| Literature DB >> 31779856 |
Adam Stys1, Maheedhar Gedela2, Udit Bhatnagar1, Marian Petrasko1, Hazem Dawoud3, Tadeusz Malinski3, Tomasz Stys1.
Abstract
BACKGROUND: We aimed to assess the decrease in contrast media volume (CMV) with ultra-low contrast delivery technique (ULCD) developed at our institution versus the usual automated contrast injector system (ACIS) contrast delivery in coronary procedures.Entities:
Keywords: Automated contrast injector system; Contrast media; Contrast-induced nephropathy; Ultra-low contrast delivery
Mesh:
Substances:
Year: 2019 PMID: 31779856 PMCID: PMC6890984 DOI: 10.1016/j.ihj.2019.09.001
Source DB: PubMed Journal: Indian Heart J ISSN: 0019-4832
Fig. 1Left coronary artery ACIST algorithm using ULCD technique (Reproduced with permission from the Journal of Invasive Cardiology). ULCD, ultra-low contrast delivery.
Fig. 2Right coronary artery ACIST algorithm using ULCD technique (Reproduced with permission from the Journal of Invasive Cardiology). ULCD, ultra-low contrast delivery.
Mean amount of contrast media volume delivered per patient using ACIST device with and without ULCD technique.
| Type of procedure | ULCD | Contrast volume (mL) | ACIS | Contrast volume (mL) | Mean percent contrast reduction (%) | |
|---|---|---|---|---|---|---|
| Diagnostic angiogram | 194 | 24.8 ± 15.8 | 200 | 42.3 ± 25.1 | 41.5 | <0.0001 |
| PCI | 52 | 23.5 ± 19.7 | 16 | 48.2 ± 30.8 | 51.2 | 0.0070 |
| Diagnostic angiogram + | 23 | 53.4 ± 32.1 | 16 | 89.7 ± 35.6 | 40.4 | 0.0024 |
| Combined diagnostic angiogram and PCI | 204 | 27.4 ± 20.5 | 181 | 44.9 ± 28.0 | 45.7 | <0.0001 |
ULCD, ultra-low contrast delivery; ACIS, automated contrast injector system; PCI, percutaneous coronary intervention.
Clinical and procedural characteristics.
| Clinical variable | ULCD | ACIS | |
|---|---|---|---|
| Age in years (mean ± SD) | 67.75 ± 11.21 | 68.56 ± 11.91 | 0.482 |
| Male | 121 (59.3%) | 113 (56.5%) | 0.567 |
| Female | 83 (40.7%) | 87 (43.5%) | |
| Weight in kkg (mean ± SD) | 94.41 ± 24.08 | 93.84 ± 24.28 | 0.814 |
| Height in cm (mean ± SD) | 171.51 ± 11.57 | 171.37 ± 11.36 | 0.905 |
| Body surface area in m2 | 2.05 ± 0.30 | 2.04 ± 0.28 | 0.828 |
| Creatinine clearance | |||
| eGFR <30% | 9 (4.4%) | 7 (3.5%) | 0.638 |
| eGFR 30–59% | 40 (19.6%) | 40 (20.0%) | 0.921 |
| eGFR 60–89% | 97 (47.5%) | 99 (49.5%) | 0.695 |
| eGFR >90% | 57 (27.9%) | 53 (26.5%) | 0.745 |
| History of CABG | 28 (13.7%) | 19 (9.5%) | 0.185 |
| No. of diagnostic coronary angiogram | 194 | 200 | N/A |
| No. of percutaneous intervention | 59 | 28 | N/A |
| No. of diagnostic coronary angiogram and | 24 | 28 | N/A |
| Indication for the procedure | |||
| Elective | 181 (88.7%) | 185 (92.5%) | 0.194 |
| Urgent | 23 (11.3%) | 14 (7.0%) | 0.136 |
| Emergent | 2 (1.0%) | 0 (0.0%) | 0.159 |
| No. of biplane used | 116 (58.0%) | 118 (59.3%) | 0.793 |
| No. of single plane used | 84 (42.0%) | 80 (40.2%) | 0.715 |
| Radial access | 173 (84.8%) | 183 (91.5%) | 0.038 |
| Ulnar access | 2 (1.0%) | 0 (0.0%) | 0.160 |
| Femoral access | 32 (15.7%) | 31 (15.5%) | 0.959 |
| Total fluoro dose in mGy | 535.04 ± 500.37 | 720.67 ± 674.82 | 0.002 |
| Total procedure time in min | 18.02 ± 11.27 | 29.35 ± 27.20 | <0.001 |
ULCD, ultra-low contrast delivery; ACIS, automated contrast injector system; eGFR, estimated glomerular filtration rate; CABG, coronary artery bypass graft surgery.
Fig. 3Bar-graph representation of contrast media volume administration using ULCD technique vs. ACIS across all coronary procedures. ULCD, ultra-low contrast delivery; ACIS, automated contrast injector system.
Comparison of contemporary studies in contrast volume reduction.
| Study | Device and/or technique used | Number of patients | Type of study | Type of procedure | Amount of contrast media volume delivered in mL |
|---|---|---|---|---|---|
| Sapontis J et al. | DyeVert | 44 | Prospective, observational | Diagnostic catheterization and/or PCI | 88.7 ± 56.9 |
| Corcione N et al. | DyeVert NG | 10 | Retrospective, observational | Diagnostic catheterization and/or PCI and peripheral angiogram and stenting | 79.9 ± 48.8 |
| Desch et al. | DyeVert | 96 | Prospective, randomized | Diagnostic catheterization | 36.9 ± 10.9 |
| Mehran et al. | AVERT | 578 | Prospective, randomized | Diagnostic catheterization and/or PCI | All procedures: 85.6 ± 50.5 |
| Gurm et al. | DyeVert Plus System | 114 | Prospective, multicenter, single-arm, observational | Diagnostic catheterization and/or PCI | 67 ± 51 |
| Our study | ULCD technique + ACIS | 204 | Prospective, single-center, observational | Diagnostic catheterization and/or PCI or | Diagnostic angiogram: 24.8 ± 15.8 |
ULCD, ultra-low contrast delivery; ACIS, automated contrast injector system; PCI, percutaneous coronary intervention.