. Eighteen patients with infected devices underwent reimplantation of their own old device after resterilization. Two patients were lost to follow-up. There was no recurrence of infection in the remaining 16 patients after a median follow-up of 15.7 months (IQR 7.0–39.7 months).

Discussion

Many previous studies from large-volume centers in the developed countries have shown that percutaneous lead extraction has high success with low complication rates [1,2,[10], [11], [12]]. The high efficacy is in part due to access to dedicated extraction tools. Although there are a growing number of CIED implants in developing countries and an accompanying increase in the indications for their removal, data about the methods used and outcomes is lacking. We report a single-center experience in lead removal from India, highlighting the issues unique to the resource-constrained setting. The key findings of the current study are (a) removal of CIED leads using manual traction and improvised snares was successful in extraction of 77.2% of the leads and 64.0% of the extraction procedures with good safety, (b) all leads with dwell time less than 33 months (≈2.7 years) could be removed safely without resorting to special extraction tools, (c) reimplantation of a resterilized device was not associated with early recurrence of infection and can be judiciously considered in some cases, and (d) extraction in EP lab was safe in appropriately selected patients, with surgical back-up. Manual traction has been reported to have a low success rate in previous studies, between 15 and 30% [[10], [11], [12]]. However, in those studies operators had access to specialized tools and a probable lower threshold to use them. The success rate with manual traction in this series is comparable to the study published by de Bie et al. which used predominantly manual traction (77.2% in our study with a median lead dwell time of 4.3 years V 75.7% in the study by de Bie et al. in the subgroup with dwell time >2.6 years) [13]. However, lead locking device was used in some cases in that study [13]. In the current study, among cases where specialized tools were used, the success was close to 90%. It is unclear at what point during the procedure manual traction should be considered unsuccessful and changed to another method. This varies between operators and will very likely to depend on their experience. However, when constrained by the availability of other tools, experienced operators are likely to exhaust all tricks during manual traction before considering it unsuccessful. This illustrates the fact that the major limitation to lead extraction in the developing world is the inability to use dedicated tools in all patients, mainly due to financial constraints. This was an independent predictor of procedural failure in our study. Right-sided implants were more difficult to extract and were independently associated with failure in this study. We speculate that due to an almost 900angle between right subclavian vein and the superior vena cava, force transmission of manual traction may be poor. Also, the cutting sheaths designed for extraction are not very flexible and difficult to maneuver from the right side. However, this was not previously reported in other series that used dedicated tools, indicating that this is probably a factor unique to manual extraction. Another problem encountered was the practice of cutting the leads flush with the pectoral fascia. This is done sometimes with an expectancy of healing, which was almost never the case in our experience. The biofilm of infecting organism frequently extends along the lead into deeper tissues, so eradication of infection is very unlikely just by cutting the lead flush. Although not analyzed statistically in this study due to the small number, in all cases it made the procedure more difficult. A cut lead poses several risks. First, it can retract into deeper tissues, or worse still, can retract into vascular space with the risk of dissemination of what was a localized infection. Second, lumen of a cut lead can be compromised due to tissue ingrowth, precluding the use of lead locking devices. Third, the short working length can make application of traction difficult. Considering the futility to control infection and the challenges it poses during extraction, cutting of the leads should be strongly discouraged. So, when feasible, it would be prudent to plan an extraction procedure with specialized tools available at hand when there are right-sided implants, cut leads and leads with long dwell time. Most procedures were performed in the EP lab with surgical backup. The safety of performing extractions in the EP lab was previously reported [14]. The arguments for extraction in the EP lab are that patients with massive bleeding are not more likely to be rescued in the operating room and performing every procedure in OR is impractical [14]. Even in the developed world, two-thirds of the procedures are performed in the EP lab [14]. The safety in this study was comparable to other large series [10,11,13]. However, there were lower number of ICD leads and dwell times were relatively short in this study. Short of a hybrid lab or a quick response surgical team, it is our opinion to perform at least the highest risk procedures in the operating room. Reuse of CIEDs, especially after infection, is an ethical grey area because of the manufacturers’ labeling of the devices as single-use and the potential for device malfunction. It is, however, an inevitable reality in many cases in the developing world due to financial constraints. Reuse of CIEDs has been previously reported to be safe and not associated with higher rates of infection or device malfunction [[15], [16], [17]]. In our series, we did not find any recurrence of infection or device malfunction. Guidelines from professional societies recommend transvenous extraction to be performed in centers with high volumes where thoracic surgery back-up is available and by operators with adequate experience [8,9]. However, such centers are very few in the developing world. The first step towards the evolution of such centers would be reporting of outcomes and sharing data towards establishing a registry. There is no standardized approach or recommendation for the selection of extraction tool or technique of extraction. The available tools are, in most cases, prohibitively expensive. So, there is a need for adapting improvised techniques to a resource-constrained setting and reporting such techniques.

Limitations

Being a retrospective single-center study with a relatively small sample size, there are limitations to its generalizability. Nevertheless, to our knowledge, this is the first case series about percutaneous lead removal from India. Inability to use specialized tools in all patients significantly decreases the success rate. However, this is the real scenario in most cases in India. The number of ICD leads in this series is small - reflecting the overall lower number of ICDs implanted.

Conclusions

In a resource-limited setting, percutaneous lead extraction of predominantly pacemaker leads, by manual traction methods achieved success in extracting three-fourths of the leads and all leads implanted less than 2.7 years could be removed with manual traction alone. The complications of lead removal were low. Inability to use specialized tools due to financial constraints was the main factor responsible for failure and it would be wise to plan a lead extraction of a right-sided implants, cut leads and leads with long dwell times with dedicated tools available at hand.

Declaration of competing interest

The authors declare no conflict of interest for this study.
  15 in total

1.  Current practice in transvenous lead extraction: a European Heart Rhythm Association EP Network Survey.

Authors:  Maria Grazia Bongiorni; Carina Blomström-Lundqvist; Charles Kennergren; Nikolaos Dagres; Laurent Pison; Jesper Hastrup Svendsen; Angelo Auricchio
Journal:  Europace       Date:  2012-06       Impact factor: 5.214

2.  Percutaneous extraction of cardiac pacemaker and implantable cardioverter defibrillator leads with evolution mechanical dilator sheath: a single-centre experience.

Authors:  Ali Oto; Kudret Aytemir; Hikmet Yorgun; Uğur Canpolat; Ergün Barış Kaya; Giray Kabakçı; Lale Tokgözoğlu; Hilmi Özkutlu
Journal:  Europace       Date:  2010-11-17       Impact factor: 5.214

3.  Transvenous lead extraction: Heart Rhythm Society expert consensus on facilities, training, indications, and patient management: this document was endorsed by the American Heart Association (AHA).

Authors:  Bruce L Wilkoff; Charles J Love; Charles L Byrd; Maria Grazia Bongiorni; Roger G Carrillo; George H Crossley; Laurence M Epstein; Richard A Friedman; Charles E H Kennergren; Przemyslaw Mitkowski; Raymond H M Schaerf; Oussama M Wazni
Journal:  Heart Rhythm       Date:  2009-05-22       Impact factor: 6.343

4.  Extraction of transvenous leads in the operating room versus electrophysiology laboratory: a comparative study.

Authors:  Frédéric Franceschi; Marc Dubuc; Jean-Claude Deharo; Julien Mancini; Pierre Pagé; Bernard Thibault; Linda Koutbi; Sébastien Prévôt; Paul Khairy
Journal:  Heart Rhythm       Date:  2011-02-09       Impact factor: 6.343

5.  A survey of the practice of lead extraction in the United States.

Authors:  Charles A Henrikson; Karl Zhang; Jeffrey A Brinker
Journal:  Pacing Clin Electrophysiol       Date:  2010-02-01       Impact factor: 1.976

Review 6.  2017 HRS expert consensus statement on cardiovascular implantable electronic device lead management and extraction.

Authors:  Fred M Kusumoto; Mark H Schoenfeld; Bruce L Wilkoff; Charles I Berul; Ulrika M Birgersdotter-Green; Roger Carrillo; Yong-Mei Cha; Jude Clancy; Jean-Claude Deharo; Kenneth A Ellenbogen; Derek Exner; Ayman A Hussein; Charles Kennergren; Andrew Krahn; Richard Lee; Charles J Love; Ruth A Madden; Hector Alfredo Mazzetti; JoEllyn Carol Moore; Jeffrey Parsonnet; Kristen K Patton; Marc A Rozner; Kimberly A Selzman; Morio Shoda; Komandoor Srivathsan; Neil F Strathmore; Charles D Swerdlow; Christine Tompkins; Oussama Wazni
Journal:  Heart Rhythm       Date:  2017-09-15       Impact factor: 6.343

7.  Reuse of pacemakers, defibrillators and cardiac resynchronisation devices.

Authors:  Raja J Selvaraj; R Sakthivel; Santhosh Satheesh; Ajith Ananthakrishna Pillai; Pascal Sagnol; Xavier Jouven; Bernard Dodinot; Jayaraman Balachander
Journal:  Heart Asia       Date:  2017-01-23

8.  Intravascular lead extraction using locking stylets, sheaths, and other techniques.

Authors:  N E Fearnot; H J Smith; L B Goode; C L Byrd; B L Wilkoff; T D Sellers
Journal:  Pacing Clin Electrophysiol       Date:  1990-12       Impact factor: 1.976

9.  Large, single-center, single-operator experience with transvenous lead extraction: outcomes and changing indications.

Authors:  Samuel O Jones; Robert E Eckart; Christine M Albert; Laurence M Epstein
Journal:  Heart Rhythm       Date:  2008-01-17       Impact factor: 6.343

10.  A single-centre experience of over one thousand lead extractions.

Authors:  Charles Kennergren; Christian Bjurman; Roger Wiklund; Jakob Gäbel
Journal:  Europace       Date:  2009-03-27       Impact factor: 5.214

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.