| Literature DB >> 31761975 |
Christian V Eisenring1,2,3, Felice Burn4,5, Michelle Baumann4, Lennart H Stieglitz4,6,7, Ralf A Kockro8, Jürgen Beck4,9, Andreas Raabe4, Markus F Oertel4,6,7.
Abstract
BACKGROUND: Currently, the trajectory for insertion of an external ventricular drain (EVD) is mainly determined using anatomical landmarks. However, non-assisted implantations frequently require multiple attempts and are associated with EVD malpositioning and complications. The authors evaluated the feasibility and accuracy of a novel smartphone-guided, angle-adjusted technique for assisted implantations of an EVD (sEVD) in both a human artificial head model and a cadaveric head.Entities:
Keywords: App; External ventricular drain; Neuronavigation; Smartphone; Ventriculostomy
Mesh:
Year: 2019 PMID: 31761975 PMCID: PMC7046572 DOI: 10.1007/s00701-019-04131-9
Source DB: PubMed Journal: Acta Neurochir (Wien) ISSN: 0001-6268 Impact factor: 2.216
Fig. 1Planning of sEVD insertion trajectories on multiplanar 3D reformatted native CT. Coronal (left) and sagittal (right) images of a human cadaveric head demonstrate determination of the sEVD insertion point located 12 cm distally to the nasion and 2 cm paramedially. The trajectory shown is tilted 12° laterally with a 5.22-cm insertion length. Note post-mortem intracranial air inclusion after anatomical fixation and loss of cerebrospinal fluid
Fig. 2Smartphone calibration. The 0° angle was calibrated according to the sagittal midline. The smartphone was placed along the midline parallel to the insertion plane defined by both the upper edge of the external acoustic meatus and Kocher’s point
Fig. 3Marking of midline in the sagittal plane, determination of the coronal insertion plan and angle-adjusted trephination. The insertion plane is represented by the curvilinear line that crosses the midline and the insertion point. The axis of the drill was routinely guided within the coronal insertion plane under visual control. It was tilted laterally toward the planned angle with the aid of the smartphone apps verbal angle specification. Placement of the twist drill was followed by angle-adjusted trephination
Fig. 4Smartphone-navigated placement of the EVD. The drain was angle-adjusted with the smartphone and inserted into the ventricle. The midline is marked in the sagittal plane. The insertion plane is represented by the curvilinear line that crosses the midline and the insertion point. The axis of the catheter was always guided within the coronal insertion plane under visual control. It was tilted laterally toward the planned angle with the aid of the smartphone apps verbal angle specification
Planned pre- and post-interventional catheter positions with mean values and standard deviation
| Planned angle (degrees) | Post-interventional angle (degrees) | Deviation (degrees) | Deviation (cm) | Intraventricular position | Kinking | |
|---|---|---|---|---|---|---|
| Head model | ||||||
| 19.60 | 19.40 | 0.20 | 0.01 | + | − | |
| 16.95 | 16.80 | 0.15 | 0.01 | + | − | |
| 14.05 | 13.10 | 0.95 | 0.07 | + | − | |
| 10.35 | 7.90 | 2.45 | 0.16 | + | − | |
| 7.70 | 10.20 | 2.50 | 0.17 | + | − | |
| 5.35 | 8.50 | 3.15 | 0.22 | + | − | |
| 6.35 | 7.90 | 1.55 | 0.11 | + | − | |
| 4.00 | 6.90 | 2.90 | 0.21 | + | − | |
| Cadaveric head | ||||||
| 18.70 | 16.50 | 2.20 | 0.21 | + | − | |
| 49.30 | 42.10 | 7.20 | 0.81 | + | + | |
| 13.00 | 13.90 | 0.90 | 0.09 | + | − | |
| 43.00 | 45.80 | 2.80 | 0.25 | + | − | |
| 10.00 | 9.30 | 0.70 | 0.06 | + | − | |
| 40.40 | 37.20 | 3.20 | 0.26 | + | + | |
| 12.00 | 18.10 | 6.10 | 0.60 | + | + | |
| 42.00 | 45.90 | 3.90 | 0.35 | + | + | |
| 13.20 | 10.80 | 2.40 | 0.21 | + | − | |
| 37.00 | 30.70 | 6.30 | 0.54 | + | + | |
| 8.00 | 12.00 | 4.00 | 0.39 | + | − | |
| 35.00 | 36.70 | 1.70 | 0.15 | + | − | |
cm centimeter, M mean, SD standard deviation
Pre-and post-interventional measurements of insertion points posterior to nasion, lateral to midline and insertion depth of the catheter, with mean values and standard deviation for both the head model and human head
| Pre-interventional measurements posterior to nasion (cm) | Deviation of post-interventional measurements posterior to nasion (cm) | Pre-interventional planned position lateral to midline (cm) | Deviation of post-interventional measurements lateral to midline (cm) | Pre-interventional planned insertion depth (cm) | Post-interventional measurement of insertion depth (cm) | Deviation of post-interventional measured insertion depth (cm) | |
|---|---|---|---|---|---|---|---|
| Head model | |||||||
| 10.00 | 0.00 | 2.50 | 0.06 | 3.90 | 3.94 | 0.04 | |
| 11.00 | 0.20 | 2.50 | 0.05 | 4.30 | 4.03 | 0.27 | |
| 12.00 | 0.20 | 2.50 | 0.03 | 4.20 | 4.14 | 0.06 | |
| 13.00 | 0.24 | 2.50 | 0.04 | 4.20 | 3.80 | 0.40 | |
| 10.00 | 0.00 | 2.50 | 0.02 | 4.00 | 3.92 | 0.08 | |
| 11.00 | 0.20 | 2.50 | 0.03 | 4.03 | 3.99 | 0.04 | |
| 12.00 | 0.20 | 2.50 | 0.01 | 4.10 | 4.15 | 0.05 | |
| 13.00 | 0.24 | 2.50 | 0.07 | 4.10 | 4.23 | 0.13 | |
| Cadaveric head | |||||||
| 10.00 | 0.60 | 2.00 | 0.10 | 6.05 | 5.60 | 0.45 | |
| 10.00 | 0.60 | 2.00 | 0.06 | 6.45 | 6.64 | 0.19 | |
| 10.00 | 0.60 | 5.00 | 0.19 | 5.50 | 5.59 | 0.09 | |
| 10.00 | 0.60 | 5.00 | 0.08 | 5.01 | 5.55 | 0.54 | |
| 12.00 | 0.70 | 2.00 | 0.03 | 5.14 | 5.30 | 0.17 | |
| 12.00 | 0.70 | 2.00 | 0.07 | 4.70 | 5.45 | 0.75 | |
| 12.00 | 0.70 | 5.00 | 0.12 | 5.22 | 5.98 | 0.76 | |
| 12.00 | 0.70 | 5.00 | 0.05 | 5.15 | 5.59 | 0.44 | |
| 14.00 | 0.81 | 2.00 | 0.01 | 5.15 | 5.00 | 0.15 | |
| 14.00 | 0.81 | 2.00 | 0.04 | 4.90 | 5.10 | 0.20 | |
| 14.00 | 0.81 | 5.00 | 0.06 | 5.60 | 6.20 | 0.60 | |
| 14.00 | 0.81 | 5.00 | 0.03 | 5.15 | 5.59 | 0.44 | |
cm centimeter, M mean, SD standard deviation
Planned pre- and post-interventional catheter positions after exclusion of catheters with kinking, with mean values and standard deviation in the cadaveric head
| Planned angle (degrees) | Post-interventional angle (degrees) | Deviation (degrees) | Deviation (cm) | Pre-interventional planned insertion depth (cm) | Post-interventional measurement of insertion depth (cm) | Deviation of post-interventional measured insertion depth (cm) | |
|---|---|---|---|---|---|---|---|
| 18.70 | 16.50 | 2.20 | 0.21 | 6.05 | 5.60 | 0.45 | |
| 13.00 | 13.90 | 0.90 | 0.09 | 5.50 | 5.59 | 0.09 | |
| 43.00 | 45.80 | 2.80 | 0.24 | 5.01 | 5.55 | 0.54 | |
| 10.00 | 9.30 | 0.70 | 0.06 | 5.13 | 5.30 | 0.17 | |
| 13.20 | 10.80 | 2.40 | 0.21 | 5.15 | 5.00 | 0.15 | |
| 8.00 | 12.00 | 4.00 | 0.39 | 5.60 | 6.20 | 0.60 | |
| 35.00 | 36.70 | 1.70 | 0.15 | 5.15 | 5.59 | 0.44 | |
cm centimeter, M mean, SD standard deviation