| Literature DB >> 31750096 |
Tamer Mohamed Naguib1, Sameh Abdelkhalik Ahmed1.
Abstract
BACKGROUND: Flexible laryngeal mask airway may be used instead of the endotracheal tube in children presented for elective repair of tongue trauma, as it may shorten the time for extubation and recovery with the amelioration of stress response to airway management.Entities:
Keywords: Endotracheal Tube; Extubation Time; Intubation Time; Laryngeal Mask Airway; Recovery Time
Year: 2019 PMID: 31750096 PMCID: PMC6820298 DOI: 10.5812/aapm.92929
Source DB: PubMed Journal: Anesth Pain Med ISSN: 2228-7523
Figure 1.CONSORT flow chart of the study
Demographic Data in the Studied Groups[a]
| ETT Group | LMA Group | P Value | CI (95%) | |
|---|---|---|---|---|
|
| 3.24 ± 0.857 | 3.13 ± 0.815 | 0.530 | -0.461 - 0.239 |
|
| ||||
| Male | 27 (60) | 24 (53.33) | 0.671 | 0.749 - 1.757 |
| Female | 18 (40) | 21 (46.67) | ||
|
| 15.33 ± 2.86 | 14.60 ± 2.42 | 0.192 | -1.842 - 0.376 |
|
| ||||
| Class I | 36 (80) | 33 (73.33) | 0.619 | 0.707 - 2.096 |
| Class II | 9 (20) | 12 (26.67) | ||
|
| ||||
| Lip trauma | 1 (2.22) | 0 (0.00) | 1.00 | 1.639 - 2.496 |
| Gum trauma | 1 (2.22) | 0 (0.00) | 1.00 | 1.639; 2.496 |
| Teeth trauma | 0 (0.00) | 0 (0.00) | ||
| Larrynx trauma | 0 (0.00) | 0 (0.00) | ||
| Gagging | 10 (22.22) | 4 (8.89) | 0.144 | 1.028 - 2.340 |
| Cough | 11(24.44) | 3 96.67) | 0.039[ | 1.212 - 2.544 |
| Laryngeospasm | 5 (11.11) | 1 (2.22) | 0.203 | 1.147 - 2.670 |
| Bronchospasm | 4 (8.89) | 1 (2.22) | 0.361 | 1.015 - 2.709 |
| Stridor | 8 (17.78) | 0 (0.00) | 0.006[ | 1.746 - 2.814 |
| Sore throat | 10 (22.22) | 2 (4.44) | 0.027[ | 1.305 - 2.643 |
Abbreviations: CI, confidence interval; ETT, Endotracheal Tube; LMA, laryngeal mask airway.
aData were presented as means ± standard deviation or as No. (%).
bSignificant difference between the two groups.
Duration of Intubation, Extubation, and Recovery in the Studied Groups
| ETT Group | LMA Group | P | CI (95%) | |
|---|---|---|---|---|
|
| 46.98 ± 10.68 | 46.60 ± 11.73 | 0.874 | -5.078 - 4.323 |
|
| 37.89 ± 6.35 | 36.77 ± 6.41 | 0.411 | -3.784 - 1.562 |
|
| 96.13 ± 11.79 | 95.20 ± 13.24 | 0.725 | -6.184 - 4.318 |
|
| 13.29 ± 2.263 | 7.47 ± 2.7 | < 0.0001[ | 4.769 - 6.876 |
|
| 60.67 ± 11.16 | 52.67 ± 11.16 | 0.001[ | 3.325 - 2.675 |
Abbreviations: CI, confidence interval; ETT, Endotracheal Tube; LMA, laryngeal mask airway.
aData were presented as means ± standard deviation or as No. (%).
bSignificant difference between the two groups.
Figure 2.Changes in the mean heart rate in the two groups. Data were presented as means ± standard deviation. *denotes a significant change. T0: before anesthesia induction, T1: after anesthesia induction and before airway securing, T2: after airway securing, T3: at the beginning of the surgery, T4: at the end of the surgery.
Figure 3.Changes in the mean systolic blood pressure in the two groups. Data were presented as means ± standard deviation. *denotes a significant change. T0: before anesthesia induction, T1: after anesthesia induction and before airway securing, T2: after airway securing, T3: at the beginning of the surgery, T4: at the end of the surgery.
Figure 4.Surgical exposure score in the two groups. Data were presented as the number of patients.