Joonas Toivonen1,2, Mikko Björkqvist3,4, Heikki Minn4, Pekka K Vallittu2,5, Jami Rekola1,2.
Abstract
PURPOSE: Radiation scattering from bone reconstruction materials can cause problems from prolonged healing to osteoradionecrosis. Glass fiber reinforced composite (FRC) has been introduced for bone reconstruction in craniofacial surgery but the effects during radiotherapy have not been previously studied. The purpose of this study was to compare the attenuation and back scatter caused by different reconstruction materials during radiotherapy, especially FRC with bioactive glass (BG) and titanium.Entities:
Keywords: bioglass; craniofacial bone reconstruction; fiber reinforced composite; radiation scattering; radiotherapy
Mesh:
Substances:
Year: 2019 PMID: 31782897 PMCID: PMC6909125 DOI: 10.1002/acm2.12776
Source DB: PubMed Journal: J Appl Clin Med Phys ISSN: 1526-9914 Impact factor: 2.102
Materials used in the FRC and FRC‐BG samples.
| Brand | Manufacturer | Lot. |
|---|---|---|
| Bis‐GMA | Esstech Europe ltd | 751‐42 |
| TEGDMA | Sigma‐Aldrich Co. LCC | STBC5193V |
| CQ | Sigma‐Aldrich Co. LCC | STBC7007V |
| DMAEMA | Sigma‐Aldrich Co. LCC | 1437599V |
| 120 g/m2 woven E‐glass fiber fabric | Ahlstrom Glassfibre Oy | NA |
| BonAlive glass granules | Vivoxid ltd | BG‐11/08, BG‐03/09 |
Bis‐GMA, bisphenol‐A‐glycidyldimethacrylate; CQ, camphorquinone; DMAEMA, N,N‐dimethylaminoethyl methacrylate; FRC, fibre‐reinforced composites; TEGDMA, triethylene glycoldimethacrylate.
Figure 1FRC (left) and FRC‐BG (right) samples. FRC, fiber‐reinforced composites.
Figure 2The setup of film dosimetry measurements.
Figure 3The setup of diamond detector measurements.
Results of the film dosimetry (relative to water equivalent material alone).
| Sample | 6 MV | 10 MV | ||||
|---|---|---|---|---|---|---|
| Film 1 | Film 2 | Film 3 | Film 1 | Film 2 | Film 3 | |
| FRC‐BG (3 mm) | 0% | 0% | 0% | 0% | 0% | 0% |
| FRC (3 mm) | 4% | 1% | 0% | 6% | 0% | 0% |
| PEEK (3.8 mm) | 10% | 0% | −1% | 7% | 1% | 1% |
| Ti (3 mm) | 18% | −15% | −9% | 17% | −5% | −4% |
FRC‐BG, glass fiber reinforced composite with bioactive glass; FRC, glass fiber reinforced composite; PEEK, polyether ether ketone; Ti, titanium.
Figure 4Change in radiation dose with increasing distance with 6 MV energy.
Figure 5Change in radiation dose with increasing distance with 10 MV energy.
Figure 6Change in dose of radiation with 6 MV energy measured with the diamond detector with different sample thicknesses at 1 mm distance.
Figure 7Change in dose of radiation with 6 MV energy measured with the diamond detector with different sample thicknesses at 10 mm distance.
Figure 8Change in dose of radiation with 10 MV energy measured with the diamond detector with different sample thicknesses at 1 mm distance.
Figure 9Change in dose of radiation with 10 MV energy measured with the diamond detector with different sample thicknesses at 10 mm distance.