Assessing the dependency of the uncertainties in the Elekta Agility MLC calibration procedure on the focal spot position.

Jacek M Chojnowski1,2, Lee M Taylor3, Jonathan R Sykes4,5, David I Thwaites4.   

Abstract

The effectiveness of radiotherapy treatments depends on the accuracy of the dose delivery process. The majority of radiotherapy courses are delivered on linear accelerators with a Multi Leaf Collimator (MLC) in 3D conformal Radiation Therapy, Intensity Modulated Radiation Therapy (IMRT) or Volumetric Modulated Arc Therapy (VMAT) modes that require accurate MLC positioning. This study investigates the MLC calibration accuracy, following manufacturer procedures for an Elekta Synergy linac with the Agility head, against the radiation focal spot offset (alignment with the collimator axis of rotation). If the radiation focal spot is not aligned ideally with the collimator axis of rotation then a systematic error can be introduced into the calibration procedure affecting absolute MLC leaf positions. Calibration of diaphrams is equally affected; however they are not investigated here. The results indicate that an estimated 0.15 mm MLC uncertainty in all MLC leaves positions can be introduced due to uncertainty of the radiation focal spot position of 0.21 mm.

Keywords:  Focal spot position; Linac quality assurance; MLC calibration

Year:  2019        PMID: 31792725     DOI: 10.1007/s13246-019-00821-x

Source DB:  PubMed          Journal:  Australas Phys Eng Sci Med        ISSN: 0158-9938            Impact factor:   1.430


  3 in total

1.  A novel method to determine linac mechanical isocenter position and size and examples of specific QA applications.

Authors:  Jacek M Chojnowski; Jonathan R Sykes; David I Thwaites
Journal:  J Appl Clin Med Phys       Date:  2021-05-30       Impact factor: 2.102

2.  Identification of a potential source of error for 6FFF beams delivered on an AgilityTM multileaf collimator.

Authors:  Friedlieb H Lorenz; Matthew I Paris
Journal:  J Appl Clin Med Phys       Date:  2021-03-06       Impact factor: 2.102

3.  Consistency of small-field dosimetry, on and off axis, in beam-matched linacs used for stereotactic radiosurgery.

Authors:  Luis Muñoz; Tomas Kron; Marco Petasecca; Joseph Bucci; Michael Jackson; Peter Metcalfe; Anatoly B Rosenfeld; Giordano Biasi
Journal:  J Appl Clin Med Phys       Date:  2021-01-13       Impact factor: 2.102

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.