Literature DB >> 31840139

Impact of different scanning strategies on the accuracy of two current intraoral scanning systems in complete-arch impressions: an in vitro study.

Leandro Passos, Sergio Meiga, Vinicius Brigagão, Alexandre Street.   

Abstract

AIM: To determine the scanning strategy that obtains the most accurate results for two intraoral scanners (IOS) in complete-arch digital impressions. Scan time was evaluated and correlated with scan strategies. MATERIALS AND
METHOD: A custom model used as the reference standard was fabricated with teeth having dentin- and enamel-identical refractive indices simulating natural dentition. A reference scan of the custom typodont was obtained using an ATOS III Triple Scan 3D optical scanner. Two IOS setups - Omnicam v 5.1.0 and Primescan v 5.0.2 - were used for complete-arch scanning, each using 13 scanning strategies, obtaining 260 digital files (n = 10 per group), recording each scan time, converting all experimental scans to standard tessellation language (STL) format, and using a comprehensive metrology program to compare the reference standard scan with the experimental scans. Statistical analyses utilized Welch's unequal variances t test.
RESULTS: Group M exhibited the lowest trueness and precision values (P < 0.05) for Primescan (47.5% of the average among all other groups) and the lowest trueness value (P < 0.05) for Omnicam (53.4% of the average among all other groups), where group B exhibited the lowest precision value (65.6% of the average among all other groups) with P < 0.05. Primescan featured a better trueness index (4.79 µm) than that of Omnicam (19.13 µm), with a statistically significant difference (P < 0.00001). Primescan, group M, also featured a better precision index (4.67 µm) than Omnicam, group B (16.75 µm), with a statistically significant difference (P < 0.00001).
CONCLUSION: For both IOS systems, group M provided the lowest scanning times. For trueness and precision of complete-arch scans, group M was the dominant scanning strategy in Primescan, while there was no dominant strategy in Omnicam. Group M had the best scanning time for both IOS systems.

Keywords:  accuracy; digital impression; full-arch dental impression; intraoral scanner; standard tessellation language; CAD/CAM

Mesh:

Year:  2019        PMID: 31840139

Source DB:  PubMed          Journal:  Int J Comput Dent        ISSN: 1463-4201            Impact factor:   1.883


  10 in total

1.  Effect of full arch two scanning techniques on the accuracy of overdenture conventional and CAD/CAM Co-Cr bars.

Authors:  Ali Alenezi; Mohammed Yehya; Mohamed Alkhodary
Journal:  Saudi Dent J       Date:  2022-08-28

2.  Impact of scanning strategy on the accuracy of complete-arch intraoral scans: a preliminary study on segmental scans and merge methods.

Authors:  Hai Yen Mai; Hang-Nga Mai; Cheong-Hee Lee; Kyu-Bok Lee; So-Yeun Kim; Jae-Mok Lee; Keun-Woo Lee; Du-Hyeong Lee
Journal:  J Adv Prosthodont       Date:  2022-04-27       Impact factor: 1.989

3.  Influence of Scanning-Aid Materials on the Accuracy and Time Efficiency of Intraoral Scanners for Full-Arch Digital Scanning: An In Vitro Study.

Authors:  Hyun-Su Oh; Young-Jun Lim; Bongju Kim; Myung-Joo Kim; Ho-Beom Kwon; Yeon-Wha Baek
Journal:  Materials (Basel)       Date:  2021-04-30       Impact factor: 3.623

4.  Trueness of ten intraoral scanners in determining the positions of simulated implant scan bodies.

Authors:  Ryan Jin Young Kim; Goran I Benic; Ji-Man Park
Journal:  Sci Rep       Date:  2021-01-28       Impact factor: 4.379

5.  Accuracy of full-arch digitalization for partially edentulous jaws - a laboratory study on basis of coordinate-based data analysis.

Authors:  Panagiotis Kontis; Jan-Frederik Güth; Christine Keul
Journal:  Clin Oral Investig       Date:  2022-01-04       Impact factor: 3.573

6.  Efficacy of Constructing Digital Hybrid Skull-Dentition Images Using an Intraoral Scanner and Cone-Beam Computed Tomography.

Authors:  Joo-Hee Lee; Soo-Hwan Byun; Sang-Min Yi; In-Young Park; Byoung-Eun Yang; Hye-Lim Lee
Journal:  Scanning       Date:  2022-03-03       Impact factor: 1.932

7.  The Effect of Scanning Strategy on Intraoral Scanner's Accuracy.

Authors:  Nikolaos A Gavounelis; Chrysoula-Maria C Gogola; Demetrios J Halazonetis
Journal:  Dent J (Basel)       Date:  2022-07-04

8.  Effect of decompression and online transmission on the dimensional changes of .stl files generated by intra-oral scanning.

Authors:  Juliano Martins Bueno; Carolina Guarniéri Gouveia; Mayara Barbosa Viandelli Mundim; Ademir Franco; José Luiz Cintra Junqueira; Monikelly do Carmo Chagas Nascimento
Journal:  PLoS One       Date:  2022-09-15       Impact factor: 3.752

9.  Accuracy of Digital and Conventional Full-Arch Impressions in Patients: An Update.

Authors:  Alexander Schmidt; Leona Klussmann; Bernd Wöstmann; Maximiliane Amelie Schlenz
Journal:  J Clin Med       Date:  2020-03-04       Impact factor: 4.241

10.  Elastic deformation of the mandibular jaw revisited-a clinical comparison between digital and conventional impressions using a reference.

Authors:  Alexander Schmidt; Leona Klussmann; Maximiliane A Schlenz; Bernd Wöstmann
Journal:  Clin Oral Investig       Date:  2021-01-13       Impact factor: 3.573

  10 in total