Elton Ho1,2, Jack Boffa2, Daniel Palanker2,3.
Abstract
Photovoltaic subretinal prosthesis is designed for restoration of central vision in patients with age-related macular degeneration (AMD). We investigated the utility of prosthetic central vision for complex visual tasks using augmented-reality (AR) glasses simulating reduced acuity, contrast, and visual field. AR glasses with blocked central 20° of visual field included an integrated video camera and software which adjusts the image quality according to three user-defined parameters: resolution, corresponding to the equivalent pixel size of an implant; field of view, corresponding to the implant size; and number of grayscale levels. The real-time processed video was streamed on a screen in front of the right eye. Nineteen healthy participants were recruited to complete visual tasks including vision charts, sentence reading, and face recognition. With vision charts, letter acuity exceeded the pixel-sampling limit by 0.2 logMAR. Reading speed decreased with increasing pixel size and with reduced field of view (7°-12°). In the face recognition task (four-way forced choice, 5° angular size) participants identified faces at >75% accuracy, even with 100 μm pixels and only two grayscale levels. With 60 μm pixels and eight grayscale levels, the accuracy exceeded 97%. Subjects with simulated prosthetic vision performed slightly better than the sampling limit on the letter acuity tasks, and were highly accurate at recognizing faces, even with 100 μm/pixel resolution. These results indicate feasibility of reading and face recognition using prosthetic central vision even with 100 μm pixels, and performance improves further with smaller pixels.Entities:
Mesh:
Year: 2019 PMID: 31770773 PMCID: PMC6880846 DOI: 10.1167/19.13.22
Source DB: PubMed Journal: J Vis ISSN: 1534-7362 Impact factor: 2.240
Figure 1Experimental setup. (a) Schematic of the experimental setup. High resolution images are presented on a monitor. The front camera of the augmented-reality (AR) glasses captures the video stream. Custom software preloaded on the AR glasses adjusts the video quality to mimic prosthetic vision and displays it in the AR glasses. (b) A subject in front of the apparatus. (c) Illustration of vision through the AR glasses.
Parameters of the image processing used for each experiment.
| Task | Equivalent pixel size on retina (μm) | Number of grayscale levels | FOV (°) |
| Letter acuity | 30, 60, 100 | 2, 8 | 7 |
| Sentence reading | 30, 60, 100 | 8 | 7, 12 |
| Face recognition | 60, 100 | 2, 4 ,8 | 7 |
Figure 2Face recognition task. (a) An example set of five faces presented. Subjects were asked to pick the face that matches the identity of the central person. Each face spanned approximately 5° × 5°. (b) Effects of the number of grayscale levels and resolution on an image.
Figure 3Letter acuity results (n = 13 for 30 and 60 μm pixels; n = 19 for natural vision and 100 μm pixels). The leftmost data point at 5 μm indicates visual acuity (VA) for natural vision of the subjects. Error bars are presented in terms of SD.
Figure 4Sentence reading speed in words per minute (WPM). (a) Simple sentences. (b) Complex sentences. Faded lines represent individual measurements, and the bold lines represent the population mean.
Reading acuity (RA), maximum reading speed (MRS), and critical print size (CPS) for reading MNREAD sentences using simulated prosthetic vision. Notes: All errors are reported as standard deviation.
| Parameters (pixels/FOV) | RA (°) | MRS (WPM) | CPS (°) |
| 100 μm /7° | 1.60 ± 0.15 | 58 ± 10 | 3.0 ± 0.6 |
| 60 μm /7° | 0.81 ± 0.08 | 77 ± 16 | 1.7 ± 0.7 |
| 30 μm /7° | 0.46 ± 0.10 | 106 ± 22 | 1.0 ± 0.6 |
| 30 μm/12° | 0.45 ± 0.06 | 141 ± 27 | 1.1 ± 0.2 |
| Natural | 0.14 ± 0.01 | 201 ± 40 | 0.36 ± 0.07 |
Reading Acuity (RA), maximum reading speed (MRS), and critical print size (CPS) for reading complex sentences using pixelated vision. Notes: All errors are reported as standard deviation. Asterisk (*) indicates p < 0.05 (two-sample t test) compared to simple sentences with the same parameters.
| Parameters (pixels/FOV) | RA (°) | MRS (WPM) | CPS (°) |
| 100 μm /7° | 1.63 ± 0.11 | 52 ± 7 | 2.7 ± 0.4 |
| 60 μm /7° | 0.71 ± 0.11* | 73 ± 10 | 1.7 ± 0.5 |
| 30 μm /7° | 0.36 ± 0.05* | 102 ± 18 | 0.80 ± 0.15 |
| 30 μm /12° | 0.37 ± 0.09* | 121 ± 33 | 0.87 ± 0.25* |
| Natural | 0.14 ± 0.01 | 173 ± 34 | 0.32 ± 0.09 |
Figure 5Face recognition. (a) Accuracy. (b) Response time. (c) Response time normalized to 100 μm pixels and eight grayscale levels. Each dot represents an independent measurement. Error bars are presented in terms of SD.